Service Manuals, User Guides, Schematic Diagrams or docs for : Agilent HP 8663A Service Vol 4

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
HP 8663A Service Vol 4


>> Download HP 8663A Service Vol 4 documenatation <<

Text preview - extract from the document
                                                     30w          MS
                 HP 8663A
      SYNTHESIZED SIGNAL GENERATOR
                      (Including               Options             001, 002, & 003)
                                                Service Manual
                                                     VOlUwlG4
                                                   SERIAL NUMBERS

                            This manual applies directly to imt.rumentS with
                            serial numbers pr&xed:
                  2234A to 2927A and all MAJOR changesthat apply to your instrument.
                                                          rev.01JULSI
                               For additional irqrtant information about serial
                               numbers, refer to "INSTRUMENTS COVERED BY
                               THIS MANUAL" in Section 1.


                                                          Third Edition

                               This material may be reproducedby or for the U.S.
                               Government pursuant to the Copyright License un-
                               der the clause at DFARS 52.227-7013(APR 1988).

                                 Copyrighl@HEWLETT-PACKARD     COMPANY   1982
              EAST 24001     MISSION AVENUE, TAF C-34, SPOKANE,   WASHINGTON,      U.S.A.   99220


Service Manual (Volume   1, 2, 3, 4) HP Pal 08663-90071

Other Documenls Available:
Operation and Calibration Manual HP Part 08663-90069
Microfiche Service Manual Parr 08663-90072
                           HP
MIcrofiche Operation and Calibration Manual HP Part 06663-90070              Printed In U.S.A. : November   1990



                                                                                                HEWLETT
                                                                                                PACKARD
                    COPYRIGHT AND DISCLAIMER NOTICE

Copyright - Agilent Technologies, Inc. Reproduced with the permission of Agilent
Technologies Inc. Agilent Technologies, Inc. makes no warranty of any kind with regard
to this material including, but not limited to, the implied warranties of merchantability
and fitness for a particular purpose. Agilent Technologies, Inc. is not liable for errors
contained herein or for incidental or consequential damages in connection with the
furnishing, performance, or use of this material or data.
Model 8663A                                                                      Service



SERVICE SHEET 38
AllA  FM PHASE DETECTOR
REFERENCEBLOCK DIAGRAM 7
    Table 4-l. Recommended Performance Tests
               After Adjustments   or Repairs
    Table 5-2. Post-Repair  Adjustment Procedures

PRINCIPLES OF OPERATION
General
The purpose of the FM Phase Detector      (AllA4) is to develop the FM
Loop Error Voltage.     This is accomplished by comparing the VCO output
with a 10 MHz reference     signal.

Phase Detector   Circuitry
The Phase Detector        Circuitry    compares the phase difference   between
the reference      signal    (FM 10 MHz Reference Signal)     and the output of
the VCO (FM Loop N-Divider          Drive),  then generates a voltage
proportional     to the phase error.        Both input signals are first
divided down to 100 kHz before they are compared by the FM Loop Phase
Detector,    U4. The 10 MHz signal from the Reference Section is
divided by a total        of 100.    This signal becomes the reference     input
to the phase detector.           The 140 MHz VCO output is divided by a total
of 1400.     This signal becomes the variable         input to the phase
detector.
The Phase Detector,   U4, is a digital     phase detector   that generates
output pulse widths proportional      to the phase difference    between the
two input signals.    If the variable    100 kHz signal is lower in
frequency  or lags the reference    input in phase, the output on pin 13
generates a pulse (See Figure 8-601).        When the variable   input is
higher in frequency   or leads the input in phase the output on pin 2
generates a pulse (see Figure 8-602).




                               Figure 8-601. Phase Detector Operation
                 (Variable   Input is Lower in Frequency than Reference S&vml)
                                                                                    8-601