Service Manuals, User Guides, Schematic Diagrams or docs for : Agilent 5991-3317EN Considerations in Making Small Signal Measurements - Application Brief c20141009 [3]
<< Back |
HomeMost service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download
Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing.
If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.
Image preview - the first page of the document
>> Download 5991-3317EN Considerations in Making Small Signal Measurements - Application Brief c20141009 [3] documenatation <<Text preview - extract from the documentKeysight Technologies
Considerations in Making
Small Signal Measurements
Application Brief
The ever-increasing demand for
battery-powered mobile devices
and more energy efficient green
product development has triggered
a rising demand for oscilloscope
measurement solution that can
accurately test and measure the
small signals with higher accuracy.
02 | Keysight | Considerations in Making Small Signal Measurements - Application Brief
Probing considerations Higher resolution and lower noise acquisition
If you are measuring the small amplitude of ripple and noise When measuring small signals with a large dynamic range, you
on your power rail or devices, you may need to use an oscillo- need a measurement resolution that is higher than the 8 bits
scope at or near its more sensitive V/div setting. provided by conventional digitizing oscilloscopes. Rather than
First, try using a probe with a lower attenuation ratio such resorting to a high-resolution digitizer or other measuring instru-
as 1:1 probe, rather than using the standard passive probe ment, you can use the built-in high-resolution mode or averaging
with 10:1 attenuation ratio that shipped with your instrument. mode on your oscilloscope to measure a parameter with a resolu-
With a 10:1 probe, not only is the oscilloscope's base-line tion higher than 8 bits and lower noise. Each acquisition mode has
noise floor increased by a factor of 10, but the minimum V/ pros and cons, but
div setting of scope is also ten times higher than with a 1:1
probe. Use averaging acquisition mode when: