Agilent 16193A
Small Side Electrode SMD Test Fixture
Operation and Service Manual
Agilent 16193A Small Side Electrode SMD Test Fixture
 Operation and Service Manual




                Agilent Part No. 16193-90040
               Printed in JAPAN August 2001
                      Seventh Edition
Notice
The information contained in this document is subject to change without notice.
This document contains proprietary information that is protected by copyright. All rights are
reserved. No part of this document may be photocopied, reproduced, or translated to another
language without the prior written consent of the Agilent Technologies.

  Agilent Technologies Japan, Ltd.
  Component Test PGU-Kobe
  1-3-2, Murotani, Nishi-ku, Kobe-shi,
  Hyogo, 651-2241 Japan





 Copyright Agilent Technologies Japan, Ltd. 1993, 1996, 1999, 2000, 2001
c
Manual Printing History
The manual printing date and part number indicate its current edition. The printing date
changes when a new edition is printed. (Minor corrections and updates that are incorporated
at reprint do not cause the date to change.) The manual part number changes when extensive
technical changes are incorporated.
   December 1993 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : First Edition (part number: 16193-90000)
   June 1996 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Second Edition
   October 1996 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Third Edition
   August 1999 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Fourth Edition (part number: 16193-90020)
   December 1999 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Fifth Edition (part number: 16193-90020)
   September 2000 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Sixth Edition (part number: 16193-90030)
   August 2001 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Seventh Edition (part number: 16193-90040)




                                                                                                                                                                        iii
Safety Summary
The following general safety precautions must be observed during all phases of operation,
service, and repair of this instrument. Failure to comply with these precautions or with specic
WARNINGS given elsewhere in this manual violates safety standards of design, manufacture,
and intended use of the instrument.
The Agilent Technologies assumes no liability for the customer's failure to comply with these
requirements.

DO NOT Operate In An Explosive Atmosphere
Do not operate the instrument in the presence of 
ammable gasses or fumes. Operation of any
electrical instrument in such an environment constitutes a safety hazard.

Keep Away From Live Circuits
Operating personnel must not remove instrument covers. Component replacement and internal
adjustments must be made by qualied maintenance personnel. Do not replace components
with the power cable connected. Under certain conditions, dangerous voltages may exist even
with the power cable removed. To avoid injuries, always disconnect power and discharge
circuits before touching them.

DO NOT Service Or Adjust Alone
Do not attempt internal service or adjustment unless another person, capable of rendering rst
aid and resuscitation, is present.

DO NOT Substitute Parts Or Modify Instrument
Because of the danger of introducing additional hazards, do not substitute parts or perform
unauthorized modications to the instrument. Return the instrument to a Agilent Technologies
Sales and Service Oce for service and repair to ensure the safety features are maintained.

Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous procedures throughout
this manual. Instructions contained in the warnings must be followed.
Warning         Dangerous voltages, capable of causing death, are present in this
                instrument. Use extreme caution when handling, testing, and adjusting
                this instrument.

The voltage levels found in this test xture when used with the intended instruments do not
warrant more than normal safety precautions for operator safety.




iv
Operating Precaution
Do not exceed the operating input power, voltage, and current level and signal type
appropriate for the instrument being used, refer to your instrument's operation manual.

Caution         Electrostatic discharge (esd) can damage the highly sensitive microcircuits in
                your instrument. ESD damage is most likely to occur as the test xtures are
                being connected or disconnected. Protect them from ESD damage by wearing a
                grounding strap that provides a high resistance path to ground. Alternatively,
                ground yourself to discharge any static charge built-up by touching the
                outer shell of any grounded instrument chassis before touching the test port
                connectors.
                Never touch the test clip contacts.
                Use a work station equipped with an anti-static work surface.




Certication
Agilent Technologies certies that this product met its published specications at the time
of shipment from the factory. Agilent Technologies further certies that its calibration
measurements are traceable to the United States National Institute of Standards and
Technology, to the extent allowed by the Institution's calibration facility, or to the calibration
facilities of other International Standards Organization members.



Warranty
This Agilent Technologies instrument product is warranted against defects in material and
workmanship for a period of one year from the date of shipment, except that in the case of
certain components listed in Instrument Specications of this manual, the warranty shall be
for the specied period. During the warranty period, Agilent Technologies will, at its option,
either repair or replace products which prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by
Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent
Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall
pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from
another country.
Agilent Technologies warrants that its software and rmware designated by Agilent
Technologies for use with an instrument will execute its programming instruction when
property installed on that instrument. Agilent Technologies does not warrant that the operation
of the instrument, or software, or rmware will be uninterrupted or error free.




                                                                                                     v
Limitation Of Warranty
The foregoing warranty shall not apply to defects resulting from improper or inadequate
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modication or
misuse, operation outside of the environmental specications for the product, or improper site
preparation or maintenance.
No other warranty is expressed or implied. Agilent Technologies specically disclaims the
implied warranties of merchantability and tness for a particular purpose.



Exclusive Remedies
The remedies provided herein are buyer's sole and exclusive remedies. Agilent Technologies
shall not be liable for any direct, indirect, special, incidental, or consequential damages,
whether based on contract, tort, or any other legal theory.



