HP 8590EM Series Calibration Guide


>> Download HP 8590EM Series Calibration Guide documenatation <<

Text preview - extract from the document
HP 8590 EM-Series EMC Analyzer
  Calibration Guide




      ABCDE
       HP Part No. 5962-0467
  Printed in USA December 1995
Notice
The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material,
including but not limited to, the implied warranties of merchantability and
tness for a particular purpose. Hewlett-Packard shall not be liable for errors
contained herein or for incidental or consequential damages in connection with
the furnishing, performance, or use of this material.

Restricted Rights Legend
Use, duplication, or disclosure by the U.S. Government is subject to restrictions
as set forth in subparagraph (c) (1) (ii) of the Rights of Technical Data and
Computer Software clause at DFARS 252.227-7013 for DOD agencies, and
subparagraphs (c) (1) and (c) (2) of the Commercial Computer Software
Restricted Rights clause at FAR 52.227-19 for other agencies.





 Copyright Hewlett-Packard Company 1995
c
All Rights Reserved. Reproduction, adaptation, or translation without prior
written permission is prohibited, except as allowed under the copyright laws.
1400 Fountaingrove Parkway, Santa Rosa CA, 95403-1799, USA
Certication
Hewlett-Packard Company certies that this product met its published
specications at the time of shipment from the factory. Hewlett-Packard
further certies that its calibration measurements are traceable to the United
States National Institute of Standards and Technology, to the extent allowed
by the Institute's calibration facility, and to the calibration facilities of other
International Standards Organization members.



Warranty
This Hewlett-Packard instrument product is warranted against defects in
material and workmanship for a period of one year from date of shipment.
During the warranty period, Hewlett-Packard Company will, at its option, either
repair or replace products which prove to be defective.
For warranty service or repair, this product must be returned to a service
facility designated by Hewlett-Packard. Buyer shall prepay shipping charges to
Hewlett-Packard and Hewlett-Packard shall pay shipping charges to return the
product to Buyer. However, Buyer shall pay all shipping charges, duties, and
taxes for products returned to Hewlett-Packard from another country.
Hewlett-Packard warrants that its software and rmware designated by
Hewlett-Packard for use with an instrument will execute its programming
instructions when properly installed on that instrument. Hewlett-Packard does
not warrant that the operation of the instrument, or software, or rmware will
be uninterrupted or error-free.
Limitation of Warranty
  The foregoing warranty shall not apply to defects resulting from improper or
  inadequate maintenance by Buyer, Buyer-supplied software or interfacing,
  unauthorized modication or misuse, operation outside of the environmental
  specications for the product, or improper site preparation or maintenance.
  NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. HEWLETT-
  PACKARD SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF
  MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Exclusive Remedies


                                                                                      iii
     THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND EXCLUSIVE
     REMEDIES. HEWLETT-PACKARD SHALL NOT BE LIABLE FOR ANY DIRECT,
     INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES,
     WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.



Assistance
Product maintenance agreements and other customer assistance agreements
are available for Hewlett-Packard products. For any assistance, contact your
nearest Hewlett-Packard Sales and Service Oce.



Compliance
This instrument has been designed and tested in accordance with IEC
Publication 348, Safety Requirements for Electronic Measuring Apparatus, and
has been supplied in a safe condition. The instruction documentation contains
information and warnings which must be followed by the user to ensure safe
operation and to maintain the instrument in a safe condition.



Safety Notes
The following safety notes are used throughout this manual. Familiarize
yourself with each of the notes and its meaning before operating this
instrument.
WARNING        Warning denotes a hazard. It calls attention to a procedure
               which, if not correctly performed or adhered to, could
               result in injury or loss of life. Do not proceed beyond
               a warning note until the indicated conditions are fully
               understood and met.




iv
CAUTION   Caution denotes a hazard. It calls attention to a procedure
          that, if not correctly performed or adhered to, would result in
          damage to or destruction of the instrument. Do not proceed
          beyond a caution sign until the indicated conditions are fully
          understood and met.




                                                                            v
General Safety Considerations
WARNING       No operator serviceable parts inside. Refer servicing to
              qualied personnel. To prevent electrical shock, do not
              remove covers.
              If this instrument is not used as specied, the protection
              provided by the equipment may be impaired. This
              instrument must be used in a normal condition (in which
              all means for protection are intact) only.
              For continued protection against re hazard, replace line
              fuse only with same type and rating ([F 5A/250V]). The
              use of other fuses or material is prohibited.

CAUTION       Before switching on this instrument, make sure that the line
              voltage selector switch is set to the voltage of the power
              supply and the correct fuse is installed.
              Always use the three-prong ac power cord supplied with this
              instrument. Failure to ensure adequate earth grounding by
              not using this cord may cause instrument damage.
              Only clean the instrument cabinet using a damp cloth.


L         The instruction documentation symbol. The product is marked
          with this symbol when it is necessary for the user to refer to the
          instructions in the documentation.
CE        The CE mark is a registered trademark of the European
          Community. (If accompanied by a year, it is when the design was
          proven.)
ISM1-A    This is a symbol of an Industrial Scientic and Medical Group 1
          Class A product.
CSA       The CSA mark is a registered trademark of the Canadian
          Standards Association.




vi
Manual Conventions
4Front-Panel Key5 This represents a key physically located on the instrument.
                  This indicates a \softkey," a key whose label is determined by
NNNNNNNNNNNNNNNNNNNNNNN

Softkey
                  the rmware of the instrument.
Screen Text This indicates text displayed on the instrument's screen.




                                                                                   vii
HP 8590 EM-Series EMC Analyzer Documentation
Description
The following documents are provided with the HP 8590 EM-Series EMC
Analyzer.
Verication Guide             provides information for verifying the
                              instrument's performance, specications and
                              characteristics, and customer support.
User's Guide                  describes instrument features and how to make
                              measurements with your EMC analyzer.
Quick Reference Guide         provides menu maps, error messages, and key
                              descriptions, remote programming codes, and
                              other helpful charts and tables.



