Instructions to replace the A/D Converter and Inguard Logic Assembly (A3)
These instructions are intended for the replacement of the A3 assembly in the 3458A that has been
diagnosed with the short term drift problem described in Service Note 3458A-18.

After completing the A3 assembly replacement the 3458A needs to be warmed up for 4 hours in
its operating environment and an ACAL ALL initiated.

The ACAL ALL will adjust the 3458A to its internal references and it will be ready to make
accurate measurements. Where necessary, a DC Gain Adjustment may be required. (see 3458A
Multimeter Calibration Manual P/N 03458-90017 on the Agilent web at:
http://cp.literature.agilent.com/litweb/pdf/03458-90017.pdf)

This document contains general precautions that should be utilized when the 3458A is opened for
internal service procedures.

WARNING
Only personnel with knowledge of electronic circuitry and an awareness of the hazards involved
should remove and install any printed circuit board assemblies.

Caution: To prevent equipment circuit damage, always remove the AC line power cord before
removing or replacing any assembly. To prevent electrostatic discharge (ESD) damage to ICs,
always observe anti-static techniques when assemblies are handled or serviced.

Static Handling
Static electricity is a familiar phenomenon which, except for an occasional shock, doesn't seem
very serious. However, it has been proven that in the electronics industry ESD is a major cause of
component failure. In many cases, the component damaged may not immediately fai, causing low
instrument reliability and future repairs. ESD damage can occur at static levels below human
perception. It has also been shown that ESD can affect both passive and active devices.

The following guidelines are the minimum requirements for static safe service environments:
              ! The workbench should be equipped with a conductive table mat. The mat
                should be grounded to earth ground through a 1 M ohm resistor. The mat should
                be equipped with at least one swivel connector for connecting wrist straps.
              ! All service and handling personnel should wear a conductive wrist strap in
                contact with bare skin. This strap should be connected to the swivel connector
                on the conductive table mat through a 1 M ohm resistor.
              ! All metal equipment at the workstation must be grounded. This includes
                soldering irons, solder removers, shelving and equipment stands.
              ! Only one common ground should be provided at the workstation.
              ! The workstation should be kept free of nonconductors. No common plastics,
                polybags, cardboards, cigarette or candy wrappers should be allowed. There
                should not be rugs or carpet on the floor, shelving or bench top.
                    !    Only proper containers should be used for shipping, storing, or transporting
                        assemblies. This is required on any assembly shipped to Agilent Technologies
                        for repair or replacement.

    Clean Handling
    Due to the accuracy of the 3458A, use the following clean handling techniques when
    removing/installing printed circuit board assemblies.
                   ! Hand the assemblies only by their edges.
                   ! Be sure to place them on a clean workbench away from dirty or dusty
                     conditions.

    A3 Replacement Service Procedure
          The following procedures describe how to remove the bottom cover and shields on the
    3458A. Removal of the cover and shields are required to replace the printed circuit board
    assemblies (A3) for this repair.

    Tools Required
                            1. #1 Pozidriv screwdriver
                            2. #TX15 Torx driver
                            3. #TX10 Torx driver

    Covers Removal Procedure
                1. Remove all connections to the 3458A.
                2. Remove AC power to the 3458A.
                3. Turn the instrument so its right side (as seen from the front) faces you. Use the
                   #1 Pozidriv to remove the right side handle strap screws. Then remove the
                   strap
                                                                                           Right handle
                                                                                           strap screw
Right handle
strap screw


                                                                                       Right handle strap

                    4. . Turn instrument so its left side faces you
                                                                                         Left handle
                                                                                         strap screw
  Left handle
  strap screw
                                                                                         Bottom cover
                                                                                         ground screw
Left handle strap
                    5. Use the #1 Pozidriv to remove the left side handle strap screws. Then, remove
                       the handle.
                    6. Remove the cover ground screw.
                    7. Turn the instrument so its rear is facing you
            8. Use the #TX15 Torx driver to remove the 4 rear bezel screws. Then remove the
               rear bezel.
            9. Turn over the 3458A so the top sits on your workbench. To remove the bottom
               cover, pull the cover toward the rear of the instrument until it clears the front
               panel. Then, slide it forward and away from the instrument. Leave the
               instrument in this position




            10. Use the #TX10 Torx driver to remove the bottom shield screw. Then remove
            the shield. Pull the shield toward the rear of the instrument until the shield
            retainers line up with the slots on the shield. Lift the shield off



Shield
retainers
                                                                                       Bottom
                                                                                       shield screw
            Removal of the A3 Assembly
                                                                                         20 pin
                                                                                         cable PS
20 pin
cable DC
Circuitry




                      1. .Locate the 20-pin cable that connects the A3 to the Inguard Power Supply
                         assembly and unplug the cable from the A3 assembly
                      2. Locate the grey 20-pin cable that connects the A3 assembly to the DC Circuitry
                         assembly and unplug this cable

                                                                                       Fiber optic
                                                                                       cables




                                                                            .
                      3. On the A3 assembly, unplug both sets (4 cables total) of the blue and grey fiber
                         optic cables that connect the A3 to the Outguard Power Supply assembly.




                      4. Use the #TX10 Torx driver to remove the 3 screws on the shield AND 2 screws
                         on the A3 assembly. Then, remove the shield.
            5. Unplug and remove the A3 assembly from the inguard chassis

Installation of the new A3 assembly.
     1. Line up the A3 assembly with the connector on the inguard chassis. Plug the board all
        the way into the connector.
                                                                        Connector




     2. Place the A3 shield on the board. Use the #TX10 Torx driver to install the 3 screws on
        the shield. Install the 2 screws that hold down the A3 assembly.




     3. Locate the grey 20-pin cable connected to the Inguard Power Supply assembly. Line
        up the cable assembly with the socket on the A3 and fully engage the cable.
     4. Locate the grey 20-pin cable connected to the DC Circuitry assembly. Line up the
        cable assembly with the socket on the A3 and fully engage the cable.




     5. Plug in both sets of grey and blue fiber optic cables into the corresponding sockets on
        the A3 assembly.

Cover Installation Procedure:
     1. The 3458A should still be upside down on the bench with the rear of the instrument
        facing. you.



                                                                                   Shield
                                                                                   screw




     2. Line up the front slots on the bottom shield with the shield retainers. Then, push the
        shield toward the front of the instrument until the shield screw hole lines up with the
        screw hole in the chassis. Use the #TX10 Torx driver to reinstall the shield screw.




     3. Install the bottom cover by placing it over the chassis with the front of the cover just
        clearing the front panel. Then, push the cover toward the front of the instrument into
        the front panel bezel. Turn instrument right side up.
              4.      With the rear of the instrument still facing you, reinstall the rear bezel with the
Warning:      bezel screws using the #TX15 Torx driver.
For safety
purposes
and proper
operation,
it is
imperative
that the      5. Turn the instrument so the left side faces you. Use the #TX10 Torx driver to reinstall
cover            the bottom cover ground screw.
grounding
screws be
reinstalled

              6. Reinstall the left side handle strap. Use the #1 Pozidriv to reinstall the side handle
                 strap screws.




              7. Turn the instrument so the right side faces you. Reinstall the right side handle strap.
                 Use the #1 Pozidriv to reinstall side handle strap screws.
              8. Your instrument is now ready to use. Agilent Technologies suggests that after you
                 apply power that you perform an automatic calibration on the instrument. To do this,
                 use the "ACAL ALL" command.

       End of A3 Assembly Installation Procedure.
                                                                                 