       Agilent 87511A/B S-PARAMETER TEST SET Operation and Service Manual

                                        Manual Change
                                                                                        Agilent Part No. N/A

                                                                                                  May 2007



Change 1
Change the part number for the Cables (page 2-1) to the following.


        Component                  Part Number for 87511A            Part Number for 87511B
     Interconnect Cable                  87511-61609                      87511-61609
   RF Connecting Cables                   8120-4387                        8120-4387




 Copyright 2007 Agilent Technologies
C
Operation and Service Manual
  Agilent 87511A/B S-PARAMETER TEST SET




                       SERIAL NUMBERS

      This manual applies directly to instruments with serial
       number prex 3026J (87511A) and 3042J (87511B).
 For additional important information about serial numbers, read
   \INSTRUMENT COVERED BY MANUAL" in Chapter 1.




                  Agilent Part No. 87511-90010
                  Printed in JAPAN   July 2001


                          Third Edition
Notice

The information contained in this document is subject to change without notice.
This document contains proprietary information which is protected by copyright. All rights
are reserved. No part of this document may be photocopied, reproduced, or translated to
another language without the prior written consent of the Agilent Technologies.

  Agilent Technologies Japan, Ltd.
  Component Test PGU-Kobe
  1-3-2, Murotani, Nishi-ku, Kobe-shi,
  Hyogo, 651-2241 Japan





 Copyright 1990, 2000, 2001 Agilent Technologies Japan, Ltd.
c
Manual Printing History

The manual printing date and part number indicate its current edition. The printing
date changes when a new edition is printed. (Minor corrections and updates which are
incorporated at reprint do not cause the date to change.) The manual part number changes
when extensive technical changes are incorporated.
  October 1990  :::::::::::::::::::::::::::::::::::: First Edition (part number: 87511-90000)
  February 2000  ::::::::::::::::::::::::::::::::: Second Edition (part number: 87511-90000)
  July 2001  :::::::::::::::::::::::::::::::::::::: Third Edition (part number: 87511-90010)




                                                                                           iii
Safety Summary

The following general safety precautions must be observed during all phases of operation,
service, and repair of this instrument. Failure to comply with these precautions or with
specic WARNINGS given elsewhere in this manual violates safety standards of design,
manufacture, and intended use of the instrument.
The Agilent Technologies assumes no liability for the customer's failure to comply with these
requirements.

DO NOT Operate In An Explosive Atmosphere

Do not operate the instrument in the presence of 
ammable gasses or fumes. Operation of any
electrical instrument in such an environment constitutes a safety hazard.
Keep Away From Live Circuits

Operating personnel must not remove the instrument covers. Component replacement and
internal adjustments must be made by qualied maintenance personnel. Do not replace
components with the power cable connected. Under certain conditions, dangerous voltages
may exist even with the power cable removed. To avoid injuries, always disconnect power and
discharge circuits before touching them.
DO NOT Service Or Adjust Alone

Do not attempt internal service or adjustment unless another person, capable of rendering
rst aid and resuscitation, is present.
DO NOT Substitute Parts Or Modify Instrument

Because of the danger of introducing additional hazards, do not substitute parts or
perform unauthorized modications to the instrument. Return the instrument to a Agilent
Technologies Sales and Service Oce for service and repair to ensure the safety features are
maintained.
Dangerous Procedure Warnings

Warnings, such as the example below, precede potentially dangerous procedures throughout
this manual. Instructions contained in the warnings must be followed.
Warning         Dangerous voltages, capable of causing death, are present in this instrument.
                Use extreme caution when handling, testing, and adjusting this instrument.




iv
Typeface Conventions

Bold                         Boldface type is used when a term is dened. For example: icons are
                             symbols.
Italics                      Italic type is used for emphasis and for titles of manuals and other
                             publications.
                             Italic type is also used for keyboard entries when a name or a variable
                             must be typed in place of the words in italics. For example: copy
                             lename means to type the word copy, to type a space, and then to
                             type the name of a le such as file1.
Computer                     Computer font is used for on-screen prompts and messages.
4HARDKEYS5                   Labeled keys on the instrument front panel are enclosed in 4 5.
                             Softkeys located to the right of the CRT are enclosed in .
NNNNNNNNNNNNNNNNNNNNNNNNNN                                                            NNNNN

SOFTKEYS




Certification

Agilent Technologies certies that this product met its published specications at the time
of shipment from the factory. Agilent Technologies further certies that its calibration
measurements are traceable to the United States National Institute for Standards and
Technology, to the extent allowed by the Institution's calibration facility, or to the calibration
facilities of other International Standards Organization members.


Warranty

This Agilent Technologies instrument product is warranted against defects in material and
workmanship for a period of one year from the date of shipment, except that in the case of
certain components listed in Chapter 1 of this manual, the warranty shall be for the specied
period. During the warranty period, Agilent Technologies will, at its option, either repair or
replace products which prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated
by Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and
Agilent Technologies shall pay shipping charges to return the product to Buyer. However,
Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent
Technologies from another country.
Agilent Technologies warrants that its software and rmware designated by Agilent
Technologies for use with an instrument will execute its programming instruction when
property installed on that instrument. Agilent Technologies does not warrant that the
operation of the instrument, or software, or rmware will be uninterrupted or error free.




                                                                                                     v
Limitation Of Warranty

The foregoing warranty shall not apply to defects resulting from improper or adequate
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modication or
misuse, operation outside of the environment specications for the product, or improper site
preparation or maintenance.
No other warranty is expressed or implied. Agilent Technologies specically disclaims the
implied warranties of merchantability and tness for a particular purpose.



Exclusive Remedies

The remedies provided herein are buyer's sole and exclusive remedies. Agilent Technologies
shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether
based on contract, tort, or any other legal theory.



