   HP 4396A Network/Spectrum Analyzer
Performance Test Program (PN: 04396-65001)
      Operation Manual




              ABCDE
            HP Part No. 04396-90106
           Printed in JAPAN   July 1999
Notice

The information contained in this document is subject to change without notice.
This document contains proprietary information which is protected by copyright. All rights are
reserved. No part of this document may be photocopied, reproduced, or translated to another
language without the prior written consent of the Hewlett-Packard Company.

  Hewlett-Packard Japan, LTD.
  Kobe Instrument Division
  1-3-2, Murotani, Nishi-ku, Kobe-shi,
  Hyogo, 651-2241 Japan





 Copyright 1993,1999 Hewlett-Packard Japan, LTD.
c
Contents
 1.   General Information
      Introduction . . . . . . . . . . . . . . . . . .         .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-1
      Initial Inspection . . . . . . . . . . . . . . . .       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-1
      Making Working Copies . . . . . . . . . . . . .          .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-2
      Required Tools . . . . . . . . . . . . . . . . .         .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-2
        Hardware Requirements . . . . . . . . . . .            .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-2
        Software Requirements . . . . . . . . . . .            .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-2
      Test Equipment . . . . . . . . . . . . . . . .           .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-3
        Calibration Data Required for Step Attenuator .        .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     1-4
 2.   Preparation
      Test Equipment Conguration . . . . . . . . . . . . . . . . . . . . . . . .                                            2-1
      System Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                                           2-2
 3.   Operation
      Keyboard and Mouse Operation . . . . . . .       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     3-1
      Program Execution . . . . . . . . . . . .        .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     3-2
      Built-in Calibration Data Editor . . . . . . .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     3-3
      Program Flow . . . . . . . . . . . . . . .       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     3-4
        Main Program Flowchart . . . . . . . . .       .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     3-4
        Built-in Calibration Data Editor Flowchart .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .     3-6




                                                                                                                       Contents-1
Figures
    2-1. Performance Test Hardware Setup . . . . . . . . . . . . . . . . . . . . .                                        2-2
    3-1. Performance Test Program Flowchart . . . . . . . . . . . . . . . . . . . .                                       3-4
    3-2. Built-in Calibration Data Editor Flowchart . . . . . . . . . . . . . . . . . .                                   3-6



Tables
    1-1.   Required Equipment . . . . . . . . . . .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   1-2
    1-2.   Required Language Extensions . . . . . .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   1-2
    1-3.   Recommended Test Equipment . . . . . .     .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   1-3
    1-4.   Calibration Data Required for HP 8496A/G   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   1-4
    1-5.   Calibration Data Required for HP 8494A/G   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   .   1-5




Contents-2
                                                                                                 1
General Information
Introduction
The HP 4396A Performance Test Program (P/N: 04396-65001) is written in HP BASIC, designed
to minimize the test time, and helps eliminate human error while performance testing the
HP 4396A Network/Spectrum Analyzer. When executing the program, the computer displays
the test instructions, and controls the HP 4396A under test and the related test equipment.
This performance test program has the \Built-in Calibration Data Editor" which edits and stores
the standards' calibration data needed in the tests.


Initial Inspection
The Performance Test Program for HP 4396A has been carefully inspected before being shipped
from the factory. Verify that the shipping container contains the following:
  2 ea. of 3-1/2 inch diskettes
  Operation Manual (this book)
The diskettes should contain the following. Check the contents of the diskette using your
computer CAT command:
Filename        Description

TEST4396A       Main Program
SUBT4396A       Subprogram
CSUBS           Compiled Utilities for BASIC 5.13
CSUBS6          Compiled Utilities for BASIC 6.2 and above
TE_T4396A       Test Equipment Conguration Program




                                                                           General Information   1-1
                                                                                           HP 4396A


Making Working Copies
Copy the contents of the performance test program diskette to a working diskette or to the
hard disk to prevent accidental deletion or destruction of the program les. Use the working
diskettes or the hard disk, and store the original diskette in a safe place.
Note               Make an exclusive directory for this program and copy the les of the diskette
                   into the directory when you use a hard disk.




Required Tools
Hardware Requirements

Table 1-1 lists the equipment required for executing the programs.
                                       Table 1-1. Required Equipment


                 Equipment               Requirements      Recommended Models         Qty.

