                   Agilent 4263B LCR Meter
                  Service Manual
                           SERIAL NUMBERS
This manual applies directly to instruments with serial number prex JP1KD.
       For additional important information about serial numbers, read
     \Instruments Covered by This Manual" in Chapter 1 of this manual.




                      Agilent Part No. 04263-90033
                      Printed in JAPAN June 2000
Notice
The information contained in this document is subject to change without notice.
This document contains proprietary information that is protected by copyright. All rights are
reserved. No part of this document may be photocopied, reproduced, or translated to another
language without the prior written consent of the Agilent Technologies Company.
  Agilent Technologies Japan, Ltd.
  Component Test PGU-Kobe
  1-3-2, Murotani, Nishi-ku, Kobe-shi,
  Hyogo, 651-2241 Japan





 Copyright 1996, 2000 Agilent Technologies Japan, Ltd.
c
Manual Printing History
 August 1996 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : First Edition (part number: 04263-90033)
 June 2000 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Second Edition (part number: 04263-90033)




                                                                                                                                      iii
Safety Summary
The following general safety precautions must be observed during all phases of operation,
service, and repair of this instrument. Failure to comply with these precautions or with specic
WARNINGS elsewhere in this manual may impair the protection provided by the equipment.
In addition it violates safety standards of design, manufacture, and intended use of the
instrument.
The Agilent Technologies Company assumes no liability for the customer's failure to comply
with these requirements.
Note            4263B complies with INSTALLATION CATEGORY II and POLLUTION DEGREE 2
                in IEC1010-1. 4263B is INDOOR USE product.



Ground The Instrument
To avoid electric shock hazard, the instrument chassis and cabinet must be connected to a
safety earth ground by the supplied power cable with earth blade.

DO NOT Operate In An Explosive Atmosphere
Do not operate the instrument in the presence of 
ammable gasses or fumes. Operation of any
electrical instrument in such an environment constitutes a denite safety hazard.

Keep Away From Live Circuits
Operating personnel must not remove instrument covers. Component replacement and internal
adjustments must be made by qualied maintenance personnel. Do not replace components
with the power cable connected. Under certain conditions, dangerous voltages may exist even
with the power cable removed. To avoid injuries, always disconnect power and discharge
circuits before touching them.

DO NOT Service Or Adjust Alone
Do not attempt internal service or adjustment unless another person, capable of rendering rst
aid and resuscitation, is present.

DO NOT Substitute Parts Or Modify Instrument
Because of the danger of introducing additional hazards, do not install substitute parts or
perform unauthorized modications to the instrument. Return the instrument to a Agilent
Technologies Sales and Service Oce for service and repair to ensure that safety features are
maintained.

Dangerous Procedure Warnings
Warnings , such as the example below, precede potentially dangerous procedures throughout
this manual. Instructions contained in the warnings must be followed.
Warning         Dangerous voltages, capable of causing death, are present in this
                instrument. Use extreme caution when handling, testing, and adjusting
                this instrument.


iv
Certication
Agilent Technologies Company certies that this product met its published specications
at the time of shipment from the factory. Agilent Technologies further certies that its
calibration measurements are traceable to the United States National Institute of Standards and
Technology, to the extent allowed by the Institution's calibration facility, or to the calibration
facilities of other International Standards Organization members.



Warranty
This Agilent Technologies instrument product is warranted against defects in material and
workmanship for a period of one year from the date of shipment, except that in the case of
certain components listed in General Information of this manual, the warranty shall be for the
specied period. During the warranty period, Agilent Technologies Company will, at its option,
either repair or replace products that prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by
Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent
Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall
pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from
another country.
Agilent Technologies warrants that its software and rmware designated by Agilent
Technologies for use with an instrument will execute its programming instruction when
property installed on that instrument. Agilent Technologies does not warrant that the operation
of the instrument, or software, or rmware will be uninterrupted or error free.



Limitation Of Warranty
The foregoing warranty shall not apply to defects resulting from improper or inadequate
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modication or
misuse, operation outside the environmental specications for the product, or improper site
preparation or maintenance.
No other warranty is expressed or implied. Agilent Technologies specically disclaims the
implied warranties of merchantability and tness for a particular purpose.




                                                                                                 v
Exclusive Remedies
The remedies provided herein are buyer's sole and exclusive remedies. Agilent Technologies
shall not be liable for any direct, indirect, special, incidental, or consequential damages,
whether based on contract, tort, or any other legal theory.



Assistance
Product maintenance agreements and other customer assistance agreements are available for
Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales and Service Oce.
Addresses are provided at the back of this manual.




vi
Safety Symbols
General denitions of safety symbols used on equipment or in manuals are listed below.
                Instruction manual symbol: the product is marked with this symbol when it is
                necessary for the user to refer to the instruction manual.
                Alternating current.
               Direct current.
               On (Supply).
               O (Supply).
               Frame or chassis terminal



               This Warning sign denotes a hazard. It calls attention to a procedure, practice,
               condition or the like, which, if not correctly performed or adhered to, could
               result in injury or death to personnel.
               This Caution sign denotes a hazard. It calls attention to a procedure, practice,
               condition or the like, which, if not correctly performed or adhered to, could
               result in damage to or destruction of part or all of the product.
               Note denotes important information. It calls attention to a procedure,
               practice, condition or the like, which is essential to highlight.




                                                                                             vii
Documentation Map
Operation Manual
Operation Manual (Agilent Part Number: 04263-90010 for English edition, 04263-97010 for
Japanese edition) explains specications, basic measurement explanations, and how to verify
conformance to published specications.

Service Manual
Service Manual (Agilent Part Number: 04263-90033) explains how to adjust, troubleshoot, and
repair the 4263B.



How To Use This Manual
This is the Service Manual for the 4263B LCR Meter. This manual contains adjustments and
repair information, and consists of the following four chapters.

Chapter 1 General Information
Chapter 1 gives general information for servicing the 4263B. This chapter lists the
recommended equipment for adjustments and troubleshooting.

