5991-3298EN In Situ Electrochemical Measurements Using the 7500 AFM - Application Note c20141020 [4]


>> Download 5991-3298EN In Situ Electrochemical Measurements Using the 7500 AFM - Application Note c20141020 [4] documenatation <<

Text preview - extract from the document
Keysight Technologies
In Situ Electrochemical Measurements
Using Keysight 7500 AFM




                                                        Application Brief




   Introduction                                         control both the potential of the sample and the
                                                        potential of the probe versus the same refer-
   EC-SPM essentially combines two valuable,            ence electrode. As with usual STM imaging, the
   independent techniques: electrochemistry (EC)        tunneling current between the tip and the sample
   and scanning probe microscopy (SPM). The             depends on the potential difference as well as
   electrochemical unit includes a potentiostat and     the distance between the two, and is used as
   a three-electrode cell that controls the electro-    the control signal for STM image formation. In
   chemical state of the working electrode, usually     EC-STM, the morphological information about
   the sample. The scanning probe microscope            the electrode surface under potential control can
   characterizes the surface of the solid electrode     be imaged in constant-current mode, and the
   with either a passive or an active probe.            changes in the localized electronic state of the
                                                        electrode surface with electrochemical poten-
   With a passive probe like that utilized in EC-       tial can be studied using "current vs. voltage"
   AFM, the potential of the probe is not controlled.   spectroscopy.
   Thus, the AFM cantilever acts as an inert probe
   that monitors the topographic changes of the         EC-SPM can deliver nanometer-scale resolution
   electrode surface caused by electrochemical          of the electrode surface in liquid. The true merit
   processes (using standard AFM imaging modes).        of EC-SPM, however, is the capability to control
                                                        the experimental conditions (e.g., temperature
   On the other hand, with an active probe like that    and humidity) and mimic the real-world environ-
   utilized in EC-STM, a bipotentiostat is used to      ment of the sample under test. This is particularly
                                                        important for energy and corrosion research.
02 | Keysight | In Situ Electrochemical Measurements Using Keysight 7500 AFM - Application Brief



Instrumentation                                            The Keysight 7500 can also be equipped
                                                           with an optional bipotentiostat that
The Keysight Technologies, Inc. 7500 AFM/                  enables stable electrochemistry control
SPM microscope is a high-performance                       in either EC-AFM or EC-STM mode. The
instrument that delivers high-resolution                   bipotentiostat offers a series of different
                                                           sensitivity settings (10nA/V, 100nA/V,
imaging with integrated environmental
                                                           1A/V, 100 A/V, and 1mA/V) covering
control functions. The standard Keysight
                                                           four orders of magnitude of currents from
7500 includes contact mode, acoustic
                                                           10 pA to 10mA. Coupled with a specially
AC mode, and phase imaging. It comes
                                                           designed liquid cell that allows the use
with a universal scanner that is capable of
                                                           of a mini-reference (Ag/AgCl) electrode
operating in both open-loop and closed-
                                                           for true potential control, instead of a
loop mode. Switching imaging modes
                                                           quasi Ag wire reference as other EC-SPM
with the 7500 AFM/SPM microscope is
                                                           systems do, the Keysight 7500 provides an
quick and simple, a result of the scanner's
                                                           outstanding imaging system for electro-
interchangeable, easy-to-load nose cones.
                                                           chemistry research.
Every 7500 AFM/SPM microscope utilizes
the industry's lowest-noise closed-loop
scanner to provide the ultimate conve-
                                                           Examples of EC-SPM
nience and performance in imaging, with-                   Measurement
out sacriicing resolution or image quality.                Electrochemical SPM has been applied to
                                                           study a wide range of systems, including
An optional EC glove box designed spe-                     surface adsorption, ilm growth and dis-
ciically for use with the 7500 features a                  solution, membrane ion transport, battery,
dual-chamber design that allows samples                    fuel cell, photovoltaic, and catalysis. The
to be prepared under environmental con-                    examples below highlight only the basic                Microscope on glove box. Inset: sample plate
                                                                                                                  with the mini reference electrode (Ag/AgCl).
trol (e.g., humidity, temperature, oxygen                  EC-SPM capabilities of the Keysight 7500.
levels, reactive gases) in one area and
then moved to an inner chamber directly
                                                           Oxygen-Free and Controlled
underneath the microscope's head/scan-
ner. An optional mini reference electrode
                                                           Environments
(Ag/AgCl) provides a true reference and                    Environmental control is critical for many
controls sample potential. The 7500 head                   EC-SPM experiments. For example, it has
on the glove box can be placed inside the                  long been recognized that electrochemi-
Keysight Pico IC isolation chamber.                        cal processes are signiicantly affected by
            Current (mA)




                                                Voltage (V)



