Text preview - extract from the document
                                            Number 2959


Application Note                            Optimizing Low Current Measurements
           Series                           with the Model 4200-SCS
                                            Semiconductor Characterization System
Introduction
                                                                                             IS
                                                                                                        FORCE HI
Many critical applications demand the ability to measure very
low currents--such as picoamps or less. These applications
include determining the gate leakage current of FETs, testing                                 RS
sensitive nano-electronic devices, and measuring leakage current                                                                 IM
of insulators and capacitors.
                                                                                              VS                       IOFFSET
    The Model 4200-SCS Semiconductor Characterization System,
                                                                                                        COMMON
when configured with the optional Model 4200-PA Remote
Preamp, offers exceptional low current measurement capability
                                                                                     Current Source                SMU or PreAmp
with a resolution of 1E