Chapter 2 Adjustments
Chapter 2 describes the adjustments, required to ensure that the 4263B is within its published
specications after it has been repaired, or it fails the performance tests.

Chapter 3 Assembly Replacement
Chapter 3 provides the information on replacing the 4263B assemblies. The information
includes replaceable assembly lists, and removal and installation procedures.

Chapter 4 Troubleshooting
Chapter 4 provides the information on troubleshooting the 4263B. The information includes
theory of operation and information on faulty assembly isolation.

Appendix A Manual Changes
Appendix A contains manual changes and provides information for using this manual with
4263Bs manufactured before the printing date of the manual.

Appendix B Default Jumper and Switch Settings
Appendix B describes the settings of the jumpers and switches on the 4263B A2 CPU Board
Assembly. The information is required when the A2 assembly is replaced.




viii
Contents
 1. General Information
    Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                                              1-1
    Instruments Covered by This Manual . . . . . . . . . . . . . . . . . . . . .                                                1-1
    Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                                                1-2
 2. Adjustments
    Introduction . . . . . . . . . . . . . . .        .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-1
      Safety Consideration . . . . . . . . . .        .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-1
    Required Equipment . . . . . . . . . . .          .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-1
    Adjustment Tools . . . . . . . . . . . .          .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-2
    Adjustable Components . . . . . . . . .           .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-2
    Factory-Selected Components . . . . . . .         .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-2
    EEPROM Write Protect Switch . . . . . .           .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-2
    Related Adjustments . . . . . . . . . . .         .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-2
    Adjustment Program . . . . . . . . . . .          .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-3
      Program Installation . . . . . . . . . .        .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-5
        Making a Working Copy . . . . . . .           .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-5
        Dening the Conguration . . . . . .          .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-5
      Program Execution . . . . . . . . . .           .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-5
    Adjustments Procedure References . . . .          .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-7
      Signal Source Level Adjustment . . . . .        .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-7
        Equipment . . . . . . . . . . . . .           .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-7
        Setup . . . . . . . . . . . . . . . .         .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-7
      Impedance Measurement 0m Adjustment             .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-8
        Equipment . . . . . . . . . . . . .           .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-8
        Setup . . . . . . . . . . . . . . . .         .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-8
      Impedance Measurement 1m Adjustment             .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-9
        Equipment . . . . . . . . . . . . .           .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-9
        Setup . . . . . . . . . . . . . . . .         .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-9
      Impedance Measurement 2m Adjustment             .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   2-10
      Impedance Measurement 4m Adjustment             .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   2-10
 3. Assembly Replacement
    Introduction . . . . . . . . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    3-1
    Ordering Information . . . . . . . .      .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    3-1
    Restored Exchange Assemblies . . . .      .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    3-1
    Replacing the A2 CPU Board Assembly       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    3-2
    Replaceable Assembly List . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    3-2
    Replaceable Mechanical Parts List . . .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    3-5
    Disassembly Procedures . . . . . . .      .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    3-9
      Tools and Fasteners . . . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    3-9
      Cover Removal . . . . . . . . . .       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    3-9
      A1 Main Board Removal . . . . . .       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   3-10
        Removal Procedure . . . . . . .       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   3-10
      A2 CPU Board Assembly Removal .         .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   3-10
        Removal Procedure . . . . . . .       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   3-10

                                                                                                                          Contents-1
         A5 AC Inlet Board Assembly Removal   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   3-10
           Removal Procedure . . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   3-10
         Keyboard Assembly Removal . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   3-11
           Removal Procedure . . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   3-11
         LCD Assembly Removal . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   3-11
           Removal Procedure . . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   3-11
         BNC Connector Removal . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   3-11
           Removal Procedure . . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   3-11
  4. Troubleshooting
       Introduction . . . . . . . . . . . . . . . . . . . . . . .                     .   .   .   .   .   .   .   .   .   .    4-1
         Safety Considerations . . . . . . . . . . . . . . . . .                      .   .   .   .   .   .   .   .   .   .    4-1
       Required Equipment . . . . . . . . . . . . . . . . . . .                       .   .   .   .   .   .   .   .   .   .    4-1
       After Service Product Safety Checks . . . . . . . . . . .                      .   .   .   .   .   .   .   .   .   .    4-2
       Theory of Operation . . . . . . . . . . . . . . . . . . .                      .   .   .   .   .   .   .   .   .   .    4-3
         Overall Measurement Theory . . . . . . . . . . . . . .                       .   .   .   .   .   .   .   .   .   .    4-3
         Overall Block Diagram . . . . . . . . . . . . . . . . .                      .   .   .   .   .   .   .   .   .   .    4-4
         Analog Section . . . . . . . . . . . . . . . . . . . .                       .   .   .   .   .   .   .   .   .   .    4-5
           Signal Source Section . . . . . . . . . . . . . . . .                      .   .   .   .   .   .   .   .   .   .    4-6
           Transducer Section . . . . . . . . . . . . . . . . .                       .   .   .   .   .   .   .   .   .   .    4-7
           Vector Ratio Detector Section . . . . . . . . . . . . .                    .   .   .   .   .   .   .   .   .   .    4-8
           Transformer Parameter Measurement (Option 001 only)                        .   .   .   .   .   .   .   .   .   .    4-8
         Digital Section . . . . . . . . . . . . . . . . . . . . .                    .   .   .   .   .   .   .   .   .   .    4-9
         Power Supply Section . . . . . . . . . . . . . . . . .                       .   .   .   .   .   .   .   .   .   .   4-10
       Troubleshooting . . . . . . . . . . . . . . . . . . . . .                      .   .   .   .   .   .   .   .   .   .   4-11
         Check Procedure References . . . . . . . . . . . . . .                       .   .   .   .   .   .   .   .   .   .   4-13
           Check 1, DC-DC Converter Output Voltage . . . . . .                        .   .   .   .   .   .   .   .   .   .   4-13
           Check 2, LCD Displays Anything? . . . . . . . . . . .                      .   .   .   .   .   .   .   .   .   .   4-14
           Check 3, LCD Operates Correctly? . . . . . . . . . .                       .   .   .   .   .   .   .   .   .   .   4-14
           Check 4, Any Error Message on LCD? . . . . . . . . .                       .   .   .   .   .   .   .   .   .   .   4-14
           Check 5, CPU Related Error Message? . . . . . . . . .                      .   .   .   .   .   .   .   .   .   .   4-14
           Check 6, CPU Power LED Lights? . . . . . . . . . . .                       .   .   .   .   .   .   .   .   .   .   4-14
           Check 7, Any Error Code on LED? . . . . . . . . . .                        .   .   .   .   .   .   .   .   .   .   4-15
           Check 8, CPU Fuse High Lead Voltage . . . . . . . . .                      .   .   .   .   .   .   .   .   .   .   4-16
           Check 9, Fuse and Line Switch . . . . . . . . . . . .                      .   .   .   .   .   .   .   .   .   .   4-16
           Check 10, DC-DC Converter Input Voltage . . . . . . .                      .   .   .   .   .   .   .   .   .   .   4-17
           Check 11, DC-DC Converter Open Output Voltages . . .                       .   .   .   .   .   .   .   .   .   .   4-18
           Check 12, Transformer Secondary Voltage . . . . . . .                      .   .   .   .   .   .   .   .   .   .   4-19
  A. Manual Changes
       Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                                         A-1
       Manual Changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                                           A-1
  B. Default Jumper and Switch Settings
       Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                                         B-1
       Jumper and Switch Settings . . . . . . . . . . . . . . . . . . . . . . . . .                                           B-1
       Index




