Text preview - extract from the document
                                 Test Instrumentation Group
                                 Keithley Instruments, Inc.
                                 28755 Aurora Road
                                 Cleveland,Ohio 44139-1891
                                 (216)X3-0400 . Fax: (216)498-2704
                                 1-800-552-1115




       INSTRUCTION MANUAL

          MODEL     227


         CURRENTSOURCE
(AND MODEL 2271 PROGRAMMING OPTION)
                INSTRUCTIONMAMlAL
                     Model 227


        (and Model 2271 Programing   Optionj




 OCOPYEICHT 1976, KEITHLEY INSTRUMENTS,
                                      INC.
FOURTHPRINTING, JAN. 1978, CLE"!%LAND,
                                     OHIO, U.S.A.
MODEL227




                                                    WARNING
                                    EXTREME CARESHOULD TAILFNTO AVOID
                                                        BE
                                    CONTACT WITR RIGH VOLTAGES WREN MAKING
                                    CONNECTIONS EXTh?2NAL
                                                 n,         TERMIIULS OR
                                    WREN SERVICING TRE MODEL227 WITX TOP
                                    AND BOTT(3.lCOVERSRZMOVEV.
           RecommendedProcedures for User Safety:



           2.   Make certain third-wire ground pin an line power card ls connected fo earth ground.
                This will ensure that the Model 227 chassis is at ground potential.

           3.   Before making connections anywhere on the front or rear panels, remove power to
                Model 227 by detaching the line power cord, or by setting the power switch to OFF,
                or by setting the OUTPUTSELECTOR  switch to "SHORT" or "OPEN". Also, remove all
                external power Sources which may be connected between LO and CASE.

           4.   If the Model 227 LO terminal is to be floated above CASE, make certain     that the
                metal link is removed between front panel LO and CASE terminals.
           5.   Make certain cable connections are well insulated and that no bare wires     are ex-
                posed. Use connectors with fully enclosed and insulated virinS.
           6.   During calibration   with top and/or bottom covers removed, extreme care must be
                taken to avoid contact with exposed circuitry.     Use insulated tools and gloves for
                maximm safety.     Various test points and adjustment circuitry  my be up to 300
                volts potential.




                                                                                                        iii
0975
 CONTENTS                                                                             MODEL227



                                                     CONTENTS



            SECTION                                                            PAGE

            SPECIFICATIONS        __-__-_----_-_____________________            vi

            1.   GE-         I&?,,mTION        __-----_________-____________     1

            2.   INITIAI.    PREPARATION
                                       __--------___------_---                   2
            3.   ,,PE&Q-ING INS'fRUCTIONS--________-__-______-----               8
            4.   *CcESSlJRIES _----__---_-__-_____----------------             24
            5.   THEORY OPEP\ATION
                       OF        __--________________-----------               25
            6.   MAINTEN)JCE ---_--___---__-_____----------------               27
            7.   REPmCE&,LE PARTS__-______--___-___-____I___                    39
            SYNDICS         --__--------_____------------------------          60




iv                                                                                       0975
 MODEL227



                                               ILLUSTRATIONS



            Fig. No.                                          Title                                       Page

                1      FrQ"f panel. ____________-__-__--______I_____________-----                           1
                2      Rear Panel Showing Line Switches. ---------------------------                        2
                3      RecommendedMounting and Ventilation.           ------------------------              3
                4      LQc*tiQ" of *ir Filter. -------------------------------------                        3
                5      Voltage Levels Present at Front and Rear Panel Terminals.                 ------     5
                6      Front Panel Terminals and Controls. ---------------__-_---------                     7
                7      Rear p**e1 Terminals. --------------------------------------                         7
            8a.b.c     Output Selector Switch Positions.         ------------------------------             9
               9       Voltage Output Versus % Rotation for Compliance Control. -------                    10
              10       Connections as a Current Source with Resistive Load. ----------                     11
              11       Connectians es a Current Sink wirh Resistive Load. ------------                     11
              12       Connections as a Current Source with Capacitive Load. ----------                    13
              13       Connecrfo"s as a Currenr Source with Inductive Load. ----------                     13
            14a,b      Power Limits when used as a Current Sink. ---------------------                     14
              15       Typical Output Impedance Versus Frequency for lOOOvA Range. ----                   15
              16       Output Noise Versus Compliance Voltage. ---------------------                      16
              17       Output Noise Versus RC Time Conetant of External Load. -------                     16
              18       Connections for Floating Operation. ---------------------------                    17
              19       Frequency Response for Voltage Programming. -------------------                    18
              20       Use of Voltage Program Inpur to Modulate Current. --------------                   18
              21       Rear panel Remote Progrenrming Connector. --------------------                     20
              22       Compliance Progranrming up to 5OV. ----------------------------                    21
              23       Compliance Programming up to 3OOV. ----------------------------                    21
              24       Resistance  Programming  of Current. ---------------------------                   22
              25       Preferred Cabling Te.=,,"iques. __-______________---------------                   23
              26       Model 1008 Single Rack Mounting Kit. ------------------------                      24
              27       Overall Block Diagram of Model 227. ----------------------------                   26
              28       Simplified  Diagram of Current Source. -------------------------                   27
              29       Test Set-up for Model 22712271 Calibration.             -------------------        36
              30       Location of Adjustments with Top Cover Removed. ---------------                    38
              31       Location of Adjustments with Bottom Cover Removed. -------------                   38
              32       Chassis Assembly.  ---------_______________________________-----                   41
              33       Compo"e"t Layout,  PC-330. ------------____-____________________                   53
              34       component Layout, PC-331. -----------------------------------                      55
              35       Component Layout, PC-342. ------------_-______________________                     56
              36       Case   Outlines,   Integrated    Circuits.     ---------_--_-----_-----            57
              37       Case Outlines,     Integrated Circuits.    -------------------------               58
              38       Case Ourlines,     Transistors. __-_-------__-------____________                   59




0975
SPECIFICATIONS                                                                                                                      MODELZL,



                                                             SPECIFICATIONS



                                                             Calibrated   et 25'C " 3*C

AS A DC CONSTANT    CURRSNT   SOURCE                                           GENERAL
OUTPUT:                                                                        STABILITY: C(O.O05% of Betting + 0.005% of =ange)/Y
  current : ?l microampere (1000 microamperes full                               Short-tefm stability     up to 30 days is masked by thi
    range) to tlOO0 milliamperes in four decade                                  temperature coefficient.
    ranges; 3-digit   in-line   readout; 11% overrange                         LOADR!.?GULATION: t0.005% of range from no load to
    to 1110.                                                                     full load.
  Voltage Compliance: *300 volts on 100 milliampere                            LINE REGULATION: +O.OOS% range for 10% change in
                                                                                                             of
    range and lauer.     f50 volts on 1000 milliampere                           line voltage.
    range. Compliance limit continuously      adjustable                       NOISE: Wideband noise less than (0.03% of range + 2
    from appraximately 3 volts t0 full voltage cam-                              millivolts)   ~llls above 5 Hz.
      plialXe.
  Floating: 1500 volts maximum off chassis ground.
RESOLUTION: 0.005% af range, 3-digit readout.                                 ENVIRONMENT:
ACCURACY: ?(0.5% of Betting + 0.12% of range).                                  operating:    O'C to 50



◦ Jabse Service Manual Search 2024 ◦ Jabse PravopisonTap.bg ◦ Other service manual resources online : FixyaeServiceinfo