                                           Agilent AN 346-2
                                           Balanced Circuit Measurement with an
                                           Impedance Analyzer/LCR Meter/Network
                                           Analyzer
                                           Application Note




Introduction                               Measuring a balanced circuit with an       Selecting a Balun
How a balanced circuit differs from an     unbalanced measurement instrument          There are several types and brands of
unbalanced circuit                         A balanced circuit cannot be directly      balun transformers. When selecting a
A balanced circuit has its electrical      measured with an unbalanced meas-          balun, ensure that frequency is com-
midpoint grounded. An unbalanced           urement instrument because of the          patible with your measurement
circuit, however, has one side ground-     difference in their configuration.         requirements. When you measure the
ed. A balanced circuit is typically                                                   impedance parameters of a balanced
used in communications equipment           When measuring balanced circuits,          circuit, you don't have to use the
because a balanced circuit has the         the unbalanced measuring instrument        balun which has the same impedance
advantage of better spurious noise         requires a balun (balanced to unbal-       with the circuit under test. However,
suppression.                               anced) transformer. A balun is a type      when you measure the transmission
                                           of impedance-matching RF trans-            or reflection of it, you have to use a
Figure 1 shows a balanced cable            former.                                    balun which has the same impedance
which is an example of a balanced                                                     with the circuit under test to keep
circuit. The voltages of the cable's two   Figure 3 shows the configuration for       impedance matching. Table 1 shows
conductors are at every point equal        measuring a balanced circuit with an       recommended balun transformers.
in amplitude and opposite in phase.        unbalanced instrument.
Figure 2 shows an unbalanced cable
which is an example of an unbal-           Note: In balanced cable measure-
anced circuit. Most measurement cir-       ments, residual current in the balun
cuits in Agilent Technologies' imped-      or the measuring instrument can
ance analyzers and LCR meters are          cause measurement errors. To reduce
unbalanced.                                the degree of error, perform
                                           open/short and load compensation at
                                           the measurement terminals of the
+V                                         balun.
     -V
                                                                           +V                      +V
                                                               High
                                                           V   terminal

                                                                             0                     -V
                                                               Low
                                                   A           terminal
Figure 1. Balanced cable                                                                                     Balanced cable


                                            Measurement Instrument

+V                                         Figure 3. Balanced cable measurement configuration




          0V

Figure 2. Unbalanced cable
  Measurement Configuration with a                          Short standard    PN 1250-0929                      Transmission measurement configuration
  Balun and Compensation                                    50  load standard PN 11593A                         with a network analyzer
  Impedance measurement configuration                                                                           Figure 7 shows transmission meas-
  with Agilent 4294A impedance analyzer                     3. Remove the connector and connect                 urement configuration (3) with a net-
  Figure 4 shows impedance measure-                         the DUT. Measure the DUT.                           work analyzer.
  ment configuration (1)/(2) with the
  4294A.                                                    Impedance measurement configuration                 To calibrate/compensate:
                                                            with the Agilent 4395A(#010)                        Short the terminals closest to the
  To calibrate/compensate for (1):                          Figure 6 shows impedance measure-                   DUT to the signal out and to the test
  1. Perform open, short, and load                          ment configuration (2) with the                     port, then perform response/thru
  compensation at the balanced termi-                       4395A.                                              calibration.
  nals of the 16314-60011. Use the
  furnished compensation standards of                       To calibrate/compensate:
  the 16314-60011.                                          Refer to "To calibrate/compensate for
                                                            (2)" of impedance measurement con-
  Standards                0  PN 04191-85300                figuration with 4294A impedance
                           0S PN 04191-85302                analyzer.
                           50  PN 04191-85301
                                                            Table 1. Recommended balun transformers
  To calibrate/compensate for (2):
  1. Assemble a female BNC connector                        Unb/Bal. (W)     Bandwidth                      Type No.              Suppliers
  as shown in Figure 5.
                                                            50:50            100 Hz to 10 MHz               16315-60011           Agilent Technologies
                                                            50:100           100 Hz to 10 MHz               16316A                Agilent Technologies
  2. Perform open, short, and load cali-
                                                            50:600           100 Hz to 3 MHz                16317A                Agilent Technologies
  bration at the BNC connector using the                    50:50            0.1