ANRITSU S113B_252C S114B_252C S331B_252C S332B Programming


>> Download ANRITSU S113B_252C S114B_252C S331B_252C S332B Programming documenatation <<

Text preview - extract from the document
                       Site Master TM
                       S113B, S114B
                       S331B, S332B
                     Antenna, Cable and
                     Spectrum Analyzer
                      Programming Manual




Hand-Held Tester For Antennas, Transmission
            Lines And Other RF Components
                                       WARRANTY
The Anritsu product(s) listed on the title page is (are) warranted against defects in materi-
als and workmanship for one year from the date of shipment.
Anritsu's obligation covers repairing or replacing products which prove to be defective
during the warranty period. Buyers shall prepay transportation charges for equipment re-
turned to Anritsu for warranty repairs. Obligation is limited to the original purchaser.
Anritsu is not liable for consequential damages.
                              LIMITATION OF WARRANTY
The foregoing warranty does not apply to Anritsu connectors that have failed due to nor-
mal wear. Also, the warranty does not apply to defects resulting from improper or inade-
quate maintenance by the Buyer, unauthorized modification or misuse, or operation
outside the environmental specifications of the product. No other warranty is expressed or
implied, and the remedies provided herein are the Buyer's sole and exclusive remedies.
                         TRADEMARK ACKNOWLEDGEMENTS
MS-DOS, Windows, and Windows for Workgroups are registered trademarks
of the Microsoft Corporation.
Site Master and Cable Mate are trademarks of Anritsu Company.
                                          NOTICE
Anritsu Company has prepared this manual for use by Anritsu Company personnel and
customers as a guide for the proper installation, operation, and maintenance of Anritsu
Company equipment and computer programs. The drawings, specifications, and informa-
tion contained herein are the property of Anritsu Company, and any unauthorized use or
disclosure of these drawings, specifications, and information is prohibited; they shall not
be reproduced, copied, or used in whole or in part as the basis for manufacture or sale of
the equipment or software programs without the prior written consent of Anritsu Com-
pany.
                                                                 Index

          How To Use This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

          Programming Manual Organization                                                    .................... 1


          Programming Overview
                     General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
                     Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
                     Serial Communication Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
                     Communications Error Checking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
                     Parameter Validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
                     Entering Remote Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
                     Exiting Remote Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
                     Lifetime of Changes to Site Master Operating Parameters . . . . . . . . . . 3
                     Write Cycle Limitation of EEPROM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3


          Functional Description

                     Initialize
                                   Enter remote mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
                                   Enter remote mode immediately . . . . . . . . . . . . . . . . . . . . . . . . 4

                     Configure
                                   System
                                           Fixed CW mode control . . . . . . . . . . . . . . . . . . . . . . . . 4
                                           LCD backlight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
                                           Watch dog timer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
                                           Serial port echo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
                                           Single sweep. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
                                           Printer type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
                                           Trace Math . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
                                           Automatically save runtime setup. . . . . . . . . . . . . . . . 4
                                   Measurement
                                           Measurement mode select. . . . . . . . . . . . . . . . . . . . . . 5
                                           Measurement unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
                                           Calibration on/off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
                                   Power Monitor
                                           Get Power levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
                                           Power Monitor Units of measure. . . . . . . . . . . . . . . . . 5
                                           Relative Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
                                           Offset Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
                                           Zero Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
                                   Frequency
                                           Start frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
                                           Stop frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
                                           Data points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

March 2000                                                                                                                               10580-00035
Copyright 1999, Anritsu Co.                                                                                                               Revision: C
               Distance
                             Start distance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
                             Stop distance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
                             Relative propagation velocity . . . . . . . . . . . . . . . . . . . 5
                             Cable loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
                             DTF Windowing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
               Display
                             Display mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
                             Scale start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
                             Scale stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
               Markers
                             On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
                             Delta Status On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
                             Value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
               Limits
                             On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
                             Beep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
                             Value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
     Utility
               Time set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
               Date set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
               Reference number set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

     Action
               Trigger sweep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
               Save system setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
               Recall system setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
               Save sweep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
               Query trace names. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
               Delete sweep trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