Assistance
Product maintenance agreements and other customer assistance agreements are available for
Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales and Service Oce.
Addresses are provided at the back of this manual.




vi
Safety Symbols
General denitions of safety symbols used on equipment or in manuals.
                Instruction manual symbol: the product is marked with this symbol when it is
                necessary for the user to refer to the instruction manual in order to protect
                against damage to the instrument.
                Indicates dangerous voltage (terminals fed from the interior by voltage
                exceeding 1000 volts must be so marked).
                Protective conductor terminal. For protection against electrical shock in case
                of a fault. Used with wiring terminals to indicate the terminal which must be
                connected to ground before operating equipment.
                Low-noise or noiseless, clean ground (earth) terminal. Used for a signal
                common, as well as providing protection against electrical shock in case of
                fault. A terminal marked with this symbol must be connected to ground in the
                manner described in the installation (Operation) manual, and before operating
                the equipment.
                Frame or chassis terminal. A connection to the frame (chassis) of the
                equipment which normally includes all exposed metal structures.
                Alternating current (power line).
                Direct current (power line).
                Alternating or direct current (power line).
                Warning denotes a hazard.     It calls attention to a procedure, practice, condition
                or the like, which, if not correctly performed or adhered to, could result in
                injury or death to personnel.
                Caution sign denotes a hazard. It calls attention to a procedure, practice,
                condition or the like, which, if not correctly performed or adhered to, could
                result damage to or destruction of part or all of the product.
                Note denotes important information. It calls attention to a procedure, practice,
                condition or the like, which is essential to highlight.




                                                                                                  vii
APC-7
 is a U.S. registered trademark of the Bunker Ramo Corporation.
     R

viii
Contents
 1. General Information
    Introduction . . . . . . . . . . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-1
      Manual Summary . . . . . . . . . . .         .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-1
    Product Description . . . . . . . . . . .      .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-1
    Applicable Instruments . . . . . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-1
    Options Available . . . . . . . . . . . .      .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-2
    Accessories Supplied . . . . . . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-2
    Specications . . . . . . . . . . . . . .      .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-2
    Supplemental Performance Characteristics       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-3
    Residual Inductance of the Shorting Bar . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-4
 2. Preparation for Use
    Introduction . . . . . . . . . . . . . . . . . . . . . . .                     .   .   .   .   .   .   .   .   .   .     2-1
    Initial Inspection . . . . . . . . . . . . . . . . . . . . .                   .   .   .   .   .   .   .   .   .   .     2-1
    Adjust the Conguration to SMD Size . . . . . . . . . . .                      .   .   .   .   .   .   .   .   .   .     2-3
    Connecting the Test Fixture for Use . . . . . . . . . . . .                    .   .   .   .   .   .   .   .   .   .     2-4
    Repackaging the Test Fixture For Shipment . . . . . . . .                      .   .   .   .   .   .   .   .   .   .     2-5
    Measurement Error Correcting Function for Each Instrument                      .   .   .   .   .   .   .   .   .   .     2-6
 3. Operation
    Introduction . . . . . . . . . . . . . . . . . . . . . .                   .   .   .   .   .   .   .   .   .   .   .     3-1
    Operating Flow . . . . . . . . . . . . . . . . . . . .                     .   .   .   .   .   .   .   .   .   .   .     3-1
    Overview . . . . . . . . . . . . . . . . . . . . . . .                     .   .   .   .   .   .   .   .   .   .   .     3-2
    How to set the SHORT condition, and the OPEN condition                     .   .   .   .   .   .   .   .   .   .   .     3-3
      Setting the SHORT condition . . . . . . . . . . . . .                    .   .   .   .   .   .   .   .   .   .   .     3-3
      Setting the OPEN condition . . . . . . . . . . . . .                     .   .   .   .   .   .   .   .   .   .   .     3-5
    How to connect the DUT . . . . . . . . . . . . . . . .                     .   .   .   .   .   .   .   .   .   .   .     3-5
 4. Service
    Introduction . . . . . . . . . . . . . .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-1
    General Information . . . . . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-1
    Assembly Replacement . . . . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-3
      Disassembly and Assembly Procedure .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-3
        Pressure Arm Assembly Removal: . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-3
        Pressure Arm Assembly Disassemble:     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-3
      Replaceable Parts . . . . . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-3
    Troubleshooting . . . . . . . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-9
      Open Impedance Check . . . . . . .       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-9
      Short Impedance Check . . . . . . .      .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   4-10
      Short Impedance Repeatability Check .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   4-10