Assistance

Product maintenance agreements and other customer assistance agreements are available for
Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales and Service Oce.
Address are provided at the back of this manual.




vi
Safety Symbols

General denitions of safety symbols used on equipment or in manuals.
                Instruction manual symbol: the product will be marked with this symbol
                when it is necessary for the user to refer to the instruction manual in order to
                protect against damage to the instrument.
                Indicates dangerous voltage (terminals fed from the interior by voltage
                exceeding 1000 volts must be so marked).
                Protective conductor terminal. For protection against electrical shock in case
                of a fault. Used with wiring terminals to indicate the terminal which must be
                connected to ground before operating equipment.
                Low-noise or noiseless, clean ground (earth) terminal. Used for a signal
                common, as well as providing protection against electrical shock in case of
                fault. A terminal marked with this symbol must be connected to ground
                in the manner described in the installation (Operation) manual, and before
                operating the equipment.
                Frame or chassis terminal. A connection to the frame (chassis) of the
                equipment which normally includes all exposed metal structures.
                Alternating current (power line).
                Direct current (power line).
                Alternating or direct current (power line).
                Warning denotes a hazard. It calls attention to a procedure, practice,
                condition or the like, which, if not correctly performed or adhered to, could
                result in injury or death to personnel.
                Caution sign denotes a hazard. It calls attention to a procedure, practice,
                condition or the like, which, if not correctly performed or adhered to, could
                result in damage to or destruction of part or all of the product.
                Note denotes important information. It calls attention to a procedure,
                practice, condition or the like, which is essential to highlight.




                                                                                              vii
                                                                                         1
General Information
INTRODUCTION

The purpose of this manual is to enable you to use your 87511A (including Option 001) or
87511B S-parameter test set eciently and condently. This manual contains both general
and specic information. When information applies to both the 87511A (the 50
 test set,
including Option 001) and the 87511B (the 75
 test set) the name 87511A/B is used. To use
the 87511A/B to perform a specic function (without reading the entire manual), follow the
directions in \USING THE 87511A/B".
This chapter
  (1) outlines how to begin using the 87511A/B as quickly as possible,
  (2) describes what instrument are covered by this manual,
  (3) describes the test set itself,
  (4) mentions operating precautions, safety considerations and warranty restrictions,
  (5) lists the additional equipment required, and
  (6) denes and lists the required instrument specications.


USING THE 87511A/B

The 87511A/B has been designed to operate specically with the 8751A network analyzer.
 To install the instrument, turn to Chapter 2.
 To oprate the 87511A/B, see the Chapter 3.
 To verify that the instrument meets its published specications, turn to Chapter 4.
 to order replaceable parts of the 87511A/B, to turn to Chapter 6.
 To troubleshoot the 87511A/B, turn to Chapter 7.




                                                                   General Information   1-1
INSTRUMENTS COVERED BY MANUAL

You will nd a two-part serial number on the rear panel of the test set. The rst four digits
and the letter are the serial number prex. The last ve digits are the sequential sux
which is unique to each test set. The contents of this manual apply directly to test sets with
the same serial number prex as the one(s) on the title page under the heading SERIAL
NUMBERS.
If the serial prex of your test set is not listed on the title page, your instrument diers from
those documented in this manual. The dierences are documented in the yellow manual
changes supplement supplied with the manual.
To keep this manual as current and accurate as possible, Agilent Technologies recommends
that you periodically request the latest manual changes supplement, as it may contain
replacement information as well as change information. The supplement for this manual is
keyed to the manual's print date and part number (on the title page) and is available on
request from Agilent Technologies.
This manual is also available on microche form (the part number appears on the title page).
With the manual (in 4 2 6 inch microlm transparency format) you will also receive the latest
manual changes supplement.


DESCRIPTION AND OPERATING CHARACTERISTICS OF THE
INSTRUMENT

The 87511A/B S-parameter test set has been designed to operate with the 8751A network
analyzer. The 87511A/B provides a convenient means of measuring the re
ection and
transmission coecients (scattering parameters) of one-port or two-port devices operating
within the frequency range of 100 kHz to 500 MHz.
Table 1-2 lists the operating specications and characteristics of the 87511A/B. The test set is
shown on the cover of this manual and shown in Figure 3-1.




1-2   General Information
OPTIONS


Option 001, 50   
 Type-N Test Port (87511A only)
Option 907, Front Handle Kit (Agilent Part Number 5062-3988)



Option 908, Rack Mount Kit (Agilent Part Number 5062-3974)

This option is a rack mount kit containing a pair of 
anges and the necessary hardware to
mount the instrument, with handles detached, in an equipment rack with 482.6 mm (19
inches) horizontal spacing.

Option 909, Rack Mount Flange and Handle Kit (Agilent Part Number 5062-3975)

This option is a rack mount kit containing a pair of 
anges and the necessary hardware to
mount the instrument with handles attached in an equipment rack with 482.6 mm (19 inches)
horizontal spacing.)

Option 910, Extra Operation Manual

This option is an extra OPERATION AND SERVICE MANUAL. (Agilent Part Number
87511-90000)




                                                                     General Information    1-3
             Figure 1-1. 87511A/B S-Parameter Test Set with Accessories Supplied




ACCESSORIES


Accessories Supplied

Figure 1-1 shows the 87511A/B with the accessories supplied:
  Interconnect Cable
  RF Connecting Cables
Accessories Available


Calibration and Accessory Kits

The calibration kits listed in Table 1-1 contain the components required to characterize the
systematic errors of various measurement systems. The calibration kits also contain precision
adapters.
The accessory kits listed contain the components generally required to measure devices with
the types of connectors indicated. Additional information is included in the 8751A manual.




