Service Manuals, User Guides, Schematic Diagrams or docs for : Keithley 2400 calsw callog_test

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
Preview not available!
Sorry, no preview image available for this document.
Possible reasons are : this is not a PDF document, it is password protected or the file is partially corrupted.
We are working to fix the issue.
You can still download the full file from the link below and it will be viewable.



>> Download callog_test documenatation <<

Text preview - extract from the document
Keithley Model 2400 - detected, S/N 0788393, Version: C32   Oct  4 2010 14:20:11/A02  /H/H

Pre-cal K2400 Calibration data
0.2 DCV Range, SENS:DATA? = -1.255853E+00,-7.547438E-05,-1.255865E+00,+0.000000E+00

0.2 DCV Range, SOUR:DATA? = +3.027005E+05,+9.805352E+02,-3.026938E+05,+9.540547E+02

2.0 DCV Range, SENS:DATA? = -1.255818E+00,-7.629395E-05,-1.255898E+00,+0.000000E+00

2.0 DCV Range, SOUR:DATA? = +3.024413E+04,+2.160469E+02,-3.024334E+04,+2.133789E+02

20.0 DCV Range, SENS:DATA? = -1.255909E+01,-1.522064E-03,-1.255986E+01,+0.000000E+00

20.0 DCV Range, SOUR:DATA? = +3.024195E+03,+2.183359E+02,-3.024123E+03,+2.110820E+02

200.0 DCV Range, SENS:DATA? = -1.255817E+02,-1.463318E-02,-1.255900E+02,+0.000000E+00

200.0 DCV Range, SOUR:DATA? = +3.024431E+02,+2.181953E+02,-3.024344E+02,+2.112461E+02

1.0e-06 DCI Range, SENS:DATA? = -6.258903E-07,-7.969447E-11,-6.258224E-07,+0.000000E+00

1.0e-06 DCI Range, SENS:DATA? = +6.068293E+10,+2.121250E+02,-6.068410E+10,+2.170313E+02

1.0e-05 DCI Range, SENS:DATA? = -6.294048E-06,-4.620233E-10,-6.294329E-06,+0.000000E+00

1.0e-05 DCI Range, SENS:DATA? = +6.034448E+09,+2.100234E+02,-6.033632E+09,+2.191016E+02

1.0e-04 DCI Range, SENS:DATA? = -6.273385E-05,-4.249159E-09,-6.273701E-05,+0.000000E+00

1.0e-04 DCI Range, SENS:DATA? = +6.054307E+08,+2.098086E+02,-6.053455E+08,+2.192617E+02

1.0e-03 DCI Range, SENS:DATA? = -6.261325E-04,-4.319008E-08,-6.261630E-04,+0.000000E+00

1.0e-03 DCI Range, SENS:DATA? = +6.065957E+07,+2.098047E+02,-6.065106E+07,+2.192695E+02

1.0e-02 DCI Range, SENS:DATA? = -6.275110E-03,-4.488975E-07,-6.275468E-03,+0.000000E+00

1.0e-02 DCI Range, SENS:DATA? = +6.052646E+06,+2.099297E+02,-6.051750E+06,+2.191328E+02

1.0e-01 DCI Range, SENS:DATA? = -6.279009E-02,-4.447997E-06,-6.279366E-02,+0.000000E+00

1.0e-01 DCI Range, SENS:DATA? = +6.048887E+05,+2.099453E+02,-6.047995E+05,+2.191445E+02

1.0e+00 DCI Range, SENS:DATA? = -6.287083E+00,-4.602075E-04,-6.287335E+00,+0.000000E+00

1.0e+00 DCI Range, SENS:DATA? = +6.044879E+04,+8.966484E+02,-6.044157E+04,+9.857813E+02

HP 3458A detected
 TEMP? = 37.5 C
DCV Calibration step -0.2 mV , measured : -2.000011150E-01 VDC, deviation 5.575 ppm
DCV Calibration step  0.2 mV , measured : 1.999950137E-01 VDC, deviation -24.932 ppm
DCV Calibration step    -2 V , measured : -2.000001226E+00 VDC, deviation 0.613 ppm
DCV Calibration step     2 V , measured : 1.999990515E+00 VDC, deviation -4.743 ppm
DCV Calibration step   -20 V , measured : -1.999985933E+01 VDC, deviation -7.033 ppm
DCV Calibration step    20 V , measured : 1.999995167E+01 VDC, deviation -2.416 ppm
DCV Calibration step  -200 V , measured : -1.999999936E+02 VDC, deviation -0.032 ppm
DCV Calibration step   200 V , measured : 1.999976683E+02 VDC, deviation -11.659 ppm
DCI Calibration step -1 uA , measured : -9.999932723E-07 ADC, deviation -6.728 ppm
DCI Calibration step  1 uA , measured : 9.999700048E-07 ADC, deviation -29.995 ppm
DCI Calibration step  -10 uA , measured : -9.999961705E-06 ADC, deviation -3.829 ppm
DCI Calibration step  10 uA , measured : 9.999932024E-06 ADC, deviation -6.798 ppm
DCI Calibration step -100 uA , measured : -9.999879347E-05 ADC, deviation -12.065 ppm
DCI Calibration step  100 uA , measured : 9.999867260E-05 ADC, deviation -13.274 ppm
DCI Calibration step -1 mA , measured : -9.999954493E-04 ADC, deviation -4.551 ppm
DCI Calibration step  1 mA , measured : 9.999928871E-04 ADC, deviation -7.113 ppm
DCI Calibration step  -10 mA , measured : -9.999891116E-03 ADC, deviation -10.888 ppm
DCI Calibration step  10 mA , measured : 9.999946257E-03 ADC, deviation -5.374 ppm
DCI Calibration step -100 mA , measured : -1.000001955E-01 ADC, deviation 1.955 ppm
DCI Calibration step  100 mA , measured : 9.999891230E-02 ADC, deviation -10.877 ppm
DCI Calibration step  -1 A , measured : -9.999968151E-01 ADC, deviation -3.185 ppm
DCI Calibration step  1 A , measured : 9.999915013E-01 ADC, deviation -8.499 ppm