Manual Change

This manual change describes the procedure when use the Agilent 16353A/16353H
Standard Resistor Set instead of the Agilent 16340A RC-Box in 4339B Performance Test.
When use the Agilent 16340A RC-Box, refer to the Chapter 9 (Maintenance) in this
manual.
          Manual Change
          Test Equipment




          Test Equipment
          Table 1 lists the recommended equipment for performing 4339B maintenance.
Table 1   Required Equipment

              Equipment                  Requirements         Recommended      Qty.   Use*1
                                                                 Model

           DC Voltmeter          Voltage Range: 0V to 1000V   3458A            1      P
                                 Accuracy: <0.04%
           Standard Resistor     No substitute                16353A/16353H    1      P, F
           Set
           10 k Resistor                                      42038A           1      P
           RC Box Adapter        No substitute                PN 04339-65005   1      P, F
           V Measurement         No substitute                PN 04339-65006   1      P
           Adapter
           BNC(F)-TXA                                                          1      P
           Ground Open (F)
           Adapter
           BNC(F)-(F)                                         PN 1250-1830     1      P
           Adapter
           Cable                 Banana-Banana Jumper         11058A           1      P
           Cable                 Dual Banana-BNC(m) Cable     PN 11001-60001   1      P
           Handler Interface     No Substitute                PN 04339-65007   1      F
           Tester
              *1.P: Performance Tests, F: Functional Tests
                                                                                  Manual Change
                                                                              Performance Tests




       Performance Tests

       Introduction
       This section provides the test procedures used to verify that the 4339B's specifications
       listed in Chapter 8, General Information, of this manual are met. All tests can be performed
       without access to the interior of the instrument. The performance tests can also be used to
       perform incoming inspection, and to verify that the 4339B meets its performance
       specifications after troubleshooting or adjustment. If the performance tests indicate that the
       4339B is NOT operating within the specified limits, check your test setup, then proceed
       with troubleshooting if necessary.

NOTE   Allow the 4339B to warm up for at least 30minutes before you execute any of the
       performance tests.


NOTE   Perform all performance tests in ambient conditions of 23