Service Manuals, User Guides, Schematic Diagrams or docs for : Agilent HP 83711A_252C B_252C 12A_252C 83711B_252C 12B User

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
HP 83711A_252C B_252C 12A_252C  83711B_252C 12B User


>> Download HP 83711A_252C B_252C 12A_252C 83711B_252C 12B User documenatation <<

Text preview - extract from the document
Notice
Hewlett-Packard to Agilent Technologies Transition
This documentation supports a product that previously shipped under the Hewlett-
Packard company brand name. The brand name has now been changed to Agilent
Technologies. The two products are functionally identical, only our name has changed. The
document still includes references to Hewlett-Packard products, some of which have been
transitioned to Agilent Technologies.




Printed in USA   March 2000
Contacting Agilent




By internet, phone, or fax, get assistance with all your test and measurement needs.
Table 1-1 Contacting Agilent
 Online assistance: www.agilent.com/find/assist

 United States                 Latin America               Canada                    Europe
 (tel) 1 800 452 4844          (tel) (305) 269 7500        (tel) 1 877 894 4414      (tel) (+31) 20 547 2323
                               (fax) (305) 269 7599        (fax) (905) 282-6495      (fax) (+31) 20 547 2390

 New Zealand                   Japan                       Australia
 (tel) 0 800 738 378           (tel) (+81) 426 56 7832     (tel) 1 800 629 485
 (fax) (+64) 4 495 8950        (fax) (+81) 426 56 7840     (fax) (+61) 3 9210 5947


                             Asia Call Center Numbers

 Country                Phone Number                      Fax Number

 Singapore              1-800-375-8100                    (65) 836-0252

 Malaysia               1-800-828-848                     1-800-801664

 Philippines            (632) 8426802                     (632) 8426809
                        1-800-16510170 (PLDT              1-800-16510288 (PLDT
                        Subscriber Only)                  Subscriber Only)

 Thailand               (088) 226-008 (outside Bangkok)   (66) 1-661-3714
                        (662) 661-3999 (within Bangkok)

 Hong Kong              800-930-871                       (852) 2506 9233

 Taiwan                 0800-047-866                      (886) 2 25456723

 People's Republic      800-810-0189 (preferred)          10800-650-0121
 of China               10800-650-0021

 India                  1-600-11-2929                     000-800-650-1101
User's Guide




HP 83711A/12A and
HP 83711B/12B
Synthesized CW
Generators
HP part number: 83711-90131
Printed in USA July 1997 Supersedes April 1995
Notice.
The information contained in this document is subject to change without
notice.
Hewlett-Packard makes no warranty of any kind with regard to this material,
including but not limited to, the implied warranties of merchantability and
tness for a particular purpose. Hewlett-Packard shall not be liable for errors
contained herein or for incidental or consequential damages in connection
with the furnishing, performance, or use of this material.





 Copyright Hewlett-Packard Company 1995, 1997
c
All Rights Reserved. Reproduction, adaptation, or translation without prior
written permission is prohibited, except as allowed under the copyright laws.
1400 Fountaingrove Parkway, Santa Rosa, CA 95403-1799, USA
The HP 83711A/12A and HP 83711B/12B
Synthesized CW Generators

The HP 83711A/12A and HP 83711B/12B synthesized CW generators are
referred to as \synthesizers" throughout this manual. The HP 83711A/11B
has a carrier frequency range of 1 GHz to 20 GHz and the HP 83712A/12B has
a carrier frequency range of 10 MHz to 20 GHz. Specication information can
be found in Chapter 4, \Specications and Options."
The HP 83711A/12A and HP 83711B/12B Synthesized CW Generators User's
Guide is written to accommodate the novice and the expert user. If you
are unfamiliar with the synthesizer, Chapter 2, \Performing Fundamental
Synthesizer Operations," is useful. If you are familiar with the synthesizer,
Chapter 3, \Generating Signals with the Synthesizer" section is helpful. If
you are looking for specic, detailed information about the synthesizer, refer
to the remaining chapters as needed.


  Notes

  1. This manual applies to instruments with rmware revision 10.0 or greater.
  2. If you have an HP 83711A/12A instrument with rmware revision number        < 10.0, refer to the
     HP 83711A/12A Synthesized Signal Generator User's Guide   (5960-7088).
  3. To view rmware revision, press 4SPCL5, 415, 4HZ5 (ENTER).




                                                                                                        iii
The Synthesizer at a Glance


The following gure and accompanying text explains some features of the
HP 83711A and HP 83711B. The HP 83712A andHP 83712B are nearly
identical.




           HP 83711B Synthesized CW Generator




iv
1. The 4MSG     5 key allows you to display      5. These keys set the carrier frequency and         10.  The data entry keys are used to enter
any error messages on the front panel display.   RF output power level of the synthesizer.           and modify various synthesizer parameters.
Error messages are generated when you
                                                 6.   The display shows the current values of
                                                                                                     The 4 BACKSPACE        5 key cancels all or
perform a keystroke sequence that is not                                                             part of an erroneous parameter entry before
                                                 synthesizer parameters as well as the status
valid, try to operate the synthesizer in a                                                           it has been terminated. The terminator keys
                                                 of many of the synthesizer functions. The
mode that is not allowed, etc.                                                                       (the right-most column of keys) are used to
                                                 display shows the current carrier frequency
                                                                                                     choose the units for the entered parameter
2. The 4SHIFT    5 key changes the function      and output power level. The annunciators
                                                                                                     as well as to terminate the parameter entry.
of some of the keys. When you press the
SHIFT
4        5 key and then press another key,
                                                 that appear below the parameters are only
                                                 visible when their associated function is active.
                                                                                                           * +
                                                                                                     The 4 5, 4 5, and 4    STEP SIZE    5 keys
                                                                                                     are used to increase or decrease a parameter
the synthesizer performs the function printed
in blue above the key.
                                                 7.      (           )
                                                    The 4 5 and 4 5 keys move the cursor             in predetermined steps.
                                                 that is over one of the digits in the display
                                                                                                         The Automatic Level Control voltage input
3. The 4RECALL       5 / SAVE key is used to     either to the right or left when pressed. The
                                                                                                     11.

                                                                                                     (ALC IN) connector is used as the feedback
save most of the synthesizer operating           digit that is under the cursor will be modied
                                                                                                     path to the synthesizer when its RF output
parameters in one of nine nonvolatile register   when the knob is rotated. If no cursor
                                                                                                     power level is being leveled externally.
locations so that they can be recalled and       appears in the display, parameter entry or
used at a later time.                            modication has been inhibited.                     12.  The 4SPCL    5 key is used to initiate the
                                                                                                     activation of several special functions available
4. The Automatic Level Control keys select       8. The RF OUTPUT connector mates with a
                                                                                                     in the synthesizer. Special functions are
the method used to regulate the synthesizer      Type-N male connector on non-Option 1E9
                                                                                                     additional functions that are not activated by
output power level. Either internal leveling,    instruments. The connector mates with a
                                                                                                     pressing a front panel key or shifted key.
external power meter leveling, or external       female APC-3.5 mm precision connector on
diode detector leveling can be selected.         instruments with Option 1E9 installed.              13. The POWER switch (LINE on an HP
Additional external equipment is required                                                            83711A/12A) turns the synthesizer either on
                                                 9. The knob is used to increase or decrease
when either external power meter leveling or                                                         or o.
                                                 the digit under the cursor ( 9 ) in the
external diode detector leveling is used.
                                                 display in steps of one.