                                                                                                                       Contents-1
Figures
    1-1.   Applicable DUT Size . . . . . . . . . . . . . . . . . . . .     .   .   .   .   .   .   .   .    1-3
    1-2.   Simulation Setup . . . . . . . . . . . . . . . . . . . . .      .   .   .   .   .   .   .   .    1-5
    2-1.   16193A Product Overview . . . . . . . . . . . . . . . . .       .   .   .   .   .   .   .   .    2-2
    2-2.   Electrodes conguration and DUT Size . . . . . . . . . . .      .   .   .   .   .   .   .   .    2-3
    2-3.   Connecting the Test Fixture with APC-7
 Connector (4291B)
                                                  R                        .   .   .   .   .   .   .   .    2-4
    3-1.   16193A Test Fixture Overview . . . . . . . . . . . . . . .      .   .   .   .   .   .   .   .    3-2
    4-1.   Replaceable Parts for Pressure Arm Assembly . . . . . . . .     .   .   .   .   .   .   .   .    4-5
    4-2.   Replaceable Parts for Test Stage Assembly . . . . . . . . . .   .   .   .   .   .   .   .   .    4-7
    4-3.   Obsolete parts manufactured earlier than 1996 . . . . . . .     .   .   .   .   .   .   .   .    4-8
    4-4.   Shorting Device Position . . . . . . . . . . . . . . . . . .    .   .   .   .   .   .   .   .    4-9




Tables
    1-1.   Furnished Accessories . . . . . . . . . . . . . . . . . . .     .   .   .   .   .   .   .   .    1-2
    1-2.   Supplied Shorting Device Residual Inductance (Typical) . . .    .   .   .   .   .   .   .   .    1-4
    1-3.   Option 010 Shorting Device Residual Inductance (Typical) . .    .   .   .   .   .   .   .   .    1-4
    2-1.   16193A Contents . . . . . . . . . . . . . . . . . . . . .       .   .   .   .   .   .   .   .    2-2
    2-2.   Measurement Error Correcting Functions for Each Instrument      .   .   .   .   .   .   .   .    2-6
    4-1.   16193A Product Overview . . . . . . . . . . . . . . . . .       .   .   .   .   .   .   .   .    4-2
    4-2.   Replaceable Parts for Main Assembly . . . . . . . . . . . .     .   .   .   .   .   .   .   .    4-4
    4-3.   Replaceable Parts for Pressure Arm Assembly . . . . . . . .     .   .   .   .   .   .   .   .    4-6
    4-4.   Replaceable Parts for Test Stage Assembly . . . . . . . . . .   .   .   .   .   .   .   .   .    4-8
    4-5.   Open Impedance Value Guideline . . . . . . . . . . . . . .      .   .   .   .   .   .   .   .   4-10
    4-6.   Short Impedance Value Guideline . . . . . . . . . . . . . .     .   .   .   .   .   .   .   .   4-10
    4-7.   Short Impedance Repeatability Guideline . . . . . . . . . .     .   .   .   .   .   .   .   .   4-10




Contents-2
                                                                                                                 1
General Information
Introduction
The purpose of this manual is to enable you to use your 16193A Small Side Electrode SMD Test
Fixture eciently and condently.

Manual Summary
This manual contains the following:
  The specications of the 16193A (see this chapter).
  Installing the 16193A (see Chapter 2).
  Operating the 16193A (see Chapter 3).
  Ordering replaceable parts for the 16193A (see Chapter 4).



Product Description
The 16193A is used to measure a small, side electrodes surface mount devise (SMD) with high
repeatability.



Applicable Instruments
The 16193A has been designed to operate specically with the following LCR meters and
impedance analyzers:
  You can directly connect the 16193A to:                                  4191A, 4291B, E4991A
  You can use the 16193A with 43961A:                                      4395A (option 010),
                                                                           4396A/B (option 010)
  You can use the 16193A with 41951A:                                      4195A
  You can use the 16193A with RF adapter 42942A:                           4294A
  You can use the 16193A with RF adapter 16085B:                           4284A1
  You can use the 16193A with RF adapter 16099A:                           4194A with 41941A2
  1 The 4284A is a typical instrument. Other 4-terminal pair instruments that can be used with the 16085B are
    also applicable.
  2 The 4194A with 41941A is a typical instrument. Other instruments that can be used with the 16099A are also
    applicable.




                                                                                           General Information   1-1
Options Available
The following option is supplied for the 16193A:
Option 001      Delete the magnifying lens and tweezers
Option 010      Add industry standard size short bar set



Accessories Supplied
The following accessories are supplied with the 16193A:
                                                   Table 1-1. Furnished Accessories
                                             Description             Part Number                                                       Quantity
                        Operation and Service Manual                                         P/N 16193-90040                                   1
                        Shorting device (4 types)                                            P/N 16191-29001                                   1
                                                                                             P/N 16191-29002                                   1
                                                                                             P/N 16191-29003                                   1
                                                                                             P/N 16191-29004                                   1
                        Case for shorting devices                                            P/N 1540-0692                                     1
                        Option 010*1                                                         P/N 16191-29005                                   1
                                                                                             P/N 16191-29006                                   1
                                                                                             P/N 16191-29007                                   1
                                                                                             P/N 16191-29008                                   1
                        Magnifying lens                                                      P/N 16193-60002                                   1
                        Tweezers                                                             P/N 8710-2081                                     1
                        Wrench                                                               P/N 8710-1181                                     1

*1 Option 010 sizes are the same as industry standard (EIA/EIAJ) SMD sizes. This short bar
set has the following SMD sizes included : 1005(mm)/0402(inch), 1608(mm)/0603(inch),
2012(mm)/0805(inch), 3216(mm)/1206(inch). Order option 010 if the SMD that is to be measured
has the same size as the EIA/EIAJ sizes.



