Service Manuals, User Guides, Schematic Diagrams or docs for : Agilent Y1225-90001 Y1225A Rack Mount Kit - Installation Note c20140923 [6]

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
Y1225-90001 Y1225A Rack Mount Kit - Installation Note c20140923 [6]


>> Download Y1225-90001 Y1225A Rack Mount Kit - Installation Note c20140923 [6] documenatation <<

Text preview - extract from the document
Keysight Y1225A Rack Mount Kit
Installation
               This installation note contains procedures for installing the Y1225A Rack Mount
               Kit for the Keysight M9502A 2-slot AXIe Chassis in a standard EIA rack.

  Parts List
               The parts included with each rack mount adapter option and the support rail are
               shown in the following table:


                        Quantity   Description                                 Part Number
                        1          Rack Flange Kit                             N5650-00020
                        4          Rack Flange Screws M4x6 Flat Head           N/A
                        2          Support Rails                               M9018-00001
                        8          Channel Nuts (for support rail-to-rack)     0590-0804
                        4          Support Rail-to-Rack Screws - 10-32 x 0.5   2680-0278
                        4          Dress Adapter Screws for rack flanges       E9900-06001


  General Guidelines
               In rack mounting the chassis, follow these guidelines:
               1 Always begin installing the chassis and heavy instruments at the bottom of
                 the rack and work up. This maintains a lower center of gravity and reduces the
                 possibility of the rack tipping.
               2 Anti-tipping feet, if available with the rack, should always be extended.
               3 The heaviest instrument of chassis should always be mounted at the bottom
                 of the rack.
               4 For maximum cooling and optimum rack thermal efficiency, place the
                 instrument or chassis with the greatest power consumption towards the top of
                 the rack. This promotes efficient cooling since heat rises when placed nearer
                 to the top of the rack, high power instruments will not unnecessarily heat
                 other instruments. However, in doing this, do not violate the guideline that
                 the heaviest instruments should be placed at the bottom of the rack.
               5 If instruments or modules in the chassis are consuming maximum power, 1U
                 of rack space is required for ventilation below the chassis.
    Keysight Y1225A Rack Mount Kit Installation




       Procedure
                      1 The M9502A AXIe chassis has a single carry handle which must be removed
                        before removing the bumpers. Remove the two #2 Phillips screws securing
                        the handle.See Figure 1.




                                                   Figure 1 . Removing the Carry Handle


                      2 Remove the four #2 Phillips screws from each side of both bumpers. See
                        Figure 2.




                                                  Figure 2 . Removing the Chassis Bumpers


                      3 Using the rack mount flanges as templates, position the flanges on the rack's
                        vertical columns where the chassis is to be mounted. Align the adapter holes
                        over the "center" holes of the EIA rack units. Slide channel nuts over the rack
                        holes to be used by the rack mount flanges. See Figure 3.




2                                                                          Keysight Y1225A Rack Mount Installation Note
                                                                                 Keysight Y1225A Rack Mount Kit Installation




                  Rack Unit
                  Center Hole




                    Channel                                                                       One
                    Nuts                                                                          Rack
                                                                                                  Unit




                                 Figure 3 . Center Holes of EIA Rack, Attaching Channel Nuts
                         4 Attach the rack mount front handle flanges to the chassis. Refer to Figure 4.




Keysight Y1225A Rack Mount Installation Note                                                                            3
    Keysight Y1225A Rack Mount Kit Installation




                                                                                                Two Flathead
                                                                                                Screws




                             Figure 4 . Attaching Rack Mount Flanges to M9502A Chassis
                      5 Position the rails on the rack's vertical columns where the chassis is to be
                        mounted. Align the adapter holes over the "center" holes of the EIA rack units.


                                       INSTALL THE SUPPORT RAILS UPSIDE DOWN IN THE RACK! To
                                       provide adequate air flow to the ventilation holes on the sides of
                                       the chassis, you must install the rack slides upside down as
                                       shown in Figure 5.


                      6 Slide channel nuts over the rack holes to be used by the support rails.
                      7 Attach the support rails UPSIDE DOWN to the rack's inside vertical columns.
                        Use four channel nuts and the four 10-32 x 0.5 support rail-to-rack pan head
                        screws as shown in Figure 5.
                      8 Set the chassis onto the support rails.


                                       To prevent injury d uring rack mounting, the chassis should
                                       be empty and two people should lift the chassis into the
                                       rack.

                      9 Slide the M9502A chassis into the rack until the rack mount front handle
                        flanges are against the rack's vertical columns. Secure the chassis to the rack
                        using the adapter dress screws.



4                                                                        Keysight Y1225A Rack Mount Installation Note
                                                                                 Keysight Y1225A Rack Mount Kit Installation




                                          If the rack will be moved excessively or over long distances, you
                                          should use a strap to secure the chassis to the rack slides.




                                                                                            10-32 x 0.5 screws




                                      Figure 5 . Attaching the Support Rails to the Rack.




Keysight Y1225A Rack Mount Installation Note                                                                            5