Series 3700 Release Notes V1.41


>> Download Series 3700 Release Notes V1.41 documenatation <<

Text preview - extract from the document
                                                                        Series 3700 System Switch/Multimeter
Keithley Instruments, Inc.
28775 Aurora Road
                                                                                      Firmware Release Notes
Cleveland, Ohio 44139-1891
1-888-KEITHLEY
http://www.keithley.com



Contents
        General Information ............................................................................................................................... 3
            Supported models ........................................................................................................................... 3
            Installation instructions.................................................................................................................... 3
            Upgrade considerations for Models 3706, 3706-S, 3706-NFP, 3706-SNFP................................ 3
        Version 1.41 Release............................................................................................................................. 4
            Overview .......................................................................................................................................... 4
            Version 1.41 installation notes........................................................................................................ 4
                 Upgrading from firmware version 1.31a or any earlier version .............................................. 4
                 Upgrading from firmware version 1.32a .................................................................................. 5
            Critical fixes ..................................................................................................................................... 6
            Enhancements................................................................................................................................. 7
            Noncritical fixes ............................................................................................................................... 8
            Known issues ................................................................................................................................ 13
        Version 1.40 Release........................................................................................................................... 14
            Overview ........................................................................................................................................ 14
            Version 1.40 installation notes...................................................................................................... 14
                 Upgrading from firmware version 1.31a or any earlier version ............................................ 14
                 Upgrading from firmware version 1.32a ................................................................................ 15
            Critical fixes ................................................................................................................................... 16
            Enhancements............................................................................................................................... 18
            Noncritical fixes ............................................................................................................................. 24
            Known issues ................................................................................................................................ 28
        Version 1.32a Release......................................................................................................................... 30
            Overview ........................................................................................................................................ 30
            Version 1.32a installation notes.................................................................................................... 30
            Critical fixes ................................................................................................................................... 31
            Enhancements............................................................................................................................... 31
            Noncritical fixes ............................................................................................................................. 31
            Known issues ................................................................................................................................ 32
        Version 1.31a Release......................................................................................................................... 33
            Overview ........................................................................................................................................ 33
            Critical fixes ................................................................................................................................... 33
            Enhancements............................................................................................................................... 34
            Noncritical fixes ............................................................................................................................. 34
            Known issues ................................................................................................................................ 34
        Version 1.30j Release.......................................................................................................................... 35
            Overview ........................................................................................................................................ 35
            Critical fixes ................................................................................................................................... 35
            Noncritical fixes ............................................................................................................................. 36
            Known issues ................................................................................................................................ 37
        Version 1.24b Release......................................................................................................................... 38
            Overview ........................................................................................................................................ 38
            Critical fixes ................................................................................................................................... 38
            Enhancements............................................................................................................................... 38
            Noncritical fixes ............................................................................................................................. 38
            Known issues ................................................................................................................................ 38
        Version 1.22d Release......................................................................................................................... 39
            Overview ........................................................................................................................................ 39


3700S Release Notes V1.40 / May 2010                                                                                                                                 1
Series 3700 Firmware Release Notes                                                                                                                            Contents


           Critical fixes ................................................................................................................................... 39
           Enhancements............................................................................................................................... 39
           Noncritical fixes ............................................................................................................................. 39
           Known issues ................................................................................................................................ 39
       Version 1.22c Release......................................................................................................................... 40
           Overview ........................................................................................................................................ 40
           Critical fixes ................................................................................................................................... 40
           Enhancements............................................................................................................................... 41
           Noncritical fixes ............................................................................................................................. 41
           Known issues ................................................................................................................................ 41
       Version 1.21 Release........................................................................................................................... 42
           Overview ........................................................................................................................................ 42
           Critical fixes ................................................................................................................................... 42
           Enhancements............................................................................................................................... 43
           Noncritical fixes ............................................................................................................................. 44
           Known issues ................................................................................................................................ 44
       Version 1.20 Release........................................................................................................................... 45
           Overview ........................................................................................................................................ 45
           Critical fixes ................................................................................................................................... 45
           Enhancements............................................................................................................................... 47
           Noncritical fixes ............................................................................................................................. 47
           Known issues ................................................................................................................................ 48
       Version 1.10 Release........................................................................................................................... 49
           Overview ........................................................................................................................................ 49
           Critical fixes ................................................................................................................................... 49
           Enhancements............................................................................................................................... 50
           Noncritical fixes ............................................................................................................................. 52
           Known issues ................................................................................................................................ 54
       Version 1.03 Release........................................................................................................................... 55
           Overview ........................................................................................................................................ 55
           Critical fixes ................................................................................................................................... 55
           Enhancements............................................................................................................................... 56
           Noncritical fixes ............................................................................................................................. 58
       Version 1.00 Release........................................................................................................................... 59
           Overview ........................................................................................................................................ 59




2                                                                                                             3700S Release Notes V1.40 / May 2010
Series 3700 Firmware Release Notes                                         General Information - Supported models



General Information
Supported models
This firmware is used on the following Keithley Instruments products:

        Models 3706, 3706-S, 3706-NFP, 3706-SNFP, and all Series 3700 cards

Installation instructions
        NOTE        Upgrading to firmware version 1.40 or greater may require special installation steps that are
                    not required for other firmware upgrades. Carefully read the Version 1.41 installation notes
                    below before upgrading your Series 3700.