            Computer                 No Substitute    HP 9000 series 200 or 300   1    1
                                                      RAM  4 MBytes
            Operating System BASIC 5.1 or higher HP 98616A                             1
            Flexible Disk Drive Dual 3-1/2 inch Drive HP 9122C                         1
            Printer             HP-IB Printer         HP 2225AJ                        1
            1 Excluding the HP 9826A


Software Requirements

Table 1-2 lists the required language extensions to be used with the performance test program.
If you already have a pre-congured BASIC system, you can verify if it contains all the
required binaries by typing LIST BIN.
                                Table 1-2. Required Language Extensions

              Name                Description            Name          Description
           CLOCK            Clock                       CRTX     CRT Extensions
           EDIT             List and Edit               FHPIB    Fast Disc Interface Driver
           GRAPHX           Graphics Extensions         IO       I/O
           MAT              Matrix Statements           PDEV     Program Development
           CRTA             Alpha CRT Driver            CS80     CS80 Disc Driver
           ERR              Error Messages              GRAPH    Graphics
           HPIB             HPIB Interface Driver       KBD      Keyboard Extensions
           MS               Mass Storage




1-2   General Information
HP 4396A


Test Equipment
Table 1-3 lists the required equipment to perform the performance test.
                                Table 1-3. Recommended Test Equipment

                   Equipment Name                                      Model or HP Part Number                Qty
   Frequency Counter                                    HP 5343A Opt. 001 or HP 5334B Opt.010 or                1
                                                        HP 5335A Opt.0101
   Frequency Standard2                                  HP 5061B                                                1
   Spectrum Analyzer                                    HP 8566A/B                                              1
   Network Analyzer                                     HP 8753A/B/C                                            1
   Power Meter                                          HP 436A Opt. 022 or HP 437B or HP 438A                  1
   Power Sensor                                         HP 8482A                                                1
   Power Sensor                                         HP 8481D                                                1
   Function Genarator                                   HP 3325A                                                1
   Signal Generator3                                    HP 8663A or HP 8642B                                    2
   Step Attenuator4                                     HP 8496A/G Option 001 and H605                          1
   Step Attenuator4                                     HP 8494A/G Option 001 and H606                          1
   Attenuator/Switch Driver                             HP 11713A 7                                             1
   50
 Type-N Calibration Kit                           HP 85032B                                               1
   T/R Test Set                                         HP 85044A                                               1
   50 MHz Low Pass Filter                               PN 0955-0306                                            1
   50
 termination, type-N(m)                           HP 909C Opt 012 or part of HP 85032B8                   3
   6 dB Fixed Attenuation, 50 
                         HP 8491A Opt 006                                        2
   6 dB Fixed Attenuation, 50 
, VSWR  1.015            HP 8491A Opt 006 & Opt H609                             2
   Two-way Power Splitter                               HP 11667A                                               1
   N(m)-N(m) cable, 50 
                                HP 11500B or part of HP 11851B 10                       4
   RF cable kit                                         HP 11851B                                               1

   1 Option 001 or 010 (optional time base) is not required, when a frequency standard in Table 1-3 is available.
   2 Required for testing an analyzer equipped with Option 1D5 (High Stability Frequency Reference).
   3 The \HPxxxx 2" type model number is assigned to the \Signal Generator 2", while the nomal \HPxxxx" is
    used for the \Signal Generator 1". This makes us possible to use two signal generators of the same model.
   4 Calibration values at 50 MHz are required in the tests. See the   Calibration Data Required for Step
    Attenuators  later in this chapter.
   5 An HP 8496A/G step attenuator with required low VSWR ( 1.02) can be purchased by specifying option H60.
   6 An HP 8494A/G step attenuator with required low VSWR ( 1.02) can be purchased by specifying option H60.
   7 Required when an HP 8494G or HP 8496G step attenuator is used in the tests.
   8 The HP 85032B includes a type-N(m) 50 
 termination.
   9 An HP 8491A Opt. 006 xed attenuator with required low VSWR ( 1.015) can be purchased by specifying
     Opt. H60.
  10 The HP 11851B includes three N(m)-N(m) cables of 61 cm and a N(m)-N(m) cable of 88 cm.




