Agilent AN 1305-4
Effective Electrolytic Capacitors Testing
Application Note




Agilent 4263B LCR Meter
Introduction                               (4) When a shorted device is measured,      Agilent 4263B LCR Meter Solution
With increased requirements for size       the internal circuit of a conventional      (1) Accurate Low Impedance Measurement
reduction and higher reliability design,   LCR meter is latched up, and it needs       The 4263B LCR Meter employs a
it is becoming necessary to evaluate       a long recovery time.                       4-terminal pair configuration and an
electrolytic capacitors employed in                                                    advanced low noise design so that it
electronic equipment. Production vol-      (5) Some inexpensive LCR meters do          makes accurate measurements up to
ume has been increasing for circuit        not range up to 100 kHz, and the            1 F. For example, when 10 mF is meas-
applications. Manufacturing and QA         equivalent series resistance of the         ured at 120 Hz, the 4263B's accuracy,
now have to improve their testing of       electrolytic capacitors need to be          0.57% can be achieved. Also, the meas-
electrolytic capacitors. Here are the      evaluated at 100 kHz.                       urement cable can be extended up to
solutions offered by the Agilent                                                       4 m while maintaining the measurement
4263B LCR Meter to meet these meas-        (6) Most LCR meters cannot detect           accuracy.
urement requirements.                      whether or not the contact between
                                           the fixture's electrodes and the device's   (2) High System Throughput
                                           electrode is good, thus decreasing the      The measurement speed of the 4263B
Current Problems in Electrolytic           reliability of a measurement.               LCR Meter at 100 Hz/120 Hz is 25 ms
Capacitor Evaluation                                                                   (Meas.Time: SHORT). Figure 1 shows
When making electrolytic capacitor         (7) There are no test fixtures for the      you the fluctuation of the 22 mF
measurements using conventional            terminals (for example, big screw           Aluminum Electrolytic Capacitor
LCR meters, there are quite a few          electrodes) of the big electrolytic         Measurement by SHORT mode, and
shortcomings.                              capacitors.                                 120 Hz of the 4263B.
(1) Impedance value of the electrolytic
capacitors is usually so low that an
accurate measurement is hard to
achieve. Some LCR meters cannot
measure above 20 mF due to a limited
measurement range.

(2) Electrolytic capacitors are cur-
rently tested at 100 Hz or 120 Hz. The
measurement speed at those frequen-
cies is quite slow, and higher through-
put cannot be achieved by automatic
measurement systems.

(3) When a charged capacitor is con-
nected to the measurement terminals
of the LCR meters, the circuit is easi-
ly damaged by the discharge energy.




                                           Figure 1. Fluctuation of the measurement value at SHORT




2
In addition, the 4263B has the follow-     (7) Various Test Fixtures
ing features which allow easy system       The Agilent 16089A Kelvin Clip Leads
integration, and improve measurement       have two large clips that can accept
throughput.                                device electrodes with a diameter of
                                           15 mm. Figure 3 shows a big capacitor