In This Book
This book helps you test the performance of the EMC analyzer. The
rst three chapters of this guide provide information for performing the
performance verication tests. The next six chapters provide specications and
characteristics for the six EMC analyzer products. The last chapter gives helpful
information when your require support.
Chapter 1        contains information on which performance verication tests
                 that you will perform and lists the equipment required to
                 perform these tests.
Chapter 2        contains the performance verication tests to ensure your EMC
                 analyzer is operating properly.
Chapter 3        contains the test records for your use when performing the
                 performance verication tests.
Chapter 4        provides the specications and characteristics for the
                 HP 8591EM EMC analyzer.
Chapter 5        provides the specications and characteristics for the
                 HP 8593EM EMC analyzer.
Chapter 6        provides the specications and characteristics for the
                 HP 8594EM EMC analyzer.

viii
Chapter 7    provides the specications and characteristics for the
             HP 8595EM EMC analyzer.
Chapter 8    provides the specications and characteristics for the
             HP 8596EM EMC analyzer.
Chapter 9    provides the characteristics for the EMC analyzer with the RF
             lter section.
Chapter 10   contains information for providing customer support to you if
             you have a problem with your EMC analyzer.
1




                                                                             ix
                                                                                1
Calibrating
This chapter identies the performance test procedures which test the electrical
performance of the analyzer.
Allow the analyzer to warm up in accordance with the temperature stability
specications before performing the tests called out in this chapter.
None of the test procedures involve removing the cover of the analyzer.



Calibration
Calibration veries that the analyzer performance is within all specications. It
is time consuming and requires extensive test equipment. Calibration consists of
all the performance tests. For a complete listing of the performance tests, see
the performance verication tests table for your specic analyzer.



Operation Verication
Operation verication only tests the most critical specications. These tests
are recommended for incoming inspection, troubleshooting, or after repair.
Operation verication requires less time and equipment than the calibration.
See the performance verication tests table for your analyzer.




                                                                    Calibrating 1-1
Calibration Cycle
The performance tests in Chapter 2 should be used to check the analyzer against
its specications once every year. Specications are listed in this calibration
guide.
The 300 MHz frequency of the CAL OUT signal must be checked at the same
time and adjusted if necessary. Refer to the \10 MHz Frequency Reference
Adjustment" procedure in the assembly-level repair service guide.




1-2 Calibrating
Performance Verication Test Tables
The tables on the following pages list the performance tests in chapter 2. Select
the analyzer option being calibrated and perform the tests marked in the option
column.
A dot indicates that the test is required for calibration. Note that some of the
tests are used for both calibration and operation verication (marked with 45).




                                                                    Calibrating 1-3
                 Table 1-1. HP 8591EM Performance Verication Tests
                                                              Calibration for Instrument Option:
                 Performance Test Name                          Std1        004       010        101
 1. 10 MHz Reference Output Accuracy                                                                    
 2. 10 MHz Precision Frequency Reference Output                               
    Accuracy
 4. Frequency Readout and Marker Count Accuracy                   
                                                                 45           
                                                                             45          
                                                                                        45           
                                                                                                    45
 6. Noise Sidebands                                               
                                                                 45           
                                                                             45          
                                                                                        45           
                                                                                                    45
 7. System Related Sidebands                                                                         
 8. Frequency Span Readout Accuracy                              45          45         45          45
10. Residual FM                                                                                      
12. Sweep Time Accuracy                                                                              
13. Scale Fidelity                                               45          45         45          45
14. Reference Level Accuracy                                     45          45         45          45
16. Absolute Amplitude Calibration and Resolution (IF)           45          45         45          45
    Bandwidth Switching Uncertainties
17. Resolution (IF) Bandwidth Accuracy                                                               
18. Calibrator Amplitude Accuracy                                45          45         45          45
19. Frequency Response                                           45          45         45          45
24. Other Input Related Spurious Responses                                                           
29. Spurious Response2                                           45          45         45          45
34. Gain Compression                                                                                 
39. Displayed Average Noise Level                                45          45         45          45
44. Residual Responses                                                                               
47. Fast Time Domain Sweeps                                                                          
49. Absolute Amplitude, Vernier, and Power Sweep                                         
    Accuracy
52. Tracking Generator Level Flatness                                                    
54. Harmonic Spurious Outputs                                                            
56. Non-Harmonic Spurious Outputs                                                        
58. Tracking Generator Feedthrough                                                       
62. CISPR Pulse Response                                                                                

1 Use this column for all other options listed in this table.
                                      not

2 \Part 2: Third Order Intermodulation Distortion, 50 MHz" is not required for operation verication.



      1-4 Calibrating
                       Table 1-2. HP 8593EM Performance Verication Tests
                                                              Calibration for Instrument Option:
          Performance Verication Test Name                    Std1      004      010     026           027     101
 1. 10 MHz Reference Output Accuracy                                                                              
 2. 10 MHz Precision Frequency Reference Output                            
    Accuracy
 3. Comb Generator Frequency Accuracy                                                                             
 5. Frequency Readout and Marker Count Accuracy                  
                                                                45         
                                                                          45         
                                                                                    45        
                                                                                             45          
                                                                                                        45        
                                                                                                                 45
 6. Noise Sidebands                                             45        45        45       45         45       45
 7. System Related Sidebands                                                                                      
 9. Frequency Span Readout Accuracy                             45        45        45       45         45       45
11. Residual FM                                                                                                   
12. Sweep Time Accuracy                                                                                           
13. Scale Fidelity                                              45        45        45       45         45       45
15. Reference Level Accuracy                                    45        45        45       45         45       45
16. Absolute Amplitude Calibration and Resolution (IF)          45        45        45       45         45       45
    Bandwidth Switching Uncertainties
17. Resolution (IF) Bandwidth Accuracy                                                                            
18. Calibrator Amplitude Accuracy                               45        45        45       45         45       45
20. Frequency Response                                          45        45        45       45         45       45
25. Other Input Related Spurious Responses                                                                        
30. Spurious Response2                                          45        45        45       45         45       45
35. Gain Compression                                                                                              
40. Displayed Average Noise Level                               45        45        45       45         45       45
46. Residual Responses                                                                                            
48. Fast Time Domain Sweeps                                                                                       
50. Absolute Amplitude Accuracy                                                     
51. Power Sweep Range                                                               
53. Tracking Generator Level Flatness                                               
55. Harmonic Spurious Outputs                                                       
57. Non-Harmonic Spurious Outputs                                                   
60. Tracking Generator Feedthrough                                                  
61. Tracking Generator LO Feedthrough Amplitude                                     
62. CISPR Pulse Response                                                                      