1-4   General Information
                                  Table 1-1. Accessories Available


                         Connector Type Calibration Kit Accessory Kit
                         7 mm                  85031B
                         Type-N 50
            85032B             11853A   1



                         Type-N 75
            85036B
                         BNC 50
                                  11854A   1




                         1   use with 85032B type-N 50
 calibration kit

Cable Sets

Use the 11857D 7 mm test port return cable set to extend the test ports of the 87511A. Use
the 11851B 50
 type-N cable to extend the test port of the 87511A with Option 001. Use the
11857B 75
 type-N test port return cable set to extend the test port of the 87511B.


OPERATING AND SAFETY PRECAUTIONS


Operating

You need observe only normal precautions in handling and operating the 87511A/B. Do not
exceed its operating level input of +23 dBm or any DC voltage.
Caution        Electrostatic discharge (ESD) can damage the highly sensitive microcircuits
               in the 87511A/B test sets. ESD damage is most likely to occur as the test
               sets are connected or disconnected. Protect them by wearing a grounding
               strap that provides a path to ground of high resistance. Alternatively, ground
               yourself by touching the outer shell of any grounded instrument chassis before
               touching the test port connectors.
               Never touch the center contacts of the connectors.
               Use a work station equipped with an anti-static surface.

Service

The voltages in this test set do not warrant more than normal caution for operator safety.
Nevertheless, service should be performed by qualied personnel.




                                                                               General Information   1-5
RECOMMENDED TEST EQUIPMENT

Table 4-1 lists equipment and accessories required for testing, and repairing the 87511A/B.
The table notes which items are required to verify the performance of the test set and which
are required to troubleshoot it. Other equipment may be substituted if its specications meet
or exceed the specications listed in the critical specications column.


SPECIFICATIONS


Definitions

The specications listed in Table 1-2 range from those guaranteed by Agilent Technologies
to those typical of most 87511A/B instruments but are not guaranteed. Codes in the far
right column of Table 1-2 reference a specication denition listed below. These denitions
are intended to clarify the extent to which Agilent Technologies supports the specied
performance of the 87511A/B.
S       This performance parameter is eld veriable using performance tests documented in
        the Chapter 4.
T       Typical but non-warranted performance characteristics intended to provide information
        useful in applying the instrument. Typical characteristics are representative of most
        instruments, though not necessarily tested in each unit and not eld tested.




1-6   General Information
These specications describe the analyzer's performance over the temperature range of 2365
 C.


                                            Table 1-2. Specifications


       Parameter                         87511A                              87511B                Code
Frequency                     100 kHz to 500 MHz                 100 kHz to 500 MHz                 S
Range
Directivity :     1
                              30 dB, 100 kHz to 300 kHz          30 dB, 100 kHz to 300 kHz
Port 1,2                      40 dB, 300 kHz to 200 MHz          33 dB, 300 kHz to 500 MHz
                              35 dB, 200 MHz to 500 MHz
                              (37 dB, 300 kHz to 200 MHz)  2



Port Match :      1
                                                                                                    T
 R, A, B                      20 dB                              20 dB
 RF IN                        20 dB                              20 dB
Switch                        60.03 dB                           60.03 dB                           S
Repeatability
Source Match              1
                              15 dB, 100 kHz to 300 kHz          15 dB, 100 kHz to 300 kHz
                              20 dB, 300 kHz to 500 MHz          20 dB, 300 kHz to 500 MHz
Frequency                                                                                           T
Response :    1


 Transmission
 Tracking
  Magnitude           3
                              61 dB                              61 dB
  Phase   4
                              610, 100 kHz to 300 kHz            615 , 100 kHz to 300 kHz
                              65 , 300 kHz to 500 MHz            65 , 300 kHz to 500 MHz
 Re
ection
 Tracking
  Magnitude           3
                              61 dB                              61 dB
  Phase   4
                              615, 100 kHz to 300 kHz            620 , 100 kHz to 300 kHz
                              65 , 300 kHz to 1 MHz              610 , 300 kHz to 1 MHz
                              65 , 1 MHz to 500 MHz              65 , 1 MHz to 500 MHz
Insertion Loss:                                                                                  Nominal
 RF IN to Port 1,2 13 dB                                         19 dB
 RF IN to R, A, B 19 dB                                          31 dB
Isolation between 110 dB                                         100 dB                             T
Port 1 and 2
1   can be greatly improved with 8751A accuracy enhancement techniques
2   Option 001 only
3   Deviation from mean value
4   Deviation from linear phase



                                                                                  General Information   1-7
                                Table 1-3. Specifications (continued)


       Parameter                  87511A                                87511B      Code
Impedance:                                                                         Nominal
 Port 1, 2            50
                               75

 RF IN, R, A, B       50
                               50

Connectors:
 Port 1, 2            7 mm (50 
)                       75
 Type-N(f)
                      50
 type-N(f)1



 RF IN R, A, B        50
 type-N(f)                     50
 type-N(f)
Maximum               +20 dBm                           +20 dBm
Input Level
Damage Level          +23 dBm                           +23 dBm
Operating
Conditions:
 Temperature          0 C to 55 C                       0 C to 55 C
 Relative Humidity    95% at 40 C                       95% at 40C
Dimensions            432 mm 2 90 mm 2 495 mm           432 mm 2 90 mm 2 495 mm
                      (17 2 3.5 2 19.5 inches)          (17 2 3.5 2 19.5 inches)
Weight                5.7 kg                            5.7 kg
Power                 +22 Vdc from 8751A                +22 Vdc from 8751A
                      via interface cable               via interface cable
                      (22 V, 20 mA typ.)                (22 V, 20 mA typ.)
1   Option 001 only




1-8    General Information
                                                                                                          2
Installation
INTRODUCTION

This chapter explains how to install the 87511A/B test set. The topics covered include initial
inspection, environment considerations, positioning and connecting the test set for use, and
packaging the instrument.