Contents-2
Figures
  1-1.   Serial Number Label . . . . . . . . . . . . .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    1-1
  2-1.   Adjustment Program Flow . . . . . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-3
  2-2.   Signal Source Level Adjustment Setup . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-7
  2-3.   Impedance Measurement 0m Adjustment Setup       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-8
  2-4.   Impedance Measurement 1m Adjustment Setup       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    2-9
  3-1.   Cover Removal . . . . . . . . . . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    3-9
  4-1.   Voltage Current Ratio Measurement Principle .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-3
  4-2.   4263B Overall Block Diagram . . . . . . . . .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-4
  4-3.   Analog Section Block Diagram . . . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-5
  4-4.   Signal Source Section Block Diagram . . . . .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-6
  4-5.   Transducer Block Diagram . . . . . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-7
  4-6.   Vector Ratio Detector Block Diagram . . . . .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-8
  4-7.   Basic Transformer Measurement Setup . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-8
  4-8.   Digital Section Block Diagram . . . . . . . .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    4-9
  4-9.   Power Supply Section Block Diagram . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   4-10
 4-10.   Troubleshooting Flow Chart . . . . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   4-12
 4-11.   DC-DC Converter Output Voltage Check . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   4-13
 4-12.   CPU Board Assembly LED Light Sequence . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   4-15
 4-13.   Location of High Lead of A2F1 Fuse . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   4-16
 4-14.   DC-DC Converter Input Voltage Check . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   4-17
 4-15.   DC-DC Converter Output Connector . . . . .      .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   4-18
 4-16.   Transformer Secondary Voltage Check . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   4-19
  B-1.   Jumper and Switch Settings . . . . . . . . .    .   .   .   .   .   .   .   .   .   .   .   .   .   .   .    B-1




Tables
 1-1.    Required Equipment . . . . . . . . . . . . . . .        .   .   .   .   .   .   .   .   .   .   .   .   .     1-2
 1-2.    Required Language Extensions . . . . . . . . . .        .   .   .   .   .   .   .   .   .   .   .   .   .     1-2
 2-1.    Related Adjustments . . . . . . . . . . . . . . .       .   .   .   .   .   .   .   .   .   .   .   .   .     2-2
 3-1.    Replaceable Major Assembly List . . . . . . . . .       .   .   .   .   .   .   .   .   .   .   .   .   .     3-3
 3-2.    A2 CPU Board Assembly, Replaceable Assembly List        .   .   .   .   .   .   .   .   .   .   .   .   .     3-4
 3-3.    Replaceable Assembly List . . . . . . . . . . . .       .   .   .   .   .   .   .   .   .   .   .   .   .     3-4
 3-4.    Replaceable Mechanical Parts List 1 . . . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .     3-5
 3-5.    Replaceable Mechanical Parts List 2 . . . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .     3-6
 3-6.    Replaceable Mechanical Parts List 3 . . . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .     3-7
 3-7.    Replaceable Mechanical Parts List 4 . . . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .     3-8
 4-1.    DC-DC Converter Output Voltage (Connected) . . .        .   .   .   .   .   .   .   .   .   .   .   .   .    4-13
 4-2.    DC-DC Converter Open Output Voltages . . . . . .        .   .   .   .   .   .   .   .   .   .   .   .   .    4-18
 A-1.    Manual Changes by Serial Number . . . . . . . .         .   .   .   .   .   .   .   .   .   .   .   .   .     A-1




                                                                                                                 Contents-3
                                                                                                  1
General Information
Introduction
This chapter provides information on instruments covered by this manual and required
equipment for the adjustments and troubleshooting.



Instruments Covered by This Manual
Agilent Technologies uses a two-section, nine character serial number which is printed on
the serial number label (Figure 1-1) attached to the instrument's rear panel. The rst ve
characters are the serial prex, and the last ve digits are the sux.