                                                                                               General Information   1-3
                                                                                               HP 4396A

                          Table 1-3. Recommended Test Equipment (continued)

                      Equipment Name                                 Model or HP Part Number     Qty
      BNC(m)-BNC(m) cable, 61 cm, 50 
                    PN 8120-1839                            1
      BNC(m)-BNC(m) cable, 122 cm, 50 
                   PN 8120-1840                            2
      BNC(f)-BNC(f) adapter, 50 
                         PN 1250-0080                            1
      BNC(f)-SMA(f) adapter, 50 
                         PN 1250-0562                            1
      Tee BNC(m)-(f)-(f) adapter, 50 
                    PN 1250-0781                            1
      N(m)-N(m) adapter, 50 
                             PN 1250-1475                            1
      N(m)-BNC(f) adapter, 50 
                           PN 1250-1476                            1
      N(f)-BNC(m) adapter, 50 
                           PN 1250-1477                            1
      APC 3.5(m)-APC 3.5(f) adapter, 50 
                 PN 1250-1866                            1
      APC 7-N(f) adapter, 50 
                            HP 11524A or part of HP 85032B 1        1

      1 The HP 85032B includes two APC 7-N(f) adapters.


Calibration Data Required for Step Attenuator

The six performance tests listed below measure the analyzer's performance against a known
standard (the attenuation values at a frequency 50 MHz of the HP 8496A/G and HP 8494A/G
step attenuators).
   3. Non-Sweep Power Linearity Test
   4. Power Sweep Linearity Test
   10. Magnitude Ratio/Phase Dynamic Accuracy Test
   14. Amplitude Fidelity Test
   15. Input Attenuator Switching Uncertainty Test
   18. IF Gain Switching Uncertainty Test
These tests require the calibrated values of the attenuators listed in Table 1-4 and Table 1-5.
Enter these attenuation values (referenced to 0 dB setting) at the beginning of the program
with following instructions. See \Built-in Calibration Data Editor" in Chapter 3 for details.
                          Table 1-4. Calibration Data Required for HP 8496A/G

                                     Frequency Attenuation Uncertainty
                                         50 MHz             10 dB     0.0060 dB
                                                            20 dB     0.0060 dB
                                                            30 dB     0.0066 dB
                                                            40 dB     0.0090 dB
                                                            50 dB     0.0165 dB
                                                            60 dB     0.0197 dB
                                                            70 dB     0.0272 dB




1-4    General Information
HP 4396A

                   Table 1-5. Calibration Data Required for HP 8494A/G

                            Frequency Attenuation Uncertainty
                               50 MHz           2 dB     0.007 dB
                                                4 dB     0.007 dB
                                                6 dB     0.007 dB
                                                8 dB     0.007 dB
                                               10 dB     0.007 dB
The calibration uncertainty is the primary source of measurement error in performance tests.
The measurement uncertainties listed in the performance test record are valid only when the
uncertainty of the step attenuation data satises that given in the third column of Table 1-4
and Table 1-5.
The calibration of step attenuators, HP 8496A/G and HP 8494A/G, are available at
Hewlett-Packard. For information about the calibration and the available uncertainties, contact
your nearest Hewlett-Packard service center.




                                                                           General Information   1-5
                                                                                              2
Preparation
Test Equipment Conguration
This procedure is used to select the equipment and to set their HP-IB addresses. When you
perform the Performance Test Program at the rst time, perform this Equipment Conguration
Program to select the equipment and the HP-IB address. After that, only when you want to
change the equipment or the HP-IB address, perform this program.
1. Set the MSUS (Mass Strage Unit Specier) to the working directory or drive using the MSI
   command.
2. Load the Test Equipment Conguration Program, \TE T4396A" by pressing LOAD
   "TE_T4396A" 4Return5 (or 4ENTER5) key.
3. Execute the program by pressing RUN, and follow the instructions on the computer's screen.
4. The BDAT les which contain the test equipment conguration information will be created
   in the working directory / drive.




                                                                                Preparation   2-1
                                                                    HP 4396A


System Setup
The system setup is shown in Figure 2-1




                      Figure 2-1. Performance Test Hardware Setup




2-2   Preparation
                                                                                             3
Operation
This chapter describes a program 
owchart, and how to operate the program.