INITIAL INSPECTION

Inspect the shipping container (including cushioning material) for damage. If it is damaged,
keep it until you have checked the contents for completeness. The contents are listed in
Table 2-1.
In addition, check the test set mechanically and electrically. If the test set and shipping
container are undamaged, performing the Directivity performance test in Chapter 4 should
suce for incoming inspection. If the test set does not pass the performance test, refer to the
troubleshooting procedures in Chapter 7.
If the shipping container is damaged, perform all electrical performance tests in Chapter 4. If
the test set fails the electrical tests, or is damaged, defective, or incomplete, keep the shipping
materials and notify both the carrier and the nearest Agilent Technologies oce. The Agilent
Technologies oce will arrange for repair or replacement of the test set without waiting for
settlement of the claim.
                                       Table 2-1. Contents of the 87511A/B


                                                                  Agilent Part Number
                        Component
                                                            87511A                  87511B
     Interconnect Cable                             08503-60051              08503-60051
     RF Connecting Cables          1
                                                    8120-4782                8120-4782
     Operation and Service Manual                   87511-90000              87511-90000
     1   Consists of four cables




                                                                                           Installation   2-1
ENVIRONMENTAL CONSIDERATIONS


Operating and Storage

The 87511A/B should be operated in temperatures between 0 C and +55 C with relative
humidity between 5% and 95% at 40C (non-condensing). It may be operated at altitudes up
to 4,500 meters (15,000) feet.
The 87511A/B may be stored in temperatures from 040 C to +70 C, with relative humidity
up to 90% at +65 C (non-condensing) and at altitudes up to 15,240 meters (50,000 feet).


PREPARATION FOR USE


Positioning the Test Set

Typically the 87511A/B is placed under the 8751A network analyzer whether it is
rack-mounted or used on a bench.
To install the 87511A/B for use on a bench, place it under or next to the 8751A. Use an
anti-static work surface to lessen the chance of ESD damage.
Connecting the Test Set


Mating Connectors

Ports 1 and 2 of the 87511A are precision 7 mm connectors and mate with precision 7
mm connectors. Ports 1 and 2 of the 87511A Option 001 are precision 50
 type-N female
connector and mate with precision 50
 male connectors. Ports 1 and 2 of the 87511B are
precision 75
 type-N female connector and mate with precision 75
 male connectors.
Caution              Do not mate type-N 50
 connectors to type-N 75
 connectors or damage may
                     result.

The RF IN, R, A and B ports of the 87511A/B are precision 50
 type-N female connectors
which mate with the corresponding precision type-N male connectors whose dimensions
conform to US specication MIL-C-39012.
The NETWORK ANALYZER INTERCONNECT connector is a series D 25-contact female
connector and mates with the corresponding male connector.
Power and Control Connections

Power for the 87511A/B is supplied by the 8751A network analyzer through the interconnect
cable. The interconnect cable also transmits control signals between the test set and the
network analyzer. Connect the interconnect cable from the NETWORK ANALYZER
INTERCONNECT connector on the rear panel of the 87511A/B to the TEST SET
INTERCONNECT connector on the rear panel of the 8751A.

2-2   Installation
Signal Path Connections

Place the test set under the network analyzer and connect its RF IN port to the network
analyzer's RF OUT port, R to R, A to A, B to B. Use the four cables supplied with the test
set.
To connect the signal path ports with the test set not positioned under the network analyzer,
use the longer cables of the 11851B RF cable set. Connect the RF IN port to the RF OUT
port with the longer cable. Connect port R to R, A to A, B to B with the three phase
matched cables.


PACKAGING

If shipping to a Hewlett Packard service center is required, each test set should be repackaged
using the original factory packaging materials.
Alteratively, comparable packaging materials may be used. Wrap the test set in heavy paper
or anti-static plastic. Use sucient shock absorbing material on all sides of the 87511A/B to
provide a thick, rm cushion and prevent movement. Seal the shipping container securely and
mark it FRAGILE .
In any correspondence with Agilent Technologies, refer to the 87511A/B by its full model and
serial number.




                                                                               Installation   2-3
                                                                                             3
Operation
INTRODUCTION

This chapter illustrates the features of the front and rear panels of the 87511A/B (Figure 3-1
and Figure 3-2). Figure 3-3 shows a typical test set measurement setup. Figure 3-4 depicts
the RF signal path through the test set for each of the four S-parameter measurements.




                                 Figure 3-1. Front Panel Features



 1.   Guard Terminal.This terminal is tied to the instrument's chassis and can be used for
    measurements that require guarding.
 2. PORT 1. The device under test is connected here or to port 2 or to both
Caution          Do not input more than +23 dBm or any DC Voltage to PORT 1 and PORT
                 2.


Note             The connector for the PORT 1 terminal is designed with a MOVABLE TEST
                 PORT; and this is not a defect.




 3.                        This indicator lights when the test set is in the S11 or S21
      S11 S21 Indicator LED.
      measurement mode and power is incident from port 1.
                                                                                 Operation   3-1
 4.   S22 S12 Indicator LED.This indicator lights when the test set is in the S22 or S12
      measurements mode and power is incident from port 2.
 5.   PORT 2. The device under test is connected here or to port 1 or to both
 6.   RF IN. This RF input connector receives the RF signal from the 8751A RF OUT
      connector.
 7.   R. This RF connector provides the reference signal to port R of the 8751A.
 8.   A. This RF connector provides an RF input signal to port A of the 8751A from the
      device under test.
 9.   B. This RF connector provides an RF input signal to port B of the 8751A from the
      device under test.




                                  Figure 3-2. Rear Panel Features



10.   NETWORK ANALYZER INTERCONNECT This connector supplies the test set with
      power from the network analyzer and also serves as a bidirectional control path.