                               Figure 1-1. Serial Number Label
An instrument manufactured after the printing date of this manual may have a serial
number prex that is not listed on the title page. This unlisted serial number prex indicates
the instrument is dierent from those described in this manual. The manual for this new
instrument may be accompanied by a yellow Manual Changes supplement or have a dierent
manual part number. This sheet contains \change information" that explains how to adapt the
manual to the newer instrument.
In addition to change information, the supplement may contain information for correcting
errors (Errata) in the manual. To keep this manual as current and accurate as possible, Agilent
Technologies recommends that you periodically request the latest Manual Changes supplement.
The supplement for this manual is identied by this manual's printing date and its part number,
both of which appear on the manual's title page. Complimentary copies of the supplement are
available from Agilent Technologies. If the serial prex or number of an instrument is lower
than that on the title page of this manual, see Appendix A, Manual Changes . For information
concerning, a serial number prex that is not listed on the title page or in the Manual Change
supplement, contact the nearest Agilent Technologies oce.




                                                                            General Information   1-1
Required Equipment
Table 1-1 lists the required equipment for adjusting and troubleshooting the 4263B. Table 1-2
lists the required language extensions to be used with the adjustment program.
                                            Table 1-1. Required Equipment
             Equipment                      Requirements                  Recommended Model         Qty. Use
      Multimeter                    AC V Accuracy: < 2.5%              3458A or 3478A                1     A1, T2
                                    (at 100 kHz)
      Standard Resistors            No Substitute                      42030A                        1       A
      Open Termination              No Substitute                      42090A3                       1       A
      Short Termination             No Substitute                      42091A3                       1       A
      Computer                      No Substitute                      HP 9000 series 200 or 3004    1       A
                                                                       RAM  4M bytes
      Operating System              BASIC 5.1 or higher                HP 98616A                     1       A
      Adjustment Program No substitute                                 Agilent PN 04263-65005        1       A
      Cable                         BNC(m)-BNC(m) Cable                Agilent PN 8120-1839          1       A
                                    Dual Banana-BNC(m) Cable Agilent PN 11001-60001                  1       A
                                    Test Leads (1 m)                   16048A                        1       A
                                    Test Leads (2 m)                   16048D                        1       A
                                    Test Leads (4 m)                   16048E                        1       A
      Adapter                       BNC(f)-BNC(f) Adapter              Agilent PN 1250-0080          4       A

      1   Adjustment

      2   Troubleshooting

      3   This model is also included in the 42100A, Four Terminal Pair Resistor Set

      4   Excluding the HP 9826A



                                     Table 1-2. Required Language Extensions
                      Name        Ver.     Name Ver.        Name Ver.        Name                   Ver.
                 CLOCK            5.0     CS80           5.01     GRAPH           5.2   KBD         5.1
                 COMPLEX          5.1     EDIT           5.1      GRAPHX          5.2   MAT         5.1
                 CRTA             5.1     ERR            5.1      GPIB            5.0   MS          5.1
                 CRTX             5.1     FGPIB          5.0      IO              5.1   PDEV        5.0
                 1   This language extension depends on the mass storage type used.




1-2       General Information
                                                                                               2
Adjustments
Introduction
This chapter describes the adjustments required to ensure the 4263B is within its published
specications after it has been repaired, or when it fails the performance tests. The
adjustments should be performed along with periodic maintenance to keep the 4263B in
optimum operating condition. If proper performance cannot be achieved after adjustments,
proceed to Chapter 4.
Note            To ensure proper results and correct instrument operation, a 30 minute
                warm-up and stabilization is required before performing the adjustments.



Safety Consideration
This chapter contains NOTEs, CAUTIONs, and WARNINGs which must be followed to ensure
operator safety and to maintain the instrument in a safe and serviceable condition. The
adjustments covered in this chapter should be performed only by a qualied person who is
aware of the hazards.
Warning         Any interruption of the protective ground conductor (inside or outside
                the instrument) or disconnection of the protective ground terminal can
                make the instrument dangerous. Intentional interruption of the protective
                ground system for any reason is prohibited.



Required Equipment
Table 1-1 lists the equipment required to perform the adjustments. Use only calibrated test
equipment when adjusting the 4263B.




                                                                                 Adjustments   2-1
Adjustment Tools
No hand tools are required for the adjustments.



Adjustable Components
There are no adjustable components, such as variable capacitors or variable resistors, in the
4263B. All adjustments are performed by updating the adjustment data stored in the internal
EEPROM.



Factory-Selected Components
No factory-selected components are used in the 4263B



EEPROM Write Protect Switch
There are no EEPROM write protect switches in the 4263B.



Related Adjustments
There are two adjustments for the 4263B. Table 2-1 lists the adjustments that must be
performed if an assembly has been replaced.
                                     Table 2-1. Related Adjustments
                               Assembly Replaced                 Required Adjustment
                    A1 Main Board Assy                           All adjustments
                    A2 CPU Board Assy without ROM None1
                    A5 AC Inlet Board Assy                       None
                    LCD Assembly                                 None
                    DC-DC Converter Unit                         None
                    Keyboard Assembly                            None
                    1 If   the EEPROM, A2U1, is replaced, all adjustments are required.




2-2   Adjustments
Adjustment Program
The adjustment program listed in Table 1-1 is required for the adjustments. Using the program,
the computer controls the 4263B and the other pieces of equipment to obtain the adjustment
data, and stores the data in the 4263B internal EEPROM, A2U1.
The adjustment program runs on the computer listed in Table 1-1. The program should be
installed into your computer according to the procedure described in \Program Installation". To
start the program after it has been installed, refer to the \Program Execution".
Figure 2-1 shows the adjustment program 
ow. The following lists the brief explanation for
each menu item. At the end of each adjustment, the program writes the obtained adjustment
data both to the 4263B internal EEPROM and to the unit-serial-number dependent le on the
disk.




                           Figure 2-1. Adjustment Program Flow
INITIAL SETUP         denes the standards' calibration values. Update the standard's
                      calibration values if necessary.
Signal Source Level   obtains the data to adjust the actual signal source output voltage to the
Adjustment            set value. This adjustment should be performed if any performance test
                      fails.
Impedance             obtains the data to adjust the actual measurement impedance to the
Measurement 0m        standards' value, when no test leads are used. This adjustment should be
Adjustment            performed if the performance test, Impedance Measurement Accuracy
                      Test, fails. Perform the Signal Source Level Adjustment prior to this
                      adjustment.