                             Synthesizer rear panel features are depicted and described in detail in
                             Chapter 5, \Front/Rear Panel," in this book.




                                                                                                                                                   v
In This Book


This book is divided into the following chapters:
 Chapter 1, \Installing and Verifying the Synthesizer," contains procedures
  for installing the synthesizer and verifying its operation.
 Chapter 2, \Performing Fundamental Synthesizer Operations," familiarizes
  you with the fundamental operation of the synthesizer.
 Chapter 3, \Generating Signals with the Synthesizer," explains how to
  generate CW signals and level signals.
 Chapter 4, \Specications and Options," contains a list of the synthesizer
  performance specications as well as the various mechanical, electrical,
  warranty, and documentation options that are available.
 Chapter 5, \Front/Rear Panel," contains entries that explain dierent
  aspects of the synthesizer front and rear panel. (For example, you turn to
  this chapter for information on the RF connectors).
 Chapter 6, \Keys/Shifted Functions," contains entries on the function of
  each key on the synthesizer front panel as well as the shifted or alternate
  function of certain keys.
 Chapter 7, \Special Functions," contains entries on the special functions
  available in the synthesizer. Special functions are hidden during normal
  instrument operation and can only be invoked by typing a specic key
  sequence on the synthesizer front panel.
 Chapter 8, \Error Messages," contains a table that lists all of the error
  messages that might be generated during use of the instrument. Each table
  entry contains a sequence that can be followed to recover from the error
  condition.
 Chapter 9, \Legal and Regulatory Information," contains information
  related to safety and SCPI conformance information. The product warranty
  is also contained in this chapter.




vi
Contents




 1. Installing and Verifying the Synthesizer
     Installing the Synthesizer . . . . .   .   .   .   .   .   .   .   .   .    .    1-3
        To Unpack the Synthesizer . . .     .   .   .   .   .   .   .   .   .    .    1-3
        To Install the Synthesizer . . .    .   .   .   .   .   .   .   .   .    .    1-5
     Verifying Synthesizer Functionality    .   .   .   .   .   .   .   .   .    .    1-7
     If You Encounter a Problem . . .       .   .   .   .   .   .   .   .   .    .    1-8
        Mechanical or Electrical Damage     .   .   .   .   .   .   .   .   .    .    1-8
        Power-up Problems . . . . . .       .   .   .   .   .   .   .   .   .    .    1-9
        Self Test Failures . . . . . . .    .   .   .   .   .   .   .   .   .    .   1-11
 2. Performing Fundamental Synthesizer Operations
     To Enter Data with the Numeric Keypad .            .   .   .   .   .   .    .    2-3
     To Modify Data with the Knob . . . . .             .   .   .   .   .   .    .    2-4
     To Modify Data with the Arrow Keys . .             .   .   .   .   .   .    .    2-6
     To Save and Recall Synthesizer States . .          .   .   .   .   .   .    .    2-8
        Programming Example . . . . . . . .             .   .   .   .   .   .    .    2-9
     To Read the Contents of the Error Queue            .   .   .   .   .   .    .   2-10
        Programming Example . . . . . . . .             .   .   .   .   .   .    .   2-11
     To Set the HP-IB Address . . . . . . .             .   .   .   .   .   .    .   2-12
        Programming Example . . . . . . . .             .   .   .   .   .   .    .   2-12
     If You Encounter a Problem . . . . . .             .   .   .   .   .   .    .   2-13
        Data Entry Problems . . . . . . . .             .   .   .   .   .   .    .   2-13
        Programming Problems . . . . . . .              .   .   .   .   .   .    .   2-15
 3. Generating Signals with the Synthesizer
     To Generate a CW Signal . . . . . . . .                .   .   .   .   .    .    3-3
     To Generate Millimeter Signals . . . . . .             .   .   .   .   .    .    3-4
     To Use External Diode Detector Leveling . .            .   .   .   .   .    .    3-7
       Programming Command . . . . . . . .                  .   .   .   .   .    .   3-10
       Related Tasks . . . . . . . . . . . . .              .   .   .   .   .    .   3-10
     To Use External Power Meter Leveling . . .             .   .   .   .   .    .   3-11
       Programming Commands . . . . . . . .                 .   .   .   .   .    .   3-13
       Related Tasks . . . . . . . . . . . . .              .   .   .   .   .    .   3-13
     To Use the Level Correct Routine . . . . .             .   .   .   .   .    .   3-14
       Related Tasks . . . . . . . . . . . . .              .   .   .   .   .    .   3-19
     To Use Previously Stored Level Correct Data            .   .   .   .   .    .   3-20


                                                                                Contents-1
         Related Tasks . . . . .    .    . . . .    .    .    .   .   .    .    .    . .   3-21
      If You Encounter a Problem   .    . . . . .    .    .   .   .   .   .    .    . .    3-22
         Annunciators Turned On     .    . . . .    .    .    .   .   .    .    .    . .   3-22
         RF Output Problems .       .    . . . .    .    .    .   .   .    .    .    . .   3-23
  4. Specications and Options
      HP 83711A/11B Specications . . . . . . . . . . .                                .    4-3
       Frequency . . . . . . . . . . . . . . . . . . .                                 .    4-3
       RF Output . . . . . . . . . . . . . . . . . . .                                 .    4-4
       Spectral Purity . . . . . . . . . . . . . . . . .                               .    4-6
      HP 83712A/12B Specications . . . . . . . . . . .                                .    4-9
       Frequency . . . . . . . . . . . . . . . . . . .                                 .    4-9
       RF Output . . . . . . . . . . . . . . . . . . .                                 .   4-10
       Spectral Purity . . . . . . . . . . . . . . . . .                               .   4-13
       General . . . . . . . . . . . . . . . . . . . .                                 .   4-17
       Front Panel Connectors . . . . . . . . . . . . .                                .   4-18
       Rear Panel Connectors . . . . . . . . . . . . . .                               .   4-19
      Options . . . . . . . . . . . . . . . . . . . . .                                .   4-20
       Electrical Options . . . . . . . . . . . . . . . .                              .   4-20
          Option 1E1 - Add Output Step Attenuator . . . .                              .   4-20
          Option 1E5 - Add High Stability Timebase . . . .                             .   4-20
          Option 1E8 - 1 Hz Frequency Resolution . . . . .                             .   4-20
          Option 1E9 - 3.5 mm RF Output Connector . . .                                .   4-21
       Mechanical Options . . . . . . . . . . . . . . .                                .   4-21
          Option 1CM - Rack Mount Kit . . . . . . . . .                                .   4-21
          Option 1CP - Rack Mount and Handle Kit . . . .                               .   4-21
          Option 1CR - Rack Slide Kit . . . . . . . . . .                              .   4-21
       Warranty Options . . . . . . . . . . . . . . . .                                .   4-22
          Option W30 - Two Additional Years Return-to-HP
             Service . . . . . . . . . . . . . . . . . .                               .   4-22
          Option W32 - Three Year Return-to-HP Calibration
             Service . . . . . . . . . . . . . . . . . .                               .   4-22
       Documentation Options . . . . . . . . . . . . .                                 .   4-22
          Option OB0 - Delete Operating Documentation . .                              .   4-22
          Option OB1 - Extra Operating Documentation . .                               .   4-23
          Option OBV - Component-Level Information Packet
             (CLIP) . . . . . . . . . . . . . . . . . .                                .   4-23
          Option OBW - Service Documentation . . . . . .                               .   4-23
          Option OBX - Component-Level Information Packet
             (CLIP)/Service Documentation . . . . . . .                                .   4-23