Specications
This section lists the complete 16193A specications. These specications are the performance
standards and limits against which the 16193A is tested. When shipped from the factory, the
16193A meets the following specications:
   Maximum DC Bias Voltage : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 6 40V
   Operating Temperature : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 055 to +85 C
   Operating Humidity (@wet bulb temperature <40 C)
       : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 15 % to 95 % RH
   Non-operating Temperature : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 055 to +85 C
   Non-operating Humidity (@wet bulb temperature <65 C)
      : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Up to 90 % RH
   Weight : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 500 g
   Dimension : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 150(W) 2 70 (H) 2 80(D) [mm]


1-2     General Information
Supplemental Performance Characteristics
This section lists supplemental performance characteristics. Supplemental performance
characteristics are not specications, but are typical characteristics included as additional
information for the operator. Supplemental performance characteristics are not guaranteed.
   Frequency Range : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : DC to 2 GHz
   Applicable DUT size (see Figure 1-1) : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 0.5 mm to 3.2 mm
   Supplied shorting device size
       P/N 16191-29001                   : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 1 2 1 2 2.4 [mm]
       P/N 16191-29002                   : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 1.6 2 2.4 2 2 [mm]
       P/N 16191-29003                   : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 2.4 2 2.4 2 3.2 [mm]
       P/N 16191-29004                   : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 2.4 2 2.4 2 4.5 [mm]
   Option 010 shorting device size
     P/N 16191-29005 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 1.0(L) 2 0.5(W) 2 0.5(H) [mm]
     P/N 16191-29006 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 1.6(L) 2 0.8(W) 2 0.8(H) [mm]
     P/N 16191-29007 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 2.0(L) 2 1.2(W) 2 0.8(H) [mm]
     P/N 16191-29008 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 3.2(L) 2 1.6(W) 2 0.8(H) [mm]
   Electrical length (when the length between electrodes is 2 mm) : : : : : : : : : : : : : : : : : : : : : 14.0 mm
   Additional Error3 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 1.5 2 f2 [%]
   Repeatability3
       (for inductive component) : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 30 + 250 2 f [m
]
                                                                                                                                           (impedance of 30 m
, 40 pH)
        : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 2 + 30 2 f [S]
                                                                                                                                                  (admittance of 2 S, 5 fF)
      3 f: frequency (GHz)




                                                         Figure 1-1. Applicable DUT Size




                                                                                                                                              General Information                1-3
Residual Inductance of the Shorting Bar
The usual method to compensate the test xture's residual inductance is to let SHORT = 0H.
In this method, the measurement result is the relative value of the measured impedance to the
shorting bar's impedance. The short bar's residual inductance as a result of its size and shape is
not estimated.
On the other hand, there is a denition method to let SHORT = x H. In this method, the
measurement result is the absolute value of the device's impedance. The short bar's residual
inductance as a result of its size and shape is estimated under specic conditions and is used as
a reference value. This method, is useful for devices with values which are close to the short
conditions of the measurement system.
The reference inductance values presented Table 1-2 and Table 1-3 were simulated as the
relative dierence to a disk-type 0 
 termination on either the 7 mm or the 3.5 mm connector.
The measurement of these short bars under other conditions than shown below cannot
reproduce the reference inductance values.
             Table 1-2. Supplied Shorting Device Residual Inductance (Typical)
         Shorting Bar l [mm] d [mm] h [mm] Oset [mm] Connector Inductance
                                                                         (Typical)
       P/N 16191-29001        1      2.4        1          0.75     3.5 mm           0.2 nH
       P/N 16191-29002      1.6      2.4        2          0.45     3.5 mm           0.2 nH
       P/N 16191-29002        2      2.4      1.6          0.25     3.5 mm           0.2 nH
       P/N 16191-29002      2.4        2      1.6           1.3       7 mm           0.7 nH
       P/N 16191-29003      3.2      2.4      2.4           0.9       7 mm           0.6 nH
       P/N 16191-29004      4.5      2.4      2.4             0       7 mm           0.6 nH

            Table 1-3. Option 010 Shorting Device Residual Inductance (Typical)
         Shorting Bar l [mm] d [mm] h [mm] Oset [mm] Connector Inductance
                                                                         (Typical)
       P/N 16191-29005      1.0      0.5      0.5          0.75     3.5 mm           0.5 nH
       P/N 16191-29006      1.6      0.8      0.8          0.45     3.5 mm           0.4 nH
       P/N 16191-29007      2.0      1.2      0.8           1.5       7 mm           0.9 nH
       P/N 16191-29008      3.2      1.6      0.8           0.9       7 mm           0.8 nH




1-4   General Information
Figure 1-2. Simulation Setup




                               General Information   1-5
                                                                                                  2
Preparation for Use
Introduction
This chapter explains how to install the 16193A. The topics include the following:
  Initial inspection.
  Connecting the test xture for use.
  Repackaging the test xture for shipment.
  Measurement error correcting function for each instrument.