                                                                                 Adjustments   2-3
Impedance                 obtains the data to adjust the actual measurement impedance to the
Measurement 1m            standards' value, when the 1m test leads are used. This adjustment
Adjustment                should be performed if the performance test, Impedance Measurement
                          Accuracy Test, fails. Perform the Impedance Measurement 0m
                          Adjustment prior to this adjustment.
Impedance                 obtains the data to adjust the actual measurement impedance to the
Measurement 2m            standards' value, when the 2m test leads are used. This adjustment
Adjustment                should be performed if the performance test, Impedance Measurement
                          Accuracy Test, fails. Perform the Impedance Measurement 0m
                          Adjustment prior to this adjustment.
Impedance                 obtains the data to adjust the actual measurement impedance to the
Measurement 4m            standards' value, when the 4m test leads are used. This adjustment
Adjustment                should be performed if the performance test, Impedance Measurement
                          Accuracy Test, fails. Perform the Impedance Measurement 0m
                          Adjustment prior to this adjustment.
Rewrite EEPROM            writes the adjustment data, which is stored in the unit-serial-number
                          dependent le, to the EEPROM. Only the adjustment data of previously
                          adjusted item(s) will be written.

Note                Make sure that the correct line frequency (50 or 60Hz) has been selected in the
                    \UUT: 4263B" window preceding the Main Menu. This selection will improve
                    the adjustment data accuracy.

Note                Because the Impedance Measurement 0m Adjustment serves the reference
                    data for the other Impedance Measurement Adjustments, it must have been
                    performed prior to the 1m, 2m, and 4m adjustments.

Note                Although performing some of the Impedance Measurement Adjustments may
                    satisfy the user's requirement (e.g. for users using the 4m test leads only for
                    the application), it is strongly recommended to perform all the Impedance
                    Measurement Adjustments. If only one of the 1m, 2m, and 4m adjustments
                    is performed after the 0m adjustment, the data for the remaining Impedance
                    Measurement Adjustments are invalid. For example, if only the 4m adjustment
                    is performed after the 0m adjustment, the measurement accuracy for the 1m
                    and 2m may be out of the specication.




2-4   Adjustments
Program Installation
The adjustment program must be installed in the computer using the following procedures.
The installation can be divided into two stages, \Making a Working Copy" and \Dening the
Conguration".
Making a Working Copy
Copy the contents of the adjustment program disk to a working disk or to the harddisk to
prevent the program les from accidental deletion or destruction. Use the working disk or the
harddisk and store the original disk in a safe place.
Dening the Conguration
Dene the conguration of the adjustment equipment. The conguration includes the type and
the GPIB address of the multimeter to be used. Because the program depends on equipment
conguration, it must be redened if the equipment conguration has changed.
Refer to the following procedure to dene the conguration.
1. Turn on the computer and bring up the BASIC system.
2. Set the Mass Storage Unit Specier to the drive/directory where the adjustment program
   exists using the MSI command.
3. Load the \TE_A4263B" le from the adjustment program disk by pressing
   LOAD "TE_A4263B" 4Return5 (or 4Enter5).
4. Run the loaded program by pressing RUN.
5. As the program instructs, select the type of the multimeter to be used and enter its GPIB
   address.
Note           Once the multimeter type and GPIB addresses are set, they cannot be changed
               unless the conguration is redened.



Program Execution
Perform the following steps to start the adjustment program.

Note           Install the program prior to execution. Refer to \Program Installation".



1. Turn on the computer and bring up the BASIC system. Refer to the Table 1-2 for the
   required language extensions.
2. Set the Mass Storage Unit Specier to the drive/directory where the adjustment program
   exists using the MSI command.
3. Load the program, \ADJ4263B" by pressing LOAD "ADJ4263B" 4Return5 (or 4Enter5).
4. Execute the program by pressing RUN.




                                                                                 Adjustments   2-5
Note                In the program, the 4263B is referred as \UUT".



Note                If the adjustment being performed follows right after EEPROM replacement,
                    change switch A2SW1-1 to \CLOSE", so that the 4263B will ignore the self test
                    errors at power-on. (Return the switch to \OPEN" when the adjustment has
                    been completed.) Refer to Figure B-1 for the switch location.




2-6   Adjustments
Adjustments Procedure References
This section provides supplemental information for the adjustments which are performed by
the program. Connection instructions and some additional explanations are provided. Use this
chapter as a reference when performing the adjustments.
A adjustment of the 4263B consists of the following ve adjustments.
  Signal Source Level Adjustment
  Impedance Measurement 0m Adjustment
  Impedance Measurement 1m Adjustment
  Impedance Measurement 2m Adjustment
  Impedance Measurement 4m Adjustment

Signal Source Level Adjustment
This adjustment updates the signal source level adjustment data stored in the EEPROM. The
data adjust the actual signal source level to the set value.
Equipment
                         Multimeter                    3458A
                         Dual Banana-BNC(m) Cable      PN 11001-60001

Setup
This setup obtains the signal source level adjustment data.




                     Figure 2-2. Signal Source Level Adjustment Setup
Perform the following steps as the program instructs.
1. Connect the Dual Banana-BNC(m) cable to the 3458A's INPUT binding posts.
2. Connect another end of the Dual Banana-BNC(m) cable to the 4263B's HCUR Connector.

                                                                               Adjustments   2-7
Impedance Measurement 0m Adjustment
This adjustment updates the data to adjust the measurement impedance to the standard values,
when no test leads are used.
Note                Perform the Signal Source Level Adjustment prior to performing this
                    adjustment.



Equipment
                                 Standard Resistors           42030A
                                 Open Termination             42090A
                                 Short Termination            42091A

Setup
This setup obtains the impedance measurement adjustment data for 0 m test leads (no test
leads).