Keyboard and Mouse Operation
The menus in \TEST4396A" use a window format. It supports keyboard and mouse operations.
The operations are as follows.
  Keyboard Operation
  1. Press 485/495 cursors until your preference is highlighted.
  2. Choose the highlighted item by pressing 4Return5 or 4Select5 (4ENTER5 or 4EXECUTE5).
  3. If QUIT or EXIT is displayed in a menu, select one of these to exit the menu. Otherwise,
       NNNNNNNNNNNNNN   NNNNNNNNNNNNNN




     press 45 (or 4CONTINUE5) key to exit. When you exit menus, the program displays next
     menu.
Note             Press 4?5 to access on-screen help information for the selection you have
                 highlighted. Help information appears in a display window.
                 Press 4Return5 or 4Select5 (4ENTER5 or 4EXECUTE5) to turn o the help screen.

 Mouse Operation
 1. Slide the mouse up or down until your preference is highlighted.
 2. Choose the highlighted item by pressing left-hand button on the mouse, or slide the mouse
    to the right.
 3. If QUIT or EXIT is displayed in a menu, select one of these to exit the menu. Otherwise,
       NNNNNNNNNNNNNN   NNNNNNNNNNNNNN




    slide the mouse to the left to exit. When you exit menus, the program displays next
    menu.
Note           Press the right-hand mouse button to access on-screen help information for the
               selection you have highlighted. Help information appears in a display window.
               Press the left-hand mouse button to turn o the help screen.




                                                                                 Operation   3-1
                                                                                                                                 HP 4396A


Program Execution
Perform the following steps to start the performance test program.
Note                          In the program, the HP 4396A under test is referred as \UUT".



Note                          Perform the Test Equipment Conguration prior to execution. Refer to \Test
                              Equipment Conguration" in Chapter 2.


Main Program Instruction

 1. Turn on the computer and bring up the BASIC system. Refer to the Table 1-2 for the
    required language extensions.
 2. Set the Mass Storage Unit Specier to the drive/directory where the adjustment program
    exists using the MSI command.
 3. Load the program, \TEST4396A" by pressing LOAD "TEST4396A" 4Return5 (or 4ENTER5).
 4. Execute the program by pressing RUN.
 5. The copyright will be displayed on the computer's screen. Press any key to continue.
 6. The UUT conguration window \UUT: HP4396A" will be opened. Enter the UUT (the
    HP 4396A under test) conguration (serial number, HP-IB address, environment, and line
    frequency), and press 45 (or 4CONTINUE5) key to continue.
 7. The program will ask \Where should Test Reports be printed?". Select CRT , PRINTER ,                         NNNNNNNNNNN   NNNNNNNNNNNNNNNNNNNNNNN




    or NO OUTPUT .
        NNNNNNNNNNNNNNNNNNNNNNNNNNNNN




                              If the performance test record is necessary, then select PRINTER and connect a
                                                                                                NNNNNNNNNNNNNNNNNNNNNNN

Note
                              printer which is addressed 701 to the computer by HP-IB cable.


 8. The used equipment list will be displayed. Select CONTINUE to go to next step, or PRINT
                                                                   NNNNNNNNNNNNNNNNNNNNNNNNNN                                       NNNNNNNNNNNNNNNNN




    to print the used equipment list to the printer.
 9. The used accessory list will be displayed. Select CONTINUE to go to next step, or PRINT to
                                                                 NNNNNNNNNNNNNNNNNNNNNNNNNN                                      NNNNNNNNNNNNNNNNN




    print the used accessories list to the printer.
10. The \TEST LIST" menu will be opened. Each test can be selected using 485/495 cursors and
    4Return5 (or 4ENTER5) key, or pressing 45 (or 4CONTINUE5) key to exit the program.


Note                          INITIAL SETUP will be performed automatically when you select any other test
                              before performing INITIAL SETUP.


Note                          INITIAL SETUP includes the Built-in Calibration Data Editor and the Standard
                              Selection screen. Refer to the next section for instruction of the Built-in
                              Calibration Data Editor.


3-2   Operation
HP 4396A


11. The program will invoke the Calibration Data Editor. Select the calibration standard's serial
    number to be used in the test among the listed serial numbers.



Built-in Calibration Data Editor
This section provides the instructions for the Built-in Calibration Data Editor.
Built-in Calibration Data Editor Instruction

1. The window \Enter the CAL FACTOR volume Msvs" will be opened. Enter the MSVS (Mass
   Storage Volume Specier) which the calibration data will be stored in.
Note            The directory must be already exist before executing the performance test
                program. Create directory rst which the calibration data should be stored in.