For other releases, firmware installation instructions are given in the "Upgrade Procedure using the USB Flash
Drive" section in the Series 3700 User's Manual (document number: 3700S-900-01). This manual is available the
Keithley Instruments website at http://www.keithley.com/support. Alternatively, you can get upgrade assistance
from Keithley by calling your local Keithley support office.

Upgrade considerations for Models 3706, 3706-S, 3706-NFP, 3706-SNFP

This release updates the firmware for the Series 3706 mainframe and the optional digital multimeter (DMM).

The following table lists the considerations that should be made when deciding whether or not to upgrade to this
version from any previous version.


                                             Upgrade considerations
         Existing           Recalibration      Requalification    Backward           Should you upgrade?
         firmware           required?          suggested?         compatibility
         version                                                  concerns?
         1.00               No                 Yes                No                 Yes
         1.03               No                 Yes                No                 Yes
         1.10               No                 Yes                No                 Yes
         1.20               No                 Yes                No                 Yes
         1.21               No                 Yes                No
         1.22c              No                 No                 No                 Yes, if using a Model
                                                                                     3740 card
         1.22d              No                 No                 No                 Yes, if using a Model
                                                                                     3740 card
         1.24b              No                 No                 No                 Yes, for web-based
                                                                                     charting feature; Model
                                                                                     3731 card supported
         1.30j              No                 No                 No                 Yes, to avoid security
                                                                                     warning
         1.31a              No                 No                 No                 Yes, if using Ethernet
         1.32a              No                 No                 No                 Yes, if using a Model
                                                                                     3732 card
         1.40               No                 No                 No                 Yes




3700S Release Notes V1.40 / May 2010                                                                                3
Version 1.41 Release - Overview                                                 Series 3700 Firmware Release Notes



Version 1.41 Release
Overview
The Series 3700 version 1.41 firmware is an enhancement and problem fix release. Some problem corrections
may impact compatibility, please review the corrections prior to installation.

NOTE        Upgrading to firmware version 1.41 from a version below 1.40 requires special installation steps that
            are not required for other firmware upgrades. Carefully read the Version 1.41 installation notes below
            before upgrading your Series 3700.

The following table lists the exact version numbers of the available Series 3700 components when the firmware
upgrade is complete.

                         Model                                                      Version
                         Series 3700 Mainframe                                      1.41c
                         Series 3700 High Performance DMM                           1.41b
                         3720 Dual 1x30 Multiplexer                                 1.40h
                         3721 Dual 1x20 Multiplexer                                 1.40h
                         3722 Dual 1x48 High Density Multiplexer                    1.40h
                         3723 Dual 1x30 High Speed Reed Relay Multiplexer           1.40j
                         3724 Dual 1x30 FET Multiplexer                             1.40j
                         3730 Dual 6x16 High Density Matrix                         1.40h
                         3731 Dual 6x16 High Speed Reed Matrix                      1.40j
                         3732 Quad 4x28 High Speed Reed Matrix                      1.40m
                         3740 32 Channel Isolated Switch                            1.40h
                         3750 Multifunction I/O Card                                1.40j


Version 1.41 installation notes
Upgrading to firmware version 1.41 requires special installation steps that are not required for other firmware
upgrades. Carefully perform the following procedures to successfully upgrade to version 1.41.

First, determine your instrument's current firmware revision. With the instrument powered on, press the SLOT
key. The instrument's current firmware version is displayed. Based on the current firmware revision, follow the
instructions in the appropriate section below.

            Upgrading from firmware version 1.31a or any earlier version

                Download the nk_5001607.nb0 and main_p37xx_0141c.CAB files from the
                Keithley Instruments website (www.keithley.com).
                To perform the upgrade:
                1. Place the nk_5001607.nb0 and main_p37xx_0141c.CAB files in the root directory of a
                   blank USB flash drive (make sure drive size is large enough for the size of the upgrade files).
                2. Ensure that the Series 3700 mainframe power is turned off, and then connect the flash drive
                   to the Series 3700 and turn on power to initiate the upgrade process.

                NOTE       The upgrade process will take three to five minutes and the instrument will reboot
                           several times during the process. Do not disturb the instrument until the normal front
                           panel display has been present for at least 30 seconds.




4                                                                            3700S Release Notes V1.40 / May 2010
Series 3700 Firmware Release Notes                           Version 1.41 Release - Version 1.41 installation notes


              3. On the front panel, press the MENU key, select LAN, select APPLY, and then press ENTER
                 to resume normal Ethernet connectivity following the upgrade.
              4. Disconnect the USB flash drive from the Series 3700 mainframe.

