U1461A Insulation Multimeter U1453A Insulation Tester Quick Start Guide U1461-90001 c20140527 [25]


>> Download U1461A Insulation Multimeter U1453A Insulation Tester Quick Start Guide U1461-90001 c20140527 [25] documenatation <<

Text preview - extract from the document
                          Agilent U1461A Insulation
                          Multimeter/U1453A
                          Insulation Tester
                          Quick Start Guide

        Standard accessories included in your purchase
        The following accessories are shipped standard with the
        U1461A and U1453A testers:
            Hard carrying case
            Alligator clips (red and black)
            Test leads (red and black)
            19 mm probes (red and black)
            4 mm probes (red and black)
            Remote switch probe and adapter
            Four 1.5 V AA lithium batteries
            IR to USB cable
            U1117A IR to Bluetooth adapter
            Printed copy of the U1117A Operating Instructions
            Printed copy of the U1461A/U1453A Quick Start Guide
             (this document)
        Model U1461A also includes the following accessories:
         Thermocouple adapter (J/K-Type)
         Thermocouple bead (J-Type)
         Thermocouple bead (K-Type)
        If any item is missing or damaged, keep the shipping materials
        and contact the nearest Agilent Sales Office.


NO TE
        The descriptions and instructions in this guide apply to the
        U1461A Insulation Multimeter and U1453A Insulation Tester.
        Model U1461A appears in all illustrations. The word tester is
        used to represent both models.
        All related documents and software are available for download
        at www.agilent.com/find/hhTechLib.


                                     Agilent Technologies
U1461A Insulation Multimeter/U1453A Insulation Tester
Install or Change the Batteries


Install or Change the Batteries
          The tester is powered by four 1.5 V AA lithium batteries
          (included in the shipment).
          Before installing or changing the batteries, pull from a corner
          and stretch the orange rubber holster to remove it.



           1                               2




           3                               4




Turn On the Tester
                                                   Turn the rotary switch
                                                   from the      position
                                                   to any other position
                                                   to begin making
                                                   measurements.




2                                    U1461A/U1453A Quick Start Guide
             U1461A Insulation Multimeter/U1453A Insulation Tester
                                            The Tester at a Glance


The Tester at a Glance

                                   Vsense/Red LED indicator
                                   (Vsense for model U1461A)


                                   Display screen



                                   Keypad



                                   Rotary switch



                                   Input terminals




                                   IR communication port
                                   Test lead/probe holders


                                   Battery access
                                   (Under the orange rubber holster)

                                   Tilt stand


                                   Fuse access
                                   (Under the orange rubber holster)




U1461A/U1453A Quick Start Guide                                    3
U1461A Insulation Multimeter/U1453A Insulation Tester
Using the Rotary Switch


Using the Rotary Switch

    NO TE
             Press [T DAR PI] to select the alternate measurement func-
             tion(s) or test methods for insulation resistance tests.



Legend      Measurement function                         U1461A   U1453A

            50 V insulation resistance, T, DAR, PI                  

            100 V insulation resistance, T, DAR, PI                 

            250 V insulation resistance, T, DAR, PI                 

            500 V insulation resistance, T, DAR, PI                 

            1000 V insulation resistance, T, DAR, PI                
            Earth-bond resistance, T                                
            Resistance, Continuity, Diode, Capacitance              

            AC V, AC V with LPF                                      -

            AC V                                            -       
            Auto (V), DC V, AC V                                    

            Auto (mV), DC mV, AC mV, Temperature                     -

            Auto (A mA), DC A mA, AC A mA                            -



WA RN ING
             Remove the test leads from the measuring source or target
             before changing the rotary switch position.
             Refer to the U1461A/U1453A User's Guide for a complete list
             and description of all rotary switch labels.



4                                         U1461A/U1453A Quick Start Guide
             U1461A Insulation Multimeter/U1453A Insulation Tester
                                                 Using the Keypad


Using the Keypad

                            Key response when pressed for:
   Legend
                   Less than 1 second                More than 1 second
              IR Test: Initiates an insulation test[1] as long as [Test] is
              held (         is shown) -- the tester sources (outputs) a
              high voltage and measures insulation resistance
              EBR Test: Initiates an earth-bond resistance test[2] as long
              as [Test] is held (      is shown)
                                                  Auto Hold: Automatically
              Trig Hold: Freezes the present      freezes the present read-
              reading in the display[3]           ing once the reading is
                                                  stable[3]
              Lock: Locks the insulation test
              or earth-bond resistance test[4]
              Hz: Displays the frequency
              (when the rotary switch is in       View: Enters the Log
              the V, mV (U1461A), or              review menu
              A mA (U1461A) position)
              Esc: Discards the changes
              made in the Setup menu
              Selects the alternate measure-
              ment function(s)
              T: Configures the tester for a
              timed test[4][5]
              DAR: Configures the tester for      Log: Starts the data Log
              a dielectric absorption ratio
              test[1][5]
              PI: Configures the tester for a
              polarization index test[1][5]
              Limit: Enables Limit compari-
              son
                                                  Max Min: Enables Max
              Trip: Configures the tester for     Min recording[6]
              Leakage Current, Scan, or
              Ramp Trip tests[4][5]
              Range: Sets a manual range
                                              Auto: Enables autorang-
              Leak: Displays the leakage cur- ing
              rent[4]