Contents-2
5. Front/Rear Panel
    Annunciators . . . . . . . . . . .          .    .   .   .   .    .    . . .        5-3
      See Also . . . . . . . . . . . . . .       .   .   .   .   .   .    . . .         5-5
    Connectors . . . . . . . . . . . .          .    .   .   .   .    .    . . .        5-6
      Coaxial Connectors . . . . . . .          .    .   .   .   .    .    . . .        5-7
      HP-IB Connector . . . . . . . .           .    .   .   .   .    .    . . .        5-8
    Display . . . . . . . . . . . . .           .    .   .   .   .    .    . . .       5-11
      See Also . . . . . . . . . . . .          .    .   .   .   .    .    . . .       5-12
    Knob . . . . . . . . . . . . . .            .    .   .   .   .    .    . . .       5-13
      Equivalent SCPI Command . . . .           .    .   .   .   .    .    . . .       5-13
      See Also . . . . . . . . . . . .          .    .   .   .   .    .    . . .       5-14
    POWER Switch . . . . . . . . . .            .    .   .   .   .    .    . . .       5-15
      See Also . . . . . . . . . . . .          .    .   .   .   .    .    . . .       5-15
    Power Cables . . . . . . . . . . .          .    .   .   .   .    .    . . .       5-16
6. Keys/Shifted Functions
    ) . . . . . . . . . . . . .     .   .   .   .    .   .   .   .   .    .   .    .    6-3
     Equivalent SCPI Command .      .   .   .   .    .   .   .   .   .    .   .    .    6-4
     See Also . . . . . . . . .     .   .   .   .    .   .   .   .   .    .   .    .    6-4
    ( . . . . . . . . . . . . .     .   .   .   .    .   .   .   .   .    .   .    .    6-5
     Equivalent SCPI Command .      .   .   .   .    .   .   .   .   .    .   .    .    6-6
     See Also . . . . . . . . .     .   .   .   .    .   .   .   .   .    .   .    .    6-6
    * . . . . . . . . . . . . .     .   .   .   .    .   .   .   .   .    .   .    .    6-7
     Equivalent SCPI Commands       .   .   .   .    .   .   .   .   .    .   .    .    6-7
     See Also . . . . . . . . .     .   .   .   .    .   .   .   .   .    .   .    .    6-8
    + . . . . . . . . . . . . .     .   .   .   .    .   .   .   .   .    .   .    .    6-9
     Equivalent SCPI Commands       .   .   .   .    .   .   .   .   .    .   .    .    6-9
     See Also . . . . . . . . .     .   .   .   .    .   .   .   .   .    .   .    .   6-10
    ADDRESS . . . . . . . . .       .   .   .   .    .   .   .   .   .    .   .    .   6-11
     Equivalent SCPI Command .      .   .   .   .    .   .   .   .   .    .   .    .   6-11
     See Also . . . . . . . . .     .   .   .   .    .   .   .   .   .    .   .    .   6-11
    BACK SPACE . . . . . . .        .   .   .   .    .   .   .   .   .    .   .    .   6-12
     Equivalent SCPI Command .      .   .   .   .    .   .   .   .   .    .   .    .   6-12
     See Also . . . . . . . . .     .   .   .   .    .   .   .   .   .    .   .    .   6-12
    ENTER (Hz) . . . . . . . .      .   .   .   .    .   .   .   .   .    .   .    .   6-13
     Equivalent SCPI Command .      .   .   .   .    .   .   .   .   .    .   .    .   6-13
     See Also . . . . . . . . .     .   .   .   .    .   .   .   .   .    .   .    .   6-14
    ENTRY OFF . . . . . . . .       .   .   .   .    .   .   .   .   .    .   .    .   6-15
     Equivalent SCPI Command .      .   .   .   .    .   .   .   .   .    .   .    .   6-15
     See Also . . . . . . . . .     .   .   .   .    .   .   .   .   .    .   .    .   6-15
    EXT DIODE . . . . . . . .       .   .   .   .    .   .   .   .   .    .   .    .   6-16

                                                                                  Contents-3
        External Diode Detector Leveling . . . .     .   .    .    .    . .   6-16
        Applications . . . . . . . . . . . . . . .   .   .   .    .    . .    6-17
        Advantages of Diode Detector Leveling . .    .   .    .    .    . .   6-18
        Disadvantages of Diode Detector Leveling     .   .    .    .    . .   6-18
        Equivalent SCPI Command . . . . . . .        .   .    .    .    . .   6-19
        See Also . . . . . . . . . . . . . . .       .   .    .    .    . .   6-19
      EXT METER . . . . . . . . . . . . . .          .   .    .    .    . .   6-20
        External Power Meter Leveling . . . . .      .   .    .    .    . .   6-20
        Applications . . . . . . . . . . . . .       .   .    .    .    . .   6-22
        External Equipment Limitations . . . .       .   .    .    .    . .   6-22
          Power Meter . . . . . . . . . . . .        .   .    .    .    . .   6-22
          Power Sensor . . . . . . . . . . . .       .   .    .    .    . .   6-22
        Advantages of Power Meter Leveling . . .     .   .    .    .    . .   6-23
        Disadvantages of Power Meter Leveling . .    .   .    .    .    . .   6-23
        Equivalent SCPI Commands . . . . . .         .   .    .    .    . .   6-24
        See Also . . . . . . . . . . . . . . .       .   .    .    .    . .   6-24
      FREQ . . . . . . . . . . . . . . . . .         .   .    .    .    . .   6-25
        Equivalent SCPI Command . . . . . . .        .   .    .    .    . .   6-26
        See Also . . . . . . . . . . . . . . .       .   .    .    .    . .   6-26
      INT LEVEL . . . . . . . . . . . . . .          .   .    .    .    . .   6-27
        Advantages of Internal Leveling . . . . .    .   .    .    .    . .   6-28
        Disadvantages of Internal Leveling . . . .   .   .    .    .    . .   6-28
        Equivalent SCPI Commands . . . . . .         .   .    .    .    . .   6-29
        See Also . . . . . . . . . . . . . . .       .   .    .    .    . .   6-29
      LOCAL . . . . . . . . . . . . . . . .          .   .    .    .    . .   6-30
        Equivalent SCPI Command . . . . . . .        .   .    .    .    . .   6-30
        See Also . . . . . . . . . . . . . . .       .   .    .    .    . .   6-30
      LVL CR . . . . . . . . . . . . . . . .         .   .    .    .    . .   6-31
        Equivalent SCPI Command . . . . . . .        .   .    .    .    . .   6-34
        See Also . . . . . . . . . . . . . . .       .   .    .    .    . .   6-34
      LVL CR ON/OFF . . . . . . . . . . . .          .   .    .    .    . .   6-35
        Equivalent SCPI Commands . . . . . .         .   .    .    .    . .   6-36
        See Also . . . . . . . . . . . . . . .       .   .    .    .    . .   6-36
      MSG . . . . . . . . . . . . . . . . .          .   .    .    .    . .   6-37
        Equivalent SCPI Command . . . . . . .        .   .    .    .    . .   6-37
        See Also . . . . . . . . . . . . . . .       .   .    .    .    . .   6-37
      MULTIPLIER . . . . . . . . . . . . .           .   .    .    .    . .   6-38
        Equivalent SCPI Command . . . . . . .        .   .    .    .    . .   6-39
        See Also . . . . . . . . . . . . . . .       .   .    .    .    . .   6-39
      Numeric Keypad . . . . . . . . . . . .         .   .    .    .    . .   6-40
        See Also . . . . . . . . . . . . . . .       .   .    .    .    . .   6-40