3-2    Operation
Figure 3-3. Typical Measurement Setup




                                        Operation   3-3
                  Figure 3-4. RF Signal Path




3-4   Operation
                                                                                          4
Performance Tests
INTRODUCTION

This chapter provides the test procedures used to verify that the 87511A/B's specications
listed in this manual are met. All tests can be performed without access to the interior of
the instrument. The performance tests can also be used to perform incoming inspection,
and to verify that the 87511A/B meets performance specications after troubleshooting or
adjustment. If the performance tests indicate that the 87511A/B is NOT operating within the
specied limits, check your test setup, and then proceed to Adjustments or Troubleshooting in
the 87511A/B service manual, if necessary.
Note           Perform all performance tests in an ambient temperature of 23  C 6 5  C.




PERFORMANCE TEST RECORD

Record the performance test results in the Test Record at the end of this chapter (Photocopy
the Test Record and use the photocopy). The Test Record lists all test specications and
their acceptable limits. Test results recorded during incoming inspection can be used for
comparison purposes during periodic maintenance, troubleshooting, and after repair or
adjustment.
Note           The test limits indicated for each test do not take into account the
               measurement error of the test equipment used. Be sure to take the test
               equipment's measurement error into account when determining whether or not
               the 87511A/B meets specications.



CALIBRATION CYCLE

The 87511A/B requires periodic performance verication. The frequency of performance
verication depends on the operating and environmental conditions under which the
87511A/B is used. Verify the 87511A/B's performance at least once a year, using the
performance tests described in this chapter.




                                                                      Performance Tests   4-1
EQUIPMENT REQUIRED

The equipment required for performance testing is listed in Table 4-1. Substitutions can be
made for the equipment in this list if the specications meet or exceed the specications listed
in the requirement column.
                                    Table 4-1. Recommended Test Equipment


         Instrument                          Critical Specications          Recommended Model
Network Analyzer              no substitute                                 8751A
Cable                         RF Cable Set ( 50 
 type-N, three 24 inch     11851B
                              phase-matched, one 35 inch)
                              Interconnect Cable                            PN 87511-61609
                              RF connecting Cable, 4 ea.                    PN 8120-4387
                              GPIB Cable                                    10833A
Calibration Kit               no substitute                                 85031B
                                                                            85032B
                                                                            85036B
Adapter                       7 mm-N(f), 50 
, 2 ea.                        PN 85054-60001
                              Minimum Loss Pad                              11852B
Load                          type-N, 50 
 (return loss  49)                909F Opt.012
Controller                    HP Technical Computer with BASIC rev 5.0 or   HP 9000 Model 310
                              above
Disk Drive                    CS80                                          9122A
DVM                           DC: up to 22 V  1
                                                                            3458A
1   Only Used For Troubleshooting




4-2     Performance Tests
DIRECTIVITY TEST

Directivity is a measure of the ability of a directive device (in this case the 87511A/B) to
discriminate between incident and re
ected signals. In this procedure directivity is tested
by using the internal bridge to measure the re
ection coecient of a standard termination.
Since the return loss of the termination is much greater than the directivity of the test set, the
measurement made in this procedure is essentially the directivity of the 87511A/B test set.
Use the highest quality termination available as that factor determines the accuracy of the
procedure.
Specification


                                                           Directivity


                                  Frequency Range                 87511A     87511A 87511B
                                                                             Opt. 001
                           100 kHz  Freq.  300 kHz  30 dB  30 dB  30 dB
                          300 kHz < Freq.  200 MHz  40 dB  37 dB  33 dB
                          200 MHz < Freq.  500 MHz  35 dB  35 dB  33 dB

Test Equipment


           Test Equipment                 87511A                         87511A              87511B
                                                                         Opt.001
       Network Analyzer          8751A                         8751A                   8751A
       Interconnect cable        PN 87511-61609        1
                                                               PN 87511-61609      1
                                                                                       PN 87511-61609  1


       RF Cable                  PN 8120-4387 , 4 ea.
                                                   2
                                                               PN 8120-4387 , 4 ea.
                                                                               2
                                                                                       PN 8120-4387 , 4 ea.
                                                                                                   2


       Calibration Kit           85031B                        85032B                  85036B
       50 
 Termination                                        909F Opt.012
       1   This cable is supplied with test set.
       2   This cable is supplied with test set.

Procedure


Note                By convention, when the connector sex is indicated in parentheses for a
                    calibration standard, it refers to the sex of the test port connector (not
                    the actual standard). For example, short (f) indicates that the test port
                    connector, note the short circuit connector, is female.

 1. Connect the equipment as shown in Figure 4-1.




                                                                                          Performance Tests   4-3
                                                                    Figure 4-1. Directivity Test Setup



 2. Set the 8751A as follows:
      Control Settings                                                                                Key Strokes


      PRESET                                                                                          4PRESET5
      Source Power = 0 dBm (87511A)                                                                         , ,
                                                                                                      4MENU5 405 4x15
      Source Power = 10 dBm (87511B)                                                                        , , ,
                                                                                                      4MENU5 415 405 4x15
      Sweep Type: LOG FREQ                                                                            4MENU5, SWEEP TYPE MENU , LOG
                                                                                                                   NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN    NNNNNNNNNNNNNNNNNNNNNNNNNN

                                                                                                                                                                                  FREQ
      IF BW = 200 Hz                                                                                     5, IF BW , 4 5, 4 5, 4 5, 4 5
                                                                                                      4AVG
                                                                                                              NNNNNNNNNNNNNNNNN

                                                                                                                                    2         0           0      x1



 3. To set the system impedance,
                                                     87511A:        Press 4                                   ,   CAL KIT , 7 mm .
                                                                                                                  NNNNNNNNNNNNNNNNNNNNN   NNNNNNNNNNNNN

                                                                                                        CAL  5

                                                     8751A Opt.001: Press 4                                  5,   CAL KIT , N 50 
 .
                                                                                                                  NNNNNNNNNNNNNNNNNNNNN   NNNNNNNNNNNNNNNNNN

