Text preview - extract from the document
Why does my 2430 fail to Source High Energy Pulses using the Pulse Mode
on the 10A Range?

In the 10A range, the duty cycle is limited to 8% with a maximum pulse width of
2.5ms.

To achieve a duty cycle of 8% with the maximum allowable pulse width of 2.5
ms, the Pulse OFF Time should be set for no less than 28.75ms.

Duty cycle      =       Pulse On Time___________
                     Pulse On time + Pulse Off Time

Pulse Off time =     (2.5ms / 0.08) 