     Data
               Recall sweep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

     Calibration
               Calibrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

     Configure Spectrum Analyzer
               Set Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
               Set Center Frequency and Span . . . . . . . . . . . . . . . . . . . . . . . . . 7
               Set Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
               Set Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
               Set Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
               Measure OCC BW % of Power . . . . . . . . . . . . . . . . . . . . . . . . . . 7
               Measure OCC BW dB Down . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
               Set Max Hold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
               Set Resolution BW Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . 7



ii                                                                                S113B, S114B, S331B, S332B PM
                         Set Video BW Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
                         Set Sweep Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
                         Set Marker to Peak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
                         Set Marker to Center. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

                Configure Custom Cable Table
                         Write to Table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
                         Read from Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

                Status
                         Query system status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
                         Self-test Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
                         Read failure counters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
                         Clear failure counters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

                Close
                         Exit remote mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8


        Control Byte Descriptions
                Control byte summary          ........................................9
                Control byte 1                Setup system . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
                Control byte 2                Set Site Master frequency . . . . . . . . . . . . . . . . 12
                Control byte 3                Select measurement mode . . . . . . . . . . . . . . . . 13
                Control byte 4                Set Site Master scale . . . . . . . . . . . . . . . . . . . . 13
                Control byte 5                Set Site Master marker . . . . . . . . . . . . . . . . . . 14
                Control byte 6                Set Site Master limit. . . . . . . . . . . . . . . . . . . . . 15
                Control byte 7                Set DTF parameters . . . . . . . . . . . . . . . . . . . . . 15
                Control byte 8                Set time/date . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
                Control byte 9                Set reference number . . . . . . . . . . . . . . . . . . . . 17
                Control byte 10               Serial port echo on/off . . . . . . . . . . . . . . . . . . . . 17
                Control byte 11               Single sweep mode on/off . . . . . . . . . . . . . . . . . 17
                Control byte 12               Watch dog timer on/off . . . . . . . . . . . . . . . . . . . 18
                Control byte 13               Sequence Site Master calibration . . . . . . . . . . 18
                Control byte 14               Set Site Master data points . . . . . . . . . . . . . . . 19
                Control byte 16               Store sweep trace . . . . . . . . . . . . . . . . . . . . . . . 19
                Control byte 17               Recall sweep trace. . . . . . . . . . . . . . . . . . . . . . . 19
                Control byte 18               Save system setup. . . . . . . . . . . . . . . . . . . . . . . 24
                Control byte 19               Recall system setup . . . . . . . . . . . . . . . . . . . . . 24
                Control byte 20               Query system status . . . . . . . . . . . . . . . . . . . . . 24
                Control byte 21               Trigger self-test . . . . . . . . . . . . . . . . . . . . . . . . . 28
                Control byte 22               Read fail counters . . . . . . . . . . . . . . . . . . . . . . . 29
                Control byte 23               Clear fail counters. . . . . . . . . . . . . . . . . . . . . . . 29
                Control byte 24               Query trace names . . . . . . . . . . . . . . . . . . . . . . . 29
                Control byte 25               Delete sweep trace . . . . . . . . . . . . . . . . . . . . . . . 30
                Control byte 30               Select printer type . . . . . . . . . . . . . . . . . . . . . . . 30
                Control byte 31               Select DTF windowing . . . . . . . . . . . . . . . . . . . . 30
                Control byte 32               Set trace math . . . . . . . . . . . . . . . . . . . . . . . . . . 31
                Control byte 39               Query power level . . . . . . . . . . . . . . . . . . . . . . . 31
                Control byte 40               Set power monitor unit . . . . . . . . . . . . . . . . . . . 32