              NOTE      Alternatively, you can have Keithley Instruments assist you with the upgrades by calling
                        your local Keithley support office.


           Upgrading from firmware version 1.32a

              Download the main_p37xx_0141c.CAB files from the Keithley Instruments website
              (www.keithley.com).

              NOTE      The following steps force the instrument to upgrade its firmware twice. This is
                        necessary to ensure everything is upgraded properly. Follow the steps below carefully.

              To perform the upgrade:
              1. Copy the main_p37xx_0141c.CAB file to a blank USB flash drive (make sure drive size is
                 large enough for the size of the upgrade file).

              NOTE      Verify that the USB flash drive is blank.

              2. Turn on the Series 3700 power.
              3. Install the USB flash drive in the front panel connector.
              4. On the front panel, press the MENU key.

              NOTE      If your model does not have a front panel, upgrade over the bus using the
                        upgrade.unit()command.

              5. Turn the navigation wheel to scroll to UPGRADE, and then press the navigation wheel.
              6. The question UPGRADE UNIT? displays. Select Yes, and then press the navigation wheel.
                 The Series 3700 upgrade status displays on the front panel, including the percentage
                 completed. When the file has been unpacked, the upgrade status is displayed as the upgrade
                 progresses (first cards installed in the slots are upgraded (including the DMM, if installed),
                 and then the Series 3700 mainframe).

              NOTE      For models without a front panel, the LAN status and clock status LEDs blink in unison
                        during the upgrade process.

              7. Wait for the Series 3700 to reboot automatically. When the reboot is complete, press the
                 MENU key on the front panel.

              NOTE      If your model does not have a front panel, upgrade over the bus using the
                        upgrade.previous() command.

              8. Turn the navigation wheel to scroll to UPGRADE and press the navigation wheel.
              9. The question UPGRADE UNIT? is displayed. Select Previous, and then press the navigation
                 wheel (note that in this step you are selecting "Previous" instead of "Yes," as was done in
                 step 6). The Series 3700 upgrade status displays on the front panel, including the percentage
                 completed. When the file has been unpacked, the upgrade status is displayed as the upgrade
                 progresses (first cards installed in the slots are upgraded (including the DMM, if installed),
                 and then the Series 3700 mainframe).
                 The Series 3700 reboots automatically when the upgrade is complete.




3700S Release Notes V1.40 / May 2010                                                                               5
Version 1.41 Release - Critical fixes                                         Series 3700 Firmware Release Notes


Critical fixes

PR42627 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           The instrument may generate a fatal error if another node on the TSP-Link network performs a
           tsplink.reset() while the instrument is powering up.

           Resolution:
           This issue has been corrected.

PR42632 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           Selecting a script to load from a sub-directory on the thumb drive may cause the unit to lock up or
           reset.

           Resolution:
           Scripts can now be loaded from any thumb drive directory level without issue.

PR41811 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           In version 1.40, single channel scans with high scan counts would sometimes yield an out of power
           error.

           Resolution:
           This problem has been resolved.

PR41968 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           The node parameter being returned by errorqueue.next() is always nil.

           Resolution:
           The node parameter is correctly returned as described in the manual.

PR42189 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           Some thumb drives are not recognized quick enough to perform an OS upgrade.




6                                                                          3700S Release Notes V1.40 / May 2010
Series 3700 Firmware Release Notes                                          Version 1.41 Release - Enhancements


          Resolution:
          If available, try another thumb drive.

PR42473 Models affected:
          3706, 3706-NFP

          Symptom:
          External reference temperature may not function properly.

          Resolution:
          In 1.40, external temperature reference only worked on slot 1. Slot 2-6 are now available. Four wire
          external temperature reference is now available on all slots.

PR42528 Models affected:
          3706, 3706-NFP

          Symptom:
          Occasionally when using the tspnet.tsp.rbtablecopy() ICL, the error "2413, TSPnet invalid
          reading buffer table" would be generated when asking for the some of the synchronous tables like
          relativetimestamps but, not for readings.

          Resolution:
          The issue has been resolved.

PR42680 Models affected:
          3706, 3706-NFP

          Symptom:
          The reading buffer channel names are not as expected according to the manual guidelines..

          Resolution:
          Channel patterns only use the first 7 characters in the channel names if the pattern name exceeds 7
          characters. Also, patterns now have a + at the end of them to indicate multiple items closed.

PR42751 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          In previous versions, TSPnet would not connect to some telnet servers.

          Resolution:
          TSPnet now connects to all Telnet servers.



Enhancements

PR42773 Models affected:



3700S Release Notes V1.40 / May 2010                                                                             7
Version 1.41 Release - Noncritical fixes                                       Series 3700 Firmware Release Notes


           3706, 3706-S, 3706-NFP, 3706-SNFP

           Error checking when using temperature with an external reference junction has been improved. This
           improvement prevents the use of an external reference junction for channel one (which is the external
           reference junction) and forces the user to measure the external reference junction before it using it for
           another channel.