1 Use this column for all other options listed in this table.
                                      not

2 \Part 2: Third Order Intermodulation Distortion, 50 MHz" is not required for operation verication.

                                                                                                        Calibrating 1-5
                  Table 1-3. HP 8594EM Performance Verication Tests
                                                              Calibration for Instrument Option:
          Performance Verication Test Name                     Std1        004       010        101
 1. 10 MHz Reference Output Accuracy                                                             
 2. 10 MHz Precision Frequency Reference Output                               
    Accuracy
 4. Frequency Readout and Marker Count Accuracy                   
                                                                 45           
                                                                             45          
                                                                                        45        
                                                                                                 45
 6. Noise Sidebands                                               
                                                                 45           
                                                                             45          
                                                                                        45        
                                                                                                 45
 7. System Related Sidebands                                                                      
 9. Frequency Span Readout Accuracy                              45          45         45       45
11. Residual FM                                                                                   
12. Sweep Time Accuracy                                                                           
13. Scale Fidelity                                               45          45         45       45
15. Reference Level Accuracy                                     45          45         45       45
16. Absolute Amplitude Calibration and Resolution (IF)           45          45         45       45
    Bandwidth Switching Uncertainties
17. Resolution (IF) Bandwidth Accuracy                                                            
18. Calibrator Amplitude Accuracy                                45          45         45       45
21. Frequency Response                                           45          45         45       45
26. Other Input Related Spurious Responses                                                        
31. Spurious Response2                                           45          45         45       45
36. Gain Compression                                                                              
41. Displayed Average Noise Level                                45          45         45       45
45. Residual Responses                                                                            
48. Fast Time Domain Sweeps                                                                       
50. Absolute Amplitude Accuracy                                                         
51. Power Sweep Range                                                                   
53. Tracking Generator Level Flatness                                                   
55. Harmonic Spurious Outputs                                                           
57. Non-Harmonic Spurious Outputs                                                       
59. Tracking Generator Feedthrough                                                      
61. Tracking Generator LO Feedthrough Amplitude                                         
62. CISPR Pulse Response                                                                         

1 Use this column for all other options listed in this table.
                                       not

2 \Third Order Intermodulation Distortion" is not required for operation verication.


      1-6 Calibrating
                  Table 1-4. HP 8595EM Performance Verication Tests
                                                              Calibration for Instrument Option:
          Performance Verication Test Name                     Std1        004       010        101
 1. 10 MHz Reference Output Accuracy                                                                
 2. 10 MHz Precision Frequency Reference Output                               
    Accuracy
 5. Frequency Readout and Marker Count Accuracy                   
                                                                 45           
                                                                             45          
                                                                                        45          
                                                                                                   45
 6. Noise Sidebands                                               
                                                                 45           
                                                                             45          
                                                                                        45          
                                                                                                   45
 7. System Related Sidebands                                                                        
 9. Frequency Span Readout Accuracy                              45          45         45         45
11. Residual FM                                                                                     
12. Sweep Time Accuracy                                                                             
13. Scale Fidelity                                               45          45         45         45
15. Reference Level Accuracy                                     45          45         45         45
16. Absolute Amplitude Calibration and Resolution (IF)           45          45         45         45
    Bandwidth Switching Uncertainties
17. Resolution (IF) Bandwidth Accuracy                                                              
18. Calibrator Amplitude Accuracy                                45          45         45         45
22. Frequency Response                                           45          45         45         45
27. Other Input Related Spurious Responses                                                          
32. Spurious Response2                                           45          45         45         45
37. Gain Compression                                                                                
42. Displayed Average Noise Level                                45          45         45         45
46. Residual Responses                                                                              
48. Fast Time Domain Sweeps                                                                         
50. Absolute Amplitude Accuracy                                                         
51. Power Sweep Range                                                                   
53. Tracking Generator Level Flatness                                                   
55. Harmonic Spurious Outputs                                                           
57. Non-Harmonic Spurious Outputs                                                       
60. Tracking Generator Feedthrough                                                      
61. Tracking Generator LO Feedthrough Amplitude                                         
62. CISPR Pulse Response                                                                            

1 Use this column for all other options listed in this table.
                                       not

2 \Third Order Intermodulation Distortion" is not required for operation verication.


