Service Manuals, User Guides, Schematic Diagrams or docs for : Keithley 197 Calibration setup

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
Preview not available!
Sorry, no preview image available for this document.
Possible reasons are : this is not a PDF document, it is password protected or the file is partially corrupted.
We are working to fix the issue.
You can still download the full file from the link below and it will be viewable.



>> Download Calibration setup documenatation <<

Text preview - extract from the document
How do I put the Model 197, 175, 485 into Cal Enable mode?

Depending on the firmware revision, there are three methods to put these instruments in the CAL ENABLE mode.

    Firmware A revision : Has an internal cal enable jumper. There is no indication on the display when cal mode has been enabled. CAL mode is enabled with the jumper to the left when looking at the instrument from the font.
    
	Firmware B revision: NO internal cal enable switch, performed digitally through sequence during power up. (see procedure below)
    
	Firmware C revision : Has external cal enable switch on rear.

Procedure for CAL ENABLE for Revision B firmware:

In Service Manual, delete paragraph 5.6.3 Calibration Jumper on page 5-4 and Figure 5-2 on page 5-5 and replace with the following:

5.6.3 Calibration Storage Enable

The Model must be in the calibration storage enable mode in order to realize storage of calibrations constants into the NVRAM. If the instrument is not placed in this mode, subsequent calibration will be lost when the instrument is turned off. Perform the following steps to place the unit in storage enable mode:

    Turn unit off
    While holding in the STO/CLR button, turn the instrument on using the ON/OFF switch.
    When the "CAL" message is displayed, release the STO/CLR button. This instrument will return to normal display mode, and the storing of calibration constants is now enabled.
    Allow 1 hour warm up before proceeding with instrument calibration.
    Press in the REL and dB pushbuttons simultaneously until the message "CAL" is displayed again. Release the buttons. The unit is now ready to be calibrated as indicated by the "C" annunciator on the display.

Note: Step 2 and 3 put the instrument in the CAL ENABLE mode, while step 5 puts the instrument in the CAL mode. 