Contents-4
POWER LEVEL . . . . . . . . .         .    .    .   .   .    .    .    . .   6-41
  Equivalent SCPI Command . . . . .    .    .   .   .   .   .    .    . .    6-42
  See Also . . . . . . . . . . . .    .    .    .   .   .    .    .    . .   6-42
PRESET . . . . . . . . . . . . .      .    .    .   .   .    .    .    . .   6-43
  Equivalent SCPI Commands . . .      .    .    .   .   .    .    .    . .   6-44
  See Also . . . . . . . . . . . .    .    .    .   .   .    .    .    . .   6-44
RCL CR . . . . . . . . . . . . .      .    .    .   .   .    .    .    . .   6-45
  Equivalent SCPI Command . . . .     .    .    .   .   .    .    .    . .   6-46
  See Also . . . . . . . . . . . .    .    .    .   .   .    .    .    . .   6-46
RECALL . . . . . . . . . . . . .      .    .    .   .   .    .    .    . .   6-47
  Equivalent SCPI Command . . . .     .    .    .   .   .    .    .    . .   6-47
  See Also . . . . . . . . . . . .    .    .    .   .   .    .    .    . .   6-47
RF ON/OFF . . . . . . . . . . .       .    .    .   .   .    .    .    . .   6-48
  Equivalent SCPI Command . . . .     .    .    .   .   .    .    .    . .   6-48
  See Also . . . . . . . . . . . .    .    .    .   .   .    .    .    . .   6-48
SAVE . . . . . . . . . . . . . .      .    .    .   .   .    .    .    . .   6-49
  Equivalent SCPI Command . . . .     .    .    .   .   .    .    .    . .   6-49
  See Also . . . . . . . . . . . .    .    .    .   .   .    .    .    . .   6-49
SAVE CR . . . . . . . . . . . .       .    .    .   .   .    .    .    . .   6-50
  Equivalent SCPI Command . . . .     .    .    .   .   .    .    .    . .   6-51
  See Also . . . . . . . . . . . .    .    .    .   .   .    .    .    . .   6-51
SHIFT . . . . . . . . . . . . . .     .    .    .   .   .    .    .    . .   6-52
  Equivalent SCPI Command . . . .     .    .    .   .   .    .    .    . .   6-52
  See Also . . . . . . . . . . . .    .    .    .   .   .    .    .    . .   6-52
SPCL . . . . . . . . . . . . . .      .    .    .   .   .    .    .    . .   6-53
  Equivalent SCPI Command . . . .     .    .    .   .   .    .    .    . .   6-53
SPCL OFF (kHz) . . . . . . . . .      .    .    .   .   .    .    .    . .   6-54
  Equivalent SCPI Command . . . .     .    .    .   .   .    .    .    . .   6-54
  See Also . . . . . . . . . . . .    .    .    .   .   .    .    .    . .   6-55
SPCL ON (MHz) . . . . . . . . .       .    .    .   .   .    .    .    . .   6-56
  Equivalent SCPI Command . . . .     .    .    .   .   .    .    .    . .   6-56
  See Also . . . . . . . . . . . .    .    .    .   .   .    .    .    . .   6-56
STEP SIZE . . . . . . . . . . . .     .    .    .   .   .    .    .    . .   6-57
  Equivalent SCPI Commands . . .      .    .    .   .   .    .    .    . .   6-57
  See Also . . . . . . . . . . . .    .    .    .   .   .    .    .    . .   6-58




                                                                       Contents-5
  7. Special Functions
      ATTEN HOLD ON/OFF . . . . .              .    .   .   .   .    .    .   .   .    7-3
        Applications . . . . . . . . . .       .    .   .   .   .    .    .   .   .    7-4
        Advantages . . . . . . . . . . .       .    .   .   .   .    .    .   .   .    7-4
        Disadvantages . . . . . . . . . .      .    .   .   .   .    .    .   .   .    7-5
        Equivalent SCPI Command . . . .        .    .   .   .   .    .    .   .   .    7-5
        See Also . . . . . . . . . . . . . .    .   .   .   .   .   .    .    . .      7-5
      CLEAR DISPLAY . . . . . . . .            .    .   .   .   .    .    .    . .     7-6
        Equivalent SCPI Command . . . .        .    .   .   .   .    .    .    . .     7-7
        See Also . . . . . . . . . . . .       .    .   .   .   .    .    .    . .     7-7
      ERASE MEMORY . . . . . . . .             .    .   .   .   .    .    .    . .     7-8
        Application . . . . . . . . . . .      .    .   .   .   .    .    .    . .     7-8
        Equivalent SCPI Command . . . .        .    .   .   .   .    .    .    . .     7-8
        See Also . . . . . . . . . . . .       .    .   .   .   .    .    .    . .     7-9
      PWR MTR ADDRS . . . . . . . .            .    .   .   .   .    .    .    . .    7-10
        Equivalent SCPI Command . . . .        .    .   .   .   .    .    .    . .    7-11
        See Also . . . . . . . . . . . .       .    .   .   .   .    .    .    . .    7-11
      PWR MTR SELECT . . . . . . .             .    .   .   .   .    .    .    . .    7-12
        Equivalent SCPI Command . . . .        .    .   .   .   .    .    .    . .    7-12
        See Also . . . . . . . . . . . .       .    .   .   .   .    .    .    . .    7-13
      REMOTE LANGUAGE . . . . . .              .    .   .   .   .    .    .    . .    7-14
        Equivalent SCPI Command . . . .        .    .   .   .   .    .    .    . .    7-14
        See Also . . . . . . . . . . . .       .    .   .   .   .    .    .    . .    7-14
      SELF TEST . . . . . . . . . . .          .    .   .   .   .    .    .    . .    7-15
        Equivalent SCPI Command . . . .        .    .   .   .   .    .    .    . .    7-16
        See Also . . . . . . . . . . . .       .    .   .   .   .    .    .    . .    7-16
      Service-Related Special Functions . .    .    .   .   .   .    .    .    . .    7-17
        See Also . . . . . . . . . . . .       .    .   .   .   .    .    .    . .    7-17
      SHOW INFO . . . . . . . . . . .          .    .   .   .   .    .    .    . .    7-18
        Equivalent SCPI Command . . . .        .    .   .   .   .    .    .    . .    7-18
        See Also . . . . . . . . . . . .       .    .   .   .   .    .    .    . .    7-18
  8. Error Messages
      Error Messages List . . . . . . . . . . . . . . . . .                            8-3
      Messages . . . . . . . . . . . . . . . . . . . . . .                             8-5




