INSTRUCTION   MANUAL

  MODEL 149
MILLI-MICROVOLTMETER
WARRANTY
     We warrant each of our products to be free
     from defects in material and workmanship. Our
     obligation under this warranty is to repair or
     replace any instrument or part thereof which,
     within a year after shipment, proves defective
     upon examination.       We will pay domestic
     surface freight costs.
     To exercise this warranty,      call your local
     field representative or the Cleveland factory,
     DDD 216-248-0400.    You will be given assist-
     ance and shipping instructions.


REPAIRS AND RECALIBRATION
     Keithley Instruments maintains a complete re-
     pair service and standards laboratory in Cleve-
     land, and has an authorized field repair facility
     in Los Angeles and in all countries outside the
     United States having Keithley field repre-
     sentatives.
     To insure prompt repair or recalibration    serv-
     ice, please contact your local field representa-
     tive or the plant directly before returning the
     instrument.
     Estimates for repairs, normal recalibrations,
     and calibrations traceable to the National Bu-
     reau of Standards are available upon request.
MODEL 149 MILLI-MICROVOLTMETER                                                                      CONTENTS



                                       TABLE OF CONTENTS



Section                                  Page    Section                                               Page

1.     INTRODUCTION                .       l-1   4.   MAINTENANCE.          . . . . . . . .             4- 1

       Specifications          .           1-2         Trouble-Shooting       . .   . .   .             4-l
                                                       Excessive     Output Noise .     . . .           4- 1
2.     OPERATION.       .                 2-1          Output not Zero with Input
                                                        Terminals     Shorted     .     . .             4-2
       Operating   Controls    . . .      2-l          220-Volt   Operation     . . . . . . .           4-3
       Preliminary    Set-Up   .          2-l
       General Precautions.      . .      2-2    5.    REPLACEABLE PARTS.          .    .       .       5-l
       Measuring Voltage.      .          2-3
       Other Applications                 2-4         Replaceable   Parts List .           .            5-1
                                                      How to Order Parts . .          .                 5-l
3.     CIRCUIT DESCRIPTION.               3-1         Model 149 Replaceable     Parts List              5-2
                                                      Models 1483, 1484 Replaceable
       Input Circuit.            .     3-l             Parts List.   . .     . . . .       .            5-6
       AC Amplifier              . .   3-l            Model 1491 Replaceable     Parts List             5-7
       DC Amplifier   . .      . . . . 3-l            Model 1501 Replaceable     Parts List             5-7
       Zero Suppression        . . . . 3-2            Model 1502 Replaceable     Parts List             5-7
       Other Controls   .      . . . . 3-2            Voltage   and Resistance   Chart .                5-9
       Power Supply . .        . . . . 3-2            Schematic Diagram.       . . . . .                5-11
                                                      Green Repair and Calibration      Form            5-13

                                                 *    Change Notice         .   . .         . .         Last
                                                                                                        Page




K    Yellow Change Notice sheet is included           only   for   instrument   modifications
     affecting the Instruction Manual.



    1068R
MODEL 149 MILLI-MICROVOLTMETER                                                                                INTRODUCTION



                                               SECTION 1.    INTRODUCTION



l-l.       GENERAL.

   a.      The Model 149 Milli-Microvoltmeter      is a stable,      versatile instrument  for measuring
low-level      dc signals.     It functions   as a voltmeter     from 100 nanovol s full  scale to 100
millivolts.       It also operates      as a dc amplifier    with gains up to 10 ' for driving   recor-
ders.

       b.    The low noise level of the Model        149, together with its long-term            stability,       makes
it       ideal for many measurements requiring         extreme power sensitivity.

   C. Typical       applications      include   measuring      the output from strain         gages, thermopilcs,
thermocouples,       bolometers,      phototubes,     ionization      chambers, scintillation        counters,    and
barrier     layer cells.        Other applications       are found in cell studies,           measurement of elec-
trochemical      potentials,      electrolytic     corrosion      studies,  molecular      weight analysis     and
Hall effect      studies.

  d.       In addition  to its use as a direct indicator   of minute potentials                 and currents,        the
Model      149 may also be used as a null detector    in Wheatstone or Mueller                 bridges.

   e. An important  feature of the instrument    is zero suppression   up to 100 times full
scale in place of the usual more limited    meter zero.   This permits   measurements of small
signals in the presence of large thermal    emf's or other masking dc signals.