                                                                                           Calibrating 1-7
                  Table 1-5. HP 8596EM Performance Verication Tests
                                                               Calibration for Instrument Option:
          Performance Verication Test Name                     Std1       004        010       101
 1. 10 MHz Reference Output Accuracy                                                            
 2. 10 MHz Precision Frequency Reference Output                               
    Accuracy
 3. Comb Generator Frequency Accuracy                             
                                                                 45           
                                                                             45          
                                                                                        45       
                                                                                                45
 5. Frequency Readout and Marker Count Accuracy                   
                                                                 45           
                                                                             45          
                                                                                        45       
                                                                                                45
 6. Noise Sidebands                                              45          45         45      45
 7. System Related Sidebands                                                                     
 9. Frequency Span Readout Accuracy                              45          45         45      45
11. Residual FM                                                                                  
12. Sweep Time Accuracy                                                                          
13. Scale Fidelity                                               45          45         45      45
15. Reference Level Accuracy                                     45          45         45      45
16. Absolute Amplitude Calibration and Resolution (IF)           45          45         45      45
    Bandwidth Switching Uncertainties
17. Resolution (IF) Bandwidth Accuracy                                                           
18. Calibrator Amplitude Accuracy                                 
                                                                 45           
                                                                             45          
                                                                                        45       
                                                                                                45
23. Frequency Response                                           45          45         45      45
28. Other Input Related Spurious Responses                                                       
33. Spurious Response2                                           45          45         45      45
38. Gain Compression                                                                             
43. Displayed Average Noise Level                                45          45         45      45
46. Residual Responses                                                                           
48. Fast Time Domain Sweeps                                                                      
50. Absolute Amplitude Accuracy                                                         
51. Power Sweep Range                                                                   
53. Tracking Generator Level Flatness                                                   
55. Harmonic Spurious Outputs                                                           
57. Non-Harmonic Spurious Outputs                                                       
60. Tracking Generator Feedthrough                                                      
61. Tracking Generator LO Feedthrough Amplitude                                         
62. CISPR Pulse Response                                                                        

1 Use this column for all other options listed in this table.
                                       not

2 \Third Order Intermodulation Distortion" is not required for operation verication.

      1-8 Calibrating
Safety
Familiarize yourself with the safety symbols marked on the analyzer, and read
the general safety instructions and the symbol denitions given in the front of
this guide before you begin verifying performance of the EMC analyzer.



Before You Start
There are four things you should do before starting a performance verication
test:
  Switch the analyzer on and let it warm up in accordance with the
  temperature stability specication.
  Read \Making a Measurement" in your analyzer user's guide.
  After the analyzer has warmed up as specied, perform the self-calibration
  procedure documented in \Improving Accuracy With Self-Calibration
  Routines" in the HP 8590 EM-Series EMC Analyzer User's Guide. The
  performance of the analyzer is only specied after the analyzer calibration
  routines have been run and if the analyzer is autocoupled.
  Read the rest of this section before you start any of the tests, and make
  a copy of the Performance Verication Test Record described below in
  \Recording the test results."

Test equipment you will need
Tables 1-6 through 1-9 list the recommended test equipment for the performance
tests. Any equipment that meets the critical specications given in the table can
be substituted for the recommended model.

Recording the test results
Performance verication test records, for each EMC analyzer, are provided in
the chapter following the tests.
Each test result is identied as a TR Entry in the performance tests and on the
performance verication test record. We recommend that you make a copy of
the performance verication test record, record the test results on the copy, and


                                                                    Calibrating 1-9
keep the copy for your calibration test record. This record could prove valuable
in tracking gradual changes in test results over long periods of time.

Frequency and amplitude self-calibration
Perform the frequency and amplitude self-calibration routines at least once
per day, or if the analyzer fails a verication test. To perform self-calibration,
press 4CAL5 then CAL FREQ & AMPTD . The instrument must be up to operating
                   NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




temperature in order for this test to be valid. Press CAL STORE when the test is
                                                                        NNNNNNNNNNNNNNNNNNNNNNNNNNNNN




complete. If the analyzer continuously fails one or more specications, complete
any remaining tests and record all test results on a copy of the test record.
Then refer to the \Customer Support" chapter for instructions on how to solve
the problem.

Periodically verifying operation
The analyzer requires periodic verication of operation. Under most conditions
of use, you should test the analyzer at least once a year with either operation
verication or the complete set of performance verication tests.




1-10 Calibrating
                      Table 1-6. Recommended Test Equipment
      Equipment                       Critical Specications for      Recommended
                                       Equipment Substitution            Model
Digital Voltmeter       Input Resistance: 10 megohms                    HP 3456A
                        Accuracy: 610 mV on 100 V range


Frequency Counter1      Frequency: 10 MHz                              HP 5334A/B
                        Resolution: 60.002 Hz
                        External Timebase


Frequency Standard      Frequency: 10 MHz                                HP 5061B
                        Timebase Accy (Aging): <1 21009 /day


Measuring Receiver      Compatible with Power Sensors                   HP 8902A
                        dB Relative Mode
                        Resolution: 0.01 dB
                        Reference Accuracy: 61.2%


Microwave Frequency     Frequency Range: 9 MHz to 7 GHz                 HP 5343A
Counter                 Timebase Accy (Aging): <5 2 10010 /day


Power Meter             Power Range: Calibrated in dBm and               HP 436A
                        dB relative to reference power 070 dBm
                        to +44 dBm, sensor dependent


Power Sensor            Frequency Range: 100 kHz to 1800 MHz            HP 8482A
                        Maximum SWR: 1.60 (100 kHz to 300 kHz)
                                       1.20 (300 kHz to 1 MHz)
                                       1.1 (1 MHz to 2.0 GHz)
                                       1.30 (2.0 to 2.9 GHz)


Power Sensor2           Frequency Range: 1 MHz to 2 GHz                 HP 8483A
                        Maximum SWR: 1.18 (600 kHz to 2.0 GHz) 75 



Power Sensor,           Frequency Range: 300 MHz                        HP 8484A
 Low Power              Amplitude Range: 020 dBm to 070 dBm
                        Maximum SWR: 1.1 (300 MHz)

1 Precision Frequency Reference only                                  Calibrating 1-11
2 HP 8591EM only
                Table 1-6. Recommended Test Equipment (continued)
      Equipment                    Critical Specications for             Recommended
                                    Equipment Substitution                   Model
Power Sensor1          Frequency Range: 50 MHz to 26.5 GHz                  HP 8485A
                       Maximum SWR: 1.10 ( 300 MHz)
                                      1.15 (50 MHz to 100 MHz)
                                      1.10 (100 MHz to 2.0 GHz)
                                      1.15 (2.0 GHz to 12.4 GHz)
                                      1.20 (12.4 GHz to 18.0 GHz)
                                      1.25 (18.0 GHz to 26.5 GHz)