                                                                                                        CAL


                                                     87511B:        Press 4                                  5,   CAL KIT , N 75 
 .
                                                                                                                  NNNNNNNNNNNNNNNNNNNNN   NNNNNNNNNNNNNNNNNN

                                                                                                        CAL




 4. Press 4 5, CALIBRATE MENU , RESPONSE to enter the response calibration menu.
                          NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN   NNNNNNNNNNNNNNNNNNNNNNNNNN

            CAL



 5. When prompted, connect the SHORT supplied with the calibration kit to the test set's
    TEST PORT 1.
 6. Press SHORT ( SHORT[F] for 87511A opt. 001 and 87511B) to measure the calibration
           NNNNNNNNNNNNNNNNN    NNNNNNNNNNNNNNNNNNNNNNNNNN




    standard on the 8751A.
 7. Press DONE: RESPONSE after the short has been measured. The status indicator \Cor"
           NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




    should appear at the left side of the display.
 8. Remove the SHORT, and connect the following Termination to 87511A/B TEST PORT 1.

4-4    Performance Tests
                                       87511A:         7 mm 50 
 Termination supplied with the 85031B
                                       87511A Opt.001: Type-N 50 
 Termination (909F opt.012)
                                       87511B:         Type-N 75 
 Termination supplied with the 85036B

 9. Perform the following key strokes to generate a single sweep.
         5, TRIGGER MENU , SINGLE
                       NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN       NNNNNNNNNNNNNNNNNNNN

    4MENU



10. Perform the following key strokes to set the marker search range from 100 kHz to 300
    kHz.
        5, ACTIVE MARKER , MARKER 2 , 4 5, 4 5, 4 5, 4  5, 4   5, 4 MODE MENU ,
                     NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNNNNNN                                                                        NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

    4MKR                                                                                                3         0           0         k/m            MKR


    4 REF MARKER , 4REF-1 , 4             5, SEARCH RANGE , SEARCH RNG STORE ,
    NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN       NNNNNNNNNNNNNNNNNNNNNN                                      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

                                                                                MKR FCTN
    NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

    PART SEARCH on OFF
11. Perform the following key strokes to move the marker to the maximum point. Record the
    maximum reading value on the Performance Test Record.
                 5, MKR SEARCH , MAX , 4   5, 4 MODE MENU , 4 MODE OFF
                                  NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN          NNNNNNNNNNN                      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

    4MKR FCTN                                                                                 MKR



12. Press 4 REF MARKER , 4REF-1 .
                   NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNN




13. Perform the following key strokes to set the marker search range from 300 kHz to 200
    MHz.
          5, ACTIVE MARKER , MARKER 1 , 4 5, 4 5, 4 5, 4    5, 4   5, ACTIVE MARKER , MARKER 2 ,
                     NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN       NNNNNNNNNNNNNNNNNNNNNNNNNN                                                                       NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNNNNNN

    4MKR                                                                                                2         0           0         k/m            MKR


    4 5, 4 5, 4 5, 4 5, 4 5, 4 5, 4      5, SEARCH RANGE , SEARCH RNG STORE
                                                                                                        NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN       NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

     1         9        9         .        7           M/u             MKR FCTN



14. Perform the following key strokes to move the marker to the maximum point. Record the
    maximum reading value on the Performance Test Record.
                 5, MKR SEARCH , MAX , 4   5, 4 MODE MENU , 4 MODE OFF
                                  NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN          NNNNNNNNNNN                      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

    4MKR FCTN                                                                                 MKR



15. Press 4 REF MARKER , 4REF-1 .
                   NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNN




16. Perform the following key strokes to set the marker search range from 200 MHz to 500
    MHz.
          5, ACTIVE MARKER , MARKER 1 , 4 5, 4 5, 4 5, 4 5, 4 5, 4  5, 4  5, ACTIVE MARKER ,
                     NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN       NNNNNNNNNNNNNNNNNNNNNNNNNN                                                                                          NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

    4MKR                                                                                                1         9           9         .       7         M/u            MKR


    MARKER 2 , 4 5, 4 5, 4 5, 4 5, 4          5, SEARCH RANGE , SEARCH RNG STORE
    NNNNNNNNNNNNNNNNNNNNNNNNNN                                                                                         NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN   NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

                                   3           0         0        M/u                MKR FCTN



17. Perform the following key strokes to move the marker to the maximum point. Record the
    maximum reading value on the Performance Test Record.
                 5, MKR SEARCH , MAX , 4   5, 4 MODE MENU , 4 MODE OFF
                                  NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN          NNNNNNNNNNN                      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

    4MKR FCTN                                                                                 MKR



18. Press 4 5, REFL:REV S22(B/R) .
                    MEAS
                                       NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




19. Press 4 5, CALIBRATE MENU , RESPONSE , to enter the calibration menu.
                    CAL
                                  NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN            NNNNNNNNNNNNNNNNNNNNNNNNNN




20. WHen prompted, connect the SHORT supplied with the calibration kit to 87511A/B
    TEST PORT 2.
21. Press SHORT ( SHORT[F] for 87511A Opt. 001 and 87511B) to measure the calibration
                   NNNNNNNNNNNNNNNNN    NNNNNNNNNNNNNNNNNNNNNNNNNN




    standard on the 8751A.

