INTRODUCTION
                   You can use this procedure to adjust the AP034 Active
                   Differential Probe to meet the warranted specifications. This
                   procedure should only be performed if the instrument fails the
                   Performance Verification tests.
                   If the probe cannot be adjusted to meet the Performance
                   Verification limits, repair may be necessary.
                   To ensure instrument accuracy, check the calibration of the
                   AP034 Active Differential Probe every year. Before calibration,
                   thoroughly clean and inspect this unit as discussed in the
                   "Cleaning" section.
                   Completion of each step in the Adjustment Procedure ensures
                   that the differential probe meets specifications. Some of the
                   adjustments interact with other parts of the circuitry. Therefore, it
                   is necessary that all adjustments be performed in the order listed.
                   For best overall instrument performance, make each adjustment
                   to the exact setting, even when adjustment is within the limits
                   stated in the procedure.
                   Adequate guard bands were designed into the AP034 Active
                   Differential Probe to ensure that it will meet or exceed published
                   specifications over the entire operating temperature range. To
                   continue to meet the environmental specifications, all
                   adjustments must be performed in a controlled environment with
                   an ambient temperature of 25 