Migrating Balanced Measurements from the HP 8903B to the U8903A Audio Analyzer - Application Note 5991-0435EN c20140819 [5]


>> Download Migrating Balanced Measurements from the HP 8903B to the U8903A Audio Analyzer - Application Note 5991-0435EN c20140819 [5] documenatation <<

Text preview - extract from the document
Keysight Technologies
Migrating Balanced Measurements from the
HP 8903B to the Keysight U8903A Audio Analyzer




                       Application Note
02 | Keysight | Migrating Balanced Measurements from the HP 8903B to the U8903A Audio Analyzer - Application Note



Abstract                                                 Hewlett-Packard introduced the HP 8903B audio analyzer in 1980. It quickly became the
                                                         industry standard because of its unprecedented performance and features. The U8903A
                                                         audio analyzer is the next-generation audio analyzer from Keysight Technologies, Inc.
                                                         (refer to Figure 1).

                                                         This application note explains the HP 8903B and U8903A audio analyzers balanced
                                                         connection architectures, and the accessories for migrating connections from the
                                                         HP 8903B to U8903A




                                                         Figure 1. The HP 8903B (left) and Keysight U8903A (right) audio analyzers



HP 8903B Output and Input                                Older generation instruments like the HP 8903B provided BNC connector pairs for both
                                                         the generator output and the analyzer input. Figure 2 shows a simplified block diagram
                                                         of the analyzer input.

                                                         The `HIGH' BNC connector is used in isolation when an unbalanced measurement is
                                                         made. When a balanced measurement is made both the `HIGH' and `LOW' BNC connectors
                                                         are used. The center conductor of the `HIGH' connector carries the signal and the `LOW'
                                                         conductor carries the inverted signal. The outer conductors of both connectors are con-
                                                         nected to ground. The generator output connections function in a similar way.


                                                                                                      BNC
                                                                                                                    `Float' = Balanced/Unbalanced
                                                                                             HIGH

                                                                                                 GND
                                                                                                                          `Float'      100 k
                                                                                             LOW

                                                                                                 GND
                                                                                                                                     GND

                                                         Figure 2. Analyzer input connections and block diagram for the HP 8903B



HP 8903B Float Switch                                    A common misunderstanding is that the `FLOAT' switch on the HP 8903B floats the
                                                         ground reference of the generator or analyzer. In actuality the `FLOAT' switch shorts the
                                                         `LOW' input line to ground when the unbalanced configuration is required (see Figure 2).
                                                         When the switch is in the `FLOAT' position the `LOW' input floats to the level of signal
                                                         that is applied to it from the source, thereby enabling a balanced measurement.

                                                         In the era of the HP 8903B, engineers referred to this as "floating the input", hence the
                                                         name float switch.
03 | Keysight | Migrating Balanced Measurements from the HP 8903B to the U8903A Audio Analyzer - Application Note



Keysight U8903A Output                                   The U8903A audio analyzer also uses BNC connectors for unbalanced measurements.
                                                         For balanced measurements the instrument uses the industry-standard XLR connectors.
and Input
                                                         EIA Standard RS-297-A describes the use of the XLR connector in balanced audio signal
                                                         applications.

                                                         Figure 3 shows a simplified block diagram of the U8903A input configuration. The same
                                                         `FLOAT' switch is used to select balanced/unbalanced configurations. The difference
                                                         between the HP 8903B and the Keysight U8903A is that the U8903A uses an internal
                                                         relay which is controlled via the user interface.


                                                                                                                    BNC


                                                                                               XLR
                                                                                                              GND




                                                                                                                                     100 k

                                                                                                                          `Float'

                                                                                                                                    GND

                                                         Figure 3. Analyzer input connections and block diagram for the U8903A
04 | Keysight | Migrating Balanced Measurements from the HP 8903B to the U8903A Audio Analyzer - Application Note



Migrating Balanced                                        While Keysight recommends users upgrade to the industry-standard XLR connec-
                                                          tors, there are situations, such as in military applications or standards testing, where
Connections from the HP 8903B
                                                          this is neither desirable nor possible. As an alternative to re-cabling from the U8903A
to Keysight U8903A                                        to your system, Keysight has made available two accessory cables: U8903A-107 and
                                                          U8903A-108 (see Figure 4). The schematic of the analyzer accessory cable alongside the
                                                          input block diagram of the U8903A is shown in Figure 5. U8903A-107 is available for the
                                                          analyzer input and U8903A-108 is available for the generator output.

                                                          Note that the BNC connectors are mounted on insulating spacers. This performs two
                                                          functions. First, it sets the connector spacing to match the HP 8903B for compatibility
                                                          with test jigs developed for this instrument. Second, the spacers stop the BNCs' ground
                                                          connections from touching each other in a random fashion. This eliminates local ground
                                                          loops within the cable assembly and ensures the highest measurement integrity. The
                                                          plastic spacers can be easily removed for applications where they are not required.




                                                          Figure 4. Accessory cables simplify the conversion of balanced measurements from HP 8903B
                                                          to U8903A


                                                                                                                    BNC

                                                    BNC
                                                                             XLR                XLR
                                                                                                            GND
                                                                         2         1        1         2
                                              GND                             3                 3

                                                                                                                                     100 k

                                                                                                                          `Float'
                                              GND                                 GND     GND
                                                       Accessory Cable                                                              GND


                                                          Figure 5. A U8903A-107 schematic alongside the input block diagram of the U8903A




Summary                                                   HP 8903B's outputs/inputs employ two BNC connectors (`HIGH' and `LOW') to enable
                                                          balanced measurements. There is a manual `FLOAT' switch which shorts the `LOW'
                                                          connection to ground where unbalanced measurements are required. The Keysight
                                                          U8903A outputs/inputs use BNC (unbalanced) and XLR (balanced) connectors. The
                                                          function of the manual `FLOAT' switch on the HP 8903B audio analyzer is replaced by
                                                          an internal relay in the Keysight U8903A. U8903A-107 and U8903A-108 accessory
                                                          cables allow the migration of balanced measurements from the HP 8903A to the
                                                          U8903A without the need to re-cable the whole measurement system.
05 | Keysight | Migrating Balanced Measurements from the HP 8903B to the U8903A Audio Analyzer - Application Note




                        myKeysight                                                                                  For more information on Keysight
                                                                                                                    Technologies' products, applications or
                        www.keysight.com/find/mykeysight
                                                                                                                    services, please contact your local Keysight
                        A personalized view into the information most relevant to you.
                                                                                                                    office. The complete list is available at:
                        www.axiestandard.org                                                                        www.keysight.com/find/contactus
                        AdvancedTCA



◦ Jabse Service Manual Search 2026 ◦ Jabse PravopisonTap.bg ◦ Other service manual resources online : FixyaeServiceinfo