Service Manuals, User Guides, Schematic Diagrams or docs for : Keithley 2000 calsw calk2000_ver2

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
Preview not available!
Sorry, no preview image available for this document.
Possible reasons are : this is not a PDF document, it is password protected or the file is partially corrupted.
We are working to fix the issue.
You can still download the full file from the link below and it will be viewable.



>> Download calk2000_ver2 documenatation <<

Text preview - extract from the document
Keithley Model 2000 - detected, S/N ???????, Version: A20  /A02  

Environment temperature = 31.87 C, humidity = 35.60 RH, pressure = 1005.5598 hPa
HP 3458A detected
 TEMP? = 38.1 C
 Ambient temp = +24.7C
HP 3458A DCV -7.834272296E-07 VDC, K2000 DUT 3.078535757E-06 VDC, Delta -1254480.471 ppm
MFC error = 1238.234 ppm 
HP 3458A DCV 9.999855743E-03 VDC, K2000 DUT 1.000182233E-02 VDC, Delta -196.623 ppm
MFC error = 60.005 ppm 
HP 3458A DCV 1.000006746E-01 VDC, K2000 DUT 1.000024920E-01 VDC, Delta -18.174 ppm
MFC error = -17.032 ppm 
HP 3458A DCV 9.999989351E-01 VDC, K2000 DUT 1.000001607E+00 VDC, Delta -2.672 ppm
MFC error = -4.894 ppm 
HP 3458A DCV 9.999971088E+00 VDC, K2000 DUT 1.000000503E+01 VDC, Delta -3.394 ppm
MFC error = 12.827 ppm 
HP 3458A DCV 1.000000769E+02 VDC, K2000 DUT 9.999921813E+01 VDC, Delta 8.588 ppm
MFC error = 21.998 ppm 
HP 3458A DCV 1.000003027E+03 VDC, K2000 DUT 1.000014860E+03 VDC, Delta -11.833 ppm
MFC error = 142.410 ppm 
Measured REF DCI 5 : 9.999802284E-05 ADC, stored DCI = 0.099986
MFC error = 4.031 ppm 
Measured REF DCI 5 : 9.999978065E-04 ADC, stored DCI = 0.999996
MFC error = 9.888 ppm 
Measured REF DCI 5 : 9.999997661E-03 ADC, stored DCI = 9.999901
MFC error = -12.991 ppm 
Measured REF DCI 5 : 9.999987479E-02 ADC, stored DCI = 100.001299
MFC error = -78.172 ppm 
Measured REF DCI 5 : 9.999943627E-01 ADC, stored DCI = 1.000078
MFC error = 19.204 ppm 
MFC error = -310.981 ppm 
MFC error = -312.800 ppm 
MFC error = -57.848 ppm 
MFC error = -20.603 ppm 
MFC error = 78.645 ppm 
MFC error = 1045.179 ppm 
Calibrations count = 3

Calibration last date = 2016,07,27

:CAL:PROT:CODE 'KI002000'
Error status =  -113,"Undefined header"

Error status =  0,"No error"

Error status =  0,"No error"

CAL MFC error = -4.150 ppm 
:CAL:PROT:DC:STEP3 9.999978
Error status =  0,"No error"

CAL MFC error = -18.864 ppm 
:CAL:PROT:DC:STEP4 -9.999971
Error status =  0,"No error"

CAL MFC error = 4.726 ppm 
:CAL:PROT:DC:STEP5 100.001143
Error status =  404,"100 vdc full scale error"