User's Guide

HP 8560 E-Series
Spectrum Analyzer




       HEWLETT
Fia    PACKARD

HP Part No. 08560-90146 Supersedes 08560-90110
Printed in USA December 1997
Notice   The information contained in this document is subject to change
         without notice.
         Hewlett-Packard makes no warranty of any kind with regard to this
         material, including, but not limited to, the implied warranties of
         merchantability and fitness for a particular purpose. Hewlett-Packard
         shall not be liable for errors contained herein or for incidental
         or consequential damages in connection with the furnishing,
         performance, or use of this material.




         @Copyright Hewlett-Packard Company 1990, 1991, 1992, 1993, 1996,
         1997
         All Rights Reserved. Reproduction, adaptation, or translation without
         prior written permission is prohibited, except as allowed under the
         copyright laws.
         1400 Fountaingrove Pkwy, Santa Rosa, CA 95403, USA
Certification   Hewlett-Packard Company certifies that this product met its
                published specifications at the time of shipment from the factory.
                Hewlett-Packard further certifies that its calibration measurements
                are traceable to the United States National Institute of Standards and
                Technology, to the extent allowed by the Institute's calibration facility,
                and to the calibration facilities of other International Standards
                Organization members.




Warranty        This Hewlett-Packard instrument product is warranted against defects
                in material and workmanship for a period of one year from date of
                shipment. During the warranty period, Hewlett-Packard Company
                will, at its option, either repair or replace products which prove to be
                defective.
                For warranty service or repair, this product must be returned to a
                service facility designated by Hewlett-Packard. Buyer shall prepay
                shipping charges to Hewlett-Packard and Hewlett-Packard shall pay
                shipping charges to return the product to Buyer. However, Buyer shall
                pay all shipping charges, duties, and taxes for products returned to
                Hewlett-Packard from another country.
                Hewlett-Packard warrants that its software and firmware designated
                by Hewlett-Packard for use with an instrument will execute
                its programming instructions when properly installed on that
                instrument. Hewlett-Packard does not warrant that the operation
                of the instrument, or software, or firmware will be uninterrupted or
                error-free.
                LIMITATION OF WARRANTY
                  The foregoing warranty shall not apply to defects resulting from
                  improper or inadequate maintenance by Buyer, Buyer-supplied
                  software or interfacing, unauthorized modification or misuse,
                  operation outside of the environmental specifications for the
                  product, or improper site preparation or maintenance.
                  NO OTHER WARRANTY IS EXPRESSED OR IMPLIED.
                  HEWLETT-PACKARD SPECIFICALLY DISCLAIMS THE IMPLIED
                  WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
                  PARTICULAR PURPOSE.
                EXCLUSIVE REMEDIES
                  THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND
                  EXCLUSIVE REMEDIES. HEWLETT-PACKARD SHALL NOT BE
                  LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR
                  CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT,
                  TORT, OR ANY OTHER LEGAL THEORY.




                                                                                        ...
                                                                                        Ill
Assistance   F'roduct maintenance agreements and other customer assistance
             agreements are available for Hewlett-Fhckard products.
             Rx- any assistance, contact gour nearest Hewlett-Rxkard Sales and
             Service OJice.




Cleaning     The instrument front and rear panels should be cleaned using a soft
             cloth with water or a mild soap and water mixture.




iv
Safety Symbols        The following safety symbols are used throughout this manual.
                      Familiarize yourself with each of the symbols and its meaning before
                      operating this instrument.

         Caution      The caution sign denotes a hazard. It calls attention to a procedure
                      which, if not correctly performed or adhered to, could result in
                      damage to or destruction of the instrument. Do not proceed beyond a
                      caution until the indicated conditions are fully understood and met.

        Warning       The warning sign denotes a hazard. It calls attention to a
                      procedure which, if not correctly performed or adhered to, could
                      result in injury or loss of life. Do not proceed beyond a warning
                      until the indicated conditions are fully understood and met.


       Instruction    The product is marked with this symbol when it is necessary for you
      Documentation   to refer to the instructions in the documentation.
          !
          A




                                                                                             V
General Safety Considerations
        Warning      w Before this instrument is switched on, make sure it has been
                       properly grounded through the protective conductor of the ac
                       power cable to a socket outlet provided with protective earth
                       contact.
                    n    Any interruption of the protective (grounding) conductor,
                        inside or outside the instrument, or disconnection of the
                        protective earth terminal can result in personal injury.
                    n   If this instrument is used in a manner not specified by
                        Hewlett-Packard Co., the protection provided by the instrument
                        may be impaired.
                    n   There are many points in the instrument which can, if
                        contacted, cause personal injury. Be extremely careful. Any
                        adjustments or service procedures that require operation of
                        the instrument with protective covers removed should be
                        performed only by trained service personnel.


        Caution     Before this instrum,ent is switched on, make sure its primary power
                    circuitry has been adapted to the voltage of the ac power source.
                    Failure to set the ac power input to the correct voltage could cause
                    damage to the instrument when the ac power cable is plugged in.
                    This product conforms to Enclosure Protection IP 2 0 according to
                    IEC-529. The enclosure protects against finger access to hazardous
                    parts within the enclosure; the enclosure does not protect against the
                    entrance of water.




vi
HP 8560 E-Series Spectrum Analyzer Documentation Description

         User's Guide   The HP 8560 E-Series User's tiide applies to the HP 85603,
                        HP 85613, HP 85623, HP 85633, HP 8564E, and HP 85653 spectrum
                        analyzers. The HP 8560 E-Series User's Guide includes information
                        about preparing the spectrum analyzer for use, spectrum analyzer
                        functions, common spectrum analyzer measurements, programming
                        fundamentals, and definitions for remote programming commands.
                        In addition to the user's guide, the following manuals are shipped
                        with your spectrum analyzer:

    Calibration Guide   The HP 8560 E-Series Calibration Guide includes the specifications
                        and characteristics, and information about how to test your
                        spectrum analyzer to determine if the spectrum analyzer meets its
                        specifications.

Quick Reference Guide
                        The HP 8560 E-Series Quick Reference Guide applies to the HP 85603,
                        HP 85613, HP 85623, HP 85633, HP 85643, and HP 85653 spectrum
                        analyzers. The HP 8560 E-Series Quick &fmence Guide includes
                        information about how to make a simple measurement with your
                        spectrum analyzer, briefly describes the spectrum analyzer functions,
                        and lists all the programming commands.




Manuals Available Separately

        Service Guide   The service guide provides information for servicing an instrument
                        to the assembly level. The manual includes instrument adjustments,
                        troubleshooting, major assembly replaceable parts lists, and
                        replacement procedures. For ordering information, contact a
                        Hewlett-Packard Sales and Service Office. This manual is not always
                        immediately available for new products.

Component-Level Information
                        The component-level information manual provides component level
                        information for the assemblies used in the instrument. Schematic
                        drawings, component locators and assembly parts list are provided for
                        the current vintage of assemblies. Component-level information is not
                        always immediately available for new assemblies.

How to Order Manuals    Each of the manuals listed above can be ordered individually. To
                        order, contact your local HP Sales and Service Office.




