Service Manuals, User Guides, Schematic Diagrams or docs for : Keithley 2306 2306 FH-B17
<< Back |
HomeMost service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download
Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing.
If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.
Image preview - the first page of the document
>> Download 2306 FH-B17 documenatation <<Text preview - extract from the documentModel 2306 Firmware Revision History
Model 2306
Keithley Instruments, Inc. Version B17 Firmware Revision History
28775 Aurora Road
Cleveland, Ohio 44139-1891
1-888-KEITHLEY
www.keithley.com
Model 2306 Firmware Revision History
Release Revision Description
Date Number
3/1999 A01 Initial Release of 2306 firmware to production
4/1999 A02 Manufacturing release: Modifications to default factory calibration values and
factory Built In Test (BIT) parameters
8/1999 B01 Optimized response of voltage protection scheme (open sense lead detect) to
reduce false failures due to transients on the load voltage lines
1/2000 B02 Added capability to turn on or off both channels with a single, new command:
BOTHOUTON or BOTHOUTOFF
B03 Not released in production
4/2000 B04 Initial release of 2306-PJ firmware version (add 500mA range, eliminate 5mA
range)
Output bandwidth compensated mode now defaults to LOW on battery channel
B05 Not released in production.
12/2000 B06 Added FAST, SEARch and DETect for pulse current measurements
6/2001 B07 Added capabilities to get readings from both channels with single new
commands: BOTHTRG, BOTHFETCH and BOTHREAD
Added new command SYSTem:AZERo:STATe to turn off internal calibration
between readings (a measurement speed-up feature for increasing rate of
multiple readings)
B08 Not released in production REQUIRES RECALIBRATION.
4/2002 B09 Added 488.1 mode to increase throughput
Added SYSTem:TRIGger:TALK:BOTH to allow triggering on talk to apply to both
channels
6/2003 B10, B11 Initial release and production release for 2306-VS version
Enhancements to setting pulse current timeout value and pulse current
digitization integration time
2306 FRP (B17