                                                                                                   TB 9-6625-2176-35
                                                                                                            Change 2


                            DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

                                         CALIBRATION PROCEDURE FOR

                                CONSTANT AMPLITUDE SIGNAL GENERATOR

                                                     AN/USM-272

                                                           AND

                                               TEKTRONIX, TYPE 191

                            Headquarters, Department of the Army, Washington, DC
                                                16 May 1988

TB 9-6625-2176-35, 15 April 1987, is changed as follows:


     1. Remove old pages and insert new pages as indicated below. New or changed material is indicated by a vertical
bar in the margin of the page.

                          Remove pages                              Insert pages
                             1 and 2                                   1 and 2
                             7 and 8                                   7 and 8

    2. File this change sheet in front of the publication for reference purposes.


By Order of the Secretary of the Army:



                                                                                             CARL E. VUONO
                                                                                          General, United States Army
                                                                                                Chief of Staff
Official:
              R. L. DILWORTH
     Brigadier General, United States Army
             The Adjutant General


Distribution:
To be distributed in accordance with DA Form 12-34C, Block No.             319, requirements for calibration procedures
publications.




                                                                                                  PIN NO: 061693-002
                                                                                                       TB 9-6625-2176-35

                                                                                                                 Change 1

                            DEPARTMENT OF THE ARMY TECHNICAL BULLETIN


                                          CALIBRATION PROCEDURE FOR

                                CONSTANT AMPLITUDE SIGNAL GENERATOR

                                                     AN/USM-272

                                                           AND

                                               TEKTRONIX, TYPE 191

                                Headquarters, Department of the Army, Washington, DC
                                                   31 March 1988


TB 9-6625-2176-35, 15 April 1987, is changed as follows:


     1. Remove old pages and insert new pages as indicated below. New or changed material is indicated by a vertical
bar in the margin of the page.


                                Remove pages                                        Insert pages

                                7 and 8                                             7 and 8


    2. File this change sheet in front of the publication for reference purposes.




By Order of the Secretary of the Army:

                                                                                                 CARL E. VUONO
                                                                                              General, United States Army
                                                                                                    Chief of Staff
Official:
               R. L. DILWORTH
      Brigadier General, United State Army
              The Adjutant General

Distribution:
To be distributed in accordance with DA Form 12-34C, Block No.             319, requirements for calibration procedures
publications.

            THIS PUBLICATION IS A COURTESY QUICK COPY FRCM THE UNITED STATES ARMY
            PUBLICATICNS DISTRIBUTION CENTER, ST. LOUIS, MISSOURI TO MEET YOUR NEEDS WHILE
            WE REPLENISH OUR RULAR STOCK.
                                                                                                                 *TB 9-6625-2176-35
                                                                                              SUPERSEDED COPY DATED 2 AUGUST 1978

                                  DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

                                                  CALIBRATION PROCEDURE FOR

                                       CONSTANT AMPLITUDE SIGNAL GENERATOR

                                           AN/USM-272 AND TEKTRONIX, TYPE 191

                                     Headquarters, Department of the Army, Washington, DC
                                                         15 April 1987

                                        Approved for public release; distribution is unlimited.

                                                         REPORTING OF ERRORS

       You can help improve this publication by calling attention to errors and by     recommending
       improvements and stating your reasons for the recommendations. Your letter or DA Form 2028,
       Recommended Changes to Publications, should be mailed directly to Commander, U.S. Army TMDE
       Support Group, ATTN: AMXTM-LPP, Redstone Arsenal, AL 35898-5400. A reply will be furnished
       directly to you.


                                                                                                             Paragraph   Page
SECTION I.        IDENTIFICATION AND DESCRIPTION
                  Test instrument identification ...................................................             1         2
                  Forms, records, and reports- ....................................................              2         2
                  Calibration description..............................................................          3         2

           II.    EQUIPMENT REQUIREMENTS
                  Equipment required..................................................................           4         3
                  Accessories required................................................................           5         3

           III.   CALIBRATION PROCESS
                  Preliminary instructions............................................................           6         4
                  Equipment setup .....................................................................          7         4
                  Amplitude accuracy..................................................................           8         5
                  Frequency accuracy.................................................................            9         7
                  Frequency response.................................................................            10        7
                  Power supply ...........................................................................       11        9
                  Final procedure........................................................................        12        9




*This bulletin supersedes TB 11-6625-2384-35, 2 August 1978, including all changes.

                                                                            1
                                                                                                 TB 9-6625-2176-35

                                                     SECTION I

                                      IDENTIFICATION AND DESCRIPTION

1.    Test Instrument Identification.     This bulletin
provides instructions for the calibration of Constant
Amplitude Signal Generator, AN/USM-272 and                         a. Forms, records, and reports required for
Tektronix, Type 191. The manufacturer's manual and            calibration personnel at all levels are prescribed by TB
TM 116625-2384-15 were used as the prime data                 750-25.
sources in compiling these instructions. The equipment             b. Adjustments to be reported are designated (R)
being calibrated will be referred to as the TI (test          at the end of the sentence in which they appear. When
instrument) throughout this bulletin.                         adjustments are in tables, the (R) follows the
                                                              designated adjustment. Report only those adjustments
                                                              made and designated with (R).
     a. Model Variations. None.
     b. Time and Technique. The time required for             3.    Calibration Description.     TI parameters and
this calibration is approximately 4 hours, using the dc       performance specifications which pertain to this
and low frequency technique.                                  calibration are listed in table 1.

2. Forms, Records, and Reports

                                         Table 1. Calibration Description
                            Test instrument                                   Performance
                              parameters                                     specifications
         Amplitude (at 50 kHz into                             Range: 0.5 to 5 V
         50W load)                                             Accuracy: 