PR40792 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           A Save button has been added to the Web page Channel Pattern Dialog to better guide the user to
           their desired operation.

PR41493 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           The tspnet.tsp.rbtablecopy() command will default and use the readings synchronous table to a
           reading buffer if one isn't specified in the name parameter.

PR41938 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           The System Web page now allows the creation of a script representing the state of the unit (see
           Create Config Script feature in the manual).

PR41976 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           A screen containing the status of memory usage has been added to the front panel display. It can be
           accessed from the menu item MENU->SYSTEM-INFO->MEMORY-USAGE.

PR 39612 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           The front panel display now indicates when a scan is being run. The selected channel will be replaced
           with the word SCAN for the duration of the scan. Any DMM attribute information will also be removed.

PR41936 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Scripts generated with the createscript command or front panel option under MAIN MENU -> SETUP -
           > CREATE-SCRIPT no longer have a warning message to not modify the file since auto generated.


Noncritical fixes

PR41691 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:




8                                                                            3700S Release Notes V1.40 / May 2010
Series 3700 Firmware Release Notes                                          Version 1.41 Release - Noncritical fixes


          In 1.40a labels were no longer allowed in channel lists when used to denote a range (for example,
          START:END).

          Resolution:
          The feature has been restored.

PR41801 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          In some cases when DMM math was enabled, the front panel units continued to follow the selected
          function.

          Resolution:
          The units now follow the math option.

PR41947 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          The eventlog.overwritemethod attribute is not stored when using the save setup feature or
          when generating a script by createscript.

          Resolution:
          The attribute is properly stored.

PR42129 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          Connecting to the web page of an instrument with a running scan will cause the scan to stop.

          Resolution:
          Connecting and disconnecting from the web page does not impact the state of a scan.

PR42302 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          After running a script generated by createscript, a scan list may be incorrect. Scan steps that were
          added to a scan with the command scan.addimagestep() are not being recreated correctly with
          createconfigscript()

          Resolution:
          A scan is accurately recreated.

PR42303 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP



3700S Release Notes V1.40 / May 2010                                                                              9
Version 1.41 Release - Noncritical fixes                                      Series 3700 Firmware Release Notes



           Symptom:
           Some commands are not included in a script generated by command createscript.

           Resolution:
           The following commands are now included: lan.trigger[N].ipaddress,
           trigger.timer[N].delaylist, tsplink.writeprotect,
           tsplink.trigger[1].pulsewidth, tspnet.tsp.abortonconnect, tspnet.timeout,
           scan.bypass, scan.mode.

PR42309 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           The script created from createscript doesn't have commands for row and column label settings.

           Resolution:
           Issue has been resolved.

PR42364 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           The Scan Start Event and Scan Complete event may be out of order when the scan count is greater
           than one.

           Resolution:
           When a scan has greater than one scan count, the Scan Start Event and Scan Complete Event on a
           scan pass greater than one are now in the correct order.


PR42548 Models affected:
           3706, 3706-NFP

           Symptom:
           DMM Auto Delay set to Once does not give expected performance.

           Resolution:
           DMM Auto Delay set to Once, for readings greater than 1, gives performance equal to DMM Auto
           Delay Off.

PR42735 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           An aperture setting is not saved with "create config script" for period and frequency. Recalling a setup
           or executing a script from "create config script" generates a data type error message (-104).

           Resolution:


10                                                                          3700S Release Notes V1.40 / May 2010
Series 3700 Firmware Release Notes                                           Version 1.41 Release - Noncritical fixes


          Issue has been resolved. In addition, the following settings are not part of a setup or create config
          script:

           schedule.alarm[N].enable
           schedule.alarm[N].seconds
           schedule.alarm[N].fractionalseconds
           schedule.alarm[N].ptpseconds

PR41753 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          The beeper.enable setting is not affected by a system reset, saved in a setup, or saved in a script
          generated by createconfigscript().

          Resolution:
          A reset() command will set the beeper.enable attribute back to 1. The setting will be maintained by
          a setup file or script generated by createconfigscript().

PR42033 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          Setting the period, repetition, seconds, fractionalseconds or ptpseconds of a schedule alarm with a
          negative value doesn't generate an error. Afterwards, the setting has an incorrect value when queried.

          Resolution:
          An out of range argument generates an error. The value for fractionalseconds neeeds to be less than
          1 and greater than or equal to 0 to not generate an error message.

PR42058 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          None

          Resolution:
          The undocumented createscript() ICL command has been changed to createconfigscript(). Calling the
          orginal function will result in an error.

          The location of the feature on front panel has changed from the CREATE-SCRIPT option of SETUP
          submenu under MAIN menu, to CREATE-CONFIG option of SCRIPT submenu under MAIN menu.

          On the System Web page, a Create Config Script button has been added.