Initial Inspection
The small side electrode SMD test xture has been carefully inspected before being shipped
from the factory. It should be in perfect physical condition, no scratches, dents or the like. It
should also be in perfect electrical condition. Verify this by carefully performing an incoming
inspection to check the small side electrode SMD test xture set for signs of physical damage
and missing contents. If any discrepancy is found, notify the carrier and Agilent Technologies.
Your Agilent Technologies sales oce will arrange for repair and replacement without waiting
for the claim to be settled.
  Inspect the shipping container for damage. Keep the shipping materials until the inspection is
  completed.
  Verify that the shipping container contains everything shown in Figure 2-1 and listed in
  Table 2-1.
  Inspect the exterior of the 16193A for any signs of damage.




                                                                            Preparation for Use   2-1
                                                          .




                                  Figure 2-1. 16193A Product Overview
                                     Table 2-1. 16193A Contents
                            Description                  Agilent Part                       Quantity
                                                            Number
          
 Test xture
          1                                                       16193A                            1
          
 Operation and Service Manual1
          2                                                       P/N 16193-90040                   1
          
 Shorting device2 (12122.4[mm])
          3                                                       P/N 16191-29001                   1
                              (1.622.422[mm])                     P/N 16191-29002                   1
                              (2.422.423.2[mm])                   P/N 16191-29003                   1
                              (2.422.424.5[mm])                   P/N 16191-29004                   1
          
 Case for shorting device
          4                                                       P/N 1540-0692                     1
          
 Magnifying lens34
          5                                                       P/N 16193-60002                   1
          
 Tweezers4
          6                                                       P/N 8710-2081                     1
          
 Wrench
          7                                                       P/N 8710-1181                     1

          1 Operation and Service Manual is not shown in Figure 2-1.
          2 Depending on your order conguration.
          3 The magnifying lens is packed separately from the 16193A body. Connect it as shown in
            Figure 2-1.
          4 Not contained if you ordered the 16193A option 001.



2-2   Preparation for Use
Adjust the Conguration to SMD Size
The 16193A's electrodes length are 0.2 mm, 0.5 mm, 1 mm, and 2 mm, change the electrodes
conguration to match the SMD as follows:




                   Figure 2-2. Electrodes conguration and DUT Size
                                                                       Preparation for Use   2-3
Connecting the Test Fixture for Use
Note      Some instruments require an adapter to be connected to the 16193A (see
                    \Applicable Instruments" in Chapter 1).


1. Calibrate your analyzer at the APC-7
 connector plane before connecting test xture. See
                                          R
   Table 2-2 for the applicable error correcting function. After calibration, disconnect any
   standards from the APC-7
 connector.
                              R

2. Place the xture on the APC-7
 connector.
                                   R

3. Tighten (turn counterclockwise) the coupling nut of the APC-7
 connector.
                                                                   R




           Figure 2-3. Connecting the Test Fixture with APC-7
 Connector (4291B)
                                                             R




2-4   Preparation for Use
Repackaging the Test Fixture For Shipment
If shipment to an Agilent Technologies service center is required, each test xture should be
repackaged using the original factory packaging materials.
If this material is not available, comparable packaging materials may be used. Wrap the small
side electrode SMD test xture in heavy paper and pack in anti-static plastic packing material.
Use sucient shock absorbing material on all sides of the 16193A to provide a thick, rm
cushion and to prevent movement. Seal the shipping container securely and mark it FRAGILE.




                                                                            Preparation for Use   2-5
    Measurement Error Correcting Function for Each Instrument
    Each LCR meter or analyzer has measurement error correcting functions. Table 2-2 shows the
    functions of the instrument that can be used for calibration, setting the electrical length, and
    Open/Short/Load compensation.
             Table 2-2. Measurement Error Correcting Functions for Each Instrument
  Instrument          Calibration        Electrical length         OPEN             SHORT            LOAD
                                                                Compensation     Compensation     Compensation
4291B                Calibration       Selecting Test Fixture       OPEN             SHORT           LOAD
                        OPEN              Port Extension         Compensation     Compensation    Compensation1
                       SHORT
                        LOAD
                 LOW-LOSS Capacitor1
4191A                Calibration         Electrical Length       No capability    No capability    No capability
                          0S
                          0

                         50 

4194A with          No capability          No capability        OPEN Correction SHORT Correction LOAD Correction1
16085B
4195A with          Calibration OPEN       Port Extension         0 S OFFSET       0 
 OFFSET      No capability
41951A                   SHORT            (Port 1 or Port 2)
                          LOAD
4194A with             Calibration         No capability         Compensation     Compensation     No capability
41941A/B,                   0S                                    ZERO OPEN       ZERO SHORT
16099A                      0

                           50 

4194A with             Calibration         No capability         Compensation     Compensation     No capability
16085B                      0S                                    ZERO OPEN       ZERO SHORT
                            0

                           50 


1 According to demand for precise measurement




    2-6   Preparation for Use
                                                                                                                 3
Operation
Introduction
This chapter describes how to use the test xture and the measurement error correcting
techniques for the test xture.

