                                                                          Model 6664C
                             8 AND 16-CHANNEL LOW THERMAL QUAD SCANNERS

                 Complements All Guildline Bridges and Fully Automates Measurements

                                                Guildline's 6664C Low Thermal Quad Scanner is the first
                                                commercial scanner capable of operating at 1000V. This has
                                                greatly improved the measurement/calibration throughput of
                                                standards laboratories, since many high ohm measurements
                                                can now be automated. High ohm measurements are typically
                                                hard to make and very time consuming because of the low
                                                signal/noise ratio. With a Guildline model 6664C scanner and
                                                software, these hard-to-make high ohm measurements (up to
                                                100 G) can be automated and scheduled at any time that is
                                                convenient to the metrologists, even at night and over the
                                                weekends.
                                                This scanner is designed to work with a Guildline 6622A
                                                Automatic Direct Current Comparator (DCC) Resistance Bridge,
                                                a model 6622T Temperature Bridge or a Teraohmmeter (model
                                                6520). Still have one of our older Guildline Automatic Bridges
                                                such as a model 6675A? Our software and scanner also support
                                                all previous Guildline' models.
                                                Both 8-channel and 16-channel models are available. Like other
           6664C FEATURES                       Guildline products, the 8-channel scanner can be expanded to
                                                a 16-channel model, giving our customers exactly what's
 Withstand up to 1000 V DC to provide fully     needed now and the flexibility for future growth.
automated high ohm measurements
                                                To manually select 4-terminal devices connected to the input
 Each channel is rated 2 A                      channels, simply press the push button switch on the Line A-
                                                output and another on the Line B-output.          All other
 8-channel and 16-channel models
                                                connections remain isolated.
 8-channel model expandable to 16-channel       The 6664C performs switching for precision resistance
 Thermoelectric Potential: <20 nV               measurements with less than 20 nanovolt of thermal offsets.
                                                Careful design in the use of latching relays, requiring only a
 Contact Life: >10,000,000 cycles               short pulse to actuate, eliminates self-heating errors.
 Front Panel or Remote GPIB Control
                                              The 6664C Low Thermal Quad Scanner Enables
 4-Terminal Switching Matrix Scanner
                                              Software Controlled, Fully Automated High Ohm
 Gold flashed Tellurium Copper Terminals      Measurements At Up to 1000 V
                                                Special low thermal latching relays are used. Leakage between
                                                the high and low circuits is reduced by switching them on
                                                different printed circuit boards.
                                                Protection circuits prevent selection of more than one 4
                                                terminal connection on a channel output at the same time.
                                                Front panel indicators show which channel is selected on each
                                                output.
6664C 8 and 16-Channel, Low Thermal Quad Scanners
SPECIFICATIONS
     Thermoelectric Potentials:                                                     <20 nV typical and <50nV max.
                                                                                    Life: >10,000,000 cycles at low levels
                                                                                    Initial Contact Resistance: 0.05  max.
     Relay Contact Ratings:                                                         Current: 2A max at 10 V
                                                                                    Voltage Switched: 100 V max. at 1 mA
                                                                                    Voltage Non-Switched: 1000 V max.
     Leakage Resistance:                                                            >1015
                                                                                    Input: 16 or 8 channels of 4-terminal gold flashed tellurium copper binding posts;
     Inputs/ Outputs:
                                                                                    Output: 2 channels of 4-terminal low thermal binding posts
     Communication                                                                  24 pin IEEE 488 connector
                                                                                    Each scanner is shipped with an external 5V DC supply that can be powered by
     Power Supply:
                                                                                    95~260V, 50/60Hz. Consumption 