Pulse Generator        Period Range: 1 ms to 980 ms 62%,                    HP 8161A
                                     single pulse mode
                       Level 02 V to +2 V
                       Transition Time: 6 ns 610%, 61 ns
                       Pulse Width: 150 ns to 3 s 61% 61 ns


Pulse Generator2       Frequency: 100 Hz                                    HP 8116A
                       Duty Cycle: 50%
                       Output: TTL


Signal Generator       Frequency Range: 1 MHz to 1000 MHz                   HP 8640B,
                       Amplitude Range: 035 to +16 dBm                      Option 002
                       SSB Noise: <0120 dBc/Hz at 20 kHz oset                   or
                                                                            HP 8642A



Spectrum Analyzer,     Frequency Range: 100 kHz to 7 GHz                   HP 8566A/B
 Microwave             Relative Amplitude Accuracy:
                       100 kHz to 1.8 GHz: <61.8 dB
                       Frequency Accuracy: <610 kHz @ 7 GHz


Synthesized Sweeper3   Frequency Range: 10 MHz to 22 GHz                   HP 8340A/B
                       Frequency Accuracy (CW): 6 0.02%                        or
                       Leveling Modes: Internal and External               HP 83630A
                       Modulation Modes: AM
                       Power Level Range: 035 to +16 dBm


Synthesizer/Function   Frequency Range:    0.1 Hz to 500 Hz                 HP 3325B
 Generator             Frequency Accuracy: 60.02%
                       Waveform: Triangle
1-12 Calibrating

Synthesizer/Level      Frequency Range: 1 kHz to 80 MHz                     HP 3335A
 Generator             Amplitude Range: +12 to 085 dBm
                       Flatness: 60.15 dB
                       Attenuator Accuracy: 60.09 dB

1 Not for HP 8591EM
2 HP 8591EM only
3 For HP 8591EM, HP 8593EM Option 026 or Option 027, HP 8594EM, HP 8595EM, and HP 8596EM
Calibrating 1-13
                        Table 1-7. Recommended Accessories
      Equipment                      Critical Specications for   Recommended
                                        Accessory Substitution        Model
Attenuator, 3 dB        Type N (m to f)                             HP 8491A
                        Attenuation: 3 dB                           Option 003
                        Frequency: dc to 12.4 GHz


Attenuator, 10 dB       Type N (m to f)                            HP 8491A
                        Frequency: 300 MHz                         Option 010



Attenuator, 1 dB Step   Attenuation Range: 0 to 12 dB               HP 355C
                        Frequency Range: 50 MHz
                        Connectors: BNC female


Attenuator, 10 dB Step Attenuation Range: 0 to 30 dB                HP 355D
                       Frequency Range: 50 MHz
                       Connectors: BNC female


Directional Bridge      Frequency Range: 0.1 to 110 MHz             HP 8721A
                        Directivity: >40 dB
                        Maximum VSWR: 1.1:1
                        Transmission Arm Loss: 6 dB (nominal)
                        Coupling Arm Loss: 6 dB (nominal)


Directional Coupler     Frequency Range: 1.7 GHz to 8 GHz           0955-0125
                        Coupling: 16 dB (nominal)
                        Max. Coupling Deviation: 61 dB
                        Directivity: 14 dB minimum
                        Flatness: 0.75 dB maximum
                        VSWR: <1.45
                        Insertion Loss: <1.3 dB


Low Pass Filter,        Cuto Frequency: 50 MHz                      0955-0306
 50 MHz                 Rejection at 80 MHz: >50 dB


Low Pass Filter,        Cuto Frequency: 300 MHz                     0955-0455
 300 MHz                Bandpass Insertion Loss: <0.9 dB
                         at 300 MHz
                        Stopband Insertion Loss: >40 dB
1-14 Calibrating         at 435 MHz


Low Pass Filter,        Cuto Frequency: 4.4 GHz                    HP 11689A
 4.4 GHz                Rejection at 5.5 GHz: >40 dB
                    Table 1-7. Recommended Accessories (continued)
      Equipment                        Critical Specications for    Recommended
                                         Accessory Substitution          Model
Modulator                Frequency 50 MHz                              0955-0533
Teletech SC35B           ON/OFF RATIO >70 dB
                         Switching Speed 2 ns
                         Insertion Loss: 5 dB


Power Splitter1          Frequency Range: 50 kHz to 1.8 GHz            HP 11667A
                         Insertion Loss: 6 dB (nominal)
                         Output Tracking: <0.25 dB
                         Equivalent Output SWR: <1.22:1


Power Splitter2          Frequency Range: 50 kHz to 22 GHz             HP 11667B
                         Insertion Loss: 6 dB (nominal)
                         Output Tracking: <0.25 dB
                         Equivalent Output SWR: <1.22:1


Termination, 50 
        Impedance: 50 
 (nominal)                      HP 908A
                          (2 required for Option 010)




Termination3                                                            HP 909D

1 HP 8591EM and HP 8593EM
2 HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM
3 HP 8595EM and HP 8596EM only




                                                                     Calibrating 1-15
                       Table 1-8. Recommended Adapters
     Equipment                     Critical Specications for   Recommended
                                     Accessory Substitution         Model
Adapter              APC 3.5 (f) to APC 3.5 (f) (2 required)      5061-5311