                                                                                              vii
How to Use This Guide

       Where to Start   If you are familiar with spectrum analyzers:
                        n    Briefly review the front and rear panel overview sections in
                            Chapter 1 for a short introduction to the HP 85603, HP 85613,
                            HP 85633, HP 85643, and HP 85653 spectrum analyzers.
                        n    If you want more information about a particular spectrum analyzer
                            function, refer to Chapter 4.
                        If you are not familiar with spectrum analyzers:
                        n    Read Chapter 1, "Quick Start Guide," which introduces you to the
                            spectrum analyzer.
                        w Continue on to Chapter 2, "Making Measurements," to gain
                          experience with spectrum analyzer measurements.

This manual uses the following conventions:
                         F r o n t - P a n e l Key)   This represents a key physically located on the
                                                      instrument.
                        Sof tkey                      This indicates a "softkey," a key whose label is
                                                      determined by the instrument's firmware.
                        Screen Text                   This indicates text displayed on the instrument's
                                                      screen.




 ...
VIII
Contents
           1. Quick Start Guide
               What You'll Find in This Chapter     . . . . . . . . . .    .      l-l
                 Introducing Your New Spectrum       Analyzer . . . . .     .     l-l
               Initial Inspection . . . . . . . .   . . . . . . . . . .    .      l-2
               Turning the Spectrum Analyzer On     for the First Time .   .      1-4
               Making a Basic Measurement . .       . . . . . . . . . .    .      1-6
               Reference Level Calibration . . .    . . . . . . . . . .    .    l-11
               Front Panel Overview . . . . . .     . . . . . . . . . .    .    1-12
                 Display Annotation . . . . . .     . . . . . . . . . .    .    1-15
               Rear Panel Overview . . . . . .      . . . . . . . . . .    .    1-17

           2. Making Measurements
              Example 1: Resolving Closely Spaced Signals (with
                  Resolution Bandwidth) . . . . . . . . . . . . .          .     2-2
                What Is Resolution Bandwidth? . . . . . . . . . .          .     2-2
                Spectrum Analyzer Function Used . . . . . . . . .          .     2-2
                Stepping Through a Measurement of Two Equal
                    Amplitude Signals . . . . . . . . . . . . . . .        .     2-2
                Stepping Through a Measurement of Two Signals of
                    Unequal Amplitude . . . . . . . . . . . . . .          .     2-3
              Example 2: Improving Amplitude Measurements with
                  Ampcor . . . . . . . . . . . . . . . . . . . . .         .     2-6
                What Is Ampcor? . . . . . . . . . . . . . . . . .          .     2-6
                Spectrum Analyzer Functions Used . . . . . . . .            .    2-6
                Stepping Through an Ampcor Measurement . . . .              .    2-6
                Ampcor Limitations . . . . . . . . . . . . . . . .         .     2-9
              Example 3: Modulation . . . . . . . . . . . . . . .          .    2-10
                What Is Modulation? . . . . . . . . . . . . . . .          .    2-10
                Spectrum Analyzer Functions Used . . . . . . . .            .   2-10
                Stepping Through Modulation Measurements . . . .            .   2-10
                  Amplitude Modulation . . . . . . . . . . . . .           .    2-10
                  Frequency Modulation . . . . . . . . . . . . .           .    2-12
              Example 4: Harmonic Distortion . . . . . . . . . . .         .    2-17
                What Is Harmonic Distortion? . . . . . . . . . . .          .   2-17
                Spectrum Analyzer Functions Used . . . . . . . .            .   2-17
                Stepping through the Fast Harmonic Measurements:
                    Procedure A . . . . . . . . . . . . . . . . .          .    2-17
                  Measure the 1 MHz fundamental and its first two
                      harmonics . . . . . . . . . . . . . . . . .          .    2-17
                  Place a second marker on the second harmonic . .          .   2-19
                  Find the harmonic distortion (method 1) . . . . .        .    2-19
                  Measure additional harmonics . . . . . . . . . .         .    2-19
                  Find the harmonic distortion (method 2) . . . . .        .    2-20
                  Plot the display . . . . . . . . . . . . . . . .         .    2-20
               An Alternative Harmonic Measurement Method:
                    Procedure B . . . . . . . . . . . . . . . . .          .    2-21

                                                                           Contents-l
                  Measure the 1 MHz fundamental . . . . . . . . .        2-21
                  Measure the second harmonic . . . . . . . . . . .      2-22
                  Find the harmonic distortion . . . . . . . . . . .     2-22
                  Measure additional harmonics . . . . . . . . . . .     2-22
               Percent of Harmonic Distortion . . . . . . . . . . .      2-23
             Example 5: Third-Order Intermodulation Distortion . . .     2-24
               What Is Intermodulation Distortion? . . . . . . . . .     2-24
               Spectrum Analyzer Functions Used . . . . . . . . .        2-24
               Stepping through the Measurement . . . . . . . . .        2-24
                  Set up the test . . . . . . . . . . . . . . . . . .    2-24
                  Reduce the frequency span . . . . . . . . . . . .      2-25
                  Center and adjust the signals on the display . . . .   2-25
                  Set the reference level . . . . . . . . . . . . . .    2-26
                  Maximize dynamic range . . . . . . . . . . . . .       2-27
                  Set a second marker and measure the distortion
                      product . . . . . . . . . . . . . . . . . . . .    2-27
                  Create a title for the measurement . . . . . . . .     2-28
                  Save the measurement information . . . . . . . .       2-29
             Example 6: AM and FM Demodulation . . . . . . . . .         2-30
               What is AM and FM Demodulation? . . . . . . . . .         2-30
               Spectrum Analyzer Functions Used . . . . . . . . .        2-30
               Stepping through the Measurement . . . . . . . . .        2-30
                  Set up the test . . . . . . . . . . . . . . . . . .    2-30
                  Set the start and stop frequencies . . . . . . . . .   2-31
                  Set a marker . . . . . . . . . . . . . . . . . . .     2-32
                  Set the time . . . . . . . . . . . . . . . . . . .     2-32
                  Demodulate the signal . . . . . . . . . . . . . .      2-32
             Example 7: Stimulus-Response Measurements . . . . .         2-33
               What Are Stimulus-Response Measurements? . . . . .        2-33
               Spectrum Analyzer Functions Used . . . . . . . . .        2-33
               Stepping through the Measurement . . . . . . . . .        2-34
                 Set up the test . . . . . . . . . . . . . . . . . .     2-34
                 Set the spectrum analyzer . . . . . . . . . . . . .     2-34
                 Calibrate . . . . . . . . . . . . . . . . . . . . .     2-38
                 Normalize . . . . . . . . . . . . . . . . . . . .       2-39
                 Measure . . . . . . . . . . . . . . . . . . . . .       2-40
                 Using Range Level versus Using Normalized
                     Reference Level . . . . . . . . . . . . . . . .     2-42
             Example 8: External Millimeter Mixers (Unpreselected) .     2-44
               Spectrum Analyzer Functions Used . . . . . . . . .        2-44
               Stepping through the Measurement . . . . . . . . .        2-44
                 Set up the equipment . . . . . . . . . . . . . . .      2-44
                 Select the Frequency Band . . . . . . . . . . . .       2-45
                 Save the average conversion-loss value . . . . . .      2-46
                 Signal Identification (unpreselected mixers only) . .   2-48
                 Identify signals with the frequency-shift method . .    2-48
                 Identify signals in wide frequency spans . . . . . .    2-49
                 Bias . . . . . . . . . . . . . . . . . . . . . . .      2-51
             Example 9: Adjacent Channel Power Measurement . . .         2-52
               What Is Adjacent Channel Power (ACP)? . . . . . . .       2-52
               Spectrum Analyzer Functions Used . . . . . . . . .        2-52
               Stepping Through a Basic ACP Measurement . . . . .        2-53
               ACP Analog Method Definition . . . . . . . . . . .        2-56
                 Adjacent Channel Power (ACP) Instrument Setup . .       2-58
                 Adjacent Channel Power (ACP) Computations . . .         2-59