S113B, S114B, S331B, S332B PM                                                                                              iii
     Control byte 41         Relative mode on/off. . . . . . . . . . . . . . . . . . . . . . 32
     Control byte 42         Offset mode on/off . . . . . . . . . . . . . . . . . . . . . . . 32
     Control byte 43         Zero mode on/off . . . . . . . . . . . . . . . . . . . . . . . . . 33
     Control byte 48         Trigger sweep . . . . . . . . . . . . . . . . . . . . . . . . . . 33
     Control byte 50         Check battery status . . . . . . . . . . . . . . . . . . . . . 33
     Control byte 64         Automatically save runtime setup . . . . . . . . . . 34
     Control byte 69         Enter remote mode . . . . . . . . . . . . . . . . . . . . . . 34
     Control byte 70         Enter remote mode immediately. . . . . . . . . . . . 34
     Control byte 80         Write custom cable . . . . . . . . . . . . . . . . . . . . . . . 35
     Control byte 81         Recall custom cable . . . . . . . . . . . . . . . . . . . . . . 35
     Control byte 96         Measure OCC BW % of Power. . . . . . . . . . . . . . 36
     Control byte 97         Measure OCC BW dB Down . . . . . . . . . . . . . . . 36
     Control byte 99         Set Spectrum Analyzer Frequency . . . . . . . . . . 36
     Control byte 100        Set Spectrum Analyzer Center Freq/Span . . . . 37
     Control byte 101        Set Spectrum Analyzer Scale . . . . . . . . . . . . . . 37
     Control byte 102        Set Spectrum Analyzer Marker . . . . . . . . . . . . 38
     Control byte 103        Set Spectrum Analyzer Limit . . . . . . . . . . . . . . 38
     Control byte 105        Set Spectrum Analyzer Max Hold. . . . . . . . . . . 39
     Control byte 106        Set Spectrum Analyzer Resolution BW Freq . . 39
     Control byte 107        Set Spectrum Analyzer Video BW Freq . . . . . . 39
     Control byte 108        Set Spectrum Analyzer Sweep Mode . . . . . . . . 40
     Control byte 109        Set SpectrumAnalyzer Marker to Peak . . . . . . 40
     Control byte 110        Set Spectrum Analyzer Marker to Center . . . . 40
     Control byte 255        Exit remote mode . . . . . . . . . . . . . . . . . . . . . . . 41
     Parameter definitions    . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42




iv                                                                         S113B, S114B, S331B, S332B PM
                                     How to Use this Manual


If you want to find out
        general concepts for programming the Site Master,
                go to the Programming Overview.

If you want to find out
        how to perform a task,
                go to the Functional Description.

If you want to find out
        what the syntax of a control byte is,
                go to the Control Byte Description.




                           Programming Manual Organization


Programming Overview
This section contains general information about programming the Site Master. It is an introduction and
helps you get started.


Functional Descriptions
This section includes a complete list of tasks you can control via the serial port. These are high level descrip-
tions. Each task refers you to the appropriate individual control byte for details.


Control Byte Descriptions
This section contains detailed explanations for each of the control bytes.




S113B, S114B, S331B, S332B PM                                                                                   1
                                    Programming Overview


This programming manual is written exclusively for the Anritsu Site Master Models S113B, S114B, S331B,
and S332B with firmware Version 1.42 and above. For information on firmware upgrade, please contact your
local Anritsu service center.


General Description
The Site Master must first be set into "remote" mode for communication with a computer. Remote mode dif-
fers from normal repetitive sweep and single-sweep modes. During remote mode, the Site Master suspends
normal operations and attends to the serial port. The front panel display indicates when the Site Master is
in remote mode.
Once in remote mode, you can send control bytes and associated data to the Site Master via the serial port.
These control byte sequences command the Site Master to perform various functions and activities. The se-
rial port supports virtually all features accessible from the keypad. The only exception is the printer, which
requires connection to the same 9 pin connector on the Site Master rear panel.
To complete the communication session, send the control byte to exit remote mode. Site Master resumes nor-
mal operations. You may also exit the remote mode by using the escape key.


Cabling
Serial communications take place via the 9 pin connector on the back of the Site Master. The Site Master is a
DTE-type serial device and therefore requires a "null modem" cable for communication with a computer,
which is also a DTE device. We provide a suitable cable with your Site Master.


Serial Communication Parameters
The Site Master communicates at a baud rate of 9600. It uses no parity bits, 8 data bits, and 1 stop bit
(N-8-1). No hardware handshaking is used.