Adapter1             BNC (f) to dual banana plug                 1251-1277

Adapter              BNC (f) to SMA (m)                          1250-1200

Adapter              BNC (m) to BNC (m)                          1250-0216

Adapter              BNC tee (m) (f) (f)                         1250-0781

Adapter              Type N (f) to APC 3.5 (f)                   1250-1745

Adapter              Type N (f) to APC 3.5 (m)                   1250-1750

Adapter              Type N (m) to APC 3.5 (m)                   1250-1743

Adapter2             Type N (m) to APC 3.5 (f)                   1250-1744

Adapter3             Type N (f) to BNC (f)                       1250-1474

Adapter              Type N (f) to BNC (m) (2 required)          1250-1477

Adapter              Type N (m) to BNC (f) (4 required)          1250-1476

Adapter              Type N (m) to BNC (m) (2 required)          1250-1473

Adapter              Type N (f) to Type N (f)                    1250-1472

Adapter3             Type N (m) to Type N (m)                    1250-1475


1 HP 8591EM only
2 HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM only
3 HP 8591EM, HP 8594EM, HP 8595EM, and HP 8596EM only




1-16 Calibrating
                          Table 1-9. Recommended Cables
        Equipment                     Critical Specications for           Recommended
                                          Cable Substitution                   Model
Cable1                 Cal Comb                                             08592-60061
                       Connectors: SMA (m) both ends
Cable1                 Connectors: SMA (m) both ends                            8120-1578
                       Length: 61 cm (18 in)
Cable2                 Frequency Range: 10 MHz to 26.5 GHz                      8120-4921
                       Maximum SWR: <1.4 at 26.5 GHz
                       Length: 91 cm (36 in)
                       Connectors: APC 3.5 (m) both ends
                       Maximum Insertion Loss 2 dB
                       (2 required)


Cable                  Type N, 183 cm (72 in)                                   HP 11500A

Cable                  Type N, 62 cm (24 in)                                HP 11500B/C

Cable                  Type N, 152 cm (60 in)                                   HP 11500D

Cable                  Frequency Range: dc to 1 GHz                             HP 10503A
                       Length: 91 cm (36 in)
                       Connectors: BNC (m) both ends
                       (2 required)


Cable                  Frequency Range: dc to 310 MHz                           HP 10502A
                       Length: 23 cm (9 in)
                       Connectors: BNC (m) both ends

1 For HP 8593EM only
2 For HP 8593EM Option 026 or 027, or HP 8594EM, HP 8595EM, or HP 8596EM only




                                                                           Calibrating 1-17
                                                                                                     2
Performance Verication Tests
These tests verify the electrical performance of the EMC analyzer. Allow
the EMC analyzer to warm up in accordance with the temperature stability
specications before performing the tests.
CAUTION        All performance verication tests (except test 62. CISPR Pulse
               Response) must be performed with the EMC analyzer set in the
               spectrum analyzer mode.
               To set the EMC analyzer to the spectrum analyzer mode, press:
                 4MODE5
                 NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

                 SPECTRUM ANALYZER


               Performance verication test 62. CISPR Pulse Response must
               be performed with the EMC analyzer set in the EMC analyzer
               mode.
               To set the analyzer to the EMC analyzer mode, press:
                 4MODE5
                 NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

                 EMC ANALYZER




                                                                         Performance Verication Tests 2-1
1. 10 MHz Reference Output Accuracy,
HP 8590 EM-Series
If your instrument is equipped with a Precision Frequency Reference, perform
\10 MHz Precision Frequency Reference Output Accuracy," instead.
This performance verication test must be performed with the EMC analyzer set
in the spectrum analyzer mode.
The settability is measured by changing the setting of the digital-to-analog
converter (DAC) which controls the frequency of the timebase. The frequency
dierence per DAC step is calculated and compared to the specication.
The related adjustment for this performance verication test is the \10 MHz
Reference, HP 8590 EM-Series".

Equipment Required
  Microwave frequency counter
  Frequency standard
  Cable, BNC, 122 cm (48 in) (2 required)




          Figure 2-1. 10 MHz Reference Output Accuracy Test Setup
Procedure
The test results will be invalid if REF UNLK is displayed at any time during this
test. REF UNLK will be displayed if the internal reference oscillator is unlocked
from the 10 MHz reference. A REF UNLK might occur if there is a hardware
failure or if the jumper between 10 MHz REF OUTPUT and EXT REF IN on the
rear panel is removed.


2-2 Performance Verication Tests
                                 1. 10 MHz Reference Output Accuracy, HP 8590 EM-Series
1. Connect the equipment as shown in Figure 2-1.
2. Set the frequency counter controls as follows:
     SAMPLE RATE : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Midrange
     50 
/1 
 SWITCH : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 50 

     10 Hz-500 MHz/500 MHz-26.5 GHz SWITCH : : : : : : : : : : : : : : : : : 10 Hz-500 MHz
     FREQUENCY STANDARD (Rear panel) : : : : : : : : : : : : : : : : : : : : : : : : : EXTERNAL
3. Wait for the frequency counter reading to stabilize. Record the frequency
   counter reading in the 10 MHz Reference Accuracy Worksheet as Counter
   Reading 1.
4. Set the EMC analyzer by pressing the following keys:
     4FREQUENCY5 037 4Hz5
                  NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

     4CAL5 More 1 of 4 More 2 of 4 VERIFY TIMEBASE

5. Record the number in the active function block of the EMC analyzer in the
   10 MHz Reference Accuracy Worksheet as the Timebase DAC Setting.
6. Add one to the Timebase DAC Setting recorded in step 5, then enter this
   number using the DATA keys on the EMC analyzer.
   For example, if the timebase DAC setting is 105, press 1,0,6 4Hz5.
7. Wait for the frequency counter reading to stabilize. Record the frequency
   counter reading in the 10 MHz Reference Accuracy Worksheet as Counter
   Reading 2.
8. Subtract one from the Timebase DAC Setting recorded in step 5, then enter
   this number using the DATA keys on the EMC analyzer.
   For example, if the timebase DAC setting is 105, press 1, 0, 4, 4Hz5.
9. Wait for the frequency counter reading to stabilize. Record the frequency
   counter reading in the 10 MHz Reference Accuracy Worksheet as Counter
   Reading 3.