Contents-2
     Stepping Through a Burst Signal ACP Measurement . .      2-62
     PDC Measurements of Mobiles According to the
         RCR-27B Standard . . . . . . . . . . . . . . . .     2-64
     PHS Measurements of Mobiles According to the RCR-28
         Standard . . . . . . . . . . . . . . . . . . . .     2-64
     NADC Measurements of Mobiles According to the IS-54
         Standard . . . . . . . . . . . . . . . . . . . .     2-66
   Example 10: Power Measurement Functions . . . . . .        2-68
     What are the Power Measurement Functions? . . . .        2-68
     Using Power Measurement Functions . . . . . . . .        2-68
       RMS Detection . . . . . . . . . . . . . . . . . .      2-68
       Making Carrier "On" Power Measurements . . . . .       2-69
       Making Carrier "Off" Power Measurements . . . . .      2-70
       Making CHAN PWR OVER BW Measurements . . .             2-70
     Spectrum Analyzer Functions Used . . . . . . . . .       2-70
     Stepping through a Carrier Power Measurement . . .       2-71
   Example 11: Time-Gated Measurement . . . . . . . .         2-72
     What Is Time-Gating? . . . . . . . . . . . . . . . .     2-72
     How Time-Gating Works . . . . . . . . . . . . . .        2-74
     Test Setup and Connection Diagram . . . . . . . . .      2-79
     Stepping through the Measurement of a Single RF-Pulse
         Train . . . . . . . . . . . . . . . . . . . . . .    2-80
     Using the Edge Mode and the Level Mode for Triggering    2-84
       Edge Mode . . . . . . . . . . . . . . . . . . . .      2-84
       Level Mode . . . . . . . . . . . . . . . . . . .       2-85
     Stepping through a Time-Gated Measurement . . . .        2-85
     Summary of Time-Gated Measurement Procedure . . .        2-91
     "Rules" for Making a Time-Gated Spectrum
         Measurement . . . . . . . . . . . . . . . . . .      2-92
   Example 12: Making Time-Domain Measurements with
      Sweep Delay . . . . . . . . . . . . . . . . . . .       2-95
   Example 13: Making Pulsed RF Measurements . . . . .        2-98
     What Is Pulsed RF? . . . . . . . . . . . . . . . . .     2-98
     Spectrum Analyzer Functions Used . . . . . . . . .       2-98
     Stepping Through Pulsed RF Measurements . . . . .        2-98
       Broadband Mode . . . . . . . . . . . . . . . . .       2-98
       Center Frequency, Sidelobe Ratio, and Pulse Width .   2-100
     Pulse Repetition Frequency (PRF) . . . . . . . . . .    2-101
     Peak Pulse Power and Desensitization . . . . . . . .    2-102

3. Softkey Menus
    Menu Trees . . . . . . . . . . . . . . . . . . . . .       3-1

4. Key Function Descriptions
    Key Descriptions . . . . . . . . . . . . . . . . . . .    4-17




                                                         Contents-3
             5. Programming
                Setup Procedure for Remote Operation         .    .    .    .    .    .    .   .     5-2
                Communication with the System . . . .       .     .    .    .    .    .    .    .    5-3
                Initial Program Considerations . . . . .    .     .    .    .    .    .    .    .    5-6
                Program Timing . . . . . . . . . . .       .     .    .    .    .    .    .    .     5-7
                Data Transfer to Computer . . . . . .       .     .    .    .    .    .    .    .   5-11
                Input and Output Buffers . . . . . . .     .     .    .    .    .    .    .    .    5-22
                Math Functions . . . . . . . . . . . .     .     .    .    .    .    .    .    .    5-25
                Creating Screen Titles . . . . . . . . .   .     .    .    .    .    .    .    .    5-29
                Generating Plots and Prints Remotely .      .     .    .    .    .    .    .    .   5-32
                Monitoring System Operation . . . . .      .     .    .    .    .    .    .    .    5-37

             6. Programming Command Cross Reference
                 Front Panel Key Versus Command . . . . . . . . .                          .    _    6-2
                 Programming Command Versus Front Panel Key . .                            .   .    6-16
                   HP 8566A and HP 8568A Compatible Commands .                             .   .    6-28
                   Mass Memory Module Commands . . . . . . . .                            .    .    6-30

             7.   Language Reference
                  Syntax Diagram Conventions . . . . . . . . . . . . .                               7-2
                    Query Responses . . . . . . . . . . . . . . . . . .                              7-3
                  Programming Commands . . . . . . . . . . . . . . .                                 7-8
                    ACPACCL Accelerate Adjacent Channel Power
                        Measurement . . . . . . . . . . . . . . . . . .                              7-9
                    ACPALPHA Adjacent Channel Power Alpha Weighting                                 7-10
                    ACPALTCH Adjacent Channel Power Alternate
                        Channels . . . . . . . . . . . . . . . . . . . .                            7-11
                    ACPBRPER Adjacent Channel Power Burst Period . .                                7-12
                    ACPBRWID Adjacent Channel Power Burst Width . .                                 7-13
                    ACPBW Adjacent Channel Power Channel Bandwidth .                                7-14
                    ACPCOMPUTE Adjacent Channel Power Compute . .                                   7-15
                    ACPFRQWT Adjacent Channel Power Frequency
                       Weighting . . . . . . . . . . . . . . . . . . . .                            7-16
                    ACPGRAPH Adjacent Channel Power Graph . . . . .                                 7-17
                    ACPLOWER Lower Adjacent Channel Power . . . . .                                 7-18
                    ACPMAX Maximum Adjacent Channel Power . . . . .                                 7-19
                    ACPMEAS Measure Adjacent Channel Power . . . . .                                7-20
                    ACPMETHOD Adjacent Channel Power Measurement
                       Method . . . . . . . . . . . . . . . . . . . . .                             7-22
                    ACPMSTATE Adjacent Channel Power Measurement
                       State . . . . . . . . . . . . . . . . . . . . . .                            7-24
                    ACPPWRTX Total Power Transmitted . . . . . . . .                                7-25
                    ACPRSLTS Adjacent Channel Power Measurement
                       Results . . . . . . . . . . . . . . . . . . . . .                            7-26
                    ACPSP Adjacent Channel Power Channel Spacing . . .                              7-29
                    ACPT Adjacent Channel Power T Weighting . . . . .                               7-30
                    ACPUPPER Upper Adjacent Channel Power . . . . .                                 7-31
                    ADJALL LO and IF Adjustments . . . . . . . . . . .                              7-32
                    ADJCRT Adjust CRT Alignment . . . . . . . . . . .                               7-33
                    ADJIF Adjust IF . . . . . . . . . . . . . . . . . .                             7-34
                    AMB Trace A Minus Trace B . . . . . . . . . . . .                               7-35
                    AMBPL Trace A Minus Trace B Plus Display Line . . .                             7-36
                    AMPCOR Amplitude Correction . . . . . . . . . . .                               7-37
                    AMPCORDATA Amplitude Correction Data . . . . . .                                7-38
                    AMPCORSIZE Amplitude Correction Data Array Size .                               7-40