PR42613 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:




3700S Release Notes V1.40 / May 2010                                                                              11
Version 1.41 Release - Noncritical fixes                                         Series 3700 Firmware Release Notes


           Some users may experience inappropriately degraded performance when using the Ethernet raw
           socket and telnet sockets. The degradation may occur when a number of small packets are
           exchanged across the socket. This is caused by the Nagle algorithm which delays Ethernet packets
           for up to 200 mS waiting to combine small packets for network efficiency.

           Resolution:
           In most cases packet immediacy is more desirable than network efficiency. The Nagle algorithm is
           now disabled by default to eliminate this delay. If necessary for Ethernet efficiency, use the lan.nagle
           ICL to enable the algorithm.

PR41324 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           An error is not logged when an empty parameter string is passed to dmm.configure.delete(),
           dmm.configure.query(), or dmm.configure.recall().

           Resolution:
           Issue has been resolved.

PR42057 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           In 1.40 the front panel calibration screen did not display the calibration information.

           Resolution:
           The information is now properly displayed. At no time was the calibration information lost.

PR42310 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           The user screen is blank at power up and after a reset.

           Resolution:
           At power up, the user screen now shows USER SCREEN and is not cleared with a reset operation.

PR42519 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           The password mode ICL is not be utilized by the web pages.

           Resolution:
           The applet will read the password mode at logout time and at powerup/initialization only. If the user
           changes the password mode via non-web interface while the applet is running, the applet will not be
           notified of this change until a reload or logout event.




12                                                                            3700S Release Notes V1.40 / May 2010
Series 3700 Firmware Release Notes                                          Version 1.41 Release - Known issues


PR41691 Models affected:
          3706, 3706-NFP

          Symptom:
          The command dmm.calibration.unlock() generates a -151, Invalid string data error message.

          Resolution:
          Now, the command generates a -150, String data error message.

Known issues

PR41421 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          The maximum number of characters that can be sent to the instrument without a newline character
          depends on which remote interface is being used. Limiting line length to less than 950 characters
          without a newline character will ensure interface independence.

PR41556 Models affected:
          3706, 3706-NFP, 3706-S, 3706-SNFP

          Symptom:
          Creating a reading buffer with the same name as a script already in the system will prevent normal
          access to the reading buffer. Choose unique names for reading buffers.




3700S Release Notes V1.40 / May 2010                                                                           13
Series 3700 Firmware Release Notes                                                  Version 1.40 Release - Overview



Version 1.40 Release
Overview
The Series 3700 version 1.40 firmware is an enhancement release including control and operation of the Model
3732 Quad 4 x 28 Ultra-High Density Reed Relay Matrix Card, as well as some problem fixes. Some problem
corrections may impact compatibility, please review the corrections prior to installation.

NOTE        Upgrading to firmware version 1.40 requires special installation steps that are not required for other
            firmware upgrades. Carefully read the Version 1.40 installation notes below before upgrading your
            Series 3700.

The following table lists the exact version numbers of the available Series 3700 components when the firmware
upgrade is complete.

                         Model                                                      Version
                         Series 3700 Mainframe                                      1.40k
                         Series 3700 High Performance DMM                           1.40k
                         3720 Dual 1x30 Multiplexer                                 1.40h
                         3721 Dual 1x20 Multiplexer                                 1.40h
                         3722 Dual 1x48 High Density Multiplexer                    1.40h
                         3723 Dual 1x30 High Speed Reed Relay Multiplexer           1.40j
                         3724 Dual 1x30 FET Multiplexer                             1.40j
                         3730 Dual 6x16 High Density Matrix                         1.40h
                         3731 Dual 6x16 High Speed Reed Matrix                      1.40j
                         3732 Quad 4x28 High Speed Reed Matrix                      1.40m
                         3740 32 Channel Isolated Switch                            1.40h
                         3750 Multifunction I/O Card                                1.40j


Version 1.40 installation notes
Upgrading to firmware version 1.40 requires special installation steps that are not required for other firmware
upgrades. Carefully perform the following procedures to successfully upgrade to version 1.40.

First, determine your instrument's current firmware revision. With the instrument powered on, press the SLOT
key. The instrument's current firmware version is displayed. Based on the current firmware revision, follow the
instructions in the appropriate section below.

            Upgrading from firmware version 1.31a or any earlier version

                Download the nk_5001607.nb0 and main_p37xx_0140k.CAB files from the
                Keithley Instruments website (www.keithley.com).
                To perform the upgrade:
                10. Place the nk_5001607.nb0 and main_p37xx_0140k.CAB files in the root directory of a
                    blank USB flash drive (make sure drive size is large enough for the size of the upgrade files).
                11. Ensure that the Series 3700 mainframe power is turned off, and then connect the flash drive
                    to the Series 3700 and turn on power to initiate the upgrade process.

                NOTE       The upgrade process will take three to five minutes and the instrument will reboot
                           several times during the process. Do not disturb the instrument until the normal front
                           panel display has been present for at least 30 seconds.