Communications Error Checking
Since there is no hardware handshaking, byte level error handling must be done by the controlling program.
The expected number of response bytes for each control byte (listed in the control byte description section of
this manual) works well for responses coming from the Site Master. For data streams going to the Site Mas-
ter, the "watch dog timer" protects against interrupted transmissions by aborting a control byte sequence if
the inter-byte time limit is exceeded.


Parameter Validation
The Site Master validates input parameters for each control byte sequence. If the input parameters are out
of range or invalid, the Site Master notifies the computer by sending Parameter Error Byte #224 (E0h). The
Site Master discards the received data and waits for the next control byte.




2                                                                          S113B, S114B, S331B, S332B PM
Entering Remote Mode
Send the Enter Remote Mode byte #69 (45h) to the Site Master to enter remote mode at the end of a sweep.
Send the Enter Remote Mode Immediately byte #70 (46h) to enter remote mode in the middle of a sweep.
The Site Master's serial port buffer is one byte wide. No internal buffer exists, so waiting is essential. If the
Site Master is not in remote, sending a second byte overwrites the original byte commanding it to go into re-
mote. If you send control byte #69, you must wait until the end of the sweep. If you send control byte #70, the
unit will enter remote mode as it receives the byte. Note that this means the data stored for the current
sweep may be incomplete.
Once you receive the response string from Site Master, you're in remote mode.


Exiting Remote Mode
Send the Exit Remote control byte #255 (FFh) to the Site Master. Site Master sends a response byte of
255 (FFh) then exits remote mode.


Lifetime of Changes to Site Master Operating Parameters
System parameters changed during remote mode remain changed for normal operation. They are not auto-
matically written to the non-volatile EEPROM. Turning off power erases the changed settings.
If you want the changes saved, you must save the change to one of the setup memories. Use either the
run-time setup (location 0, which holds the power-on defaults) or one of the ten saved setups. See control
byte #18 (12h) for details.


Write Cycle Limitation of EEPROM
The EEPROM, used to store setups, has a guaranteed lifetime of at least 100,000 write cycles and an unlim-
ited number of read cycles. The write cycle limitation is for a specific location. For example, you can store
setup #1 100,000 times and setup #2 100,000 times, etc.
It is for this reason we do not automatically store the changed system parameters to EEPROM. Instead, we
provide a means of changing the operating parameters independent of this limitation.
Be aware of the EEPROM write cycle limitation when programming the Site Master. Keep the number of
write cycles to a minimum.




S113B, S114B, S331B, S332B PM                                                                                  3
                                     Functional Description


INITIALIZE

Enter remote mode - see control byte #69 (45h)

       Site Master must be in remote mode for further control byte command sequences to be accepted.
       Remember to wait until Site Master responds before issuing further control byte sequences.
Enter remote mode immediately - see control byte #70 (46h)


CONFIGURE SYSTEM

Fixed CW mode control - see control byte #1 (01h)

       Site Master "dithers" the CW signal for improved immunity to interfering signals. Turning on the
       fixed CW signal bit "cleans up" the output spectrum by disabling the dithering. The drawback is
       decreased immunity to interfering signals.
LCD backlight -see control byte #1 (01h)

       LCD backlight helps in adverse lighting conditions.
Watch dog timer - see control byte #12 (0Ch)

       Used to guard against interrupted data transfers to the Site Master.
Serial port echo - see control byte #10 (0Ah)

       Use this to synchronize the Site Master sweep and computer requests for current trace data. This
       feature is especially useful for synchronizing distance to fault sweeps.
       Site Master goes into single sweep mode. It sends the sweep complete byte #192 (C0h) after the com-
       pletion of each sweep. You must send the sweep trigger byte #48(30h) for the Site Master to perform
       another sweep. Retrieve the trace data using control byte #17 (11h).
Single sweep - see control byte #11 (0Bh)

       This puts the Site Master into single sweep mode. Normally the Site Master sweeps continually.
       Once in the single sweep mode, send the sweep trigger byte #48 (30h) to cause Site Master to per-
       form another sweep.
Printer type - see control byte #30 (1Eh)