Contents-6
9. Legal and Regulatory Information
    Safety Symbols . . . . .      .    .   .   .   .   .   .    .   .   .   .    .   .    .    9-3
      Miscellaneous Symbols .     .    .   .   .   .   .   .    .   .   .   .    .   .    .    9-4
    Safety Considerations . .     .    .   .   .   .   .   .    .   .   .   .    .   .    .    9-5
      Before Applying Power .     .    .   .   .   .   .   .    .   .   .   .    .   .    .    9-5
      Safety Earth Ground . .     .    .   .   .   .   .   .    .   .   .   .    .   .    .    9-5
      Cleaning the Product . .   .    .    . . . . .        .   .   .   .   .   .    . .       9-6
    Certication . . . . . .      .    .    . . . .        .    .   .   .   .    .    . .      9-7
    Warranty . . . . . . . .      .    .    . . . .        .    .   .   .   .    .    . .      9-8
      Limitation of Warranty      .    .    . . . .        .    .   .   .   .    .    . .      9-8
      Exclusive Remedies . .      .    .    . . . .        .    .   .   .   .    .    . .      9-9
    Assistance . . . . . . .      .    .    . . . .        .    .   .   .   .    .    . .     9-10
   Index




                                                                                         Contents-7
Figures




   1-1.   Line Fuse Removal and Replacement . . . . . . . . . . .          1-10
   3-1.   Millimeter-Wave Equipment Setup . . . . . . . . . . . . .         3-4
   3-2.   External Diode Detector Leveling Setup . . . . . . . . . .        3-8
   3-3.   External Power Meter Leveling Setup . . . . . . . . . . .        3-11
   3-4.   Level Correct Routine Equipment Setup . . . . . . . . . .        3-15
   4-1.   Typical Maximum Available Output Power from 1 to 20
              GHz, at 25  C with Output Step Attenuator (Option 1E1)
              Installed . . . . . . . . . . . . . . . . . . . . . .         4-4
   4-2.   Typical Output Level Accuracy and Flatness at +10 and 085
              dBm . . . . . . . . . . . . . . . . . . . . . . . .           4-5
   4-3.   Typical single-sideband phase noise at 1 GHz, 10 GHz and 20
              GHz, 25  C, CW mode. Osets less than 100 Hz require
              the high stability timebase, Option 1E5. . . . . . . . .      4-6
   4-4.   Typical 2nd Harmonic Levels Measured at Output Power of +6
              dBm . . . . . . . . . . . . . . . . . . . . . . . .           4-7
   4-5.   Typical Residual FM Measured in 50 Hz - 15 kHz Bandwidth,
              CW Mode, with High Stability Timebase (Option 1E5) . .        4-8
   4-6.   Typical Maximum Available Output Power from 1 to 20
              GHz, at 25  C with Output Step Attenuator (Option 1E1)
              Installed . . . . . . . . . . . . . . . . . . . . . .        4-10
   4-7.   Typical Maximum Available Output Power from 0.01 to 1 GHz
              at 25  C. . . . . . . . . . . . . . . . . . . . . . .        4-11
   4-8.   Typical Output Level Accuracy and Flatness at +10 and 085
              dBm . . . . . . . . . . . . . . . . . . . . . . . .          4-12
   4-9.   Typical single-sideband phase noise at 50 MHz, 1 GHz, 10 GHz
              and 20 GHz, 25  C, CW mode. Osets less than 100 Hz
              require the high stability timebase, Option 1E5. . . . . .   4-14
  4-10.   Typical 2nd Harmonic Levels Measured at Output Power of +6
              dBm . . . . . . . . . . . . . . . . . . . . . . . .          4-15
  4-11.   Typical Residual FM Measured in 50 Hz - 15 kHz Bandwidth,
              CW mode, with High Stability Timebase (Option 1E5) . .       4-16
   5-1.   Synthesizer Display Annunciators . . . . . . . . . . . . .        5-3
   5-2.   Synthesizer Connectors - Front Panel . . . . . . . . . . .        5-6
   5-3.   Synthesizer Connectors - Rear Panel . . . . . . . . . . . .       5-6
   5-4.   HP-IB Connector and Cable . . . . . . . . . . . . . . . .         5-8
   5-5.   Synthesizer Display . . . . . . . . . . . . . . . . . . .        5-11
   5-6.   Power Cable and Line (Mains) Plug Part Numbers . . . . . .       5-17


Contents-8
                                                              Contents




6-1. External Diode Detector Leveling Setup . . . . . . . . . .   6-17
6-2. External Power Meter Leveling Setup . . . . . . . . . . .    6-21




                                                           Contents-9
Tables




   4-1.   Osets (HP 83711A only) . . . . . . . .      .   .   .   .   .   .   .   .   .    4-6
   4-2.   Osets (HP 83711B only) . . . . . . . .      .   .   .   .   .   .   .   .   .    4-6
   4-3.   Osets (HP 83712A only) . . . . . . . .      .   .   .   .   .   .   .   .   .   4-13
   4-4.   Osets (HP 83712B only) . . . . . . . .      .   .   .   .   .   .   .   .   .   4-13
   5-1.   HP-IB Interface Cables Available . . . .    .   .   .   .   .   .   .   .   .    5-8
   6-1.   PRESET Conditions . . . . . . . . . .       .   .   .   .   .   .   .   .   .   6-43
   7-1.   Service-Related Special Functions . . . .   .   .   .   .   .   .   .   .   .   7-17
   9-1.   Hewlett-Packard Sales and Service Oces      .   .   .   .   .   .   .   .   .   9-11




Contents-10
1




Installing and Verifying the
Synthesizer
Installing and Verifying the Synthesizer


This chapter contains procedures for properly installing your HP 83711A/12A
or HP 83711B/12B synthesizer and procedures for functional verication of
the instrument.




1-2
Installing the Synthesizer




This procedure explains how to inspect, install, and power-up the
synthesizer.




T Unpack the Synthesizer
 o


1. Inspect the shipping container for damage.
   Look for signs of damage such as a dented or torn shipping container or
   cushioning material that shows signs of unusual stress or compacting.
2. Carefully remove the contents from the shipping container and inspect
  each item for damage.
  If the instrument or any accessories appear to be damaged, refer to
  \Mechanical or Electrical Damage" at the end of this chapter. The
  following items should have been received in the shipment.




