Tensile Testing of Fibers using Keysight T150 UTM Quasi-static Tensile Test 5991-2269EN c20141030 [4]


>> Download Tensile Testing of Fibers using Keysight T150 UTM Quasi-static Tensile Test 5991-2269EN c20141030 [4] documenatation <<

Text preview - extract from the document
Keysight Technologies
Tensile Testing of Fibers using Keysight T150 UTM
Quasi-static Tensile Test

Application Note

Introduction                                  the electrospinning setup to collect
                                              aligned nano/micro-ibers those can be
Fibers for advanced applications, such        easily transferred to templates for tensile
as bulletproof vests, biomedical scaf-        testing using the Keysight Technologies,
folds, microelectronics, etc. require a       Inc. T150 UTM.
good combination of strength, stiffness
and toughness. However, characterizing        Tensile Test of Thin
these properties for micro/nano-ibers is
challenging because of their low stiff-       Fibers: Challenges and
ness. Moreover, because of the iber           The T150 Advantage
drawing process, most of the chemical
bonds in a iber are oriented along the        The most effective method for mechani-
iber axis. This gives rise to a big differ-   cal characterization of ibers is uniaxial
ence in mechanical properties along the       tensile test. During a tensile test, a iber
iber axis compared to the perpendicular       specimen is pulled at a speciied rate and
direction and, hence, it is not reasonable    the resulting reaction force is measured.
to interpret iber properties from their       As we know the initial sample dimensions
bulk counterparts. Such differences are       (iber length and iber diameter), sev-
especially evident in ibers with smaller      eral key engineering parameters can be
diameters (few hundred nanometers to a        calculated using the measured force and
few 10's of microns).                         displacement. Some of these parameters
                                              are: ultimate tensile strength, Young's
For example, one of the popular methods       modulus, yield stress, yield strain and
to fabricate ultra-thin                       failure strain. As more and more ibers,
nano/micro-ibers is electrospinning. De-      especially micro and nano-ibers are be-
spite the importance, most of the studies     ing proposed for structural applications,
on mechanical properties have been            the tensile parameters are important for
limited to characterizing the ibermats        better simulation and design of struc-
instead of individual ibers. However, in      tures using these ibers.
a recent study, Chen et al. clearly shows
that the mechanical behavior of colla-        As the iber diameter decreases, the
gen-chitosan ibermats is signiicantly         absolute stiffness of a iber drops sig-
different from that of the correspond-        niicantly, and it becomes challenging
ing individual ibers 1. In recent years,      to measure the tensile parameters using
multiple studies, including the above         conventional test equipment. The mass
reference, have shown ways to design          and inertia of the grips and the driveshaft
02 | Keysight | Tensile Testing of Fibers using Keysight T150 UTM, Quasi-static Tensile Test - Application Note


for a conventional tensile test system are                                                                                          for thin polymeric ibers, which requires
                                                                                         P
much greater than the small-diameter                                                                                                small forces for deformation but can ex-
                                                                      Grip
iber samples. Hence, although one may                                                                Spring                         tend to large strains. A stress-strain curve
accommodate a iber sample in a conven-                          Capacitive                                                          for a polymer iber, such as PET (Figure 2),
tional tensile testing system, the accu-                        Sensor                                                              is extremely important for understanding
racy of the results is doubtful. One way                                                                                            its mechanical deformation 2-4. A polymer
researchers try to avoid the challenges is                                                                                          iber consists of many polymer chains,
by interpreting the properties of thin ibers                                                                                        and their physical properties govern the
from a larger sample of same material, or                                                                                           properties of the iber. During a tensile
a yarn (bundle of multiple ibers) or woven/                                                                                         test, extension in a polymer iber is a result
nonwoven ibermat. However, this is not                                                                                              of elastic extension of the polymer chains,
a convincing characterization method                                                                                                and elastic and non-elastic rotation of the
because of the differences in molecular                                                                                             chain axis towards the iber axis. Hence,
alignment, and various elastic and non-                                                                                             a polymer iber with small orientation
elastic interactions between ibers. The                                                           Coil (Magnet)                     distribution of the polymer chains exhibits
force range and the gripping arrangement                                                                                            high modulus. However, the strength of a
of the Keysight T150 UTM are more suit-                      Figure 1. Schematic of nanomechanical actuating                        polymer iber also depends on the nature
able for measuring these small-diameter                      transducer (NMAT).                                                     of overlap between the chains, since the
iber samples. There are several critical                                                                                            strength of the physical hydrogen or van
components behind the accurate mea-                          nano/micro-iber materials. The technical                               der Waals bonds between the chains are
surements using a nanomechanical tensile                     speciications for the T150 UTM can be                                  much weaker compared to the covalent
tester like the T150 UTM. These compo-                       found in Table 1.                                                      bonds within the chains.
nents include a patented nano-mechanical
actuating transducer (NMAT) -- shown
                                                             Applications                                                           Spider Silk
in Figure 1 -- that can measure small
changes in force to provide a uniquely                                                                                              Spider silk is known to be a wonderful ma-
large dynamic force range (0.0005 to                         Polymer Fibers                                                         terial for many structural and biomedical
500mN) for tensile measurements. A pre-                                                                                             applications. Different protein glands in a
cise extension axis (35nm resolution) also                   Figure 2 shows an engineering stress vs.                               spider produce a variety of silk with diver-
enables characterization of small diameter                   engineering strain curve for a PET iber,                               siied arrangement of protein molecules.
ibers at a wide range of strain rates. In                    obtained by extending the iber at a speci-                             Hence, it provides researchers a complete
addition, the Keysight NanoSuite control                     ied strain rate. In this case, the large ex-                           toolkit to study the protein molecules
software for the T150 not only facilitates                   tension range (up to 200mm) of the T150                                and their arrangements that gives rise to
precise test control and analysis of data,                   UTM makes it possible to characterize the                              the unique combination of high strength
but also adds the capability to add new                      ibers over the complete range of strain,                               and high ductility 5. However, mechani-
test protocols for novel characterization of                 through breaking. This is especially useful                            cal testing of the individual silk strands


 Maximum load                                     500 mN (50.8 gm)
 Load resolution                                  50 nN (5.1 