Contents-4
AMPCORRCL Amplitude Correction Recall . . . . . .        7-41
AMPCORSAVE Amplitude Correction Save . . . . . .         7-42
ANNOT Annotation On/Off . . . . . . . . . . . . .        7-43
APB Trace A Plus Trace B . . . . . . . . . . . . .       7-44
AT Input Attenuation . . . . . . . . . . . . . . . .     7-45
AUNITS Absolute Amplitude Units . . . . . . . . . .      7-46
AUTOCPL Auto Coupled . . . . . . . . . . . . . .         7-47
AXB Trace A Exchange Trace B . . . . . . . . . . .       7-48
BLANK Blank Trace . . . . . . . . . . . . . . . .        7-49
BML Trace B Minus Display Line . . . . . . . . . .       7-50
CARROFF Carrier Off Power . . . . . . . . . . . .        7-51
CARRON Carrier On Power . . . . . . . . . . . . .        7-52
CF Center Frequency . . . . . . . . . . . . . . . .      7-53
CHANPWR Channel Power . . . . . . . . . . . . .          7-55
CHANNEL Channel Selection . . . . . . . . . . . .        7-56
CHPWRBW Channel Power Bandwidth . . . . . . . .          7-57
CLRW Clear Write . . . . . . . . . . . . . . . . .       7-58
CNVLOSS Conversion Loss . . . . . . . . . . . . .        7-59
CONTS Continuous Sweep . . . . . . . . . . . . . .       7-60
COUPLE Input Coupling . . . . . . . . . . . . . .        7-61
DELMKBW Occupied Power Bandwidth Within Delta
   Marker . . . . . . . . . . . . . . . . . . . . .       7-62
DEMOD Demodulation . . . . . . . . . . . . . . .          7-63
DEMODAGC Demodulation Automatic Gain Control . .          7-65
DEMODT Demodulation Time . . . . . . . . . . . .          7-66
DET Detection Modes . . . . . . . . . . . . . . . .       7-67
DL Display Line . . . . . . . . . . . . . . . . . .       7-69
DLYSWP Delay Sweep . . . . . . . . . . . . . . .          7-70
DONE Done . . . . . . . . . . . . . . . . . . . .         7-72
ERR Error . . . . . . . . . . . . . . . . . . . . .       7-73
ET Elapsed Time . . . . . . . . . . . . . . . . . .       7-75
EXTMXR External Mixer Mode . . . . . . . . . . .          7-76
FA Start Frequency . . . . . . . . . . . . . . . . .      7-77
FB Stop Frequency . . . . . . . . . . . . . . . . .       7-78
FDIAG Frequency Diagnostics . . . . . . . . . . . .       7-80
FDSP Frequency Display Off . . . . . . . . . . . .        7-81
FFT Fast Fourier Transform . . . . . . . . . . . . .      7-82
FOFFSET Frequency Offset . . . . . . . . . . . . .        7-84
FREF Frequency Reference . . . . . . . . . . . . .        7-85
FS Full Span . . . . . . . . . . . . . . . . . . . .      7-86
FULBAND Full Band . . . . . . . . . . . . . . . .         7-87
GATE Gate . . . . . . . . . . . . . . . . . . . . .       7-89
GATECTL Gate Control . . . . . . . . . . . . . . .        7-90
GD Gate Delay . . . . . . . . . . . . . . . . . . .       7-91
GL Gate Length . . . . . . . . . . . . . . . . . .        7-92
GP Gate Polarity . . . . . . . . . . . . . . . . . .      7-93
GRAT Graticule On/Off . . . . . . . . . . . . . . .       7-94
HD Hold . . . . . . . . . . . . . . . . . . . . . .       7-95
HNLOCK Harmonic Number Lock . . . . . . . . . .           7-96
HNUNLK Unlock Harmonic Number . . . . . . . . .           7-98
ID Output Identification . . . . . . . . . . . . . . .    7-99
IDCF Signal Identification to Center Frequency . . . .   7-100
IDFREQ Signal Identified Frequency . . . . . . . . .     7-101
IP Instrument Preset . . . . . . . . . . . . . . . .     7-102
LG Logarithmic Scale . . . . . . . . . . . . . . . .     7-105

                                                     Contents-5
             LN Linear Scale . . . . . . . . . . . . . . .    .     .     . 7-106
             MBIAS Mixer Bias . . . . . . . . . . . . . .     .     .     . 7-107
             MEANPWR Mean Power Measurement . . . . .         .     .     . 7-109
             MEAS Measurement Status . . . . . . . . . .      .     .     . 7-110
             MINH Minimum Hold . . . . . . . . . . . . .      .     .     . 7-111
             MKA Marker Amplitude . . . . . . . . . . . .     .     .     . 7-112
             MKBW Marker Bandwidth . . . . . . . . . .        .     .     . 7-113
             MKCF Marker to Center Frequency . . . . . .       .     .     . 7-114
             MKCHEDGE Marker to Channel Edges . . . . .        .     .     . 7-115
             MKD Marker Delta . . . . . . . . . . . . . .     .     .     . 7-116
             MKDELCHBW Delta Markers to Channel Power
                Bandwidth . . . . . . . . . . . . . . . .     .     .     .     7-117
             MKDR Reciprocal of Marker Delta . . . . . . .      .     .     .   7-118
             MKF Marker Frequency . . . . . . . . . . . .      .     .     .    7-119
             MKFC Frequency Counter . . . . . . . . . . .      .     .     .    7-120
             MKFCR Frequency Counter Resolution . . . . .      .     .    .     7-121
             MKMCF Marker Mean to the Center Frequency .        .    .     .    7-123
             MKMIN Marker to Minimum . . . . . . . . . .       .     .     .    7-124
             MKN Marker Normal . . . . . . . . . . . . .       .     .     .    7-125
             MKNOISE Marker Noise . . . . . . . . . . . .      .     .     .    7-127
             MKOFF Marker Off . . . . . . . . . . . . . .     .     .     .     7-128
             MKPK Peak Search . . . . . . . . . . . . . .     .     .     .     7-129
             MKPT Marker Threshold . . . . . . . . . . .       .     .     .    7-130
             MKPX Peak Excursion . . . . . . . . . . . .       .     .     .    7-132
             MKRL Marker to Reference Level . . . . . . .      .     .     .    7-134
             MKSP Marker Delta to Span . . . . . . . . . .     .     .     .    7-135
             MKSS Marker to Center Frequency Step-Size . .     .     .     .    7-136
             MKT Marker Time . . . . . . . . . . . . . .      .     .     .     7-137
             MKTRACK Signal Track . . . . . . . . . . . .      .     .     .    7-138
             ML Mixer Level . . . . . . . . . . . . . . .     .     .     .     7-139
             MXMH Maximum Hold . . . . . . . . . . . .         .     .     .    7-140
             MXRMODE Mixer Mode . . . . . . . . . . . .       .     .     .     7-141
             NORMLIZE Normalize Trace Data . . . . . . .       .     .     .    7-142
             NRL Normalized Reference Level . . . . . . .      .     .     .    7-144
             NRPOS Normalized Reference Position . . . . .     .     .     .    7-146
             OCCUP Percent Occupied Power Bandwidth . .        .     .     .    7-148
             OP Output Display Parameters . . . . . . . . .    .     .     .    7-149
             PLOT Plot Display . . . . . . . . . . . . . .    .     .     .     7-150
             PLOTORG Display Origins . . . . . . . . . . .     .     .     .    7-151
             PLOTSRC Plot Source . . . . . . . . . . . . .    .     .     .     7-152
             PP Preselector Peak . . . . . . . . . . . . .    .     .     .     7-154
             PRINT Print . . . . . . . . . . . . . . . . .    .     .     .     7-155
             PSDAC Preselector DAC Number . . . . . . . .      .     .     .    7-157
             PSTATE Protect State . . . . . . . . . . . . .   .     .     .     7-158
             PWRBW Power Bandwidth (Full Trace) . . . .        .     .     .    7-159
             RB Resolution Bandwidth . . . . . . . . . . .    .     .     .     7-160
             RBR Resolution Bandwidth to Span Ratio . . .      .     .     .    7-161
             RCLOSCAL Recall Open/Short Average . . . . .     .     .     .     7-162
             RCLS Recall State . . . . . . . . . . . . . .    .     .     .     7-164
             RCLT Recall Trace . . . . . . . . . . . . . .    .     .     .     7-165
             RCLTHRU Recall Thru . . . . . . . . . . . .      .     .     .     7-166
             REV Revision Number . . . . . . . . . . . .      .     .     .     7-168
             RL Reference/Range Level . . . . . . . . . .      .     .     .    7-169
             RLCAL Reference Level Calibration . . . . . .     .     .     .    7-172
             ROFFSET Amplitude Reference Offset . . . . .      .     .     .    7-174