                                                                        1-3
Installing and Verifying the Synthesizer
Installing the Synthesizer




                                 Item                       Model/Part Number


       Instrument                            HP 83711A/12A    or   HP 83711B/12B
       Power Cable                           Depends on shipment destination. Refer to
                                             \Power Cables"in Chapter 5 of this manual.
       User's Guide                          83711-90131
       Quick Start Guide                     83711-90133
       Programming Guide                     83711-90132
       Calibration Guide                     83731-90125
       Calibration Kit                       5063-1635
          Calibration Guide                     83731-90125
          Calibration Software                  5010-7723
       Also includes one of the following:


       Handle Kit                            5062-3989 (standard)
       Rack Mount Kit                        5062-3977 (Option 1CM)
       Rack Mount Kit with Handles           5062-3983 (Option 1CP)
       Rack Slide Kit                        1494-0059 (Option 1CR)



3. Keep the shipping materials for future use.
   If undamaged, shipping materials are useful for shipment or storage of the
   instrument. If damaged, shipping materials should be kept for the carrier's
   inspection.




1-4
                                                          Installing and Verifying the Synthesizer
                                                                   Installing the Synthesizer




          T Install the Synthesizer
           o


          The following provides a general procedure for installation and initial power
          up of the HP 83711A/12A or HP 83711B/12B synthesizer.
          1. Check to make sure that the power cable is undamaged.
             Do not use the power cable if the plug contacts are bent or broken or if
             the wire insulation is damaged or if wire is exposed. Never use a power
             cable if the grounding contact has been removed.
CAUTION   Always use the three-prong AC power cord supplied with this product.
          Failure to ensure adequate earth grounding by not using this cord may cause
          product damage.

          2. Set the POWER switch (LINE on an HP 83711A/12A) on the
             synthesizer to standby ( ).
          3. Set the line voltage selector switch to match the mains voltage.
             The line voltage selector switch is located on the rear panel of the
             synthesizer to the right of the power module. Use a small 
at blade
             screwdriver to set the switch up for mains voltages in the range of 90 V to
             132 V; 50, 60, or 400 Hz or down for mains voltages in the range of 198 V
             to 264 V; 50 or 60 Hz.




                                                                                              1-5
          Installing and Verifying the Synthesizer
          Installing the Synthesizer




WARNING
          This is a Safety Class I product (provided with a protective earthing
          ground incorporated in the power cord). The mains plug shall only be
          inserted in a socket outlet provided with a protective earth contact. Any
          interruption of the protective conductor inside or outside of the product
          is likely to make the product dangerous. Intentional interruption is
          prohibited.
          4. Push the module end of the power cable into the power module on the
             rear panel of the synthesizer until it is rmly seated.
          5. Connect the plug end of the power cable to a suitable mains power
             receptacle.
          6. Set the POWER switch (LINE on an HP 83711A/12A) to on ( ).
              When you turn the synthesizer on, the displays and annunciators light
              momentarily and the ventilation fan starts. In addition, it is normal for the
              LED annunciators (including the MSG LED) to turn on and then o during
              a power-up test. (The MSG LED should be o when the power-up test is
              complete.)


            NOTE
            Error message 940 TIMEBASE OVEN COLD will remain on after the power-up test if the
            instrument has the Option 1E5 installed, and has been disconnected from the mains power. This error
            should go away after a short warm-up period.




              Once the power-up test is complete, frequency and power level will be
              displayed in the right-most display and pulse parameters will be shown in
              the display if pulse modulation is enabled.




          1-6
Verifying Synthesizer Functionality




The verication procedure is suitable for incoming inspection; however,
you can refer to the HP 83711A/12A and 83711B/12B Synthesized CW
Generators Service Guide or the HP 83711A/12A, HP 83711B/12B, HP
83731A/32A, and HP 83731B/32B Synthesized CW Generators Calibration
Guide for procedures that test all warranted specications.
1. Activate the SELF TEST special function.
   To activate the SELF TEST special function, perform the following
   procedure.
   a. Press the 4SPCL 5 key.

   b. Press 4 5 on the synthesizer numeric keypad.
            5


   c. Terminate the special function entry by pressing the 4 5 (ENTER) key.
                                                             Hz


       The display will read SELF TEST?, PRESS ENTER.
2. Press the ENTER key again to initiate the synthesizer self test routine.
   When the self test routine is running, the display will alternately read
   SELF TESTING and PRESETTING INSTRUMENT. After the test completes,
   the display momentarily reads SELF TEST PASSED.
3. If the self test indicates an error condition, refer to the section entitled
   \If You Encounter a Problem" at the end of this chapter.
   If an error condition exists, the display will momentarily read
   TEST XX = YY where XX and YY are numbers indicating the error
   condition.




                                                                            1-7
If You Encounter a Problem




If you have a problem while installing or verifying the synthesizer, check
the following list of commonly encountered problems and troubleshooting
procedures.
If the problem that you encounter is not in the following list, contact the
nearest Hewlett-Packard oce for assistance.


  NOTE
  When transporting the product, use original packaging or comparable only.




Mechanical or Electrical Damage


If the instrument is mechanically or electrically damaged:
  Contact the nearest Hewlett-Packard oce.
If the shipping materials are damaged and the instrument is mechanically
or electrically damaged:
  Contact the carrier as well as the nearest Hewlett-Packard oce.
  Keep the shipping materials for the carrier's inspection.




1-8
                                                         Installing and Verifying the Synthesizer
                                                                 If You Encounter a Problem




          Power-up Problems

          If the synthesizer has no power:
            Check that the power cord is fully seated in both the mains power
            receptacle and the synthesizer power module.
            Check that the synthesizer line fuse is not open.
WARNING
          For continued protection against re hazard, replace line fuse only with
          same type and ratings (type nA/nV). The use of other fuses or materials is
          prohibited.
            Refer to Figure 1-1 to remove the fuse from the power module. You can use
            a continuity light or an ohmmeter to check the fuse. An ohmmeter should
            read very close to zero ohms if the fuse is good. The 6.3A, 250 V fuse is
            HP part number 2110-0703.




                                                                                             1-9
Installing and Verifying the Synthesizer
If You Encounter a Problem




                           Figure 1-1. Line Fuse Removal and Replacement

   Contact the nearest Hewlett-Packard oce for service, if necessary.
If the display lights, but the ventilation fan does not start:
   Check that the fan is not stuck. To check the fan, follow these steps:
   1. Set the POWER switch (LINE on an HP 83711A/12A) to standby ( ).
   2. Check that the fan blades are not jammed.
   Contact the nearest Hewlett-Packard oce for service, if necessary.




1-10
                                                           Installing and Verifying the Synthesizer




If the synthesizer MSG LED annunciator remains on after the power-up
test is complete:


  NOTE
  Error message 940 TIMEBASE OVEN COLD will remain on after the power-up test if the
  instrument has the Option 1E5 installed, and has been disconnected from the mains power. This error
  should go away after a short warm-up period.




If the synthesizer MSG annunciator is on, there is a problem with the
synthesizer. To determine the error and turn o the MSG annunciator, refer
to \To Read the Contents of the Error Queue" in Chapter 2 and the listing of
error messages in Chapter 8.




Self Test Failures


If the Self Test Fails:
  Check the contents of the error queue for error messages.
  To determine errors and clear them, refer to \To Read the Contents of the
  Error Queue" in Chapter 2 and the listing of error messages in Chapter 8.