3700S Release Notes V1.40 / May 2010                                                                                 14
Series 3700 Firmware Release Notes                           Version 1.40 Release - Version 1.40 installation notes


              12. On the front panel, press the MENU key, select LAN, select APPLY, and then press ENTER
                  to resume normal Ethernet connectivity following the upgrade.
              13. Disconnect the USB flash drive from the Series 3700 mainframe.

              NOTE      Alternatively, you can have Keithley Instruments assist you with the upgrades by calling
                        your local Keithley support office.


           Upgrading from firmware version 1.32a

              Download the main_p37xx_0140k.CAB files from the Keithley Instruments website
              (www.keithley.com).

              NOTE      The following steps force the instrument to upgrade its firmware twice. This is
                        necessary to ensure everything is upgraded properly. Follow the steps below carefully.

              To perform the upgrade:
              14. Copy the main_p37xx_0140k.CAB file to a blank USB flash drive (make sure drive size is
                  large enough for the size of the upgrade file).

              NOTE      Verify that the USB flash drive is blank.

              15. Turn on the Series 3700 power.
              16. Install the USB flash drive in the front panel connector.
              17. On the front panel, press the MENU key.

              NOTE      If your model does not have a front panel, upgrade over the bus using the
                        upgrade.unit()command.

              18. Turn the navigation wheel to scroll to UPGRADE, and then press the navigation wheel.
              19. The question UPGRADE UNIT? displays. Select Yes, and then press the navigation wheel.
                  The Series 3700 upgrade status displays on the front panel, including the percentage
                  completed. When the file has been unpacked, the upgrade status is displayed as the upgrade
                  progresses (first cards installed in the slots are upgraded (including the DMM, if installed),
                  and then the Series 3700 mainframe).

              NOTE      For models without a front panel, the LAN status and clock status LEDs blink in unison
                        during the upgrade process.

              20. Wait for the Series 3700 to reboot automatically. When the reboot is complete, press the
                  MENU key on the front panel.

              NOTE      If your model does not have a front panel, upgrade over the bus using the
                        upgrade.previous() command.

              21. Turn the navigation wheel to scroll to UPGRADE and press the navigation wheel.
              22. The question UPGRADE UNIT? is displayed. Select Previous, and then press the navigation
                  wheel (note that in this step you are selecting "Previous" instead of "Yes," as was done in
                  step 6). The Series 3700 upgrade status displays on the front panel, including the percentage
                  completed. When the file has been unpacked, the upgrade status is displayed as the upgrade
                  progresses (first cards installed in the slots are upgraded (including the DMM, if installed),
                  and then the Series 3700 mainframe).
                  The Series 3700 reboots automatically when the upgrade is complete.




3700S Release Notes V1.40 / May 2010                                                                             15
Version 1.40 Release - Critical fixes                                       Series 3700 Firmware Release Notes


Critical fixes

PR38770 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           Functions on the advanced trigger web page do not operate correctly.

           Resolution:
           The advanced trigger web page now operates correctly.

PR39078 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           Recalling a saved setup takes too long to complete.

           Resolution:
           This issue is resolved.

PR39569 Models affected:
           3706, 3706-NFP

           Symptom:
           Measurement timestamps show excessive jitter.

           Resolution:
           Measurement timestamp jitter has been significantly reduced.

PR39802 Models affected:
           3706, 3706-NFP

           Symptom:
           When measuring temperature, the internal measured reference junction value is not used when using
           the dmm.close() command for a configuration that includes an internal reference junction or when
           changing the DMM transducer type with a channel closed on the temperature function.

           Resolution:
           This issue has been corrected.

PR39903 Models affected:
           3706, 3706-S, 3706-NFP, 3706-SNFP

           Symptom:
           Having a reset command in the autoexec script causes the instrument to generate a -222,
           parameter data out of range error message, and the reset command is not executed.




16                                                                        3700S Release Notes V1.40 / May 2010
Series 3700 Firmware Release Notes                                             Version 1.40 Release - Critical fixes


          Resolution:
          The issue has been corrected.

PR39944 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          The lan config DNS verify setting and the lan config DNS dynamic setting operate incorrectly when
          accessed from the instrument's front panel.

          Resolution:
          This issue has been corrected.

PR40259 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          An instrument downgraded below version 1.30b may no longer be capable of upgrading back to a
          higher revision. The upgrade process now warns of this possibility. Instruments without front panels
          will no longer downgrade past 1.30b.

          Resolution:
          Do not downgrade an instrument's firmware to 1.30b or earlier unless you are certain you will not want
          to upgrade to a newer firmware version.

PR40353 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          Using os.date with invalid format specifiers may cause the instrument to stop responding.

          Resolution:
          Invalid format specifiers now return an Invalid Format message.

PR40953 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          Changing the TSP-N termination in the middle of a connection to a TSP instrument will confuse the
          processing of the connection.

          Resolution:
          TSP-Net termination will generate an error if modified during a connection to a TSP instrument. The
          termination is always set to linefeed.