Operating Flow
Before performing a measurement, you have to compensate for the residual error of the test
xture by using the measurement error correcting functions. Perform the following steps when
measuring the DUT:
   The 16193A has inherent stray capacitance, residual inductance, and residual resistance that aect the
   measurement. To compensate for these residuals and thus minimize their eect on measurement accuracy, the
   measurement instrument's OPEN and SHORT compensation capabilities must be used.

                                              ( 7 indicates procedures in the following pages.)

                                  Described in Chapter 2.



                                  The 16193A's electrical length is 14.0 mm (typical value).
                                  To set the electrical length, see your analyzer's Operation Manual.


                                  To set the xture and devices for xture compensation, see:
                                  7   \How to set the SHORT condition, and the OPEN condition"

                                  These procedures show how to set the test xture for the
                                  OPEN/SHORT/LOAD compensation procedure.
                                  For how to measure OPEN/SHORT/LOAD, see the Operation Manual
                                  supplied with your analyzer. (You can omit LOAD compensation.)


                                  To connect the DUT, see:
                                  7 \How to connect the DUT"




 3 The compensating functions are dierent for each instrument. See Table 2-2 and perform the listed functions.
                                                                                                     Operation   3-1
Overview




                                 Figure 3-1. 16193A Test Fixture Overview
Arm lifter                         Tilts the pressure arm up.
Positioner                         Slides to set the pressure arm atop the DUT.
Locking knob                       Fixes the positioner.
(Pressure) arm                     Fixes the DUT.
(Device) holder knob               Fixes the holder.
(Device) holder                    Holds the DUT.
(Size selection) frame             Fixes the electrode.
Electrodes                         These are connected to the device.
Arm latch                          Latches the pressure arm tilted up. (Push the arm latch, tilt the arm
                                   up by pushing the arm lifter. To set the latch to o, push the arm
                                   lifter.)
Pressure adjustment                Adjusts the pressure load on the device.
Shorting device                    (Not shown in Figure 3-1) Used for short compensation.
      Four types of shorting devices are supplied with the 16193A. Use one that has dimensions similar to the DUT.
      When making a shorting device of same dimensions as your DUT, very low residual impedance and a high
      conductivity metal (that is not easily corroded) must be used to construct the shorting device. (It must also be
      clean.)

Magnifying lens                      (Not shown in Figure 3-1) Used to set a small device within the
                                     electrodes.
Warning              DO NOT look at the sun through the magnifying glass.
                     DO NOT operate or leave the magnifying glass under direct sunlight.

3-2    Operation
How to set the SHORT condition, and the OPEN condition
Setting the SHORT condition




   The names of each part are listed above.           1. 
 Push the arm lifter. 
 Push the arm latch
                                                         1                       2
                                                      button. 
 Release the arm lifter while pushing the
                                                              3
                                                      arm latch button so that the pressure arm is held.




   2. 
 Loosen the locking knob. 
 Rotate the
      1                             2                 3. Position the size selection frame depending on the
   positioner until the nose of the pressure arm is   size of the DUT. (This procedure shows the case for D
   located outside the stage.                         as an example.)




                                                                                                 Operation    3-3
      4. Place the shorting device on the electrode.       5. 
 Loosen the knob. 
 Position the DUT holder so
                                                              1                    2
                                                           that the short bar contacts both electrodes. 
                                                                                                        3
                                                           Tighten the knob to x the DUT holder.




      6. 
 Reposition the positioner so that the nose of the 7. 
 Push the arm lifter so that the arm latch is
         1                                                       1
      pressure arm is just above the shorting device. 
                                                      2      released. 
 Let down the pressure arm to press down
                                                                        2
      Tighten the locking knob.                              the shorting device.




3-4   Operation
Setting the OPEN condition




   1. 
 Push the arm lifter. 
 Push the arm latch
      1                       2                         2. Remove the shorting device.
   button. 
 Release the arm lifter while pushing the
           3
   arm latch button so that the pressure arm is held.




How to connect the DUT




   1. 
 Loosen the locking knob. 
 Rotate the
      1                             2                   2. Place the DUT on the electrode.
   positioner until the nose of the pressure arm is
   located outside the frame.




                                                                                             Operation   3-5
      3. 
 Loosen the knob. 
 Position the DUT holder so 
 Reposition the positioner so that the nose of the
         1                   2                              1
      that the DUT contacts both electrodes. 
 Tighten the pressure arm is just above the DUT. 
 Tighten the
                                             3                                                 2
      knob.                                                locking knob.




      3. 
 Push the arm lifter so that the arm latch is
          1
      released. 
 Let down the pressure arm to press down
                2
      the DUT.




3-6   Operation
                                                                                              4
Service
Introduction
This chapter covers the following subjects:
  General Information
  Assembly Replacement
     Disassembly Procedure for the Main Assembly
     Replaceable Parts
  Troubleshooting
Warning        These servicing instructions are for use by qualied personnel only. Do
               NOT perform any servicing (other than that contained in the operating
               section) unless you are qualied to do so.