Figure 1. Oxygen level control by glove box. CV of Au in 0.1M H2SO 4 recorded after 1 hr in N2 (red) and 30 min
after exposure to air (blue).
03 | Keysight | In Situ Electrochemical Measurements Using Keysight 7500 AFM - Application Brief


the amount of oxygen existing in the experimental environment. Therefore, the elec-
trolyte and the cell in regular electrochemical experiments are often deoxygenated by
purging inert gases such as N2 and Ar before or during the experiment. The integrated
environmental control system for the 7500 is capable of temperature control, humidity
control, liquid exchange, and more (see Figure 1). This control is particularly useful for
the sample preparation and transportation often required for battery studies. With the
help of the 7500's environmental control system, clean and "oxygen-free" conditions
can be achieved to ensure proper electrochemical experiments. The oxygen level can                                                                          -0.4V
be reduced from ambient (~21%) to less than 1% in about 5 minutes by N2 purging.                                                                 -0.3V
                                                                                                                                        -0.2V
Copper Electrodeposition on Au
                                                                                                                             -0.1V
Electrodeposited copper is used extensively in microelectronics and microelectro-
mechanical systems (MEMS), particularly in chip metallization such as contacts and                          Figure 2. Three-dimensional EC-AFM topog-
interconnects and the illing of vias. As these processes and devices are continually                        raphy image of Cu ilms deposited at different
                                                                                                            overpotentials.
driven toward the nanoscale, the need for more research into conditions affecting the
electrochemical plating process arises. Electrochemical SPM techniques have been ex-
tremely beneicial in understanding the dynamics of nucleation and growth processes,
and how additives and adsorption can inhibit or enhance these processes.

The simple experiment presented here examines the effect of overpotential on the
kinetics and ilm properties of Cu deposition on epitaxial Au(111)/mica surface. The
electrolyte used was 15mM CuSO4 in 0.1M H2SO4 and the potential was controlled
against a quasi-reference electrode (Cu/Cu2+). The electrode potential was pulsed for
2 sec to -0.1, -0.2, -0.3, and -0.4V sequentially. After each pulse, the electrodeposited
Cu ilm was kept at 0.0 V for imaging by AFM for about 2 minutes, then stripped off at
0.4V before the next pulse. Figure 2 shows the ilms deposited at each of the different
overpotential values, and imaged by contact AFM.

Since the rate of Cu deposition is highly dependent on the overpotential applied, which
in turn affects surface properties such as the roughness of the ilm, the kinetics and
surface roughness for ilm growth can be evaluated by EC-SPM. For example, the
thickness of the Cu ilms deposited during the pulse experiment can be measured from
the averaged line proile shown in Figure 3(A). The estimated thickness and surface
roughness of each ilm are presented in Figure 3(B).




  A



       Voltage                        0.4 V           -0.1 V        -0.2 V        -0.3 V           -0.4 V
       Cu/Au
       Thickness (nm)                 0               10.4          53.0          77.2             84.2
       Roughness (nm)                 2.78            3.29          3.81          5.79             10.3

  B


Figure 3. (A) Averaged topography line proile from the AFM image. (B) The estimated thickness and
surface roughness of the Cu ilms deposited at different overpotentials.
04 | Keysight | In Situ Electrochemical Measurements Using Keysight 7500 AFM - Application Brief




                                                                                                        For more information on Keysight
Summary                                                   AFM Instrumentation from                      Technologies' products, applications or

EC-SPM has been proven to be an indis-
                                                          Keysight Technologies                         services, please contact your local Keysight
                                                                                                        office. The complete list is available at:
pensable tool for direct, high-resolution                                                               www.keysight.com/find/contactus
                                                          Keysight Technologies offers high-preci-
studies of physicochemical processes,
                                                          sion, modular AFM solutions for research,
providing an atomic and molecular                                                                       Americas
                                                          industry, and education. Exceptional
basis for understanding the kinetics and                                                                Canada                     (877) 894 4414
                                                          worldwide support is provided by experi-
thermodynamics of electriied interfaces.                                                                Brazil                     55 11 3351 7010
                                                          enced application scientists and technical    Mexico                     001 800 254 2440
Application of electrochemical SPM will
                                                          service personnel. Keysight's leading-edge    United States              (800) 829 4444
continue to impact and drive research in
                                                          R&D laboratories are dedicated to the time-
diverse disciplines, including energy, cor-
                                                          ly introduction and optimization of innova-   Asia Paciic
rosion, and life sciences.                                                                              Australia                  1 800 629 485
                                                          tive and easy-to-use AFM technologies.
                                                                                                        China                      800 810 0189
                                                                                                        Hong Kong                  800 938 693
                                                          www.keysight.com/find/afm
                                                                                                        India                      1 800 112 929
                                                                                                        Japan                      0120 (421) 345
                                                                                                        Korea                      080 769 0800
                                                                                                        Malaysia                   1 800 888 848
                                                                                                        Singapore                  1 800 375 8100
                                                                                                        Taiwan                     0800 047 866
                                                                                                        Other AP Countries         (65) 6375 8100

                                                                                                        Europe & Middle East
                                                                                                        Austria                    0800 001122
                                                                                                        Belgium                    0800 58580
                                                                                                        Finland                    0800 523252
                                                                                                        France                     0805 980333
                                                                                                        Germany                    0800 6270999
                                                                                                        Ireland                    1800 832700
                                                                                                        Israel                     1 809 343051
                                                                                                        Italy                      800 599100
                                                                                                        Luxembourg                 +32 800 58580
                                                                                                        Netherlands                0800 0233200
                                                                                                        Russia                     8800 5009286
                                                                                                        Spain                      0800 000154
                                                                                                        Sweden                     0200 882255
                                                                                                        Switzerland                0800 805353