         FIGURE 1.    Keithley   Instruments     Model   149 Milli-Microvoltmeter.

                                                                                                                     l-l
  INTRODUCTION                                                                                      MODEL 149 MILLI-MICROVOLTMETER



  l-2.        SPECIFICATIONS.

  RANGE: 0.1 micro"olt                (10 x 10-8 volt)   full           scale        to 100 millivolts          on zero-center      meter.
  13 overlapping ranges               in lx and 3x steps.

  ACCURACY: *2% of full                scale    on all       ranges   exclusive         of noise       and drift.

  ZERO DRIFT:      Less than 10 nanovolts   per hour or less than 30 nanovolts   in any *-hour                                            per-
  iod after  approximately     Z-hour warm-up with reasonably   constant ambient temperature.
  Long-term  drift    is non-cumulative.

  INPUT NOISE (with              input shorted):            Less than    0.6     nanovolt          rms (3 nanovolts       peak-to-peak)
  on most sensitive              range.

  INPDT CHARACTERISTICS:

                                               Input Resistance                                        Maximum Source1
                                                  Greater   than,                                        Resistance,
                  m                                    ohms                                                  ohms
              0.1 P"                                   10 k                                                  100
              0.3 ,I"                                  30 k                                                  300
              1.0 &I"                                 100 k                                                     lk
              3.0 ,1"                                300 k                                                      3k
             10.0 1"                                     1M                                                   10 k
             30.0 )A"                                   3M                                                    30 k
         100 pv and above                              10 M                                                   30 k

         Note:l        Source resistances          higher      than   the recommended maximum will                    increase   noise
                       and rise time.

 LINE FREQUENCYREJECTION: Greater                        than 5O:l on the most sensitive                     r~ange. (Ratio        of im-
 pressed peak-to-peak line frequency                      voltage at input to indicated                     dc voltage.)

 ISOLATION:            Circuit    ground to chassis    ground:   Approximately     10' ohms shunted by 0.05 mi-
 crofarad.            Circuit    ground may be floated     up to *400 volts    with respect  to chassis ground.

 RISE TIME (10% to 90%):                                                                                                                         -
   O.l-microvolt  Range:   mess than 2 seconds when source resistance   is less than 10% of
   maximum; 4 seconds using maximum source resistance.
   0.3-microvolt  to lOO-millivolt  Ranges:   Less than 1 second when source resistance  is
   less than 10% of maximum; 2 seconds using maximum source resistance.

 ZERO SUPPRESSION: Up to at                    least 1 millivolt        on the microvolt                ranges and up to at least
 10 millivolts  on the millivolt                   ranges.    Stability      is such that               100 times full  scale may
 be suppressed.

 RECORDEROUTPUT:
   output: *lO volts                dc at up to 5 milliamperes                 for    full-scale        meter   deflection.

      Resistance:          Less than     10 ohms within          the amplifier           pass band,

                         10 volts
      Gain:       Range setting   in volts



1-2                                                                                                                                1167R
MODEL 149 MILLI-MICROVOLTMETER                                                                                               INTRODUCTION



  Noise:       Input   noise    times     gain      plus   modulation       products.

  Modulation       Products:      Less than          2% peak-to-peak         of full       scale     with       input      shorted.

CONNECTORS: Input:             Special      connector.         Front    Output:         Binding      posts.         Rear Output:
Amphenol 80-PC2F.

POWER: 105-125         or 210-250        volts,      60 cps,    50 watts.         SO-cps models             available.

DIMENSIONS, WEIGHT:            7 inches      high     x 19 inches       wide   x 13 inches           deep;       net weight,          24 pounds.

ACCESSORIES SUPPLIED: Model 1501 Low Thermal Input                              Cable with          alligator           clips;    mating
output connector; length of low-thermal solder.*


-'t The solder     is screwed to the right  side of the copper                          input      chassis,        located       inside    the
    Model 149.       Remove the top cover to reach the solder.