                    Figure 2-3. Impedance Measurement 0m Adjustment Setup
Perform the following steps as the program instructs.
1. Connect the specied termination or standard resistor directly to the 4263B UNKNOWN
   Connectors.
2. Repeat step 1 as instructed.




2-8   Adjustments
Impedance Measurement 1m Adjustment
This adjustment updates the data to adjust the measurement impedance to the standard values,
when the 1m test leads are used.
Note           Perform the Impedance Measurement 0m Adjustment prior to performing this
               adjustment.



Equipment
                            Standard Resistors           42030A
                            Open Termination             42090A
                            Short Termination            42091A
                            Test Leads (1 m)             16048A

Setup
This setup obtains the impedance measurement adjustment data for 1 m test leads.




                Figure 2-4. Impedance Measurement 1m Adjustment Setup
Perform the following steps as the program instructs.
1. Connect the 16048A directly to the 4263B UNKNOWN Connectors.

                                                                              Adjustments   2-9
2. Connect four BNC(f)-BNC(f) Adapters directly to another end of the 16048A.
3. Connect the specied termination or standard resistor directly to the 4263B UNKNOWN
   Connectors.
4. Repeat step 3 as instructed.

Impedance Measurement 2m Adjustment
This adjustment updates the data to adjust the measurement impedance to the standard values,
when the 2m test leads are used.
Note                 Perform the Impedance Measurement 0m Adjustment prior to performing this
                     adjustment.


Perform the same setup procedures for the 16048D as used for the 16048A.

Impedance Measurement 4m Adjustment
This adjustment updates the data to adjust the measurement impedance to the standard values,
when the 4m test leads are used.
Note                 Perform the Impedance Measurement 0m Adjustment prior to performing this
                     adjustment.


Perform the same setup procedures for the 16048E as used for the 16048A.




2-10   Adjustments
                                                                                                  3
Assembly Replacement
Introduction
This chapter provides the 4263B replaceable parts lists, and the step by step disassembly
procedures for the major assemblies. Table 3-1 to Table 3-7 describe all the 4263B replaceable
parts. The gure section of Table 3-7 also shows the cable connections.



Ordering Information
To order a part listed in the replaceable parts lists, quote the Agilent Technologies part number,
indicate the quantity required, and address the order to the nearest Agilent Technologies oce.
To order a part that is not listed in the replaceable parts list, include the instrument model
number, complete instrument serial number, the description and function of the part, and the
quantity of parts required. Address the order to the nearest Agilent Technologies oce.



Restored Exchange Assemblies
Some replacement assemblies are available as either new or restored assemblies. The restored
assembly is more economical than a new assembly and, as with new assemblies, a 90-day
warranty applies through the instrument's support life. The defective assembly must be
returned for credit. For this reason, new assemblies must be ordered for spare parts. The part
numbers for both new and restored assemblies are given in the replaceable parts lists.
Note            The Exchange Program may not apply in your country. Contact the nearest
                Agilent Technologies oce for further details.




                                                                           Assembly Replacement   3-1
Replacing the A2 CPU Board Assembly
The A2 CPU Board Assembly has a ROM for the rmware and an EEPROM for the adjustment
data and user data. Both ROM and EEPROM are mounted in sockets for easy replacement. If
the A2 assembly is replaced, the ROM and EEPROM may be transferred from the defective
assembly to the new assembly as long as the ROM and EEPROM are not defective. The
ROM transfer reduces parts cost for the repair, and the EEPROM transfer saves time from
readjusting the 4263B. For this purpose, the A2 assembly can only be ordered without the
ROM.
When the A2 assembly is replaced, make sure that the jumpers and switches on the A2
assembly are in the default position as shown in Appendix B.



Replaceable Assembly List
The replaceable electrical assemblies are listed in Table 3-1 and Table 3-3.




3-2   Assembly Replacement
                              Table 3-1. Replaceable Major Assembly List




     Reference            Agilent Part        Qty.                      Description
    Designation            Number
           1            04263-66505            1         A5 AC Inlet Board Assembly
           2            04263-60004            1         Transformer Assembly
           3            04263-80901            1         DC-DC Converter Unit
           4            04263-66532            1         A2 CPU Board Assembly w/o ROM
                        04263-00604            1         Case Shield1
                        0515-1550              4         Screw M3
           5            04263-66504            1         A4 Keyboard Assembly
                        04263-25151            1         Rubber Key
                        0515-1550              1         Screw M3
           6            04263-61010            1         LCD Assembly
                        04263-25003            1         Filter Transparent
                        0515-1550              2         Screw M3
           7            04263-66501            1         A1 Main Board Assembly
                        04263-69501                      A1 Main Board Assembly (Exchange
                                                         Assembly)

1   Included in 04263-66532




                                                                              Assembly Replacement   3-3
                Table 3-2. A2 CPU Board Assembly, Replaceable Assembly List




   Reference           Agilent Part        Qty.                      Description
  Designation           Number
         1           2110-1123              1         A2F1, Fuse SMT 2 A
         2           1818-5146              1         A2U1, EEPROM
         3           04263-85041            1         A2U11, ROM Rev. 1.01

                              Table 3-3. Replaceable Assembly List
   Reference           Agilent Part        Qty.                      Description
  Designation           Number
      A1C500         0180-4334              1         Capacitor, 22000 F (for power supply)
      A1C501         0180-4334              1         Capacitor, 22000 F (for power supply)
      A5SW1          3101-3173              1         Line Switch
      A5VR1          0837-0237              1         Varistor
      A5VR2          0837-0237              1         Varistor




3-4   Assembly Replacement
Replaceable Mechanical Parts List
The replaceable mechanical parts are listed in Table 3-4 to Table 3-7.
                               Table 3-4. Replaceable Mechanical Parts List 1




       Reference             Agilent Part               Qty.                                          Description
      Designation              Number

            1            04263-04004                        1       Cover

            2            5041-9167                          4       Foot

                         1460-1345                          2       Tilt Stand

            3            04263-65020                        1       Front Panel Assembly
                                                                                                        1
            4            5041-9170                          2       Trim Side