Caution        When you repair the 16193A, put on lint-free groves to avoid contaminating
               inner parts of the 16193A. Agilent Technologies supplies lint-free groves
               (Agilent part number 9300-0163).




General Information
Table 4-1 shows all items included with the 16193A Small Side Electrode SMD Test Fixture.




                                                                                    Service   4-1
                                Table 4-1. 16193A Product Overview




    Reference           Agilent Part
    Designator           Number           Qty.                  Description
        1              (not assigned)      1     Main Assembly
        2               16193-60002        1     Magnifying Glass
        3                8710-1181         1     Wrench
        4                8710-2081         1     Tweezers
        5               16191-29001        1     Shorting Device (12122.4 mm)
        6               16191-29002        1     Shorting Device (1.622.422 mm)
        7               16191-29003        1     Shorting Device (2.422.423.2 mm)
        8               16191-29004        1     Shorting Device (2.422.424.5 mm)
        9                1540-0692         1     Case for shorting devices
        -               16191-29021        1     Styrofoam Holder1
        -               16193-90040        1     Operation and Service Manual1

1     Not shown in the gure.




4-2   Service
Assembly Replacement
This section includes Disassembly and Assembly Procedures and Replacement Parts for the
Main Assembly.

Disassembly and Assembly Procedure
This section contains the information required to disassemble and assemble the Main Assembly.
Pressure Arm Assembly Removal:

1. Rotate the Locking Knob counter-clockwise and remove it (see Table 4-2).
2. Remove the Pressure Arm Assembly (including Spring, Bushing, and Washer) from the Test
   Stage Assembly.
Pressure Arm Assembly Disassemble:

1. Remove two screws (marked with \8" and \12" in Figure 4-1).
2. Remove bottom cover (marked with \1" in Figure 4-1).
3. Remove other screws and other parts.

Replaceable Parts
Table 4-2 through Table 4-4 show and list the replaceable parts for the Main Assembly. The
parts listed can be ordered from your nearest Agilent Technologies Oce. Ordering information
must include the Agilent part number and the quantity required.




                                                                                   Service   4-3
                   Table 4-2. Replaceable Parts for Main Assembly




 Reference      Agilent Part
 Designator      Number           Qty.                    Description
        1        16193-23004        1     Locking Knob
        2         1460-2368         1     Spring
        3        16193-25003        1     Bushing
        4       (not assigned)      1     Pressure Arm Assembly
        5        16193-87101        1     Label
        6        16193-25004        1     Washer
        7       (not assigned)      1     Test Stage Assembly


4-4   Service
Figure 4-1. Replaceable Parts for Pressure Arm Assembly


                                                          Service   4-5
                Table 4-3. Replaceable Parts for Pressure Arm Assembly
 Reference       Agilent Part
 Designator        Number           Qty.                    Description
        1        16193-04002         1      Bottom Cover
        2         1460-2367          1      Spring
        3        16193-23005         1      Shaft
        4         2950-0054          1      Nut
        5         2190-0054          1      Washer
      6, 16       0515-0914          2      Screw Flat Head M3L6
        7        16193-25006         1      Arm Latch
      8, 12       0515-0914          2      Screw Flat Head M3L6
        9        16193-04001         1      Top Cover
       10        16193-24008         1      Screw
       11         0515-1052          1      Screw Socket Head M3L12
       13        16193-25007         1      Screw
       14         3050-1139          1      Flat Washer, Te
on
       15         1460-2316          1      Spring
       17        16193-25008         1      Spacer
       18        16193-01201         1      Angle
       19        16193-25005         2      Bushing
       20         0515-0976          1      Screw Pan Head M2L6
       21         3050-1066          1      Flat Washer M2
       22        16193-24007         2      Nut
       23         3030-0285          1      Set Screw 6-32
       24        16193-23001         1      Rod
       25        16193-23002         1      Holder




4-6   Service
Figure 4-2. Replaceable Parts for Test Stage Assembly




                                                        Service   4-7
                  Table 4-4. Replaceable Parts for Test Stage Assembly
 Reference        Agilent Part
 Designator        Number            Qty.                     Description
        1          16193-24011           1       Device Holder Knob
        2          16193-25009           1       Device Holder
        3           0515-1050            2       Screw Socket Head M3L8
        4          16193-24004           1       Size Selection Frame
        5          16193-00613           1       Plate
        6          16193-23003           1       Flange
        7           0515-0914            6       Screw Flat Head M3L6
        8          16193-00201           1       Table
        9          16193-00604           1       Ground Plate
       10          16193-60011           1       Contact Assembly
       11           0515-0952            3       Screw Flat Head M2L4
       12           0515-0976            2       Screw Pan Head M2L6
       13          16193-24005           2       Spacer
       14           1250-0907            1       Contact-RF Connector

Caution         If your 16193A test xture was manufactured earlier than 1996 and you need
                to replace either the plate (Figure 4-3: 9) or the contact assembly (Figure 4-3:
                10), it will be necessary to buy replacements for both parts (Figure 4-2: 9
                and 10) since they were modied at the same time. Neither of the new parts
                (Figure 4-2: 9 and 10) can be tted to the other obsolete parts (Figure 4-3: 9
                and 10).