                                                                                                                                           1-3
MDDEL 149 KLLI-MICROVOLB                                                              OPERATION




                               SECTION 2           - OPERATION


  A.   OPERATINGCONTROLS

       The controls   of the Model 149 are simple                  and conveniently
       placed.   Their functions  are as follows:

            ON-OFF switch      is located         to the right       of the panel
                 meter.
            FUNCTION switch selects              the function       which Is to be
                used: MilliVOltS,                Or !dCrOVOltB.
            2AN'm switch selects the full scale multiplier    of
                 the function   selected by the FUNCTION switch.
            ZERC SUPPiWSS controls    consists    of the ZERO RANGE
                 switch which selects the coa.188 range of sup-
                 pressing voltage in discreet        steps and the ZERO
                 SET potentiometer    which gives continuously
                 variable   fine control   Including     settings through
                  TieI-0.


       B.   PRELIMINARY SET-UP


            Connect the instrument       to the power line.     Unless otherwise
            marked the unit may be used on 117 volt, 60 cps line.            To con-
            vert to 220-volt operation,         refer to the MAIN'lXK4NC!E section.
            A three-wire    line cord is furnished,      which grounds the cabinet.
            If a three-wire     receptable    is not available,    use the two-pin
            adapter furnished,      and ground the third lead to an external
            ground.
            Set controls      *a followa:
                  Function:         Millivolts

                  Range :     100
                  zero suppress:           OFF

                  Input:      Short     the input       leads     together.




                                                                                            2-1
OPE3ATION                                     KNEL 149 MILLI-MICROVOL-R
  C.   GENERAL PRECAUTIONS

       1.      Source Resistance        - 1n Section luncler the Input Re-
        sistance    SpeCifiCatYiOg      the maximum source resistance         for use with
       each voltage       range is specified.         Reasonable operation       is possible
       with source resistance           up to ten times greater      than those specified,
       however, the measurement will            suffer    from a considerable      decrease in
       speed of response,        and measuring       accuracy.    If the instrument      is
       left    completely     open-circuited,      the meter will    generally     drift   off
       scale on any range.

       2.     Shielding     - Since the instrument       operates      with a modulator
       frequency     of 120 cps, it is not generally           sensitive      to 60 cps pick-
       up unless it is large enough to overload              the amplifier.          The pickup
       may be a source of difficulty          when using the amplifier            with high
       impedances on ~the more sensitive         voltage     ranges.        In these cases it
       is desirable      to shield   the leads and the sources as completely                as
       possible.       In some cases a simple low-pass          filter      at the input to
       eliminate     frequencies    of about 1 cps and above will             be helpful.
       No use is made of an input filter          in this instrument           since any input
       series    impedance due to the filter        will   increase       the input noise and
       the thermal drift.         When operating    above ground, the case of the in-
       strument    must be grounded.

       3.     Determination      of Excessive    AC Pickup - A terminal             attached      to
       the output of the AC amplifier           at the point of the demodulator                 is
       provided    at the rear of the instrument.             It is labeled         DEMOD. OUTPUT.
       If an inability        to make consistent     readings       persists,     it is possible
       to check for the presence of excessive              pick-up      by observing       the
       wave-form at this point.          With the input shorted            the picture       should
       be approximately        as shown in figure      2. If excessive         pickup is ob-
       served it will       look as shown in figure        3. The circuit           will   operate
       reasonably     well as long as the wave-form does not clip as shown
       in figure    4. At this point the operation              will    be erratic.



                                                                         --I.-
                                                                    lo v P-P

                                                                      t
                                        FIOUIlE2


                                                                         c


                                                                    40 v P-P



                                                                     t
MODEL 149 MILLI-MICROVOLl'METI                                                            OPERATION




                                          FIGURE 4

      4.      Thermal EMF - Extreme precautions            have been taken in the input
      circuit      to minimize    thermal    EMF's so the residual        RMF is usually     less
      than 0.5 uv.         The material     used in the input circuit         is pure copper.
      Any other metal will          generate    a thermocouple    potential.      Lead solder is
      particularly       troublesome.       Where thermal     W's    are a problem,      soldering
      should be done with the cadmium-tin              solder supplied       with the instrument.

      5.     Input   Noise - The noise at the input          is a function     of input
      resistance     and is approximately given by
                             E = 1.29 x 10-loum
      where E is the RMS noise,      and R is the source resistaixe.            It is assumed
      that the bandwidth    of the instrument      is about 1 cps and the temperature
      is 80