Contents-6
     RQS Request Service Conditions . .     . . . . . . . . .                7-175
     SAVES Save State . . . . . . . .      . . . . . . . . .                 7-177
     SAVET Save Trace . . . . . . . .      . . . . . . . . .                 7-178
     SER Serial Number . . . . . . . .     . . . . . . . . .                 7-179
     SIGID Signal Identification . . . .    . . . . . . . . .                7-180
     SNGLS Single Sweep . . . . . . .      . . . . . . . . .                 7-181
     SP Frequency Span . . . . . . . .     . . . . . . . . .                 7-182
     SQUELCH Squelch . . . . . . . .       . . . . . . . . .                 7-184
     SRCALC Source Leveling Control .       . . . . . . . . .                7-186
     SRCCRSTK Coarse Tracking Adjust .      . . . . . . . . .                7-187
     SRCFINTK Fine Tracking Adjust . .      . . . . . . . . .                7-189
     SRCPOFS Source Power Offset . . .      . . . . . . . . .                7-191
     SRCPSTP Source Power Step . . .        . . . . . . . . .                7-192
     SRCPSWP Source Power Sweep . .         . . . . . . . . .                7-193
     SRCPWR Source Power . . . . . .       . . . . . . . . .                 7-194
     SRCTKPK Source Tracking Peak . .       . . . . . . . . .                7-196
     SRQ Service Request . . . . . . .     . . . . . . . . .                 7-197
     SS Center Frequency Step-Size . .     . . . . . . . . .                 7-198
     ST Sweep Time . . . . . . . . . .     . . . . . . . . .                 7-200
     STB Status Byte Query . . . . . .     . . . . . . . . .                 7-201
     STOREOPEN Store Open . . . . .        . . . . . . . . .                 7-202
     STORESHORT Store Short . . . . .      . . . . . . . . .                 7-203
     STORETHRU Store Thru . . . . .         . . . . . . . . .                7-205
     SWPCPL Sweep Couple . . . . . .       . . . . . . . . .                 7-207
     SWPOUT Sweep Output . . . . . .       . . . . . . . . .                 7-208
     TDF Trace Data Format . . . . . .     . . . . . . . . .                 7-209
     TH Threshold . . . . . . . . . .      . . . . . . . . .                 7-211
     TITLE Title Entry . . . . . . . .     . . . . . . . . .                 7-212
     TM Trigger Mode . . . . . . . . .     . . . . . . . . .                 7-214
     TRA/TRB Trace Data Input/Output .      . . . . . . . . .                7-215
     TRIGPOL Trigger Polarity . . . . .     . . . . . . . . .                7-2 17
     TS Take Sweep . . . . . . . . . .     . . . . . . . . .                 7-218
     TWNDOW Trace Window . . . . .         . . . . . . . . .                 7-219
     VAVG Video Average . . . . . . .      . . . . . . . . .                 7-220
     VB Video Bandwidth . . . . . . .      . . . . . . . . .                 7-221
     VBR Video Bandwidth to Resolution      Bandwidth Ratio                  7-223
     VIEW View Trace . . . . . . . .       . . . . . . . . .                 7-224
     VTL Video Trigger Level . . . . .     . . . . . . . . .                 7-225

8. Options and Accessories
    Options . . . . . . . . . . . . . . . . . . . . . . .                      8-l
    Accessories Available . . . . , . . . , . . . . . . . ,                    8-4

9. If You Have a Problem
   What You'll Find in This Chapter . . .     . . . .    .    .    .    .      9-l
      Read the Warranty . . . . . . . . .     . . . .    .    .    .    .      9-l
   Spectrum Analyzer Problems . . . . .       . . . .    .    .    .    .      9-2
      Blank Display . . . . . . . . . . .     . . . .    .    .    .    .      9-2
      Check the Basics . . . . . . . . . .    . . . .    .    .    .    .      9-2
      Hardware Problems . . . . . . . . .     . . . .    .    .    .    .      9-3
   HP 85629B Test and Adjustment Module        . . . .    .    .    .    .     9-4
      Running the Automatic Fault Isolation   Routine     .    .    .    .     9-4
   HP 85620A Mass Memory Module . . .          . . . .    .    .    .    .     9-5
   Replacing the Battery . . . . . . . . .    . . . .    .    .    .    .      9-6
   Power Requirements . . . . . . . . .       . . . .    .    .    .    .      9-7