       Printer type is selectable.
Trace Math - see control byte #32 (20h)

Automatically Save Runtime Setup - see control byte #64 (40h)




4                                                                       S113B, S114B, S331B, S332B PM
CONFIGURE MEASUREMENT

Measurement mode select - see control byte #3 (03h)

       Measurement mode selectable as Site Master in frequency or distance domain, power monitor mode,
       or spectrum analyzer mode.
Measurement unit - see control byte #1 (01h)

       The Site Master uses Metric or English units.
Calibration on/off - see control byte #1 (01h)

       Calibration corrects for test port uniqueness. It should always be used to ensure accurate readings.


CONFIGURE POWER MONITOR

The power monitor is available as an option.
Get Power levels - see control byte #39 (27h)

       Querying the Site Master gives the raw and corrected power levels plus information about the config-
       uration.
Power Monitor Units of measure - see control byte #40 (28h)

Relative Mode - see control byte #41 (29h)

Offset Mode - see control byte #42 (2Ah)

Zero Mode - see control byte #43 (2Bh)


CONFIGURE FREQUENCY

Start frequency - see control byte #2 (02h)

Stop frequency - see control byte #2 (02h)

Data points - see control byte #14 (0Eh)


CONFIGURE DISTANCE

Start distance - see control byte #7 (07h)

Stop distance - see control byte #7 (07h)

Relative propagation velocity - see control byte #7 (07h)

       This affects the distance to fault calculatons.
Cable loss - see control byte #7 (07h)

       This affects the measurement magnitude.



S113B, S114B, S331B, S332B PM                                                                                 5
DTF Windowing - see control byte #31 (1Fh)

       Allows selection of a window to lessen the side lobes.


CONFIGURE DISPLAY

Display mode - see control byte #3 (03h)

       Selectable among SWR, return loss, cable insertion loss, transmission loss, transmission gain, power
       monitor, and spectrum analyzer.
Scale start - see control byte #4 (04h)

Scale stop - see control byte #4 (04h)


CONFIGURE MARKERS

Marker (on/off) - see control byte #5 (05h)

Marker Delta Status (on/off) - see control byte #5 (05h)

Marker value - see control byte #5 (05h)


CONFIGURE LIMITS

Limit (on/off) - see control byte #6 (06h)

Limit beep (on/off) - see control byte #6 (06h)

Limit value - see control byte #6 (06h)


UTILITY

ASCII text strings, which are stored with each sweep, help identify the stored traces. Use these control bytes
to modify the strings before storing the trace.
Time set - see control byte #8 (08h)

Date set - see control byte #8 (08h)

Reference number set - see control byte #9 (09h)


ACTION

Trigger sweep - see control byte #48 (30h)

       Both serial port echo and single sweep modes use this byte to trigger a Site Master sweep. If the Site
       Master is not in one of those modes, this byte is ignored.
Save system setup - see control byte #18 (12h)



6                                                                        S113B, S114B, S331B, S332B PM
Recall system setup - see control byte #19 (13h)

Save sweep - see control byte #16 (10h)

Query trace names - see control byte #24 (18h)

Delete sweep trace - see control byte #25 (19h)


DATA

Recall sweep - see control byte #17 (11h)

       Use this control byte to obtain stored traces or the current measurement data from the Site Master.
       All traces and modes use the same trace data format.


CALIBRATION

Calibrate - see control byte #13 (0Dh)

       Computer sequencing of the open-short-load calibration is possible.


