U1451-90001 U1452A U1452AT U1451A Insulation Tester Quick Start Guide c20131121 [18]


>> Download U1451-90001 U1452A U1452AT U1451A Insulation Tester Quick Start Guide c20131121 [18] documenatation <<

Text preview - extract from the document
                          Agilent
                          U1452A/U1452AT/U1451A
                          Insulation Tester
                          Quick Start Guide

        Standard accessories included in your purchase
        The following accessories are shipped standard with the
        U1452A, U1452AT, and U1451A testers:
            Hard carrying case
            Alligator clips (red and black)
            Test leads (red and black)
            19 mm probes (red and black)
            4 mm probes (red and black)
            IR to USB cable
            Four 1.5 V AA alkaline batteries
            Printed copy of the U1452A/U1452AT/U1451A
             Quick Start Guide (this document)
        Model U1452A also includes the following accessories:
         Remote switch probe and adapter
         U1117A IR to Bluetooth adapter
         Printed copy of the U1117A Operating Instructions
        If any item is missing or damaged, keep the shipping materials
        and contact the nearest Agilent Sales Office.


NO TE
        The descriptions and instructions in this guide apply to the
        U1452A, U1452AT, and U1451A Insulation Tester.
        Model U1452A appears in all illustrations. The word tester is
        used to represent all models.
        All related documents and software are available for download
        at www.agilent.com/find/hhTechLib.




                                     Agilent Technologies
U1452A/U1452AT/U1451A Insulation Tester
Install or Change the Batteries


Install or Change the Batteries
         The tester is powered by four 1.5 V AA alkaline batteries
         (included in the shipment).
         Before installing or changing the batteries, pull from a corner
         and stretch the orange rubber holster to remove it.



           1                              2




           3                              4




Turn On the Tester
                                                  Turn the rotary switch
                                                  from the      position
                                                  to any other position
                                                  to begin making
                                                  measurements.




2                        U1452A/U1452AT/U1451A Quick Start Guide
                       U1452A/U1452AT/U1451A Insulation Tester
                                        The Tester at a Glance


The Tester at a Glance

                                 Red LED indicator



                                 Display screen



                                 Keypad



                                 Rotary switch



                                 Input terminals




                                 IR communication port
                                 Test lead/probe holders


                                 Battery access
                                 (Under the orange rubber holster)

                                 Tilt stand


                                 Fuse access
                                 (Under the orange rubber holster)




U1452A/U1452AT/U1451A Quick Start Guide                          3
U1452A/U1452AT/U1451A Insulation Tester
Using the Rotary Switch


Using the Rotary Switch

    NO TE
             Press [T DAR PI] to select the alternate measurement func-
             tion(s) or test methods for insulation resistance tests.



                                                           Model
Legend      Measurement function
                                                 U1452A U1452AT U1451A
            50 V insulation resistance test,
            T, DAR, PI[1]
                                                                     -

            100 V insulation resistance test,
            T, DAR, PI[1]
                                                                     -

            250 V insulation resistance test,
            T, DAR, PI[1]
                                                             -      
            500 V insulation resistance test,
            T, DAR, PI[1]
                                                             -      
            1000 V insulation resistance test,
            T, DAR, PI[1]
                                                             -      
            Earth-bond resistance, T                                
            Resistance, Continuity                                  
            Capacitance                                             
            Auto (V), DC V, AC V                                    
[1] DAR and PI tests for models U1452A and U1452AT only.




WA RN ING
             Remove the test leads from the measuring source or target
             before changing the rotary switch position.
             Refer to the U1452A/U1452AT/U1451A User's Guide for a
             complete list and description of all rotary switch labels.




