Text preview - extract from the document
Keysight Technologies
DSOX3PWR/DSOX4PWR/DSOX6PWR
Power Measurement Options
Data Sheet



             For InfiniiVision 3000, 4000 and
             6000 X-Series Oscilloscopes
Achieve cost-effective analysis of your switching mode power supply (SMPS)
characteristics
Today's power supply designers are            constraints translate into more time      The Keysight Technologies, Inc.
facing an increasing number of                dedicated to power device measure-        DSOX3PWR, DSOX4PWR, and
constraints in the development of             ment and analysis for today's power       DSOX6PWR are power measurement
high-efficiency, low-cost power sup-          supply designers.                         and analysis options that are integrated
plies. Cost-effective solutions used to                                                 into InfiniiVision 3000, 4000, and 6000
be the designer's key target. Today,          In spite of the increasing analysis       X-Series scopes. The embedded appli-
rising energy costs bring power supply        capability offered by many oscil-         cation provides a quick and easy way of
efficiency to the forefront. Additionally,    loscopes over recent years, it is not     analyzing the reliability and efficiency
other constraints such as design              uncommon to see designers perform         of your switching power supply.
compactness, migration to digital             measurements and analysis manually.
control, tighter voltage tolerances, and      These measurements typically take         These power measurement options
regulations for power quality and EMI         a considerable amount of time to          also come with the user license for
force the need for quick and thorough         capture, analyze and report.              the U1881A-003 PC-based power mea-
power supply testing. Increasing design                                                 surement and analysis software that
                                                                                        provides even more powerful insight
                                                                                        into your power supply measurements.

 Type of analysis           Measurements/feature                                Keysight power measurement application for
                                                                                InfiniiVision 3000, 4000, and 6000 X-Series
                                                                                DSOX3PWR/DSOX4PWR/ U1881A-003 PC based
                                                                                DSOX6PWR integrated
 Input line analysis        Current harmonics                                                                 
                            Power factor                                                                      
                            True power                                                                        
                            Apparent power                                                                    
                            Crest factor                                                                      
                            Phase angle                                         
                            Pre-compliance IEC61000-3-2 (class A, B, C and D)                                 
                            RTCA D0-160E                                                                      
                            Inrush current                                                                    
 Power device analysis      Switching loss                                                                    
                            Safe Operating Area (SOA)                                                         
                            SOA mask editor                                                                   
                            Dynamic on resistance (Rds)                                                       
                            dV/dt slew rate                                                                   
                            dI/dt slew rate                                                                   
 Modulation analysis        Pulse width vs. time                                                              
                            Duty cycle vs. time                                                               
                            Period vs. time                                                                   
                            Frequency vs. time                                                                
 Output analysis            Output ripple                                                                     
 Deskew                     Auto deskew                                                                       
 Turn on/off analysis       Turn on time                                                                      
                            Turn off time                                                                     
 Transient response         Transient response                                                                
 analysis
 Efficiency analysis        Pout/pin analysis                                                            
 PSRR analysis              PSRR analysis                                       
 On/off line analysis       On/off line analysis                                                         
 Report generation          Report generation                                                            

2
Power device analysis
The switching loss in a power supply determines its effi-
ciency. You can easily characterize for instantaneous power
loss and conduction power loss at the switching device over
a designated switching cycle. To determine the efficiency of
the power supply it is very important to measure the power
loss during dynamic load changes.

By measuring the switching loss and conduction loss, you
can characterize the instantaneous power dissipation in your
switching power supply. Locating peak switching loss helps
you analyze the reliability of the power supply. The di/dt and
dv/dt represent the rate at which the current and voltage         Figure 1. By measuring the switching loss and conduction loss,
change at switching. This helps in analysis of reliable           you can characterize the instantaneous power dissipation in a
operation of the switching mode power supply.                     switching power supply.




Line power analysis
Power supply designers need to characterize the line power
for power quality, harmonics and conducted emissions
under different operating conditions of the power supply.
Some of the implicit measurements are real power, apparent
power, reactive power, power and crest factor and graphical
display of harmonics with respect to standards such as EN
61000-3-2 (Class A,B,C,D). By using a current probe and
power measurement option, conducted power line harmonics
can be measured. Also, line power analysis includes the
inrush current measurement that shows the peak inrush
                                                                  Figure 2. Perform the pre-compliance line harmonic testing of
current value when the power supply is first turned on.
                                                                  your power supply to the IEC 61000 3-2 standards. This analysis
                                                                  presents up to 40 harmonics.




                                                                 Figure 3. The Inrush current analysis measures the peak inrush
                                                                 current of the power supply when the power supply is first turned on.



                                                                                                                                         3
Power quality analysis
The Power Quality analysis shows the quality of the AC
input line. Some AC current may flow back into and back
out of the load without delivering energy. This current, called
reactive or harmonic current, gives rise to an "apparent"
power which is larger than the actual power consumed.
Power quality is gauged by these measurements: power
factor, apparent power, true power, reactive power, crest
factor, and phase angle of the current and voltage of the
AC line.

                                                                  Figure 4. The power measurements option provides a results
                                                                  table with the following power quality measurements: Power
                                                                  Factor, Real Power, Apparent Power, Reactive Power, Crest
                                                                  Factor and Phase Angle.




Modulation analysis
Modulation analysis allows designers to quickly see the
on-time and off-time information of the PWM signal, which
is difficult to visualize because the information bandwidth
is much lower than the pulse switching frequency. Plotting
the embedded variation of on time or off time in the PWM
signal over a long period of time can reveal the control loop
response of the feedback loop system. This measurement
performs data trending on the switching variation of the
acquired waveform in the following format.