CONFIGURE SPECTRUM ANALYZER

Set Frequency - see control byte #99 (63h)

Set Center Frequency and Span - see control byte #100 (64h)

Set Scale - see control byte #101 (65h)

Set Markers - see control byte #102 (66h)

Set Limits - see control byte #103 (67h)

Measure OCC BW % of Power - see control byte #96 (60h)

Measure OCC BW dB Down - see control byte #99 (63h)

Set Max Hold - see control byte #105 (69h)

Set Resolution BW Frequency - see control byte #106 (6Ah)

Set Video BW Frequency - see control byte #107 (6Bh)

Set Sweep Mode - see control byte #108 (6Ch)

Set Marker to Peak - see control byte #109 (6Dh)

Set Marker to Center - see control byte #110 (6Eh)




S113B, S114B, S331B, S332B PM                                                                                7
CONFIGURE CUSTOM CABLE TABLE

Write to Table - see control byte #80 (50h)

Read from Table - see control byte #81 (51h)


STATUS

Query system status - see control byte #20 (14h)

        The system status contains virtually all information about the current Site Master system state. It is
        a powerful and complete "snapshot" of the system.
        Some control bytes affect multiple system parameters (such as distance to fault parameters). Use the
        response from this control byte for preserving any of the system parameters you wish to remain un-
        changed.
Self-test Trigger - see control byte #21 (15h)

        Upon receiving this control byte the Site Master runs self-test. After completion of the self-test, the
        Site Master sends the results via the serial port.
Read failure counters -see control byte #22 (16h)

        In order to provide absolute confidence about the validity of a measurement, internal counters
        monitor the analog integrator and phase lock loops for error conditions. If an error occurs, it is
        logged. These counters perform the same function as the icons on the front panel.
Clear failure counters - see control byte #23 (17h)

        Use this control byte to reset the failure counters.


CLOSE

Exit remote mode - see control byte #255 (FFh)

        Ends communication with the Site Master. "Exiting remote mode" returns the Site Master to normal
        operation.




8                                                                          S113B, S114B, S331B, S332B PM
                                          Control Byte Summary
                                           Control Byte Summary (1 of 3)

                                                                                                                 Watch-
 Control
                        Name                                             Description                              dog
 Byte #
                                                                                                                 Timer

  1 (01h)   Setup system                   System status flags and switches                                       Yes

  2 (02h)   Set frequency                  Sets Site Master frequency range                                       Yes

  3 (03h)   Select measurement mode        Measurement domain and measurement mode                                Yes

  4 (04h)   Set scale                      Sets Site Master scale values                                          Yes

  5 (05h)   Set marker                     Set position and on/off status of a Site Master marker                 Yes

  6 (06h)   Set limit                      Set position and on/off status of a Site Master limit line             Yes

  7 (07h)   Set DTF parameters             Sets Distance to Fault parameters                                      Yes

  8 (08h)   Set time/date                  Sets Time and date for a sweep trace                                   Yes

  9 (09h)   Set reference number           Sets reference number for a sweep trace                                Yes

 10 (0Ah)   Serial port echo On/Off        Allows synchronization of the Site Master and request from computer    Yes
                                           for sweep trace

 11 (0Bh)   Single sweep On/Off            Enable or disable the single sweep mode                                Yes

 12 (0Ch)   Watch-dog timer On/Off         Enable or disable the watch-dog timer                                  ---

 13 (0Dh)   Sequence Site Master           Trigger a calibration step                                             Yes
            calibration

 14 (0Eh)   Set Site Master data points    Sets number of measurement data points for Site Master mode            Yes

 16 (10h)   Store sweep trace              Save current trace data to EEPROM                                      Yes

 17 (11h)   Recall sweep trace             Site Master sends sweep data                                           Yes

 18 (12h)   Save system setup              Save system setup parameters to EEPROM                                 Yes

 19 (13h)   Recall system setup            Recall system setup parameters from EEPROM                             Yes

 20 (14h)   Query system status            Get the current system settings                                        ---

 21 (15h)   Trigger self-test              Trigger a self test                                                    ---

 22 (16h)   Read fail counters             Same as icons on front panel                                           ---

 23 (17h)   Clear fail counters            Reset the fail counters                                                ---

 24 (18h)   Query trace names              Returns list of all saved traces                                       ---

 25 (19h)   Delete sweep trace             Delete single or all stored sweep traces                               Yes

 30 (1Eh)   Select printer type            Select printer type                                                    Yes

 31 (1Fh)   Select DTF Windowing           Select DTF Windowing Methods                                           Yes

 32 (20h)   Set Trace Math                 Select Trace Math operation                                            ---