4                              U1452A/U1452AT/U1451A Quick Start Guide
                          U1452A/U1452AT/U1451A Insulation Tester
                                                Using the Keypad


Using the Keypad

                           Key response when pressed for:
   Legend
                   Less than 1 second              More than 1 second
              IR Test: Initiates an insulation test[1] as long as [Test] is
              held (       is shown) -- the tester sources (outputs) a high
              voltage and measures insulation resistance
              EBR Test: Initiates an earth-bond resistance test[2] as long
              as [Test] is held (     is shown)
                                                 Auto Hold: Automatically
              Trig Hold: Freezes the present     freezes the present read-
              reading in the display[3]          ing once the reading is
                                                 stable[3]
              Lock: Locks the insulation test
              or earth-bond resistance test[4]
              Hz: Displays the frequency         View (U1452A and
              (when the rotary switch is in      U1452AT): Enters the Log
              the V position)                    review menu
              Esc: Discards the changes
              made in the Setup menu
              Selects the alternate measure-
              ment function(s)
              T/Time: Configures the tester
              for a timed test[4][5]
              DAR (U1452A and U1452AT):          Log (U1452A and
              Configures the tester for a        U1452AT): Starts the data
              dielectric absorption ratio        Log
              test[1][5]
              PI (U1452A and U1452AT):
              Configures the tester for a
              polarization index test[1][5]
              Limit: Enables Limit compari-
                                                 Smooth: Enables Smooth
              son
              Range: Sets a manual range
                                                 Auto: Enables autorang-
              Leak (U1452A and U1452AT):         ing
              Displays the leakage current[4]




U1452A/U1452AT/U1451A Quick Start Guide                                       5
U1452A/U1452AT/U1451A Insulation Tester
Using the Input Terminals


                                      Key response when pressed for:
    Legend
                          Less than 1 second                      More than 1 second

                    Null: Enables Null                       -

                      : Enable or disable the LCD            Setup: Enters the Setup
                    backlight                                menu

[1] When the rotary switch is in one of the 
                                                   Mega position.
[2] When the rotary switch is in the 
                                            EB position.
[3] When the rotary switch is NOT in one of the 
                                                           Mega or the  EB position.
[4] When the rotary switch is in one of the 
                                                   Mega or the  EB position.
[5] Press [Test] to start the test.




Using the Input Terminals
WA RN ING
             To avoid damaging this device, do not exceed the input limit.




     Rotary position                  Input terminals            Overload protection

                                                                 1000 Vrms

                                                                 1000 Vrms for
                                                                 short circuit <0.3 A


                                                                 440 mA/1000 V, 30 kA
                                                                 fast-acting fuse




6                                U1452A/U1452AT/U1451A Quick Start Guide
                                  U1452A/U1452AT/U1451A Insulation Tester
                                             Insulation Resistance (IR) Test


Insulation Resistance (IR) Test
                 Ensure that the device-under-test (DUT) is de-energized before
                 performing any resistance measurement.
                 The tester automatically detects if the circuit is energized. If
                 the external voltage is detected to be greater than 30 V, the
                 test is inhibited and the voltage hazard symbol ( ) is shown on
                 the display.


CAU TI O N
                 DO NOT perform insulation resistance test in distribution sys-
                 tems with voltages higher than 600 V.




                   Select a
                 test voltage                                   3


                  2




             1                                                  Remote
                                                                Switch
                                                                Probe




U1452A/U1452AT/U1451A Quick Start Guide                                         7
U1452A/U1452AT/U1451A Insulation Tester
Insulation Resistance (IR) Test


                          4      Hold




                                                      Test will continue until
                                                        [Test] is released.


    NO TE
            When an insulation test is in progress, the red LED indicator at
            the top of the tester will blink every 2 seconds (if the Limit
            function is not enabled).



            Using the Remote Switch Probe



                                    =




            Locking the Test for IR/EBR Tests

                          1




                          2   Press once



                                                    Press [Test] or [Lock]
                                                    again to unlock and
                                                    stop the test.


8                             U1452A/U1452AT/U1451A Quick Start Guide
                             U1452A/U1452AT/U1451A Insulation Tester
                                        Insulation Resistance (IR) Test


          Performing PI/DAR/T Tests

                         1




                         2




                         3
                                      OR




          Polarization Index (PI) = IR 10 minutes/IR 1 minute
          Dielectric Absorption Ratio (DAR) = IR 60 seconds/IR 30 seconds
          (default; can be changed to IR 60 seconds/IR 15 seconds in Setup)
          Timed (T) = IR 1 minute (default; can be changed in Setup)


  NO TE
          