Service Manuals, User Guides, Schematic Diagrams or docs for : Agilent Power Supply Connections for Your CET Signal Conditioner Card Application Note 5990-6549EN c20140929 [9]

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
Power Supply Connections for Your CET Signal Conditioner Card Application Note 5990-6549EN c20140929 [9]


>> Download Power Supply Connections for Your CET Signal Conditioner Card Application Note 5990-6549EN c20140929 [9] documenatation <<

Text preview - extract from the document
Keysight Technologies
Power Supply Connections for Your
CET Signal Conditioner Card


                                Application Note
Power Requirement for CET Card

The CET signal conditioner card                Power is connected via a wire wrap of AWG28 wires to pins 11 and 12 of J6 of
(referenced in this application note           the CET card. There is another pair of power connections on J6 at pins 3 and 4.
as CET card) can be powered from               These are the power connections for the USB and do not provide power to the
various sources. The recommended               CET card itself.
sources are discussed in the follow-
ing sections.

The input supply voltage to the CET
card is +12 V @ 0.3 A. The selected
source to provide the power needs
to be controllable through the test
plan execution. Power to the CET
card must be uninterrupted for the                                                                Connector J6
duration of powered testing. If power
to the CET card is disconnected, an
                                                                                            19                          1
"unpowered" command must be
issued before turning the power
back on.
                                                                                            20                          2

                                               Figure 1. Connections on CET card (J6 connector)




Table 1. Pin layout of J6 connector on CET card
 CET card (J6 connector)                 Wire color                      Wire type                Remarks
 Pin 1                                   White                                                    USB D