                         0515-0914                          4       Screw M3 (Hidden)

            5            04263-40001                        1       Front Bezel
                                                                                     2
            6            04263-00605                        1       Sub Panel (Hidden)
                                                                                                  2
                         0515-0914                          4       Screw M3        (Hidden)
                                                                                             2
                         0515-1550                          2       Screw M3 (Hidden)
                                                                                                  2
                         0160-4808                          4       Capacitor 470 pF (Hidden)
                                                                                                              2
                         1901-1250                          8       Diode (Hidden)
                                                                                          2
            7            1250-0252                          4       BNC Connector
                                                                                          2
                         2950-0035                          4       Nut
                                                                        2
                         5040-3324                          4       Insulator
                                                                                2
                         5040-3325                          4       Insulator (Hidden)
                                                                                                  2
                         04263-01221                        4       Lug Terminal (Hidden)
                                                                                                        2
            8            1510-0130                          1       Binding Post
                                                                                      2
                         2190-0067                          1       Washer (Hidden)
                                                                                              2
                         2950-0006                          1       Nut    (Hidden)
                                                                                    2
            9            04263-87121                        1       Panel Label
                                                                                     2

1   The assembly consists of the parts designated by 5 through 9.

2   This part is included in the Front Panel Assembly, 3.



                                                                                                            Assembly Replacement   3-5
                              Table 3-5. Replaceable Mechanical Parts List 2




   Reference              Agilent Part            Qty.                   Description
  Designation              Number
         1              04263-61615                1       GPIB Connector with 
atcable1
         2              1250-0083                  1       BNC Connector, Ext DC Bias
                        0360-1190                  1       Lug Terminal (Hidden)
                        2190-0016                  1       Washer (Hidden)
                        2950-0001                  1       Nut
         3              2110-0565                  1       Fuse Holder
                        2110-0202                  1       Fuse 0.25 A, Slow Blow (230V)
                        2110-0201                          Fuse 0.5 A, Slow Blow (115V)
         4              0515-1551                  2       Screw M3
         5              04263-00102                1       Chassis
         6              2190-0054                  1       Washer
                        2950-0054                  1       Nut

1 Includes   two sets of screw, washer and nut.
                                      ,




3-6   Assembly Replacement
                  Table 3-6. Replaceable Mechanical Parts List 3




 Reference     Agilent Part         Qty.                     Description
Designation     Number
    1         0515-2079              1         Screw M4 (on the Chassis)
    2         04263-24001            2         Nut Rod (for Transformer)
              2190-0586              2         Washer Spring M4
    3         04263-00607            1         Plate Shield
    4         04263-24002            2         Nut Rod (for DC-DC Conv.)
              2190-0584              2         Washer Spring M3
    5         0515-1550              7         Screw M3 (for A1 and A2)
              04339-24006            4         Stando (Hidden, for A1)
    6         5040-3347              2         Plate (Hidden, Chassis Reverse side)
              0460-0616         150 mm 2 2     Adhesive Tape (for Plate)
    7         04263-07001            2         Gasket
              0515-1550              4         Screw M3
    8         0370-2862              1         Button, Line Switch
              04339-23001            1         Shaft, Line Switch
              0535-0031              1         Nut (for Shaft)
    9         04339-25002            2         Support, Line Switch Plate
              0515-0914              2         Screw M3 Flat Head
    10        04263-01201            1         Plate, Line Switch
    11        0515-0914              5         Screw M3 Flat Head (for A5)




                                                                   Assembly Replacement   3-7
                         Table 3-7. Replaceable Mechanical Parts List 4




   Reference          Agilent Part         Qty.                     Description
  Designation          Number
         1          04263-61608             1         Wire Assy (V Selector)
         2          04263-61607             1         Wire Assy (Ext DC Bias)
         3          04263-61602             1         Cable Assembly, \A"-\A"
                    04339-25004             1         Core Holder
                    0515-1551               1         Screw M3
         4          04263-61604             1         Cable Assembly, \B"-\B"
                    04339-25004             1         Core Holder
                    0515-1551               1         Screw M3
         5          04263-61605             1         Cable Assembly, \C"-\C"
                    04339-25004             1         Core Holder
                    0515-1551               1         Screw M3
         6          04263-61606             1         Cable Assembly, \D"-\D"
                    04339-25004             1         Core Holder
                    0515-1551               1         Screw M3
         7          04339-61614             1         Wire Assy (Binding Post)




3-8   Assembly Replacement
Disassembly Procedures
This section provides the procedures to disassemble the 4263B's major assemblies. Tools and
fasteners required for the replacement are described.
Caution         SUSCEPTIBLE TO DAMAGE FROM ESD.
                Perform the following procedures only at a static-safety workstation and wear
                a grounding strap.


Tools and Fasteners
The 4263B mechanical components are secured using metric threaded fasteners. Many
fasteners in the 4263B may appear to be Phillips type, but they are in fact, Pozidrive type. To
avoid damaging them, use only pozidrive screwdrivers to remove or tighten pozidrive type
fasteners.

Cover Removal
The following procedure is common for gaining access to any of the assemblies.




                                  Figure 3-1. Cover Removal
1. Remove the two screws which fasten the cover to the chassis rear panel.
2. Slide the cover toward the rear while holding the front panel bezel.




                                                                          Assembly Replacement   3-9
A1 Main Board Removal
Removal Procedure
1. Remove the A2 CPU Board Assembly referring to \A2 CPU Board Assembly Removal".
2. Disconnect the following cable assemblies and wire assemblies from the A1 assembly.
   a. The four cable assemblies marked \A", \B", \C", and \D" which are connected to the
       UNKNOWN connectors on the Front Panel
   b. The wire assembly which is connected to the transformer
   c. The two wire assemblies which are connected to the DC-DC Converter
   d. The wire assembly which is connected to the Ext DC Bias Connector
3. Remove the nut which fasten the Ext Trigger connector to the rear panel.
4. Remove the four studs, that secure the A1 assembly to the chassis, by rotating them with a
   
at bladed screwdriver.
5. Remove the three screws that secure the A1 assembly to the chassis.
6. Remove the A1 assembly from the chassis by sliding the assembly toward forward before
   lifting.