S113B, S114B, S331B, S332B PM                                                                                           9
                                        Control Byte Summary (2 of 3)

                                                                                                              Watch-
 Control
                     Name                                             Description                              dog
 Byte #
                                                                                                              Timer

 39 (27h)   Query Power Level           Return Power Level at Detector Port                                    ---

 40 (28h)   Set Power Unit              Set Power Monitor displaying unit                                      Yes

 41 (29h)   Set Relative Mode           Enable or disable Power Monitor Relative Mode                          Yes

 42 (2Ah)   Set Offset Mode             Set/Reset Power Level offset                                           Yes

 43 (2Bh)   Set Zero Mode               Set/Reset Power Level zeroing mode                                     Yes

 48 (30h)   Trigger sweep               Starts the next sweep                                                  ---

 50 (32h)   Check Battery Status        Return smart battery status                                            ---

 64 (40h)   Auto Save Runtime Setup     Automatically save the runtime setup when exiting remote mode          ---

 69 (45h)   Enter remote mode           Enters remote mode at the end of the sweep and returns model number    ---
                                        and firmware version

 70 (46h)   Enter remote mode           Enters remote mode immediately and returns model number and firm-      ---
            immediately                 ware version

 80 (50h)   Write Custom Cable          Input Custom Cable data to Site Master                                 ---

 81 (51h)   Recall Custom Cable         Recall Custom Cable data from Site Master                              ---

 96 (60h)   Measure OCC BW % of         Measure OCC BW with % of Power method                                  Yes
            Power

 97 (61h)   Measure OCC BW dB           Measure OCC BW with dB down method                                     Yes
            Down

 99 (63h)   Set Spectrum Analyzer       Sets the Spectrum Analyzer frequency range                             Yes
            Frequency

100 (64h)   Set Spectrum Analyzer       Sets the Spectrum Analyzer center frequency and frequency span         Yes
            Center Freq./Span

101 (65h)   Set Spectrum Analyzer       Sets the Spectrum Analyzer scale value                                 Yes
            Scale

102 (66h)   Set Spectrum Analyzer       Set position and on/off status of a Spectrum Analyzer marker           Yes
            Marker

103 (67h)   Set Spectrum Analyzer       Set position and on/off status of Spectrum Analyzer limit line         Yes
            Limit

105 (69h)   Set Spectrum Analyzer       Enable or disable the Spectrum Analyzer Max Hold                       Yes
            Max Hold

106 (6Ah)   Set Spectrum Analyzer       Sets the Spectrum Analyzer resolution BW frequency                     Yes
            Resolution Bandwidth Freq

107 (6Bh)   Set Spectrum Analyzer       Sets the Spectrum Analyzer video BW frequency                          Yes
            Video Bandwidth Freq




10                                                                               S113B, S114B, S331B, S332B PM
                                     Control Byte Summary (3 of 3)

                                                                                                         Watch-
 Control
                      Name                                        Description                             dog
 Byte #
                                                                                                         Timer

 108 (6Ch)   Set Spectrum Analyzer   Sets the Spectrum Analyzer sweep mode                                Yes
             Sweep Mode

 109 (6Dh)   Set Spectrum Analyzer   Sets specified marker to peak value of the sweep                     Yes
             Marker to Peak

 110 (6Eh)   Set Spectrum Analyzer   Sets the center frequency equal to the frequency of the specified    Yes
             Marker to Center        marker

 255 (FFh)   Exit remote mode        End serial communications                                            ---




S113B, S114B, S331B, S332B PM                                                                                   11
                                   Control Byte Descriptions


Setup system - Control byte #1 (01h)
Description: Sets system status flags and switches. The Site Master acts on the entire byte. You must ac-
count for the state of each of the bits. For example, if you wanted to turn the LCD backlight on without dis-
turbing the other switches you would do the following: query the Site Master with control byte #20. You
would mask in the LCD status with the data from response byte #41 and send this to the Site Master after
control byte #1.
See control byte #20 (14h) response byte 159 for current Site Master configuration. It is important to set the
Metric/English flag to the proper value before sending distance information.