                                                                        Contents-7
               Setting the Line-Voltage Selector Switch . . . .    .    .    .     9-7
               Checking the Fuse . . . . . . . . . . . . . .       .    .    .      9-8
               Power Cable . . . . . . . . . . . . . . . . .       .    .    .      9-8
             Procedures . . . . . . . . . . . . . . . . . . .      .    .    .    9-10
               Trace Alignment . . . . . . . . . . . . . . .       .    .    .    9-10
               Reference Level Calibration . . . . . . . . . .      .    .    .   9-10
               HP-IB Address Selection . . . . . . . . . . .        .    .    .   9-11
               Plotting and Printing Directly . . . . . . . . .     .    .    .   9-11
             Servicing the Spectrum Analyzer Yourself . . . .       .    .    .   9-12
             Calling Hewlett-Packard Sales and Service Offices .    .    .    .   9-13
             Returning Your Spectrum Analyzer for Service . .       .    .    .   9-13
               Original Packaging . . . . . . . . . . . . . .      .    .    .    9-13
               Other Packaging . . . . . . . . . . . . . . .       .    .    .    9-15
             Serial Numbers . . . . . . . . . . . . . . . . .      .    .    .    9-17
             Electrostatic Discharge . . . . . . . . . . . . .     .    .    .    9-18
               Reducing Potential for ESD Damage . . . . . .        .    .    .   9-19
               Static-Safe Accessories . . . . . . . . . . . .     .    .    .    9-19
             Error Messages . . . . . . . . . . . . . . . . .      .    .    .    9-20
               Viewing Multiple Messages . . . . . . . . . .       .    .    .    9-20
               Eliminating Error Messages . . . . . . . . . .      .    .    .    9-21
               Realigning the LO and IF . . . . . . . . . . .      .    .    .    9-21
               Error Code Listing . . . . . . . . . . . . . .      .    .    .    9-22

             Index




Contents-8
Figures
            l-l. Accessories Supplied . . . . . . . . . . . . . . . .        l-3
            l-2. Selecting the Correct Line Voltage . . . . . . . . . .      l-4
            l-3. 300 MHz Calibration Signal Connection . . . . . . . .       l-6
            l-4. Softkey Menu . . . . . . . . . . . . . . . . . . .          l-7
            l-5. 300 MHz Center Frequency . . . . . . . . . . . . .          l-7
            l-6. 20 MHz Frequency Span . . . . . . . . . . . . . .           l-8
            l-7. Activated Normal Marker . . . . . . . . . . . . . .         l-9
            l-8. -10 dBm Reference Level . . . . . . . . . . . . . .       l-10
            l-9. Peaked Signal to Reference Level . . . . . . . . . .      l-11
          l-10. Front Panel of an HP 8560 E-Series Spectrum Analyzer       1-12
          l-l 1. Display Annotation . . . . . . . . . . . . . . . . .      l-15
          1-12. Rear Panel Functions . . . . . . . . . . . . . . . .       l-17
            2-l. 1 kHz Signal Separation . . . . . . . . . . . . . . .      2-3
            2-2. 2 kHz Signal Separation . . . . . . . . . . . . . . .      2-3
            2-3. Bandwidth Shape Factor . . . . . . . . . . . . . .         2-4
           2-4. 100 kHz Bandwidth Resolution . . . . . . . . . . .          2-5
           2-5. 300 kHz Bandwidth Resolution . . . . . . . . . . .          2-5
           2-6. Ampcor Measurement Setup . . . . . . . . . . . .            2-7
           2-7. An Amplitude-Modulated Signal . . . . . . . . . . .        2-11
           2-8. Percentage of Modulation . . . . . . . . . . . . . .       2-11
           2-9. FM Deviation Test Setup . . . . . . . . . . . . . .        2-12
          2-10. Bessel Functions for Determining Modulation Index . .      2-13
          2-l 1. Markers Show Modulating Frequency . . . . . . . .         2-14
          2-12. A Frequency-Modulated Signal . . . . . . . . . . .         2-15
          2-13. FM Signal with Carrier at a Null . . . . . . . . . . .     2-15
          2-14. FM Signal with First Sidebands at a Null . . . . . . .     2-16
          2-15. Input Signal and Harmonics . . . . . . . . . . . . .       2-18
          2-16. Peak of Signal is Positioned at Reference Level for
                     Maximum Accuracy . . . . . . . . . . . . . . .        2-18
          2-17. Harmonic Distortion in dBc (marker threshold set to
                     -70 dB) . . . . . . . . . . . . . . . . . . . .       2-19
          2-18. Percentage of Distortion versus Harmonic Amplitude .       2-20
          2-19. Input Signal Displayed in a 1 MHz Span . . . . . . .       2-21
          2-20. Second Harmonic Displayed in dBc . . . . . . . . .         2-22
          2-21. Third-Order Intermodulation Test Setup . . . . . . .       2-25
          2-22. Signals Centered on Spectrum Analyzer Display . . .        2-26
          2-23. Signal Peak Set to Reference Level . . . . . . . . .       2-27
          2-24. Intermodulation Distortion Measured in dBc . . . . .       2-28
          2-25. Display with Title . . . . . . . . . . . . . . . . .       2-28
          2-26. AM and FM Demodulation Test Setup . . . . . . . .          2-30
          2-27. FM Band . . . . . . . . . . . . . . . . . . . . . .        2-31
          2-28. Place a marker on the signal of interest, then
                     demodulate. . . . . . . . . . . . . . . . . . .       2-32
          2-29. Block Diagram of a Spectrum Analyzer and Tracking
                     Generator Measurement System . . . . . . . . .        2-33
          2-30. Transmission Measurement Test Setup . . . . . . . .        2-34

                                                                      Contents.9
               2-31. Tracking-Generator Output Power Activated . . . . .        2-35
               2-32. Adjust spectrum analyzer settings according to the
                         measurement requirement. . . . . . . . . . . .         2-35
               2-33. Decrease the resolution bandwidth to improve
                         sensitivity. . . . . . . . . . . . . . . . . . . .     2-36
               2-34. Manual tracking adjustment compensates for tracking
                         error. . . . . . . . . . . . . . . . . . . . . . .     2-37
               2-35. Guided calibration routines prompt the user. . . . . .     2-38
               2-36. The thru trace is displayed in trace B. . . . . . . . .    2-38
               2-37. Normalized Trace . . . . . . . . . . . . . . . . . .       2-39
               2-38. Measure the rejection range with delta markers. . . .      2-40
               2-39. NORM REF LVL adjusts the trace without changing
                         analyzer settings. . . . . . . . . . . . . . . . .     2-41
               2-40. Increase the dynamic measurement range by using
                         RANGE LVL. . . . . . . . . . . . . . . . . . .         2-41
               2-41. Normalized Frequency Response Trace of a Preamplifier      2-42
               2-42. NORM KEF LVL is a trace function. . . . . . . . . .        2-43
               2-43. RANGE LVL adjusts analyzer settings for
                         compression-free measurements. . . . . . . . . .       2-43
               2-44. External Mixer Setup (a) without Bias; (b) with Bias .     2-45
               2-45. Select the band of interest. . . . . . . . . . . . . .     2-46
               2-46. Store and correct for conversion loss. . . . . . . . .     2-47
               2-47. Signal Responses Produced by a 50 GHz Signal in U
                         Band . . . . . . . . . . . . . . . . . . . . . .       2-48
               2-48. Response for Invalid Signals . . . . . . . . . . . . .     2-49
               2-49. Response for Valid Signals . . . . . . . . . . . . . .     2-49
               2-50. SIG ID AT MKR Performed on an Image Signal . . .           2-50
               2-51.   SIG ID AT MKA Performed on a True Signal . . . . .       2-50
               2-52.   Adjacent Channel Power Measurement Test Setup . .        2-53
               2-53.   Adjacent Channel Power Parameters . . . . . . . . .      2-54
               2-54.   Adjacent Channel Power Measurement Results . . . .       2-55
               2-55.   ACP Graph Display . . . . . . . . . . . . . . . . .      2-56
               2-56.   Trigger Configuration for Gated Method, Non-Option
                           OOl.......................                           2-66
               2-57.   Trigger Configuration for Gated Method, Option 001 .     2-67
               2-58.   Simplified Digital Mobile-Radio Signal in Time Domain    2-72
               2-59.   Frequency of the Combined Signals of the Radios . . .    2-73
               2-60.   Time-Gated Spectrum of Signal Number 1 . . . . . .       2-73
               2-61.   Time-Gated Spectrum of Signal Number 2 . . . . . .       2-73
               2-62.   Block Diagram of the Spectrum Analyzer with Time
                           Gate . . . . . . . . . . . . . . . . . . . . . .     2-74
               2-63.   Timing Relationship of Signals During Gating . . . . .   2-75
               2-64.   Using Time-Gating to View Signal 1 . . . . . . . . .     2-76
               2-65.   Using Time-Gating to View Signal 2 . . . . . . . . .     2-77
               2-66.   Connection Diagram for Time-Gated Spectrum
                           Measurements     . . . . . . . . . . . . . . . . .   2-79
               2-67.   Connection Diagram for Example . . . . . . . . . .       2-80
               2-68.   Frequency Spectrum of Signal without Gating . . . .      2-83
               2-69.   Oscilloscope Display . . . . . . . . . . . . . . . .     2-83
               2-70.   Spectrum Analyzer Display . . . . . . . . . . . . .      2-84
               2-71.   Using Positive or Negative Triggering . . . . . . . .    2-85
               2-72.   Time-domain Parameters . . . . . . . . . . . . . .       2-86
               2-73.   Positioning the Gate . . . . . . . . . . . . . . . .     2-87