                                                                                                 1-11
Installing and Verifying the Synthesizer
2




Performing Fundamental
Synthesizer Operations
          Performing Fundamental Synthesizer Op-
          erations

          This chapter describes fundamental synthesizer operations. The purpose
          of this chapter is to familiarize you with the fundamental operation of the
          synthesizer. Procedures cover how to enter or modify data, how to set the
          HP-IB address, how to save and recall instrument states, etc.


            NOTE
            This product has been designed and tested in accordance with IEC Publication 1010, Safety
            Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The
            instruction documentation contains information and warnings which must be followed by the user to
            ensure safe operation and to maintain the product in a safe condition.




WARNING   If this product is not used as specied, the protection provided by the
          equipment could be impaired. This product must be used in a normal
          condition (in which all means for protection are intact) only.




          2-2
To Enter Data with the Numeric Keypad




The synthesizer numeric keypad and units terminator keys provide one way
to enter function parameters.
1. Select the desired function key or shifted function key.
   The function must have a numeric value (parameter) associated with
   it in order to enter a new value with the numeric keypad. 4        5 and      FREQ

   4POWER LEVEL   5 are examples of functions with parameters.

   For example, to select power level so that its parameter is active, press the
   4POWER LEVEL   5 key.

   Notice that a cursor ( 9 ) appears over one of the digits in the power level
   display. The cursor will always appear over one of the digits of the active
   parameter.


  NOTE
  Two cursors will appear ( 9 9 ) when the cursor is positioned o the display.




2. Enter the desired value of the parameter by pressing the numeric keys
   and, if necessary, the negative and decimal keys.
   Notice that, as you press the keys, the display shows the numbers that are
   entered.
3. Press the appropriate units terminator key to enter the value.
   The units terminator keys appear to the right side of the numeric keypad.
   The following steps show how to enter a value of 09.5 dBm for power
   level:
   a. Press the 4           5 key.
                     POWER LEVEL

   b. Press 40 5 on the numeric keypad.
                  9.5

   c. Press the dBm (4 5) key to terminate the entry.
                             GHz


   Once you terminate the entry, the synthesizer updates the power level
   value to 09.5 dBm.

                                                                                        2-3
To Modify Data with the Knob




The knob on the synthesizer front panel is used to modify data. You turn the
knob in order to increase or decrease the parameter value of the currently
active function. Additionally, you can modify the position of the cursor ( 9 )
that is over the active parameter in order to increase or decrease the rate at
which the function parameter changes.
1. Select the function key of the parameter to be modied.
   When the function key is selected, the cursor appears over one of the
   digits of the selected parameter.
   For example, to select the frequency function, press the 4
                                                            FREQ  5 key. The
   frequency parameter is selected when the cursor appears over one of the
   digits of the frequency display.
2. Turn the knob clockwise to increase the parameter or
  counterclockwise to decrease the parameter.
  When you turn the knob, the digit under the cursor increases or decreases
  in steps of one.
  For example, when frequency is in its preset state and is the active
  parameter, the frequency display shows the following:
  3.000000000 GHz
  where the cursor appears over the third \0" to the right of the decimal.
  When you turn the knob in this case, the output frequency changes in 1
  MHz steps.
3. If you wish to move the position of the cursor one position to the right,
  press the 4)5 key.
  Pressing 4)5 once moves the position of the cursor to the right one digit in
  order to decrease the knob resolution by a factor of ten.
4. If you wish to move the position of the cursor one position to the left,
  press the 4(5 key.
  4(5 moves the position of the cursor to the left one digit in order to
  increase the knob resolution by a factor of ten.




2-4
                                                      Performing Fundamental Synthesizer Operations
                                                                   To Modify Data with the Knob



5. Continue to press the 4)5 or 4(5 key until the cursor is positioned over
   the desired digit.
   For example, suppose that frequency is the current active parameter, and
   you wish to change the cursor position from 1 MHz to 1 GHz. The steps
   that follow illustrate how to accomplish this:
   a. Press the 4        5 key.
                    PRESET


      When the instrument has been set to the preset state, the frequency
      display indicates the following:
      3.000000000 GHz
      where the cursor appears over the third \0" to the right of the decimal.
   b. Press the 4(5 key three times. This moves the cursor from the position
      corresponding to MHz to the position corresponding to GHz. (The
      cursor is now over the \3".)
   c. Rotate the knob slowly so that you can see the frequency change in 1
      GHz steps.



  NOTE
                * +                                                    * +
  The Knob and 4 5, 4 5 keys operate independently of each other. The 4 5, 4 5 keys increase or
  decrease parameters in dened steps.




                                                                                                  2-5
To Modify Data with the Arrow Keys




The synthesizer 4*5 and 4+5 (arrow) keys increase or decrease the value of the
currently active parameter by an amount set with the 4 STEP SIZE   5 key. The
amount set with the 4           5 key is referred to as the increment value.
                        STEP SIZE


1. Select the function key of the parameter to be modied.
   When the function key is selected, a cursor
   ( 9 ) appears over one of the digits of the selected parameter.
   For example, to select the frequency function, press the 4
                                                            FREQ    5 key. The
   frequency parameter is selected when the cursor appears over one of the
   digits of the frequency display.
2. Press the 4*5 key to increase the parameter by the current increment
  value or press the 4+5 key to decrease the parameter by the current
  increment value.
  For example, when frequency is in its preset state, the frequency display
  shows the following:
  3.000000000 GHz
  where the cursor appears over the third \0" to the right of the decimal.
  The preset frequency increment value is 100 MHz. When you press the 4*5
  key, the value of frequency increases to 3.100000000 GHz.
3. If you wish to change the increment value, you can do so using the
  4STEP      5
          SIZE   key.
  The 4 STEP SIZE5 key allows you to change the increment value of the
  currently active function. For example, suppose that frequency is the
  current active parameter, and you wish to change the increment value
  from its preset value of 100 MHz to 250 MHz. The steps that follow
  illustrate how to accomplish this:
  a. Press the 4          5 key.
                   STEP SIZE


  b. Type 4 5 on the synthesizer numeric keypad.
             250




2-6
                                                     Performing Fundamental Synthesizer Operations
                                                           To Modify Data with the Arrow Keys



 c. Terminate the frequency increment value entry by pressing the 4 5                     MHz

    key.
    When either the 4*5 or 4+5 key is pressed, the frequency will be either
    increased or decreased by 250 MHz instead of 100 MHz.



NOTE
              * +
The knob and 4 5, 4 5 keys operate independently of each other. The knob increases or decreases
the digit in the display that is under the cursor.




                                                                                                  2-7
To Save and Recall Synthesizer States




When you use the synthesizer for a specic application, you can save and
then recall the instrument state for future use. You can save up to ten
dierent instrument states.
1. Press the 4       5 key and then the 4
                  SHIFT                          5 (SAVE) key.
                                                      RECALL


   When 4       5, 4
             SHIFT         5 (SAVE) is pressed, the text SAVE STATE IN REG
                        RECALL

   XXX is shown on the synthesizer display where XXX is the last register
   number entered.
2. Use the numeric keypad to enter the desired register number.
   Valid register numbers are 0 through 9.
3. Press the 4 5 (ENTER) key to terminate the entry.
                  Hz


4. To recall the instrument state from memory, press the 4          5 key.         RECALL


   When 4         5 is pressed, the text RECALL STATE FROM REG XX is shown
             RECALL

   on the synthesizer display where XX is the last register recalled.
5. Use the numeric keypad to enter the desired register number.
6. Press the 4 5 (ENTER) key to terminate the entry.
                  Hz




  Notes

  1. When an instrument state is saved to an instrument state register, it will write over any instrument
     state previously stored to that register.