                                                                                                                                                                                      Performance Tests                                    4-5
22. Press DONE: RESPONSE after the short has been measured. The status indicator \Cor"
                     NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




    should appear at the left side of the display.
23. Remove the SHORT, and connect the following Termination to 87511A/B TEST PORT 2.
                                        87511A:         7 mm 50 
 Termination supplied with the 85031B
                                        87511A Opt.001: Type-N 50 
 Termination (909F opt.012)
                                        87511B:         Type-N 75 
 Termination supplied with the 85036B

24. Perform the following key strokes to generate a single sweep.
           5, TRIGGER MENU , SINGLE
                         NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNN

    4  MENU



25. Perform the following key strokes to set the marker search range from 100 kHz to 300
    kHz.
          5, ACTIVE MARKER , MARKER 1 , 4 5, 4 5, 4 5, 4    5, ACTIVE MARKER , MARKER 2 , 4 5, 4 5,
                      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNNNNNN                                                    NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN   NNNNNNNNNNNNNNNNNNNNNNNNNN

    4  MKR                                                                                            1         0        0        k/m                                                                                           3         0


    4 5, 4    5, 4     5, 4 MODE MENU , 4 REF MARKER , 4REF-1 , 4              5, SEARCH RANGE ,
                                                NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN       NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN   NNNNNNNNNNNNNNNNNNNNNN                                    NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

       0         k/m               MKR                                                                                                                               MKR FCTN


    SEARCH RNG STORE .
      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




26. Perform the following key strokes to move the marker to the maximum point. Record the
    maximum reading value on the Performance Test Record.
                 5, MKR SEARCH , MAX , 4   5, 4 MODE MENU , 4 MODE OFF
                                    NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNN                      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

    4  MKR FCTN                                                                             MKR



27. Press 4 REF MARKER , 4REF-1 .
                     NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNN




28. Perform the following key strokes to set the marker search range from 300 kHz to 200
    MHz.
          5, ACTIVE MARKER , MARKER 1 , 4 5, 4 5, 4 5, 4    5, 4   5, ACTIVE MARKER , MARKER 2 ,
                       NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNNNNNN                                                                    NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN     NNNNNNNNNNNNNNNNNNNNNNNNNN

    4  MKR                                                                                            2         0        0         k/m              MKR


    4 5, 4 5, 4 5, 4 5, 4 5, 4 5, 4      5, SEARCH RANGE , SEARCH RNG STORE
                                                                                                      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN   NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

       1         9        9         .       7         M/u               MKR FCTN



29. Perform the following key strokes to move the marker to the maximum point. Record the
    maximum reading value on the Performance Test Record.
                 5, MKR SEARCH , MAX , 4   5, 4 MODE MENU , 4 MODE OFF
                                    NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNN                      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

    4  MKR FCTN                                                                             MKR



30. Press 4 REF MARKER , 4REF-1 .
                     NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNN




31. Perform the following key strokes to set the marker search range from 200 MHz to 500
    MHz.
          5, ACTIVE MARKER , MARKER 1 , 4 5, 4 5, 4 5, 4 5, 4 5, 4  5, 4  5, ACTIVE MARKER ,
                       NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN     NNNNNNNNNNNNNNNNNNNNNNNNNN                                                                                       NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

    4  MKR                                                                                            1         9        9        .        7           M/u             MKR


    MARKER 2 , 4 5, 4 5, 4 5, 4 5, 4          5, SEARCH RANGE , SEARCH RNG STORE
      NNNNNNNNNNNNNNNNNNNNNNNNNN                                                                                 NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

                                     3        0            0        M/u            MKR FCTN



32. Perform the following key strokes to move the marker to the maximum point. Record the
    maximum reading value on the Performance Test Record.
                 5, MKR SEARCH , MAX , 4   5, 4 MODE MENU , 4 MODE OFF
                                    NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNN                      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

    4  MKR FCTN                                                                             MKR




4-6      Performance Tests
SWITCH REPEATABILITY TEST

The test checks the repeatability of the switch by toggling it ten times and comparing the
initial trace to the resultant trace after toggling.
Specification

Switch Repeatability:                       6 0.03 dB
Test Equipment


         Test Equipment                   87511A                8751A                87511B
                                                                Opt.001
      Network Analyzer           8751A                    8751A                8751A
      RF Cable Set               11851B                   11851B               11851B
      Adapters                   PN 85054-60001 , 2 ea.
                                                    1
                                                                               11852B, 2 ea.
                                                                                           2


      Interconnect Cable     3
                                 PN 87511-61609     3
                                                          PN 87511-61609   3
                                                                               PN 87511-61609   3


      RF Cable                   PN 8120-4387 , 2 ea.
                                                3
                                                          PN 8120-4387 , 2 ea. PN 8120-4387 , 2 ea.
                                                                       3                    3



     1   A part of the 85032B calibration kit
     2   Minimum Loss Pad
     3   This cable is supplied with test set

Procedure

 1. Connect the equipment as shown in Figure 4-2.




                                   Figure 4-2. Switch Repeatability Test Setup



                                                                                 Performance Tests    4-7
 2. Set up the 8751A as follows:
      Control Settings                                                                                         Key Strokes


      PRESET                                     4      5                                                        PRESET

      Source Power = 0 dBm (87511A)              4    5, POWER , 4 5, 4 5                                        MENU
                                                                                                                                     NNNNNNNNNNNNNNNNN

                                                                                                                                                             0         x1


      Source Power = 10 dBm (87511B)             4    5, POWER , 4 5, 4 5, 4 5                                   MENU
                                                                                                                                     NNNNNNNNNNNNNNNNN

                                                                                                                                                             1         0        x1


      IF Bandwidth = 200 Hz                      4  5, IF BW , 4 5, 4 5, 4 5, 4 5                                AVG
                                                                                                                                NNNNNNNNNNNNNNNNN

                                                                                                                                                         2        0         0           x1


      Scale Ref. = 0.01                          4        5, 4 5, 4 5, 4 5, 4 5                                  SCALE REF                       .           0        1         x1


 3.   Press 4 5, TRIGGER MENU , SINGLE to generate a single sweep with the network
                      MENU
                                        NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN    NNNNNNNNNNNNNNNNNNNN




      analyzer.
 4.   At the end of the sweep, press 4       5, DEFINE TRACE , DATA->MEMORY , DATA/MEM to
                                                                                                           NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN        NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN   NNNNNNNNNNNNNNNNNNNNNNNNNN