2. The menu for selecting ADD, EDIT or DELETE the standard will be opened. Select ADD for
                                                                                                                       NNNNNNNNNNN




   adding the new standard data, EDIT for editing an existing standard data, or DELETE for
                                                      NNNNNNNNNNNNNN                                      NNNNNNNNNNNNNNNNNNNN




   deleting an existing standard data.
                The SERIAL                                       eld cannot be modied when editing the calibration
                     NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

Note                                      NUMBER
                data.




                                                                                                                Operation            3-3
                                                                                   HP 4396A


Program Flow
Figure 3-1 shows the performance test program 
owchart. The following information gives a
brief explanation for each menu item. Figure 3-2 shows the Built-in calibration data editor

owchart.

Main Program Flowchart




                     Figure 3-1. Performance Test Program Flowchart




3-4   Operation
HP 4396A


UUT Conguration Input Input the UUT serial number, HP-IB address, installed options,
                           environment, and line frequency.
Destination of Test Report Select where the Test Report will be output. The output medium
Selection                  can be selected as CRT , PRINTER , or NO OUTPUT .
                                                      NNNNNNNNNNN   NNNNNNNNNNNNNNNNNNNNNNN   NNNNNNNNNNNNNNNNNNNNNNNNNNNNN




Equipment / Accessories Conrm equipment and accessories are correct. If necessary,
Conrmation                select PRINT to print out the list. Otherwise, select CONTINUE to
                                  NNNNNNNNNNNNNNNNN                                                                           NNNNNNNNNNNNNNNNNNNNNNNNNN




                           continue.
TEST LIST Menu             Each test can be selected by using 485/495 cursors and 4Return5 (or
                           4ENTER5) key. If necessary, you can invoke the Built-in Calibration
                           Data Editor by selecting \INITIAL SETUP".
                           Press 45 (or 4CONTINUE5) key to quit the program.
Built-in Calibration Data Input standards' calibration data which are needed to perform the
Editor                     performance test.
Standards Selection        Select the standards' serial numbers used in the performance test.
                           This list must be lled before performing each test.
                           Press 45 (or 4CONTINUE5) key to exit this screen.
Performing Each Test       Each test can be selected in the `TEST LIST' Menu.
Test Report Output         Outputs the Test Report to the medium that was selected at
                           \Destination of Test Report Selection".




                                                                                                                                      Operation            3-5
                                                                           HP 4396A


Built-in Calibration Data Editor Flowchart




                  Figure 3-2. Built-in Calibration Data Editor Flowchart




3-6   Operation
HP 4396A


Power Sensor or Step        Select
                                     NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

                                     POWER SENSOR                             or   STEP ATTENUATOR .
                                                                                   NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN



Attenuator
                           Selecting EXIT or pressing 45 (or 4CONTINUE5) key quits the Built-in
                                            NNNNNNNNNNNNNN




                           Calibration Data Editor.
Entering the MSVS to       Enter the MSVS (Mass Storage Volume Specier) which the
store the Calibration Data calibration data will be stored in. For example:
                                :CS80,702,1                                                   Flexible disk based system
                                /CAL_FACTORS:REMOTE                                           SRM based system
                           Default MSVS is current MSVS.
                           Press 45 (or 4CONTINUE5) key to exit this screen.
Add / Edit / Delete Menu You can select NEW CAL FACTOR FILE to add the new calibration
                                                        NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




                           data, or EDIT CAL FACTOR FILE to modify the existing calibration
                                         NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




                           data, or PURGE CAL FACTOR FILE to delete the existing calibration
                                         NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




                           data.
Entering / Editing the Tag Enter / edit the serial number, cal due date, trace number, and
Data                       press 45 (or 4CONTINUE5) key to exit this screen.
Existing Data Selection    A list of existing data is displayed. Select one of the data to be
                           edited / deleted using 485/495 cursors, and press 4Return5 (or 4ENTER5)
                           key to go to next step.
Editing the Calibrated     A screen that corresponds to the Calibration Report is displayed.
Values                     You can select the eld that you want to edit by using 485/495
                           cursors, and pressing 4Return5 (or 4ENTER5) key to start modify. Each
                           value is adjusted by 485/495 cursors.
                           Press 45 (or 4CONTINUE5) key to back to previous screen.