Contents-l 0
2-74. Best Position for Gate . . . . . . . . . . . . . . . .    2-87
2-75. Setup Time for Interpulse Measurement . . . . . . .       2-88
2-76. Resolution Bandwidth Filter Charge-Up Effects . . . .     2-89
2-77. Gate Positioning Parameters . . . . . . . . . . . . .     2-92
2-78. Pulsed-RF Signal in Time Domain . . . . . . . . . .       2-95
2-79. Display of Zero-Span without Sweep Delay . . . . . .      2-96
2-80. Display of Zero-Span with Sweep Delay . . . . . . .       2-96
2-81. Main Lobe and Side Lobes . . . . . . . . . . . . . .      2-99
2-82. Trace Displayed as a Solid Line . . . . . . . . . . .     2-99
2-83. Center Frequency at Center of Main Lobe . . . . . .      2-100
2-84. Markers Show Sidelobe Ratio . . . . . . . . . . . .      2-100
2-85. Markers Show Pulse Width . . . . . . . . . . . . .       2-101
2-86. Measuring Pulse Repetition Frequency . . . . . . . .     2-102
 4-l. Channel Bandwidth Parameters . . . . . . . . . . .        4-23
 4-2. ACP Graph Display . . . . . . . . . . . . . . . . .       4-26
 4-3. CRT Alignment Pattern . . . . . . . . . . . . . . .       4-42
 4-4. Tracking Error . . . . . . . . . . . . . . . . . . .      4-60
 4-5. PEAK EXCURSN defines the peaks on a trace. . . . .        4-72
 5-l. HP 85603 connected to an HP 9000 Series 300
          computer. . . . . . . . . . . . . . . . . . . . .       5-2
 5-2. Invalid Trace Information . . . . . . . . . . . . . .       5-8
 5-3. Updated Trace Information . . . . . . . . . . . . .         5-9
 5-4. Update trace with TS before executing marker
          commands. . . . . . . . . . . . . . . . . . . .       5-10
 5-5. Data Transferred in TDF M Format . . . . . . . . .        5-16
 5-6. Buffer Summary . . . . . . . . . . . . . . . . . .        5-24
 5-7. Display Units . . . . . . . . . . . . . . . . . . . .     5-28
 5-8. Screen Titles Appear in the Upper-Right Corner of the
          Graticule . . . . . . . . . . . . . . . . . . . .     5-29
 5-9. PI and P2 Coordinates . . . . . . . . . . . . . . .       5-32
 7-l. Command Syntax Figure . . . . . . . . . . . . . .          7-2
 7-2. Numeric Value Query Response . . . . . . . . . . .         7-3
 7-3. Binary State Query Response . . . . . . . . . . . .        7-3
 7-4. CRT Alignment Pattern . . . . . . . . . . . . . . .       7-33
 7-5. MKPX Determines Which Signals are Considered Peaks       7-132
 9-l. Voltage Selection Switch and Line Fuse Locations . .       9-7
 9-2. AC Power Cables Available . . . . . . . . . . . . .         9-9
 9-3. CRT Adjustment Pattern . . . . . . . . . . . . . .        9-10
 9-4. Shipping Container and Cushioning Materials . . . . .     9-14
 9-5. Serial Number Label Example . . . . . . . . . . . .       9-17
 9-6. Example of a Static-Safe Workstation . . . . . . . .      9-18




                                                          Contents.1 1
Ihbles
                 l-l. Spectrum Analyzer Operating Range . . . . . . . . .          l-l
                 l-2. Front-Panel Connector Data . . . . . . . . . . . . .       1-14
                 2-l. Carrier Nulls and Modulation Indexes . . . . . . . .       2-13
                 2-2. Sideband Nulls and Modulation Indexes . . . . . . .        2-13
                 2-3. External Mixer Frequency Ranges . . . . . . . . . .        2-46
                 2-4. Adjacent Channel Power Measurement Parameters . .          2-62
                 2-5. Instrument Connections . . . . . . . . . . . . . . .       2-79
                 2-6. Determining Spectrum Analyzer Settings for Viewing a
                          Pulsed RF Signal . . . . . . . . . . . . . . . .       2-92
                2-7. Suggested Initial Settings for Known Pulse Width (7)
                          and Zero Signal Delay . . . . . . . . . . . . . .      2-93
                2-8. Suggested Sweep Times for a Known Pulse Repetition
                          Interval (PRI) or Pulse Repetition Frequency (PRF)     2-93
                2-9. If You Have a Problem with the Time-Gated
                          Measurement . . . . . . . . . . . . . . . . . .        2-94
                4-l. Fundamental Functions . . . . . . . . . . . . . . .          4-2
                4-2. Instrument State Functions . . . . . . . . . . . . .         4-4
                4-3. Marker Functions . . . . . . . . . . . . . . . . . .        4-12
                4-4. Control Functions . . . . . . . . . . . . . . . . .         4-14
                4-5. Flatness Points and Conversion Losses for External
                         Mixing Bands . . . . . . . . . . . . . . . . . .        4-40
                4-6. Mixing Harmonics for Unpreselected External Mixing .        4-58
                4-7. Instrument State after c-1 Is Executed
                                              .     .      .     .    .     .    4-79
                5-1. Units for Returned Data . . . . . . . . . . . . . .         5-12
                5-2. HP-IB Transmission Sequence for Different Trace-Data
                         Formats (TDF) . . . . . . . . . . . . . . . . .         5-21
                5-3. Scaling Points for Various Plotters . . . . . . . . . .     5-33
                5-4. Status Register . . . . . . . . . . . . . . . . . . .       5-39
                6-l. Front Panel Key Versus Command . . . . . . . . . .           6-2
                6-2. Programming Command Functional Index . . . . . .            6-17
                6-3. Mass Memory Module Commands . . . . . . . . . .             6-30
                7-l. Syntax Elements . . . . . . . . . . . . . . . . . .          7-4
                7-2. Characters and Secondary Keywords (Reserved Words)           7-5
                7-3. ACP Methods Versus Results (with 0 Alternate
                         Channels) . . . . . . . . . . . . . . . . . . . .       7-27
                7-4. Alternate Channels . . . . . . . . . . . . . . . . .        7-27
                7-5. Unpreselected External-Mixer Frequency Bands . . .          7-87
                7-6. Frequency Bands and the Corresponding LO Harmonic
                         For Unpreselected Mixers . . . . . . . . . . . .        7-96
                7-7. HP 8560 E-Series Preset States . . . . . . . . . . .       7-102
                7-8. Frequency Ranges and Minimum Reference Level (0 dB
                         Input Attenuation) . . . . . . . . . . . . . . .       7-170
                7-9. Status Register . . . . . . . . . . . . . . . . . . .      7-175
               7-10. Special Printing Characters . . . . . . . . . . . . .      7-213
                9-1. Problems and Possible Causes . . . . . . . . . . . .          9-3
                9-2. Operating Power Requirements . . . . . . . . . . .           9-7