A2 CPU Board Assembly Removal
Removal Procedure
1. Remove the cover referring to \Cover Removal".
2. Disconnect the two 
atcable assemblies and the wire assembly from the front side of the A2
   assembly.
3. Disconnect the 
atcable assembly, which is connected to the GPIB connector on the rear
   panel, from the rear right side of the A2 assembly.
4. Remove the four screws that secure the A2 assembly to the stud.
5. Remove the 
at cable assembly which is connected to the A1 main board assembly on the
   bottom, while sliding the A2 assembly toward front.
6. Remove the A2 assembly from the chassis.
Note              The ROM and EEPROM on the A2 assembly may be transferred to the new
                  assembly to minimize the repair time and parts cost. Refer to \Replacing the
                  A2 CPU Board Assembly" for more information.


A5 AC Inlet Board Assembly Removal
Removal Procedure
1. Disconnect the cable assembly with the 4 pin connector from the A5 assembly.
2. Disconnect the cable assembly with the 3 pin connector from the A5 assembly.
3. Remove the screw which xes the green and yellow wire connected to the neutral terminal
   of the power cable connector to the chassis bottom.
4. Remove the ve screws which secure the A5 assembly to the chassis, from the side.



3-10   Assembly Replacement
5. Slide the A5 assembly toward the right to remove the switch on the A5 assembly from the
   line switch plate slot.
6. Remove the A5 assembly from the chassis.

Keyboard Assembly Removal
Removal Procedure
1. Remove the cover referring to \Cover Removal".
2. Disconnect the two 
atcable assemblies and the wire assembly from the front side of the A2
   assembly.
3. Remove the two 
at head screws that secure the front panel from the bottom.
4. Remove the side trim label from both sides of the front bezel.
5. Remove four screws that secure the front panel bezel to the chassis from both sides.
6. Place the front panel assembly, so that the front panel faces down.
7. Remove the screw that secures the keyboard to the front panel bezel.
8. Remove the keyboard assembly from the front panel bezel.

LCD Assembly Removal
Removal Procedure
1. Remove the cover referring to \Cover Removal".
2. Disconnect the two 
atcable assemblies and the wire assembly from the front side of the A2
   assembly.
3. Remove the two 
at head screws that secure the front panel from the bottom.
4. Remove the side trim label from both sides of the front bezel.
5. Remove the four screws that secure the front panel bezel to the chassis from both sides.
6. Place the front panel assembly, so that the front panel faces down.
7. Remove the two screws that secure the LCD to the front panel bezel.
8. Remove the LCD assembly from the front panel bezel.

BNC Connector Removal
Removal Procedure
 1. Remove the cover referring to \Cover Removal".
 2. Disconnect the two 
atcable assemblies and the wire assembly from the front side of the
    A2 assembly.
 3. Remove the two 
at head screws that secure the front panel from the bottom.
 4. Remove the side trim label from both sides of the front bezel.
 5. Remove four screws that secure the front panel bezel to the chassis from both sides.
 6. Place the front panel assembly, so that the panel faces down.


                                                                      Assembly Replacement   3-11
 7. Disconnect the cable assembly which is connected to the BNC connector being replaced
    from the A1 main board assembly.
 8. Remove the core holder which holds the cable assembly which is connected to the BNC
    connector being replaced, by unscrewing the screw secures the holder from the front side.
 9. Disconnect the cable from the BNC connector housing being replaced, by unscrewing the
    nut which fastens, from the back of the front panel, the cable to the BNC connector.
10. Unscrew the nut which connects the BNC connector housing to the sub panel.
11. Remove the BNC connector.




3-12   Assembly Replacement
                                                                                               4
Troubleshooting
Introduction
This chapter provides assembly level troubleshooting information for the 4263B. The
information includes theory of operation, and troubleshooting.

Safety Considerations
This chapter contains WARNINGs and CAUTIONs that must be followed for your protection and
to avoid damaging the equipment.
Warning         The servicing procedures described herein are performed when power is
                supplied to the instrument and its protective covers are removed. This
                type of servicing must be performed only by service trained personnel
                who are aware of the hazards involved (for example, re, and electrical
                shock). When serving can be performed without power applied to the
                instrument, remove power from the instrument. Before any repair is
                completed, ensure that all safety features are intact and functioning,
                and that all necessary parts are properly connected to the protective
                grounding system.



Required Equipment
Table 1-1 lists the equipment required for troubleshooting.




                                                                             Troubleshooting   4-1
After Service Product Safety Checks
Warning            Whenever it appears likely that protective safety provisions have been
                   impaired, the apparatus shall be marked as inoperative and should be
                   secured against any unintended operation. The protection provisions will
                   have likely been compromised if, for example:
                     The instrument shows visible signs of damage.
                     The instrument fails to perform the intended measurement.
                     The unit has undergone prolonged storage under unfavorable conditions.
                     The instrument was severely stressed in transport.

Perform the following ve safety checks to verify the safety condition of the 4263B (these
checks may also be used for safety checks after troubleshooting and repair).
1. Visually inspect the interior of the instrument for any signs of abnormal internally generated
   heat, such as discolored printed circuit boards and components, damaged insulation, or
   evidence of arcing. Determine the cause, and repair.
2. Use an ohmmeter which can accurately measure 0.1 
 to check the resistance from the
   instrument enclosure to the power cord's ground pin. The resistance must be less than
   0.1 
. Flex the power cord while making this measurement to determine if any intermittent
   discontinuities exist.
3. Unplug the 4263B's power plug from the power source. Set the line switch to ON. Tie the
   line and neutral pins of the power connector together and check the resistance between
   them and enclosure. The minimum accepta