U1461A/U1453A Quick Start Guide                                               5
U1461A Insulation Multimeter/U1453A Insulation Tester
Using the Input Terminals


                                      Key response when pressed for:
    Legend
                          Less than 1 second                   More than 1 second
                                                             Vsense (U1461A):
                    Null: Enables Null                       Enables the non-contact
                                                             voltage detector
                       : Increases or decreases the
                    OLED brightness -- this option Setup: Enters the Setup
                    must first be enabled in the    menu
                    Setup
[1] When the rotary switch is in one of the 
                                                   Mega position.
[2] When the rotary switch is in the 
                                            EB position.
[3] When the rotary switch is NOT in one of the 
                                                           Mega or the  EB position.
[4] When the rotary switch is in one of the 
                                                   Mega or the  EB position.
[5] Press [Test] to start the test.

[6] Max Min is disabled when Trip is enabled.




Using the Input Terminals
WA RN ING
             To avoid damaging this device, do not exceed the input limit.




     Rotary position                  Input terminals         Overload protection

                                                               1000 Vrms


                                                               1000 Vrms for
                                                               short circuit <0.3 A


                                                               440 mA/1000 V, 30 kA
                                                               fast-acting fuse




6                                              U1461A/U1453A Quick Start Guide
                      U1461A Insulation Multimeter/U1453A Insulation Tester
                                               Insulation Resistance (IR) Test


Insulation Resistance (IR) Test
                 Ensure that the device-under-test (DUT) is de-energized before
                 performing any resistance measurement.
                 The tester automatically detects if the circuit is energized. If
                 the external voltage is detected to be greater than 30 V, the
                 test is inhibited and the voltage hazard symbol ( ) is shown
                 on the display.


CAU TI O N
                 DO NOT perform insulation resistance test in distribution sys-
                 tems with voltages higher than 600 V.




                   Select a
                 test voltage                                     3


                  2




             1
                                                                  Remote
                                                                  Switch
                                                                  Probe




U1461A/U1453A Quick Start Guide                                                     7
U1461A Insulation Multimeter/U1453A Insulation Tester
Insulation Resistance (IR) Test


                         4      Hold




                                                     Test will continue until
                                                       [Test] is released.


    NO TE
            When an insulation test is in progress, the red LED indicator at
            the top of the tester will blink every 2 seconds (if the Limit
            function is not enabled).



            Using the Remote Switch Probe




                                    =




            Locking the Test for IR/EBR Tests

                          1




                          2   Press once



                                                   Press [Test] or [Lock]
                                                   again to unlock and
                                                   stop the test.


8                                       U1461A/U1453A Quick Start Guide
              U1461A Insulation Multimeter/U1453A Insulation Tester
                                       Insulation Resistance (IR) Test


          Performing PI/DAR/T for IR Tests
                        1




                        2




          Polarization Index (PI) = IR 10 minutes/IR 1 minute
          Dielectric Absorption Ratio (DAR) = IR 60 seconds/IR 30 seconds
          (default; can be changed to IR 60 seconds/IR 15 seconds in Setup)
          Timed (T) for IR and EBR = IR 1 minute (default; can be changed
          in Setup)


  NO TE
          Error is shown on the display if the IR is greater than the maxi-
          mum range or less than 0.001 M after t1/t15/t30; if the test
          is interrupted by the user; or if the tester's battery is low.



          Viewing Leakage Current




U1461A/U1453A Quick Start Guide                                           9
U1461A Insulation Multimeter/U1453A Insulation Tester
Insulation Resistance (IR) Test


             Performing Leakage Current/Scan/Ramp Trip Tests
             Use the Leakage Current Trip Test, Scan Trip Test, and Ramp
             Trip Test to test MOVs (Metal Oxide Varistors), gas discharge
             tubes, voltage arresters, or sparking gaps.
             The voltage source will be stopped when the current is greater
             than the trip current you set in the Setup.


                           1




                           2
                           1




                           3




                           4




     NO TE
             For more information on the Leakage Current Trip Test, Scan
             Trip Test, and Ramp Trip Test, refer to the respective sections in
             the U1461A/U1453A User's Guide.


10                                       U1461A/U1453A Quick Start Guide
                      U1461A Insulation Multimeter/U1453A Insulation Tester
                                            Earth-Bond Resistance (EBR) Test


Earth-Bond Resistance (EBR) Test
CAU TI O N
                 The tester automatically detects if the circuit is energized. If
                 the external voltage is detected to be greater than 2 V, the test
                 is inhibited and            is shown on the display. Discon-
                 nect the tester and remove power before proceeding.




                                                              3




                  2




             1                                                    Remote
                                                                  Switch
                                                                  Probe




                           4
                                 Hold
                                           OR          >




U1461A/U1453A Quick Start Guide                                                 11
U1461A Insulation Multimeter/U1453A Insulation Tester
Voltage Measurement


Voltage Measurement