                             TB 9-6625-2099-35
                                          CHANGE 1
      DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
            CALIBRATION PROCEDURE FOR
          DUAL TRACE AMPLIFIERS AM-6880/U
               (TEKTRONIX, TYPE 7A18),
       TEKTRONIX, TYPES 7A18A, 7A18A W/OPTION 6,
         AND AM-6785/U (TEKTRONIX, TYPE 7A26)
                  Headquarters, Department of the Army, Washington, DC
                                   30 October 2002
                        Approved for public release; distribution is unlimited.
TB 9-6625-2099-35, 19 July 2002, is changed as follows:

1. Remove old pages and insert new pages as indicated below.                      New or changed
material is indicated by a vertical bar in the margin of the page.

               Remove Pages                                           Insert Pages
                  1 thru 6                                               1 thru 6
                 11 thru 14                                             11 thru 14


2. File this change sheet in front of the publication for reference purposes.
By Order of the Secretary of the Army:




                         0224712
Distribution:
    To be distributed in accordance with IDN 342232, requirements for TB 9-6625-2099-35.
                             *TB 9-6625-2099-35
               CALIBRATION PROCEDURE FOR
             DUAL TRACE AMPLIFIERS AM-6880/U
                  (TEKTRONIX, TYPE 7A18),
                 TEKTRONIX, TYPES 7A18A,
              7A18A W/OPTION 6, AND AM-6785/U
                  (TEKTRONIX, TYPE 7A26)
                    Headquarters, Department of the Army, Washington, DC
                                        19 July 2002
                            Approved for public release; distribution is unlimited
                 REPORTING OF ERRORS AND RECOMMENDING IMPROVEMENTS
        You can improve this manual. If you find any mistakes or if you know of a way to improve these
        procedures, please let us know. Mail your letter, DA Form 2028 (Recommended Changes to
        Publications and Blank Forms), or DA Form 2028, directly to Commander, U.S. Army Aviation
        and Missile Command, ATTN: AMSAM-MMC-MA-NP, Redstone Arsenal, AL 35898-5000. A
        reply will be furnished to you. You may also provide DA Form 2028 information to AMCOM via
        e-mail, fax, or the World Wide Web. Our FAX number is: DSN 788-6546 or Commercial 256-
        842-6546. Our e-mail address is: [email protected]. Instructions for sending an
        electronic 2028 may be found at the back of this manual. For the World Wide Web, use:
        https://amcom2028.redstone.army.mil.

                                                                                                Paragraph     Page
  SECTION           I.      IDENTIFICATION AND DESCRIPTION
                            Test instrument identification ............................             1           2
                            Forms, records, and reports ...............................             2           2
                            Calibration description .......................................         3           2
                    II.     EQUIPMENT REQUIREMENTS
                            Equipment required ...........................................          4           3
                            Accessories required...........................................         5           3
                   III.     CALIBRATION PROCESS FOR AM-6880/U
                            (TEKTRONIX, TYPE 7A18) AND
                            TEKTRONIX, TYPES 7A18A AND 7A18A
                            W/OPTION 06
                            Preliminary instructions ....................................           6           4
                            Equipment setup................................................         7           4
                            Vertical gain    ................................................       8           5
                            Risetime         ................................................       9           8
                            Final procedure ................................................       10          10




__________
  *This bulletin supersedes TB 9-6625-2099-35, dated 29 March 1993, including all changes.                  CHANGE 1
TB 9-6625-2099-35
                                                                                             Paragraph   Page
                 IV.     CALIBRATION PROCESS FOR AM-6785/U
                         (TEKTRONIX, TYPE 7A26)
                         Preliminary instructions ....................................          11        10
                         Equipment setup................................................        12        10
                         Vertical gain    ................................................      13        11
                         Risetime         ................................................      14        16
                         Final procedure ................................................       15        18

                                     SECTION I
                          IDENTIFICATION AND DESCRIPTION

1. Test Instrument Identification. This bulletin provides instructions for the
calibration of Dual Trace Amplifiers, AM-6880/U (Tektronix, Type 7A18), Tektronix,
Types 7A18A, 7A18A w/Option 06, and AM-6785/U (Tektronix, Type 7A26). The
manufacturer's manuals were used as the prime data sources in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test
instrument) throughout this bulletin.

   a. Model Variations. Variations among models are described in text.
   b. Time and Technique. The time required for this calibration is approximately 1
hour, using the dc and low frequency technique.
2. Forms, Records, and Reports

   a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
   b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear.   When adjustments are in tables, the (R) follows the designated
adjustments. Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
                                     Table 1. Calibration Description
    Test instrument
       parameters                                    Performance specifications
             AM-6880/U (Tektronix, Type 7A18), Tektronix, Types 7A18A, and 7A18A w/option 06
  Vertical gain          Range: 5 mV/div to 5 V/div
                         Accuracy: 