Contents-l 2
9-3. Hewlett-Packard Sales and Service Offices . . . . . .     9-16
9-4. Static-Safe Accessories . . . . . . . . . . . . . . .     9-19




                                                        Contents-13
                                                                                               1
Quick Start Guide
What You'll Find in This Chapter
                       n    Initial Inspection
                       w Turning the Spectrum Analyzer On for the First Time
                       n    Making a Basic Measurement
                       n    Setting Reference Level Calibration
                       n    Front Panel Overview
                       w Rear Panel Overview



Introducing Your New Spectrum Analyzer


                                    `Ihble l-l. Spectrum Analyzer Operating Range

                            Spectrum               Amplitude Range       Frequency Range
                            Analyzer
                           HP 85603              -145 dBm to +30 dBm     30 Hz to 2.9 GHz

                           HP 85613              - 145 dBm to + 30 dBm   30 Hz to 6.5 GHz

                           HP 85623              -148 dBm to +30 dBm     30 Hz to 13.2 GHz

                           HP 85633              -148 dBm to +30 dBm     9 kHz to 26.5 GHz
                                                                         30 Hz to 26.5 GHz
                                                                         (with Option 006)
                           HP 85643              -147 dBm to +30 dBm      9 kHz to 40 GHz
                                                                          30 Hz to 40 GHz
                                                                         (with Option 006)
                           HP 85653              -147 dBm to +30 dBm      9 kHz to 50 GHz
                                                                          30 Hz to 50 GHz
                                                                         (with Option 006)




                                                                           Quick Start Guide   l-1
Initial Inspection                    Inspect the shipping container upon receipt. Retain it and the
                                      cushioning materials. If the container or cushioning material is
                                      damaged, verify that the contents are complete and that the analyzer
                                      functions correctly mechanically and electrically.
                                      If the contents are incomplete or the analyzer fails the verification
                                      tests in the calibration guide, notify one of the Hewlett-Packard
                                      Sales and Service Offices listed in `Pable 9-3. Show any container or
                                      cushioning materials damages to the carrier. The Hewlett-Packard
                                      Sales and Service Office will arrange for repair or replacement without
                                      waiting for a claim settlement.
                                      The shipping container and cushioning materials are shown in
                                      the "If You Have A Problem" chapter, Figure 9-4. Instructions for
                                      repackaging the analyzer are included in that chapter.


                                             Fart Numbers of Accessories Supplied
                                             Item                                 HP Part Number
                        Front cover                                                  5063-0274
                        Mass memory module                                          HP 85620A
                          (not included with Option 104)
                        BNC cable, 23 cm (9 in.)                                   HP 10502A
                        4 mm hex (Allen) wrench                                     8710-1755
                        Power cord                                                see Figure 9-2
                        Fuse: 5 A, 250 V (supplied in fuse holder)                  2110-0709
                        Adapter, Type-N (m) to SMA (f) (HP 85633 only)              1250-1250
                        Adapter, Type-N (m) to BNC (f)                              1250-0780
                          (not included with Option 026)
                          (2 per Option 002)
                        Adapter, SMA (m) to BNC (f)                                 1250-1200
                          (Option 026, HP 85643 or HP 85653 only)
                        Adapter, APC 3.5 (f) to APC 3.5 (f)                         5061-5311
                           (Option 026 only)
                        Adapter, K (f) to 2.4 mm (f)                                1250-2187
                           (HP 85643 or HP 85653 only)
                        Adapter, 2.4 mm (f) to 2.4 mm (f)                           1250-2188
                           (HP 85643 or HP 85653 only)
                        Termination, 50 D SMA (m)                                   1810-0118
                        Power holes will occur when using adapters and cables beyond their specified range.




1-2 Quick Start Guide
                            w   ADAPTER




                                                            BNC CABLE




HEX WRENCH




                                                                        50 OHM TERMINATION




                            MASS MEMORY MODULE




             * POWER CORD




                                                               ADAPTER




                               FRONT PANEL COVER




                     Figure 1-1. Accessories Supplied

                    * See Figure 9-2 for HP part numbers.




                                                                           Quick Start Guide   1-3
Turning the Spectrum Analyzer On for the First Time
                          The spectrum analyzer requires no installation other than connection
                          to an ac power source. If you want to install your spectrum analyzer
                          into an HP System II cabinet or a standard 19 inch (486.2 mm)
                          equipment rack, complete instructions are provided with the
                          Option 908 and Option 909 Rack mounting Kits.

                Caution   Do not connect the analyzer to an ac power source before verifying
                          that the line voltage is correct, the line-voltage selector switch located
                          on the analyzer rear panel is set to the correct voltage, and the proper
                          fuse is installed. Failure to verify that these items are correct could
                          result in equipment damage.




                                             FUSE IN USE

                                       \ SPARE   FUSE

                                     Figure 1-2. Selecting the Correct Line Voltage
                Caution   When installing the instrument in a cabinet, the convection into
                          and out of the instrument must not be restricted. The ambient
                          temperature (outside the cabinet) must be less than the maximum
                          operating temperature of the instrument by 4 degrees C for every
                          100 watts of power dissipated in the cabinet. If the total power
                          dissipated in the cabinet is greater than 800 watts, then forced
                          convection must be used.




1-4 Quick Start Guide
1. Press (LINE] to turn the analyzer on.
2. The analyzer takes about half a minute to perform a series of
   self-diagnostic and adjustment routines. At completion, the screen
   displays the analyzer model number and the firmware date (for
   example, 890802 indicates August 2, 1989). If you should ever need
   to call Hewlett-Packard for service or have any questions regarding
   your analyzer, it is helpful to know the firmware date to get the
   most accurate information.
3. Allow the analyzer a 5-minute warmup. Refer to the warmup
   specification in the calibration guide.
4. The reference level calibration adjustment procedure in this
   chapter should be performed before checking the specifications.
5. The trace alignment procedure can be per