Service Manuals, User Guides, Schematic Diagrams or docs for : Keithley 3706 SPEC-3723

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
SPEC-3723


>> Download SPEC-3723 documenatation <<

Text preview - extract from the document
                                                                                                              Model 3723
Keithley Instruments, Inc.                             Dual 1x30, High Speed Multiplexer Specifications
28775 Aurora Road
Cleveland, Ohio 44139                                       60 differential channels, long life reed relays
1-888-KEITHLEY
www.keithley.com

MULTIPLEXER CONFIGURATION: Two independent 1x 30 2-pole Multiplexers. Banks can be connected together
via relay creating a single 1 x 60 Multiplexer. Banks can be isolated from the backplane by relays. Card can be
configured for 1,2 and 4 wire.
CONTACT CONFIGURATION: 2 pole form A.
CONNECTOR TYPE: Two 78 pin male D-shells.
Model 3723-ST screw terminal option:
 #22AWG typical wire size with .062 inch O.D 124 conductors maximum. 16 AWG maximum wire size with .092
inch O.D. 36 conductor per card maximum.
MAXIMUM SIGNAL LEVEL: 200V DC or RMS, 1A switched (1.25A carry), 15W.
COMMON MODE VOLTAGE: 300V DC or RMS between any terminal and chassis.
VOLT-HERTZ LIMIT : 8x107
CONTACT LIFE: REED > 1x109 operations, No Load
                              1x107 operation @100V, 10mA
                     EMR > 1x108 operations @5V,10mA
                             1x105 operations @ max signal level

                                               Dual 1x301                                   Single 1x601, 2

    Channel Resistance
                                               <1.5                                          <2.0
    (end of contact life)