PR41359 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP




3700S Release Notes V1.40 / May 2010                                                                             17
Version 1.40 Release - Enhancements                                           Series 3700 Firmware Release Notes


          Symptom:
          For the Models 3723, 3724, and 3731 cards, the user settle delay (channel.setdelay() command)
          is incorrectly applied to open actions, rather than close actions only.

          Resolution:
          This issue has been corrected.

PR 41891 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Symptom:
          Disabling the Ethernet Ping protocol also disables most other Ethernet functionality

          Resolution:
          The ability to disable Ethernet Ping protocol has been removed. Ping is now always enabled.

Enhancements

PR33446, Models affected:
PR39259
         3706, 3706-S, 3706-NFP, 3706-SNFP

          Firmware updates can now be initiated from the web interface by pressing the Upgrade Firmware
          button on the system web page. Download the firmware image file from www.keithley.com before
          initiating the update from the instrument's web interface. Follow the dialog box prompts after selecting
          this button. During the upgrade, the instrument will disconnect from the web interface. After the
          upgrade finishes and the instrument reboots, close all browser windows to clear any cached
          information, and then open a new browser window to reconnect to the web interface.

PR 36048 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Scripts that utilize the channel.*() ICL commands may exhibit performance improvements. This
          improvement may be most noticeable when the channel list contains a high number of channels.

PR37683 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          The instrument will no longer support specifying a pseudocard using a constant, for example,
          slot[2].pseudocard = slot.PSEUDO_3722. Pseudocards will now be specified by using the
          model number only, for example, slot[2].pseudocard = 3722. For consistency, Keithley
          recommends using this new convention for all cards. However, to ensure backward compatibility,
          constants for existing cards will continue to be available, but constants will not be added for future
          cards.

PR 38853 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          The Log menu item on the instrument's main web page has been moved to the main LXI web page.

PR 39230 Models affected:



18                                                                         3700S Release Notes V1.40 / May 2010
Series 3700 Firmware Release Notes                                           Version 1.40 Release - Enhancements


          3706, 3706-S, 3706-NFP, 3706-SNFP

          The system web interface now contains the card serial numbers in the report list.

PR 39612 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          Front panel and web interface enhancements have been made to support the large number of
          channels available on the Model 3732 card. For details about these enhancements, refer to PA-1021,
          Model 3732 Reed Relay Matrix Card Connection Information, which is supplied with the Model 3732.

PR39911 Models affected:
PR39915
        3706, 3706-S, 3706-NFP, 3706-SNFP

          The following ICL commands have been added:

          trigger.blender[N].reset()
          trigger.timer[N].reset()
          digio.trigger[N].reset()
          tsplink.trigger[N].reset()

          These commands reset the associated N blender, timer, or trigger attributes back to the factory default
          settings, as indicated below.

               NOTE       For details, including default settings for each command, refer to the Series 3700
                          Reference Manual (document number 3700S-901-01).

          The trigger.blender[N].reset() command resets the following back to factory defaults:

               trigger.blender[N].orenable
               trigger.blender[N].stimulus[M]

           It also clears the overrun:

               trigger.blender[N].overrun

          The digio.trigger[N].reset() command resets the following back to factory defaults:

               trigger.timer[N].count
               trigger.timer[N].delay
               trigger.timer[N].delaylist
               trigger.timer[N].passthrough
               trigger.timer[N].stimulus

          It also clears the overrun:

               trigger.timer[N].overrun

          The digio.trigger[N].reset() command resets the following back to factory defaults:

               digio.trigger[N].mode
               digio.trigger[N].pulsewidth
               digio.trigger[N].stimulus

          It also clears the overrun:


3700S Release Notes V1.40 / May 2010                                                                            19
Version 1.40 Release - Enhancements                                            Series 3700 Firmware Release Notes



               digio.trigger[N].overrun

          The tsplink.trigger[N].reset() command resets the following back to factory defaults:

               tsplink.trigger[N].mode
               tsplink.trigger[N].stimulus

          It also clears the overrun:

               tsplink.trigger[N].overrun

PR 40328 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          The eventlog.next() command takes an optional integer parameter ranging from 1 to 50,
          indicating the number of events desired in the response. If the parameter is not specified, then it
          defaults to 1 to mimic the behavior of the command before the enhancement.

PR 40661 Models affected:
          3706, 3706-S, 3706-NFP, 3706-SNFP

          A new ICL function that creates a string channel descriptor from a series of card-dependent integer
          arguments has been added:

          channel.createspecifier()

          Usage:

          There are five variants of this function that can be used, depending on the type of card in the specified
          slot:

              =    channel.createspecifier(,             , , )
              =    channel.createspecifier(,             , )
              =    channel.createspecifier(,             , )
              =    channel.createspecifier(,             )
              =    channel.createspecifier(,             )

               Where:

                     slot                 Specifies the slot number to use
                     bank                 Specifies the bank number to use (if applicable)
                     row                  Specifies the row number to use
                     column               Specifies the column number to use
                     index                Specifies the index to use
                     backplane            Specifies the backplane to use


          This function creates a string channel descriptor from a series of card-dependent integer arguments.