                        Figure 4-3. Obsolete parts manufactured earlier than 1996




4-8   Service
Troubleshooting
This section includes the functional check procedure used to troubleshoot the 16193A Small
Side Electrode SMD Test Fixture.

Open Impedance Check
1. Set the Size Selection Frame to the position for small size DUT.
2. Adjust the Device Holder for connecting the 12122.4 mm Shorting Device (PN
   16191-29001).
   (Refer to Figure 4-4.)




                             Figure 4-4. Shorting Device Position
3. Place the xture on the calibrated APC-7
 terminal of the 4291B RF Impedance/Material
                                            R
   Analyzer.
4. Set the 4291B RF Impedance/Material Analyzer as follows:
   Measurement Parameter              Cp (parallel capacitance)
   Start Frequency                    100 MHz
   Stop Frequency                     1 GHz
   OSC Level                          0.12 V
   Number of Point                    2
   Point Averaging Factor             16
   Point Averaging                    ON
5. Perform a single sweep measurement.

                                                                                   Service   4-9
6. Read Cp value at 100 MHz and 1 GHz.
The guideline for the open impedance value is as follows:
                         Table 4-5. Open Impedance Value Guideline
                 Parameter       Frequency       Guideline (absolute value)
                    Cp            100 MHz             750 fF 6 400 fF
                    Cp              1 GHz             750 fF 6 400 fF


Short Impedance Check
1. After the Open Impedance Check is completed, put the 12122.4 mm Shorting Device (PN
   16191-29001) between the center electrode and the ground electrode (refer to Figure 4-4).
2. Set the 4291B RF Impedance/Material Analyzer as follows:
   Measurement Parameter             Ls (series inductance)
   Start Frequency                   100 MHz
   Stop Frequency                    1 GHz
   OSC Level                         0.12 V
   Number of Point                   2
   Point Averaging Factor            16
   Point Averaging                   ON
3. Perform a single sweep measurement.
4. Read Ls value at 100 MHz and 1 GHz.
The guideline for the short impedance value is as follows:
                         Table 4-6. Short Impedance Value Guideline
                 Parameter       Frequency        Guideline (absolute value)
                     Ls           100 MHz                2 nH 6 1 nH
                     Ls             1 GHz                2 nH 6 1 nH


Short Impedance Repeatability Check
1. After the Short Impedance Check is completed, remove the Shorting Device and put it back
   on the same place again (refer to Figure 4-4).
2. Perform a single sweep measurement with the same measurement conditions.
3. Read Ls value at 100 MHz and 1 GHz and check the value is within the Short Impedance
   Value Guideline described in Table 4-6.
4. Calculate the dierence between the rst Ls measurement value and second Ls
   measurement value at each frequency.
The guideline for the short impedance repeatability is as follows:
                      Table 4-7. Short Impedance Repeatability Guideline
                 Parameter        Frequency         Guideline (dierence)
                     Ls            100 MHz                 6 45 pH
                     Ls             1 GHz                  6 20 pH

4-10   Service
              REGIONAL SALES AND SUPPORT OFFICES
For more information about Agilent Technologies test and measurement products, applications, services, and
for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one
of the following centers and ask for a test and measurement sales representative.                          11/29/99


United States:                           (fax) (61 3) 9272 0749
Agilent Technologies                     (tel) 0 800 738 378 (New Zealand)
Test and Measurement Call Center         (fax) (64 4) 802 6881
P.O.Box 4026
Englewood, CO 80155-4026                 Asia Pacific:
(tel) 1 800 452 4844                     Agilent Technologies
                                         24/F, Cityplaza One, 1111 King's Road,
Canada:                                  Taikoo Shing, Hong Kong
Agilent Technologies Canada Inc.         (tel) (852)-3197-7777
5150 Spectrum Way                        (fax) (852)-2506-9284
Mississauga, Ontario
L4W 5G1
(tel) 1 877 894 4414

Europe:
Agilent Technologies
Test & Measurement
European Marketing Organization
P.O.Box 999
1180 AZ Amstelveen
The Netherlands
(tel) (31 20) 547 9999

Japan:
Agilent Technologies Japan Ltd.
Call Center
9-1, Takakura-Cho, Hachioji-Shi,
Tokyo 192-8510, Japan
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840

Latin America:
Agilent Technologies
Latin American Region Headquarters
5200 Blue Lagoon Drive, Suite #950
Miami, Florida 33126
U.S.A.
(tel) (305) 267 4245
(fax) (305) 267 4286

Australia/New Zealand:
Agilent Technologies Australia Pty Ltd
347 Burwood Highway
Forest Hill, Victoria 3131
(tel) 1-800 629 485 (Australia)
16193-90040



◦ Jabse Service Manual Search 2024 ◦ Jabse PravopisonTap.bg ◦ Other service manual resources online : FixyaeServiceinfo