Text preview - extract from the document
                                             0$18$/ &+$1*(6                          -XO\ 
                                              Manual for Model Number       3324A
                                              Manual printed on             October 1990
                                              Manual Part Number            03324-90001

Make all ERRATA corrections.

Check the following table for your instrument serial prefix/serial number/EDC and make the listed
changes to your manual

New Item

Serial Prefix or                                 Manual
Serial Number                                    Changes

ERRATA

3009G00806                                       1
3009G02706                                       2




14 July, 1998                                                                           Page 1 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB
INDEX OF MANUAL CHANGES

MANUAL      FRAME
CHANGE
ERRATA
    1       W4,5
    2       MP30




__________________________________________________________________________________
07/14/98                                                                Page 2 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB
ERRATA

On Page 1-9, General Information (Specifications)

           Frequency Sweep
           .
           .
            Linear Sweep (settable for each interval)
            .
            .
              ADD: Minimum sweep rate: 0.2 Hz/sec.

On Page 2-9, Installation
                 step 6: change to read:

                    ...Refer to Fig. 2.7

On Page 2-10, Installation
                 step 8: change to read:

                    ...Refer to Fig. 2.7

On Page 4-1, Replaceable Parts, change to read:

                    Introduction
                     .
                     .
                    Repair philosophy .... However Appendix B ....


Performance Test

           Triangle Linearity

           Page 12-37 in the 03324-90011 Manual
           Page C-26 in the 03324-90001 Manual

           step 13:
           change to read:
           .
           .
           .
           y=a1x+ao


step 14:
           change to read:

           For each delay (x), subtract the calculated voltage (y') from

           the measured voltage (y) [y-y'= y]. Find the largest positive

           or negative voltage difference ( y max). Using the following

           Formular, computer the % linearity:

            % linearity = (|y max | / 4 Volt) x 100%


__________________________________________________________________________________
07/14/98                                                                Page 3 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

Performance Test

         Amplitude Accuracy

         Page 12-24 and 12-49, in the 03324-90011 Manual

         Page C-18 and C-39, in the 03324-90001 Manual

         step 32, or Test Record List

         change to read:

        Amplitude 1 V                                                                  New Limits!
        DC Oddset 1 mV

Test                       Freq                   Function                   Minimum                 Maximum
13                         99.9Hz                 Square                     0.958 V                 1.042 V
15                         99.9Hz                 Triangle                   0.953 V                 1.047 V
18                         99.9Hz                 + Ramp                     0.953 V                 1.047 V
19                         99.9Hz                 - Ramp                     0.953 V                 1.047 V
21                         1kHz                   Square                     0.958 V                 1.042 V
22                         2kHz                   Triangle                   0.953 V                 1.047 V
23                         500Hz                  + Ramp                     0.953 V                 1.047 V
24                         500Hz                  - Ramp                     0.953 V                 1.047 V
25                         100kHz                 Square                     0.880 V                 1.120 V
26                         10kHz                  Triangle                   0.918 V                 1.082 V
27                         10kHz                  + Ramp                     0.868 V                 1.132 V
28                         10kHz                  - Ramp                     0.868 V                 1.132 V


Specifications

         page 1-8 in the 03324-90001 manual

         DC Offset

         change to read:
         Accuracy:

         a)   DC only:

DC Offset Range                         Accuracy
1.5000 V to 5.0000 V                    +/- 50 mV
0.5000 V to 1.4999 V                    +/- 17 mV
0.0150 V to 0.4999 V                    +/- 5 mV
0.0500 V to 0.1499 V                    +/- 1.7 mv
0.0150 V to 0.0499 V                    +/- 0.52 mV
0.0050 V to 0.0149 V                    +/- 0.19 mV
0.0015 V to 0.0049 V                    +/- 0.07 mV
0.0005 V to 0.0014 V                    +/- 0.037 mV


The accuracy given to each range is the same for each value in this range.




__________________________________________________________________________________
07/14/98                                                                Page 4 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

b) DC + AC,  < 1 MHz:

         Sinewave + Squarewave

Amplitude Range                       Accuracy
1.0000 V to 10.0000 V                 +/- 60 mV
0.3000 V to 0.0000 V                  +/- 20 mV
0.1000 V to 0.2999 V                  +/- 6 mV
0.0300 V to 0.0999 V                  +/- 2 mV
0.0100 V to 0.0299 V                  +/- 0.6 mV
0.0030 V to 0.0099 V                  +/- 0.2 mV
0.0010 V to 0.0029 V                  +/- 0.06 mV

The accuracy given to each range is the same for each value in this range


Ramp + Triangle, upto 11 kHz

Amplitude Range                       Accurancy
1.0000 V to 10.0000 V                 +7- 120 mV
0.3000 V to 0.9999 V                  +/- 40 mV
0.1000 V to 0.2999 V                  +/- 12 mV
0.0300 V to 0.0999 V                  +/- 4 mV
0.0100 V to 0.0299 V                  +/- 1.2 mV
0.0030 V to 0.0099 V                  +/- 0.4 mV
0.0010 V to 0.0029 V                  +/- 0.12 mV

The accuracy given to each range is the same for each value in this range.


c) DC + AC,  < 1 MKz

         Sinewave up to 21 MHz

         Squarewave up to 11 MHz


Amplitude Range                       Accuracy
1.0000 V to 10.0000 V                 +/- 150 mV
0.3000 V to 0.9999 V                  +/- 48 mV
0.1000 V to 0.2999 V                  +/- 15 mV
0.0300 V to 0.0999 V                  +/- 4.8 mV
0.0100 V to 0.0299 V                  +/- 1.5 mV
0.0030 V to 0.0099 V                  +/- 0.49 mV
0.0010 V to 0.0029 V                  +/- 0.15 mV

The accurancy given to each range is the same for each value in this range.