Bytes to Follow: 1 byte
       1) Status byte (bit data)
               (LSB) bit 0 : Fixed CW mode On/Off
                       bit 1 : Currently Unused
                       bit 2 : LCD Back Light On/Off
                       bit 3 : Measurement Unit Metric/English (0 = English, 1 = Metric)
                       bit 4 : Cal On/Off
                       bit 5-7 : Currently Unused

Site Master Returns:        255 (FFh) Operation Complete Byte
                            224 (E0h) Parameter Error : Invalid Printer Selection or unsuccessful attempt to
                            turn calibration on.
                            238 (EEh) Time-out Error

Notes: Cal can be turned on ONLY if a complete calibration has been performed at the current start and
       stop frequencies.

       Bit Value: 0 = Off
                  1 = On



Set Site Master frequency - Control byte #2 (02h)
Description: Sets the Site Master frequency range.
See control byte #20 (14h) response bytes 4-11 for current Site Master configuration.


Bytes to Follow: 8 bytes
       1) Start Frequency (highest byte)
       2) Start Frequency
       3) Start Frequency
       4) Start Frequency (lowest byte)
       5) Stop Frequency (highest byte)
       6) Stop Frequency
       7) Stop Frequency
       8) Stop Frequency (lowest byte)




12                                                                        S113B, S114B, S331B, S332B PM
Site Master Returns:      255 (FFh) Operation Complete Byte
                          224 (E0h) Parameter Error : Invalid frequency range
                          238 (EEh) Time-out Error

Notes: Start and stop frequencies are given in terms of 1 Hz steps.
       (e.g. 100.3 MHz would be sent as 100030000 = 1,000,300,000 Hz)



Select Measurement mode - Control Byte #3 (03h)
Description: Sets measurement domain and display graph type. You must have a valid calibration (current
frequencies and calibration frequencies being the same) to switch to the Distance domain.
See control byte #20 (14h) response byte 1 (01h) for current Site Master domain.


Bytes to Follow: 1 byte
        1) Measurement Mode
               00h = RL Frequency
               01h = SWR Frequency
               02h = Cable Loss Frequency
               10h = RL Distance
               11h = SWR Distance
               20h = Transmission Mode
               30h = Spectrum Analyzer Mode
               40h = Power Monitor Mode

Site Master Returns:      255 (FFh) Operation Complete Byte
                          224 (E0h) Parameter Error : Invalid domain or graph type or combination
                          238 (EEh) Time-out Error



Set Site Master scale - Control byte #4 (04h)
Description: Sets the Site Master graph boundaries.
The scale settings require knowing the graph type for proper configuration. For example, the scale start in
return loss is the top of the graph while scale start is the bottom of the graph for SWR. Think of the scale
start as being the smallest value on the graph.
See control byte #20 (14h) response bytes 12 to 19 for current Site Master scaling.


Bytes to Follow: 8 bytes
       1) Scale Top (highest byte)
       2) Scale Top
       3) Scale Top
       4) Scale Top (lowest byte)
       5) Scale Bottom (highest byte)
       6) Scale Bottom
       7) Scale Bottom
       8) Scale Bottom (lowest byte)




S113B, S114B, S331B, S332B PM                                                                                  13
Site Master Returns:      255 (FFh) Operation Complete Byte
                          224 (E0h) Parameter Error : Invalid scale range
                          238 (EEh) Time-out Error

Notes:
         Return Loss &    Scaling is in thousandths of a dB
         Cable Loss:      Maximum value sent is 54000 which represents 54.00 dB
                          Minimum value sent is 0 which represent 0.00 dB
                          Scale Start Value is the top of the graph
                          Scale Stop Value is the bottom of the graph

         SWR:             Scaling is in thousandths (of ratio)
                          Maximum value sent is 65535 which represents 65.53
                          Minimum value sent is 1000 which represents 1.00
                          Scale Start Value is the bottom of the graph
                          Scale Stop Value is the top of the graph

         Gain/Insertion   Scaling is in hundredths (of dB)
         Loss:            Maximum value sent is 22000 which represents 