                           Series 9334A

                          Resistance Standards
                      Very High Stability Calibration Laboratory Standards


                                                    G     uildline 9334A series of Resistance Standards are designed
                                                          as very high stability calibration laboratory standards for
                                                    high accuracy resistance calibration in air, without the need for
                                                    stabilization in a temperature controlled bath.
                                                    They can be used as working standards or highly reliable and
                                                    rugged transportable transfer standards. They are extremely
                                                    useful for the calibration of resistance ranges of multi-function
                                                    calibrators and high accuracy DVMs, as well as being used in
                                                    more classical standards and calibration laboratory
                                                    applications.
                                                    Hysteresis error, is typically better than negligible to 0.1 ppm
                                                    when stressed at three times the maximum voltage, and less
                                                    than 0.3 ppm over a temperature cycle between 0 0C and 40 0C.
                                                    Connections to these resistance standards are by conventional
                                                    4-terminal connection up to 1M  and two terminal
                                                    connection for values above 1M . The model 9334AH-1G and
                                                    the 9334AW-1G are designed to answer a difficult calibration
                                                    need for owners of the HP/Agilent 3458A and the
                                                    Wavetek/Datron 1271 and 1281 model DMMs.


9334A Features                                   The 9334A Series Precision Resistance Standards
                                                 are available in a wide range of off the shelf and
>   Stability < 2.5 ppm
                                                 custom values to satisfy demanding applications
>   Oil or Air Baths not required
                                                 between 1