Service Manuals, User Guides, Schematic Diagrams or docs for : Fluke 5700a testlog_5700a

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
Preview not available!
Sorry, no preview image available for this document.
Possible reasons are : this is not a PDF document, it is password protected or the file is partially corrupted.
We are working to fix the issue.
You can still download the full file from the link below and it will be viewable.



>> Download testlog_5700a documenatation <<

Text preview - extract from the document
Fluke 5700A - detected, S/N 5635315, Version: JA*

This is Series I unit
-i- Reading initial calibration data from 5700A
Unit last calibrated : 1308 days ago
Unit running time : 111438 hr
CAL CONST 6.5V reference voltage : +6.891366504559953E+00

CAL CONST 13V reference voltage : +1.379482200889601E+01

CAL CONST 10KOHM standard resistance : +1.000044087249608E+04

CAL CONST, Zero calibration temperature : +2.245000076293945E+01

CAL CONST, All calibration temp : +2.245000076293945E+01

Calibrator UNDER REMOTE CONTROL
Calibrator STABLE (settled within spec)
Calibrator is cooking report over SERIAL
User string PUD : #2055700A

Fatal errors history: "
 3/17/93  8:18:20  Fault   230: 5725 RESET (power-up or watchdog timer)
 3/30/93  7:07:49  Fault  4014: Sequencer Timed Out Waiting for 5725
10/05/94 18:53:17  Fault   218: 5725 +400V Supply Too Large
10/14/94 20:32:02  Fault   218: 5725 +400V Supply Too Large
11/07/94 12:27:47  Fault   218: 5725 +400V Supply Too Large
11/07/94 12:28:06  Fault   218: 5725 +400V Supply Too Large
 1/06/95  7:56:16  Fault   218: 5725 +400V Supply Too Large
 1/06/95  7:56:39  Fault   218: 5725 +400V Supply Too Large
 1/06/95 
12:34:42  Fault   218: 5725 +400V Supply Too Large
 1/25/95  7:59:37  Fault   230: 5725 RESET (power-up or watchdog timer)
 1/25/95  8:00:18  Fault   230: 5725 RESET (power-up or watchdog timer)
 2/03/95  7:02:33  Fault   230: 5725 RESET (power-up or watchdog timer)
 4/07/95 12:53:33  Fault   225: 5725 Fan Not Working
 4/07/95 12:53:34  Fault   225: 5725 Fan Not Working
 4/07/95 12:53:35  Fault   225: 5725 Fan Not Working
 4/07/95 12:53:36  Fault   225: 5725 Fan Not Working
"
HP 3458A detected
 TEMP? = 37.5 C
Fluke 5700A readback output = 10.000000000 V
10V test result: 1.000001492E+01 VDC [deviation 1.4920 ppm]
Program completed!



◦ Jabse Service Manual Search 2024 ◦ Jabse PravopisonTap.bg ◦ Other service manual resources online : FixyaeServiceinfo