                                                                                    DISPLAY

      store the sweep in memory.
 5.   Press 4 5, TRIGGER MENU , CONTINUOUS to put the 8751A in continuous sweep in
                                        NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN    NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

                      MENU

      memory.
 6.   Press 4 5 and then toggle the 87511A/B switch by alternately pressing
                      MEAS


      Refl:REV S22(B/R) and Refl:FWD S11 (A/R) ten times on the 8751A. You should be
      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN            NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




      able to hear the switch cycle and see two LEDs light alternately each time you press the
      keys. Finish with the instrument in the \Re
: FWD S11" mode.
 7.   Press 4 5, TRIGGER MENU , SINGLE to generate one more sweep.
                                        NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN    NNNNNNNNNNNNNNNNNNNN

                      MENU



 8.   Note the amount the trace deviates from the reference line and enter that value in the
      Performance Test Record (S11).
 9.   Press 4 5, Refl:REV S22(B/R) , 4 5, TRIGGER MENU , SINGLE to generate a single
                      MEAS
                                        NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

                                                                                                 MENU
                                                                                                                     NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN      NNNNNNNNNNNNNNNNNNNN




      sweep in the S22 measurement mode.
10.   At the end of the sweep, press 4       5, DEFINE TRACE , DATA->MEMORY , DATA/MEM to
                                                                                                           NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN        NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN   NNNNNNNNNNNNNNNNNNNNNNNNNN

                                                                                    DISPLAY

      store the sweep measurement data into memory.
11.   Press 4 5, TRIGGER MENU , CONTINUOUS to put the 8751A in continuous sweep in
                                        NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN    NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN

                      MENU

      memory.
12.   Press 4 5 and then toggle the 87511A/B switch by alternately pressing
                      MEAS


      Refl:FWD S11 (A/R) and Refl:REV S22 (B/R) on the 8751A ten times. You should be
      NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN            NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN




      able to hear the switch cycle and see two LEDs light alternately each time you press the
      keys. Finish with the instrument in the \Re
: REV S22(B/R)" mode.
13.   Press 4 5, TRIGGER MENU , SINGLE to generate one more sweep.
                      MENU
                                        NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN    NNNNNNNNNNNNNNNNNNNN




14.   Note the amount the trace deviates from the reference line and enter that value in the
      Performance Test Record (S22).




4-8      Performance Tests
EFFECTIVE SOURCE MATCH TEST

This test checks the source match of the 87511A/B. This procedures determines the eective
source match of port 1 and port 2 by examining the calibration coecient arrays. These
arrays are generated during measurement calibration. The array data are the dierence
between the measured known calibration standards and ideal modeled data. Thus the
calibration coecient arrays are a direct indication of performance because they represent the
systematic errors of the test set.
Specification


                                   Frequency Range           87511A 87511A 87511B
                                                                    Opt.001
                            100 kHz  Freq.  300 kHz  15 dB  15 dB  15 dB
                            300 kHz < Freq.  500 MHz  20 dB  20 dB  20 dB

Test Equipment


           Test Equipment                      87511A                 87511A                 87511B
                                                                      Opt.001
 Network Analyzer                        8751A                  8751A                  8751A
 RF cable                                PN 8120-4387 , 4 ea.
                                                     1
                                                                PN 8120-4387 , 4 ea.
                                                                            1
                                                                                       PN 8120-4387 , 4 ea.
                                                                                                   1


 Interconnect cable                      PN 87511-61609  1
                                                                PN 87511-61609  1
                                                                                       PN 87511-61609  1


 Calibration Kit                         85031B                 85032B                 85036B
 50 
 Termination                                               909F Opt.012
 Controller                              HP 9000 Model 310      HP 9000 Model 310      HP 9000 Model 310
 Disk Drive                              9122A                  9122A                  9122A
 GPIB Cable                              10833A, 2 ea.          10833A, 2 ea.          10833A, 2 ea.
 BASIC 5.1 Operating System              98616A                 98616A                 98616A
 1   This cable is supplied with test set.




                                                                                    Performance Tests      4-9
Procedure

 1. Connect the equipment as shown in Figure 4-3.




                                    Figure 4-3. Effective Source Match Test Setup



 2. Set the 8751A as follows:
   Control Settings                                                                  Key Strokes


   PRESET                                                                            4PRESET5
   Sweep Type: LOG FREQ                                                              4MENU5            ,   SWEEP TYPE MENU , LOG FREQ
                                                                                                           NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN          NNNNNNNNNNNNNNNNNNNNNNNNNN




   Source Power = 0 dBm (87511A)                                                     4MENU            5,       ,
                                                                                                           POWER ,
                                                                                                           NNNNNNNNNNNNNNNNN

                                                                                                                                   405 4x15
   Source Power = 10 dBm (87511B)                                                    4MENU         POWER , 4 5, 4 5, 4 5
                                                                                                      5,
                                                                                                           NNNNNNNNNNNNNNNNN

                                                                                                                                   1          0           x1


   IF Bandwidth = 200 Hz                                                             4AVG      5, IF BW , 4 5, 4 5, 4 5, 4
                                                                                                      NNNNNNNNNNNNNNNNN

                                                                                                                               2         0         0            5
                                                                                                                                                               x1



 3. To set the system impedance,
                                    87511A:         Press 4                                            ,   CAL KIT , 7 mm                                 .
                                                                                                           NNNNNNNNNNNNNNNNNNNNN       NNNNNNNNNNNNNNNN

                                                                                            CAL       5

                                    87511A Opt.001: Press 4                                           5,   CAL KIT , N 50 
 .
                                                                                                           NNNNNNNNNNNNNNNNNNNNN       NNNNNNNNNNNNNNNNNN

                                                                                            CAL


                                    87511B:         Press 4             