Agilent
High Accuracy and
Fast RF Inductor Testing
Application Note 369-10




Agilent 4285A
Precision LCR Meters
Introduction                                                   Problems and the solutions offered
                                                               by the Agilent 4285A

RF inductors are employed in electronic equipment such         1. Measurement accuracy and measurement speed
as VCRs and intermediate frequency (IF) circuits for
                                                               Stable inductance (L) and quality factor (Q) low
AM/FM radios, mobile radios, pagers and code-less
                                                               inductance measurements are difficult to make, because
telephone, which require RF inductors to be smaller,
                                                               measuring low inductance values means measuring low
lower cost, and have higher reliability. This application
                                                               impedance, values. If a higher measurement frequency
note describes solutions offered by the Agilent 4285A
                                                               (MHz) is issued to increase the measured impedance,
precision LCR meter for realizing these requirements.
                                                               the LCR meter's measurement accuracy is worse,
Information for accurate and fast RF inductor testing,
                                                               so accurate measurement data cannot be obtained.
and for practical simple test systems are discussed.
                                                               The 4285A can perform reliable and fast measurements
                                                               with 0.5% inductance measurement accuracy even in the
                                                               MHz frequency range, and with a 30 ms measurement
                                                               speed. Formerly LCR meters could not measure
                                                               accurately enough for testing high Q inductors, and
                                                               Q meters required long measurement time and high
                                                               operator skill. Table 1 lists the 4285A's key specifications.

                                                               2. Measurement condition setup
                                                               There are many types of RF inductors, so the measurement
                                                               frequency range for an inductor is selected from
                                                               across a frequency range spanning kHz through MHz,
                                                               depending on the inductance value to be measured and
                                                               the application in which it will be used. The 4285A
                                                               covers a measurement frequency range of from 75 kHz
                                                               to 30 MHz, including frequencies defined in the MIL
                                                               C-15305D standard (79.6 kHz, 252 kHz, 796 kHz, ..25.5 MHz)

                                                               A constant current test signal gives the best inductance
                                                               measurement results, because inductive devices are
                                                               current mode devices and L and Q values depend on
                                                               current levels rather than voltage levels. Q meters do
                                                               not offer current test signals, and former LCR meters
                                                               used voltage test signals, not current test signals.

                                                               The 4285A offers both voltage and current test signal
                                                               level selections. Furthermore, its auto level control (ALC)
                                                               function allows constant test signal level measurements.




Table 1. 4285A key specifications


                                                  Agilent 4285A (standard)
Measurement method                                Auto balancing bridge
Measurement frequency                             75 kHz to 30 MHz with 100 Hz resolution
Basic accuracy (typical)            10 mH         L: 