20                                                                           3700S Release Notes V1.40 / May 2010
Series 3700 Firmware Release Notes                                                   Version 1.40 Release - Enhancements


          The arguments are dependent on the card type in the specified slot. The channel.createspecifier
          command can only create valid channel descriptors; if an illegal argument is sent for the type of card in
          the specified slot, an error is generated.

          Variants of this function can be used, depending on the type of card in the specified slot:

                                 Type of card in slot           Function variants to use
                                 Matrix card containing banks   Variant 1 or 5
                                 Matrix card without banks      Variant 2 or 5

                                 Multiplexer cards              Variant 3, 4, or 5


          Example 1:

          To create a channel descriptor on the Model 3732 card configured as a single 4x112 matrix in Slot 3,
          Bank 1, Row 2, Column 101:

               cd = channel.createspecifier(3, 1, 2, 101)
               print(cd)
               312A1

          Example 2:

          To set the pole setting to 2 for all channels in Bank 1 on a Model 3732 card configured as a dual 8x28
          matrix in Slot 1:

               for row = 1,8 do
                     for col = 1,28 do
                           ch = channel.createspecifier(1,1,row,col)
                                 channel.setpole(ch, 2)
                           end
               end

          Example 3:

          To create a channel descriptor on a Model 3724 multiplexer card in Slot 2, Bank 2, Index 1:

               cd = channel.createspecifier(2, 2, 1)
               print(cd)
               2031

          Example 4:

          To create a channel descriptor on a Model 3724 multiplexer card in Slot 1, Backplane 911:

               cd = channel.createspecifier(1, 911)
               print(cd)
               1911

PR40699 Models affected:
          3706, 3706-S

          The front-panel display functionality has been improved to increase usability. The plus (+) sign
          indicator has moved to the second line of the display and is no longer after the selected channel
          description. Now, the plus (+) sign denotes that a channel other than the selected channel is closed.
          To see which channels are closed, use the DISPLAY button to cycle to a new front-panel screen that


3700S Release Notes V1.40 / May 2010                                                                                 21
Version 1.40 Release - Enhancements                                            Series 3700 Firmware Release Notes


          lists the closed channels. If the closed channel list exceeds the screen size, use the navigation wheel
          to scroll through the list.


PR 40707 Models affected:
          3706, 3706-S

          A new ICL function that allows inclusion of multiple channels in a single scan step has been added:

          scan.addimagestep()

          Usage:

          scan.addimagestep(, [])

               Where:

                                String specifying a list of channels
                                  Optional string specifying a DMM configuration

          This function adds a list of channels to be closed simultaneously in a single step of a scan. An optional
          DMM configuration can be added to force the scan to take a measurement during the same step.

          The scan.addimagestep function is an advanced command; the channel list parameter must
          specify appropriate relays to support the requested DMM configuration, or an invalid measurement will
          result.

          Unlike the scan.add() function, paired channels and backplanes necessary for measurement are
          not automatically added to the step. Use the channel.setpole() command to indicate if the paired
          channel should be added or not. Backplanes assigned to channels by the
          channel.setbackplane() command will not be added to the image step automatically. For
          example, if a measurement is taken on a 4-wire ohms configuration without designating 4-pole with the
          channel.setpole() command, the corresponding paired channel will not be added, and the
          specified "dmm config" will not cause additional relay closures as it normally would. Backplane
          channels that are needed for the desired measurement must be specified in the channel list
          parameter.

          If you have changed the pole setting on any of the channels in the list (using channel.setpole()),
          an additional paired channel will be added or removed appropriately. For example, to ensure that the
          proper channels close to enable a 4-wire measurement, set the pole setting in addition to using the
          4-wire ohms DMM configuration.

          When a DMM configuration (other than "nofunction") is specified, the instrument will take the
          appropriate measurement based on the function set in the configuration; if no DMM configuration is
          specified with the command, no measurement will be taken.

          Example 1:

          This example adds a single step that closes Channel 1001 and takes a DC voltage measurement.
          Note that the voltage measurement will be inaccurate if this is the only step in the scan (because the
          backplane channels are not closed).

               scan.addimagestep("1001", "dcvolts")



22                                                                          3700S Release Notes V1.40 / May 2010
Series 3700 Firmware Release Notes                                                Version 1.40 Release - Enhancements


          Example 2:

          To add a single step that closes Channels 1001 and 1911, and then takes a DC voltage
          measurement:

               scan.addimagestep("1001, 1911", "dcvolts")

          Example 3:

          This example first sets Channel 1001 to 4-pole, then adds a single step that closes Channels 1001,
          1031, and 1911, and then takes a DC voltage measurement.

               channel.setpole("1001", 4)
               scan.addimagestep("1001, 1911", "dcvolts")

          Example 4:

          This example adds three steps with the following actions:

           