Text preview - extract from the document
Cost Effective Techniques used to Validate the Performance
 of the microK Resistance Thermometry Instrument with
                   sub mK Uncertainty
                                               Paul Bramley
                                        Metrosol Limited, Towcester, UK

                                            Neil Robinson
                                 Isothermal Technology Limited, Southport, UK

This article describes techniques used to validate the performance of the microK thermometry instrument, which
was designed for use in secondary temperature calibration and high accuracy temperature measurement
applications (to <0.4mK uncertainty). The microK is the first precision thermometer that can work with both
resistance thermometers and thermocouples and provides sub mK uncertainties with SPRTs and mK
uncertainties with thermocouples. With a resistance ratio accuracy of 0.4ppm and voltage accuracy of 0.25