  2. If an instrument state has not been previously stored to an instrument state register, the
     synthesizer will be set to the preset state if you attempt to recall the instrument state from that
     register.




2-8
                                             Performing Fundamental Synthesizer Operations
                                                 To Save and Recall Synthesizer States




Programming Example

Use the following commands to store the instrument state to register 9 and
then recall it from register 9:
   OUTPUT 719; "*SAV 9"   Saves the current instrument state to register #9.
   OUTPUT 719; "*RCL 9"   Recalls the previously stored instrument state
                          from register #9.




                                                                                       2-9
To Read the Contents of the Error Queue




When one or more error messages are stored in the synthesizer error queue,
the front panel MSG LED annunciator will light. Once all error messages
have been read and all error conditions have been corrected, the MSG
annunciator will turn o.
1. Press the 4 5 key.
                 MSG


   When the 4 5 key is pressed, the most recent uncleared manual error
                 MSG

   number and the front panel error message will appear on the display.
2. Look up the manual error number in the table in Chapter 8.
   Chapter 8 is organized in ascending manual error number order.
3. Perform the instructions following the error message in the list to
   correct or clear the error condition.
   After you have completed the procedure in the list, return to this
   procedure to continue.
4. If the MSG LED annunciator is still lit, perform steps 1 through 3 again
   until the MSG annunciator turns o. If the MSG annunciator is turned
   o, continue with the next step.
5. Press the 4 5 key one more time.
                 MSG


   Pressing the 4    MSG  5   key again returns the display to normal operation.



  NOTE
  In the case of unleveled power or unlocked frequency, the MSG LED annunciator may remain lit
  continuously until the problem is corrected.




2-10
                                           Performing Fundamental Synthesizer Operations
                                            To Read the Contents of the Error Queue




Programming Example


To read the entire contents of the error queue, run the following program:
   10 DIM B$[160]                  Dimensions array B$ to accept 160 characters.
   20    OUTPUT 719; "SYST:ERR?"   Queries the oldest uncleared error number and
                                   message in the HP-IB error queue.
   30       ENTER 719; A,B$        Enters the error number into variable A and the
                                   HP-IB error message into variable B$.
   40       PRINT A,B$             Prints the error number and HP-IB error mes-
                                   sage to the controller screen.
   50 IF A<>0 THEN 20              Returns to line 20 if there are more errors in the
                                   queue.
   60 END




                                                                                   2-11
 To Set the HP-IB Address




 The synthesizer default HP-IB address is preset to 19. You can, however,
 change the HP-IB address of the synthesizer from the front panel.
 1. Press the 4SHIFT5 key on the synthesizer front panel.

 2. Press the ADDRESS (4    LOCAL5) key.

    The display indicates HPIB ADDRESS XX when the 4          5 key is pressed
                                                           LOCAL

    where XX is the current HP-IB address.
 3. Enter the desired HP-IB address using the numeric keypad.
    For example, if you want to set the synthesizer HP-IB address to 12, type
    4 5 on the numeric keypad.
     12


    The display indicates HPIB ADDRESS 12.
 4. Terminate the HP-IB address entry by pressing the 4 5 (ENTER) key.
                                                              Hz


    Note that when 4 5 (ENTER) is pressed, the cursor appears over the
                       Hz

    address parameter in the display. The synthesizer HP-IB address is now
    set to the new value (12 in the example), but you can still change it with
    the numeric keypad, knob, or arrow keys at this point. Pressing 4 5     Hz

    (ENTER) again returns the display to normal operation.




 Programming Example

 To set the synthesizer HP-IB address to 12, send the following command:
OUTPUT 719; "SYST:COMM:GPIB:ADDR 12"   Sets the synthesizer HP-IB address to 12.




 2-12
If You Encounter a Problem




If you have a problem operating the synthesizer, check the following list of
commonly encountered problems and troubleshooting procedures. If the
problem that you encounter is not in the following list, contact the nearest
Hewlett-Packard oce for assistance.


  NOTE
  When transporting the product, use original packaging or comparable only.




Data Entry Problems


If the data entry controls (keypad, knob, 4*5, 4+5, 4(5, 4)5 keys) do not
respond:
   Check that the ENTRY OFF function is not enabled.
   The ENTRY OFF function is not enabled when the cursor ( 9 ) appears
  over any of the parameters in the display. To return to normal entry mode,
  press the desired function key which has a numeric parameter associated
  with it (for example, press 4      5 if you want to enter frequency).
                                        FREQ


  Check that the function key which is selected accepts data.
  For instance, 4      5 accepts data, but, 4
                     FREQ                         5 does not.
                                                        LOCAL


  Check that the synthesizer is in the local (not remote) operating mode.
  If the synthesizer is in the remote operating mode, the RMT annunciator
  will be lit. Press the 4      5 key on the front panel to return the
                                LOCAL

  synthesizer to local operating mode.


                                                                              2-13
Performing Fundamental Synthesizer Operations
If You Encounter a Problem



If no cursor ( 9 ) appears over a parameter in the display:
   Check that the ENTRY OFF function is not enabled.
   The ENTRY OFF function is enabled when the cursor ( 9 ) doesn't appear
   over any of the parameters in the display. To return to normal entry mode,
   press the desired function key which has a numeric parameter associated
   with it (for example, press 4   5 if you want to enter frequency).
                                       FREQ


If the parameter you are trying to enter is not accepted by the
synthesizer:
   Ensure that you are not trying to set the parameter greater than or less
   than its limit. Refer to the specication table in this manual for the
   parameter limits.
   Check that the MSG LED annunciator is o.
   If the synthesizer MSG annunciator is on, there is a problem with the
   synthesizer. To determine the error and turn o the MSG annunciator,
   refer to \To Read the Contents of the Error Queue" in this chapter and the
   listing of error messages in Chapter 8.
If the synthesizer does not display/output the carrier frequency entered:
   Check that the frequency multiplier value entered is the expected value.
   When a multiplier value other than one is entered, the frequency
   resolution of the signal before multiplication must be obeyed. For example,
   for a desired frequency of 40 GHz using a multiplier value of three, the
   synthesizer would have to output a frequency of 13.333333333 GHz. Since
   frequency resolution at that frequency is 1 kHz, the closest the synthesizer
   can set the frequency is 13.333333000 GHz. This yields 39.999999000 GHz
   after multiplication, not 40 GHz.




2-14
                                     Performing Fundamental Synthesizer Operations




Programming Problems


If the synthesizer does not respond to programming commands:
  Refer to the procedure, \To Set the HP-IB Address," in this chapter to check
  and, if necessary, change the HP-IB address.




                                                                             2-15
Performing Fundamental Synthesizer Operations
3




Generating Signals with
the Synthesizer
Generating Signals with the Synthesizer





◦ Jabse Service Manual Search 2024 ◦ Jabse PravopisonTap.bg ◦ Other service manual resources online : FixyaeServiceinfo