                                                                                                                    Performance Verication Tests 2-3
1. 10 MHz Reference Output Accuracy, HP 8590 EM-Series
                  10 MHz Reference Accuracy Worksheet
                                Description     Measurement
                         Counter Reading 1              Hz
                         Timebase DAC Setting
                         Counter Reading 2              Hz
                         Counter Reading 3              Hz


10. Calculate the frequency settability by performing the following steps:
    a. Calculate the frequency dierence between Counter Reading 2 and
       Counter Reading 1.
    b. Calculate the frequency dierence between Counter Reading 3 and
       Counter Reading 1.
    c. Divide the frequency dierence with the greatest absolute value by two
       and record the value as TR Entry 1 of the performance verication test
       record. The settability should be less than 6150 Hz.
    d. Press 4PRESET5 on the EMC analyzer. The timebase DAC will be reset
       automatically to the value recorded in step 5.




2-4 Performance Verication Tests
2. 10 MHz Precision Frequency Reference Output Accuracy, HP 8590 EM-Series Option 004

              2. 10 MHz Precision Frequency Reference Output
              Accuracy,
              HP 8590 EM-Series Option 004
              If the EMC analyzer is not equipped with a Precision Frequency Reference,
              perform \10 MHz Reference Output Accuracy," instead.
              This performance verication test must be performed with the EMC analyzer set
              in the spectrum analyzer mode.
              This test measures the warmup characteristics of the 10 MHz reference
              oscillator. The ability of the 10 MHz oscillator to meet its warmup characteristics
              gives a high level of condence that it will also meet its yearly aging
              specication.
              A frequency counter is connected to the 10 MHz REF OUTPUT. After the EMC
              analyzer has been allowed to cool for at least 60 minutes, the EMC analyzer
              is powered on. A frequency measurement is made ve minutes after power
              is applied and the frequency is recorded. Another frequency measurement is
              made 25 minutes later (30 minutes after power is applied) and the frequency is
              recorded. A nal frequency measurement is made 60 minutes after power is
              applied. The dierence between each of the rst two frequency measurements
              and the last frequency measurement is calculated and recorded.
              The related adjustment for this procedure is \10 MHz Precision Frequency
              Reference for
              Option 004, HP 8590 EM-Series".

              Equipment Required
                Frequency counter
                Frequency standard
                Cable, BNC, 122 cm (48 in) (two required)

              Procedure
              The EMC analyzer must have been allowed to sit with the power o for at least
              60 minutes before performing this procedure. This adequately simulates a cold
              start. A cold start is dened as the EMC analyzer being powered on after being
              o for at least 60 minutes.
               1. Allow the EMC analyzer to sit with the power o for at least 60 minutes
                  before proceeding. Connect the equipment as shown in Figure 2-2.
                                                                   Performance Verication Tests 2-5
2. 10 MHz Precision Frequency Reference Output Accuracy, HP 8590 EM-Series Option 004
 2. Set the EMC analyzer LINE switch on. Record the Power On Time below.
                                                   Power On Time




                                 Figure 2-2.
       10 MHz Precision Frequency Reference Output Accuracy Test Setup




2-6 Performance Verication Tests
2. 10 MHz Precision Frequency Reference Output Accuracy, HP 8590 EM-Series Option 004
                3. Set the frequency counter controls as follows:
                      FUNCTION/DATA : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : FREQ A
                      INPUT A
                      X10 ATTN : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : OFF
                      AC : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : OFF
                      50 
 Z : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : OFF
                      AUTO TRIG : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : ON
                      100 kHz FILTER A : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : OFF
                4. On the frequency counter, select a 10 second gate time by pressing
                   4GATE TIME5 10 4GATE TIME5. Oset the displayed frequency by 010.0 MHz by
                   pressing 4MATH5, 4SELECT/ENTER5, 4CHS/EEX5 10 4CHS/EEX5 6 4SELECT/ENTER5,
                   4SELECT ENTER5. The frequency counter should now display the dierence
                   between the INPUT A signal and 10.0 MHz with 0.001 Hz resolution.
                5. Proceed with the next step 5 minutes after the Power On Time noted in step
                   2.
                6. Wait at least two periods for the frequency counter to settle. Record the
                   frequency counter reading in the 10 MHz Reference Accuracy Worksheet as
                   Counter Reading 1 with 0.001 Hz resolution.
                7. Proceed with the next step 30 minutes after the Power On Time noted in
                   step 2.
                8. Record the frequency counter reading in the 10 MHz Reference Accuracy
                   Worksheet as Counter Reading 2 with 0.001 Hz resolution.
                9. Proceed with the next step 60 minutes after the Power On Time noted in
                   step 2.
               10. Wait at least two periods for the frequency counter to settle. Record the
                   frequency counter reading in the 10 MHz Reference Accuracy Worksheet as
                   Counter Reading 3 with 0.001 Hz resolution.
                                                                  10 MHz Reference Accuracy Worksheet
                                                                             Description                      Measurement
                                                                      Counter Reading 1                                       Hz
                                                                      Counter Reading 2                                       Hz
                                                                      Counter Reading 3                                       Hz




                                                                                                                        Performance Verication Tests 2-7
2. 10 MHz Precision Frequency Reference Output Accuracy, HP 8590 EM-Series Option 004
11. Calculate the 5 Minute Warmup Error by subtracting Reading 3 from
    Reading 1 and dividing the result by 10 MHz.
        5 Minute Warmup Error = (Reading 1 0 Reading 3) / (10.0 2 106 )
12. Record the results as TR Entry 1 of the performance verication test
    record.
13. Calculate the 30 Minute Warmup Error by subtracting Reading 3 from
    Reading 2 and dividing the result by 10 MHz.
        30 Minute Warmup Error = (Reading 2 0 Reading 3) / (10.0 2 106 )
14. Record the results as TR Entry 2 of the performance verication test
    record.




