Service Manuals, User Guides, Schematic Diagrams or docs for : Tektronix 7A13, 7A22 Differential Amplifiers - Calibration Procedure (Mil TB 9-6625-2132-35) (2005) WW

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
7A13, 7A22 Differential Amplifiers - Calibration Procedure (Mil TB 9-6625-2132-35) (2005) WW


>> Download 7A13, 7A22 Differential Amplifiers - Calibration Procedure (Mil TB 9-6625-2132-35) (2005) WW documenatation <<

Text preview - extract from the document
                                 TB 9-6625-2132-35
                                          CHANGE 2
      DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
     CALIBRATION PROCEDURE FOR
  DIFFERENTIAL AMPLIFIERS AM-6881/U
        (TEKTRONIX, TYPE 7A13)
  AND AM-6786/U (TEKTRONIX, TYPE 7A22)
                   Headquarters, Department of the Army, Washington, DC
                                    29 November 2005
          Distribution Statement A: Approved for public release; distribution is unlimited.
TB 9-6625-2132-35, 23 March 2005, is changed as follows:

1. Remove old pages and insert new pages as indicated below. New or changed material is
indicated by a vertical bar in the margin of the page.

                 Remove Pages                                       Insert Pages
                    5 and 6                                            5 and 6


2. File this change sheet in front of the publication for reference purposes.


     By Order of the Secretary of the Army:

                                                                       PETER J. SCHOOMAKER
                                                                      General, United States Army
     Official:                                                               Chief of Staff

          SANDRA R. RILEY
      Administrative Assistant to the
         Secretary of the Army
                       0527602




Distribution:
    To be distributed in accordance with IDN 342240, requirements for calibration procedure
TB 9-6625-2132-35.
                                 TB 9-6625-2132-35
                                          CHANGE 1
      DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
     CALIBRATION PROCEDURE FOR
  DIFFERENTIAL AMPLIFIERS AM-6881/U
        (TEKTRONIX, TYPE 7A13)
  AND AM-6786/U (TEKTRONIX, TYPE 7A22)
                   Headquarters, Department of the Army, Washington, DC
                                    13 September 2005
          Distribution Statement A: Approved for public release; distribution is unlimited.
TB 9-6625-2132-35, 23 March 2005, is changed as follows:

1. Remove old pages and insert new pages as indicated below. New or changed material is
indicated by a vertical bar in the margin of the page.

                 Remove Pages                                       Insert Pages
                    7 and 8                                            7and 8


2. File this change sheet in front of the publication for reference purposes.


     By Order of the Secretary of the Army:

                                                                       PETER J. SCHOOMAKER
                                                                      General, United States Army
     Official:                                                               Chief of Staff

          SANDRA R. RILEY
      Administrative Assistant to the
         Secretary of the Army
                       0519607




Distribution:
    To be distributed in accordance with IDN 342240, requirements for calibration procedure
TB 9-6625-2132-35.
                                *TB 9-6625-2132-35
        DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
      CALIBRATION PROCEDURE FOR
   DIFFERENTIAL AMPLIFIERS AM-6881/U
         (TEKTRONIX, TYPE 7A13)
   AND AM-6786/U (TEKTRONIX, TYPE 7A22)
                       Headquarters, Department of the Army, Washington, DC
                                          23 March 2005
            Distribution Statement A: Approved for public release; distribution is unlimited.
            REPORTING OF ERRORS AND RECOMMENDING IMPROVEMENTS
         You can improve this manual. If you find any mistakes or if you know of a way
         to improve these procedures, please let us know. Mail your letter or DA Form
         2028 (Recommended Changes to Publications and Blank Forms) directly to:
         Commander, US Army Aviation and Missile Command, AMSAM-MMC-MA-NP,
         Redstone Arsenal, AL 35898-5000. A reply will be furnished to you. You may
         also provide DA Form 2028 information to AMCOM via e-mail, fax, or the World
         Wide Web. Our fax number is DSN 788-6546 or Commercial 256-842-6546. Our
         e-mail address is [email protected]. Instructions for sending an
         electronic 2028 may be found at the back of this manual. For the World Wide
         Web, use https://amcom2028.redstone.army.mil.
                                                                                                       Paragraph   Page
   SECTION               I.     IDENTIFICATION AND DESCRIPTION
                                Test instrument identification ............................                1        2
                                Forms, records, and reports.................................               2        2
                                Calibration description ........................................           3        2

                         II.    EQUIPMENT REQUIREMENTS
                                Equipment required .............................................           4        3
                                Accessories required.............................................          5        3

                        III.    CALIBRATION PROCESS FOR
                                 DIFFERENTIAL AMPLIFIER
                                 TEKTRONIX, TYPE 7A13
                                Preliminary instructions......................................             6        4
                                Equipment setup ..................................................         7        4
                                Gain and deflection accuracy...............................                8        6
                                Comparison voltage and linearity.......................                    9        7
                                +INPUT and -INPUT attenuator
                                 compensation .....................................................       10        8
                                High-frequency response and risetime ...............                      11        9
                                Final procedure ....................................................      12       11



_________
  *This bulletin supersedes TB 9-6625-2132-35, dated 25 Apr 1985, including all changes.
TB 9-6625-2132-35
                                                                                              Paragraph   Page
                 IV.   CALIBRATION PROCESS FOR
                         DIFFERENTIAL AMPLIFIER
                         TEKTRONIX, TYPE 7A22
                       Preliminary instructions......................................            13        11
                       Equipment setup ..................................................        14        11
                       Gain and attenuator accuracy .............................                15        12
                       Attenuation compensation...................................               16        13
                       Input attenuator differential balance .................                   17        14
                       HF and LF-3 dB point..........................................            18        15
                       Final procedure ....................................................      19        16




                                  SECTION I
                       IDENTIFICATION AND DESCRIPTION

1. Test Instrument Identification.           This bulletin provides instructions for the
calibration of Differential Amplifiers AM6881/U (Tektronix, Type 7A13) and AM-6786/U
(Tektronix, Type 7A22). The manufacturers' manuals were used as the prime data source
in compiling these instructions. The equipment being calibrated will be referred to as the
TI (test instrument) throughout this bulletin.

    a. Model Variations. Variations among models are described in text, tables, and
figures.
   b. Time and Technique. The time required for this calibration is approximately 4
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports

   a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
   b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.




2
                                                                                                     TB 9-6625-2132-35
                                             Table 1. Calibration Description

            Test instrument parameters                                Performance specifications
                                              Tektronix, Type 7A13
Deflection factor                                         Range: 1 mV/div to 5 V/div
                                                          Accuracy: 



◦ Jabse Service Manual Search 2024 ◦ Jabse PravopisonTap.bg ◦ Other service manual resources online : FixyaeServiceinfo