__________________________________________________________________________________
07/14/98                                                                Page 5 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

Section: General Information

        Add: on page 1-3, after "Tools and Test Equipment:

        Recalibration, cleaning and lubrication

        The Performance Tests should be performed at 12-month intervals, if necessary using the
        adjustment procedure described in Chapter 3 in this manual should any of the
        performance tests fail.
        At the same intervals the instrument should be inspected mechanically and for cleanliness.
        Dust is best removed with a small vacuum cleaner or finebrush. Accumulation of dirt on the
        front panel or covers can be removed with a soft cloth if necessary slightly moistened with
        water.

        The instrument requires no lubrication.


on page 1-14, General Characteristics

        add: Calibration cycle: 12 months recommended

page 3-7, Adjustments

        Amplitude

        Negative setting

        change to read:

        step 9: ... Specification requirement is -10V +/- 0.2V.

        Positive setting

        change to read:

        step 2: ... Specification requirement is 10V +/- 0.2V.


page Contents-4, Table of Contents

        add after line "General Fault Finding":

         Access - Cover Removal ....... 5-27


page 5-27, General Fault Finding:

        change the heading "Access" to read:

        Access - Cover Removal

page 5-32, Service

         removal and refitting procedure

        add: Cover removal is described on page 5-27, Access - Cover Removal.




__________________________________________________________________________________
07/14/98                                                                Page 6 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

page 4-1, Replaceable Parts

        Introduction

        change to read:

        ... However Appendix B to this manual is reserved for an optional information package,
         CLIP p/n 03324-90031, which contains a full parts listing and schematics for those
         customers who require it.

In Table 4-2, Page 4-5, add the following lines after the linebegining "MP28":
         MP29              Screw M M3X0.58 mm                    0515-0897
                           Securing screw for board assemblies.




__________________________________________________________________________________
07/14/98                                                                Page 7 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

Page 1-14 add:

                         'HFODUDWLRQ RI &RQIRUPLW\

                                  (similar to ISO/IEC Guide 22)


        Manufacturer:    Hewlett-Packard GmbH
                         Boeblingen Instruments Division
                         Herrenberger Str. 130
                         D-71034 Boeblingen Germany

        We declare that the product


                          +3 $ )XQFWLRQ6ZHHS *HQHUDWRU


        conforms to the following standards:

        Safety: IEC 348 (1978)

        EMC:    EN 55011 (1991)/CISPR 11 Group1, Class B
                EN 50082-1 (1991)
                IEC 801-2 ESD: 4kV cd, 8kV ad
                IEC 801-3 Radiated Immunity: 3V/m
                IEC 801-4 Fast Transients: 0.5kV, 1kV
Supplementary Information:

        During the measurements against EN 55011, the I/O ports were terminated with their
        nominal impedance, the HP-IB connection was terminated with the cable HP 10833B.
        When the product is connected to other devices, the user must ensure that the connecting
        cables and the other devices are adequately shielded to prevent radiation.




        Boeblingen, 29th April 1993
        Robert Hofgaertner
        Quality Assurance Manager




__________________________________________________________________________________
07/14/98                                                                Page 8 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

Page C-17, Performance Tests

            change to read:
        .
        .

            Test              Frequenz        Function     Minimum         Maximum
            25                101kHz          Square       2.850 V         3.150 V
        .
        .

Page C-38 Test Record

        Amplitude: 3 V (p-p)
        change to read:

        .
        .                                    Min              Measured             Max
        Square, 101kHz                       2.850 V          -------              3.150 V
        .
        .


page 1-14 Operating Environment:
add errata:
Altitude up to 2000m
Installation Category II
Pollution Degree 2
Warning: To prevent electrical shock, disconnent the HP model 3324A from mains before cleaning. Use a dry cloth or
           one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.


========================================================================

                      ! CHANGE TO NEW SERIAL NUMBER FORMAT !

                        last serial number old format                    3009G02755
                        first serial number new format                   DE30202756

========================================================================




__________________________________________________________________________________
07/14/98                                                                Page 9 of 10
02'(/ $
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB

MANUAL CHANGE 1

On Page 4-6, change the Table of Replaceable Parts to read:

                                    W4                   03324-61601   CBL AY PWR SS
                                    W5                   03324-61602   CBL AY PWR UC


MANUAL CHANGE 2

On Page 4-4 Repl.Parts, Table 4-2 add:

                                     MP30                 7121-5585      Ce-Label




__________________________________________________________________________________
07/14/98                                                               Page 10 of 10