2-8 Performance Verication Tests
      3. Comb Generator Frequency Accuracy, HP 8593EM and HP 8596EM

3. Comb Generator Frequency Accuracy,
HP 8593EM and HP 8596EM
This performance verication test must be performed with the EMC analyzer set
in the spectrum analyzer mode.
A 100 MHz signal from a synthesized source and the output from a comb
generator are applied to the input of the EMC analyzer. The source frequency
is adjusted until the two signals appear at the same frequency. The frequency
setting of the source is then equal to the comb generator frequency and this
frequency is compared to the specication.
The related adjustment procedure for this performance verication test is the
\Comb Generator, HP 8593EM and HP 8596EM" adjustment.

Equipment Required
  Synthesized sweeper
  Power splitter
  Cable, APC 3.5 mm (m) 91 cm (36 in)
  Cable, SMA 61 cm (18 in) (m) to (m)
  Adapter, Type N (m) to APC 3.5 (m)
  Adapter, 3.5 mm (f) to 3.5 mm (f)




        Figure 2-3. Comb Generator Frequency Accuracy Test Setup


                                                  Performance Verication Tests 2-9
3. Comb Generator Frequency Accuracy, HP 8593EM and HP 8596EM
Procedure
 1. Connect the equipment as shown in Figure 2-3.
    Option 026 only: Omit the Type N to APC 3.5 mm adapter.
 2. Press instrument preset on the synthesized sweeper, then set the controls as
    follows:
       CW : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 100.025 MHz
       POWER LEVEL : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 0 dBm
       RF : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : OFF
 3. Press 4PRESET5 on the EMC analyzer, then wait for preset routine to nish.
    Set the EMC analyzer by pressing the following keys:
       4FREQUENCY5 100 4MHz5
       4AUX/USER5 COMB GEN ON OFF (ON)
                                    NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




       4SPAN5 10 4MHz5
       4AMPLITUDE5 REF LVL 117 4dBV5
                                       NNNNNNNNNNNNNNNNNNNNNNN




       4BW5 IF BW AUTO MAN (MAN) 10 4kHz5
                     NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




       4MKR !5 MKR ! HIGH
                               NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




       4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
                        NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




       4SPAN5 100 4kHz5

 4. Press 4AMPLITUDE5 and adjust the reference-level setting until the signal peak
    is 10 dB below the reference level.
 5. Set the synthesized sweeper RF on. Adjust the synthesized sweeper power
    level until the two signals are the same amplitude.
 6. Set SCALE LOG LIN (LOG) to 2 dB on the EMC analyzer.
             NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




 7. If necessary, readjust the synthesized sweeper power level until the two
    signals are the same amplitude.
 8. Set the synthesized sweeper CW to 100 MHz. A very unstable signal will
    probably appear. The peak amplitude should be at least 3 dB greater in
    amplitude than either of the individual signals.
 9. Adjust the synthesized sweeper CW setting until a single signal appears to
    rise and fall in amplitude at the slowest rate (1 Hz frequency resolution
    will be necessary). The signal peak should be displayed approximately 6 dB
    above the amplitude of the individual signals.

2-10 Performance Verication Tests
      3. Comb Generator Frequency Accuracy, HP 8593EM and HP 8596EM
10. Record the synthesized sweeper CW frequency setting as TR Entry 1 of the
    performance verication test record. The frequency should be between
    99.993 MHz and 100.007 MHz.




                                                Performance Verication Tests 2-11
4. Frequency Readout and Marker Count Accuracy,
HP 8591EM and HP 8594EM
This performance verication test must be performed with the EMC analyzer set
in the spectrum analyzer mode.
The frequency readout accuracy of the EMC analyzer is tested with an input
signal of known frequency. By using the same frequency standard for the
EMC analyzer and the synthesized sweeper, the frequency reference error is
eliminated.

Equipment Required
  Synthesized sweeper
  Adapter, Type N (f) to APC 3.5 (m)
  Adapter, APC 3.5 (f) to APC 3.5 (f)
  Cable, BNC, 122 cm (48 in)
  Cable, APC 3.5 mm (m) 91 cm (36 in)




                             Figure 2-4.
   Frequency Readout Accuracy Test Setup, HP 8591EM and HP 8594EM
Procedure
This performance verication test consists of two parts:
  Part 1: Frequency Readout Accuracy
  Part 2: Marker Count Accuracy
Perform \Part 1: Frequency Readout Accuracy" before performing \Part 2:
Marker Count Accuracy."

2-12 Performance Verication Tests
4. Frequency Readout and Marker Count Accuracy, HP 8591EM and HP 8594EM
   Part 1: Frequency Readout Accuracy
   1. Connect the equipment as shown in Figure 2-4. Remember to connect the 10
      MHz REF OUT of the synthesized sweeper to the EXT REF IN of the EMC
      analyzer.
   2. Perform the following steps to set up the equipment:
        Press INSTRUMENT PRESET on the synthesized sweeper, then set the
        controls as follows:
           CW : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 1.5 GHz
           POWER LEVEL : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 010 dBm
        On the EMC analyzer, press 4PRESET5, wait for the preset routine to nish,
        then press 4FREQUENCY5 1.5 4GHz5.
   3. Set the EMC analyzer to measure the frequency readout accuracy by pressing
      the following keys:
         4SPAN5 20 4MHz5
         4MKR !5 MARKER ! HIGH
                            NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




   4. Record the MKR frequency reading as TR Entry 1 in the performance
      verication test record. The reading should be within the limits shown in
      Table 2-1.
   5. Change to the next EMC analyzer span setting listed in Table 2-1.
   6. Repeat steps 3 through 5 for each EMC analyzer span setting/TR entry listed
      in Table 2-1.
                                         Table 2-1. Frequency Readout Accuracy
                                              EMC Analyzer                          MKR Reading
                                                        Span                Min. TR Entry          Max.
                                                       (MHz)               (MHz)                  (MHz)
                                                           20             1.49918        1    1.50082
                                                           10             1.49958        2    1.500