Text preview - extract from the document
                                                                                                             Model 6517B
Keithley Instruments, Inc.                                                             Version A10 Firmware Release Notes
28775 Aurora Road
Cleveland, Ohio 44139-1891
1-888-KEITHLEY
www.keithley.com



Contents
Contents ............................................................................................................................................................................. 1
General Information ........................................................................................................................................................... 2
  Supported models.......................................................................................................................................................... 2
  Installation instructions .................................................................................................................................................. 2
  Upgrade considerations for the Model 6517B.............................................................................................................. 2
A10 Release....................................................................................................................................................................... 2
  Overview ........................................................................................................................................................................ 2
  Critical fixes.................................................................................................................................................................... 2
  Enhancements ............................................................................................................................................................... 2
A09 Release....................................................................................................................................................................... 3
  Overview ........................................................................................................................................................................ 3
  Critical fixes.................................................................................................................................................................... 3
  Enhancements ............................................................................................................................................................... 3
A08 Release....................................................................................................................................................................... 3
  Overview ........................................................................................................................................................................ 3
  Critical fixes.................................................................................................................................................................... 3
  Enhancements ............................................................................................................................................................... 4
A07 Release....................................................................................................................................................................... 4
  Overview ........................................................................................................................................................................ 4
  Critical fixes.................................................................................................................................................................... 4
  Enhancements ............................................................................................................................................................... 4
A06 Release....................................................................................................................................................................... 4
  Overview ........................................................................................................................................................................ 4
  Critical Fixes .................................................................................................................................................................. 4
  Enhancements ............................................................................................................................................................... 5
A05 Release....................................................................................................................................................................... 5
  Overview ........................................................................................................................................................................ 5
  Critical fixes.................................................................................................................................................................... 5
  Enhancements ............................................................................................................................................................... 6
A04 Release....................................................................................................................................................................... 6
  Overview ........................................................................................................................................................................ 6
  Critical fixes.................................................................................................................................................................... 6
  Enhancements ............................................................................................................................................................... 6
A03 Release....................................................................................................................................................................... 7
  Overview ........................................................................................................................................................................ 7
  Critical fixes.................................................................................................................................................................... 7
  Enhancements ............................................................................................................................................................... 9
  Non-critical fixes ............................................................................................................................................................ 9
  Known issues................................................................................................................................................................. 9




6517B-801-FRP Rev. K / May 13, 2010                                                                                                                                                   1
General Information - Supported models                                                            6517B Firmware Release Notes




General Information
Supported models
This firmware is used on the following Keithley Instruments products:

           Model 6517B


Installation instructions
Use Keithley Flash Wizard version C09 or later to upgrade the Model 6517B.

Upgrade considerations for the Model 6517B
The following table outlines the considerations that should be made when deciding whether to upgrade to this
version from any previous version.


                                            Consideration                          From
                                                                                    A09
                                            Recalibration required                  No
                                                                                       1
                                            Requalification suggested              Yes
                                                                                         2
                                            Should you upgrade?                   Review



A10 Release

Overview

Critical fixes
PR 41677             Models affected:

                     Model 6517B

                     Symptom:

                     Units may have slight offsets in measurements when calibrated while operating on 60 Hz power.

                     Resolution

                     The issue has been repaired.


Enhancements

No enhancements have been added in version A10.

1
    If any features are used related to any of the corrected issues, requalification is suggested to ensure compatibility.
2
    As needed to correct problems addressed by version A09.


2                                                                                         6517B-801-FRP Rev. K / May 13, 2010
6517B Firmware Release Notes



A09 Release
Overview

Critical fixes
PR 41581         Models affected:

                 Model 6517B

                 Symptom:

                 Noise and offsets may appear in lower current ranges when operating on 50Hz power due to
                 aperture being set incorrectly.

                 Resolution:

                 The issue has been repaired.

PR 41462         Models affected:

                 Model 6517B

                 Symptom:

                 Selecting scientific notation for the number format in charge measurements causes the unit to
                 stop responding.

                 Resolution:

                 The issue has been repaired.

Enhancements
No enhancements have been added in version A09.

A08 Release
Overview

Critical fixes
PR40525          Models affected:

                 Model 6517B

                 Symptom:

                 Resistance, current, and voltage measurements may be different from expected, due to relays
                 not latching properly.

                 Resolution:

                 The issue has been repaired.


6517B-801-FRP Rev. K / May13, 2010                                                                               3
A07 Release - Enhancements                                                          6517B Firmware Release Notes


Enhancements
No enhancements have been added in version A08.


A07 Release
Overview

Critical fixes
PR40525          Models affected:

                 Model 6517B

                 Symptom:

                 Resistance, voltage, and current measurements may be different from expected, due to relays
                 not latching properly.

                 Resolution:

                 The issue has been repaired.

Enhancements
No enhancements have been added in version A07.

A06 Release
Overview

Critical Fixes
PR39774          Models affected:

                 Model 6517B

                 Symptom:

                 Alternating polarity resistance test sequence delays are shorter than expected.

                 Resolution:

                 The issue has been repaired.

PR39898          Models affected:

                 Model 6517B

                 Symptom:

                 Voltage source output may be zero when turned on.

                 Resolution:

4                                                                            6517B-801-FRP Rev. K / May 13, 2010
6517B Firmware Release Notes



                 The issue has been repaired.


Enhancements

No enhancements have been added in version A06.




A05 Release
Overview

Critical fixes
PR38989          Models affected:

                 Model 6517B

                 Symptom:

                 The menu option to configure the buffer count to use the trigger model count causes the front
                 panel to stop responding.

                 Resolution:

                 The issue has been repaired.

PR39438          Models affected:

                 Model 6517B

                 Symptom:

                 Real-time clock can jump ahead by several hours.

                 Resolution:

                 The issue has been repaired.

PR39637          Models affected:

                 Model 6517B

                 Symptom:

                 Using the trigger count setting for the buffer count causes "Buffer Data Lost" errors if no readings
                 are stored.

                 Resolution:

                 The issue has been repaired.

PR39641          Models affected:
6517B-801-FRP Rev. K / May13, 2010                                                                                      5
A04 Release - Enhancements                                                             6517B Firmware Release Notes



                 Model 6517B

                 Symptom:

                 "Buffer Data Lost" errors occur if the buffer count is set, but no readings are stored.

                 Resolution:

                 The issue has been repaired.

PR39651          Models affected:

                 Model 6517B

                 Symptom:

                 Readings are noisier unless line sync is turned off and then turned back on.

                 Resolution:

                 The issue has been repaired.


Enhancements
No enhancements have been made in version A05.




A04 Release
Overview

Critical fixes
PR38548 Models affected:
          Model 6517B

          Symptom:
          Clock time reported incorrectly over GPIB and RS232 between 12M and 1PM.

          Resolution:
          The issue has been repaired.

Enhancements
No enhancements have been made in A04.




6                                                                              6517B-801-FRP Rev. K / May 13, 2010
6517B Firmware Release Notes



A03 Release
Overview

Critical fixes
PR37667 Models affected:
          Model 6517B

          Symptom:
          Hi limit settings do not allow values beyond 9.9E8.

          Resolution:
          The issue has been repaired.

PR37767 Models affected:
          Model 6517B

          Symptom:
          Buffer readings do not include temperature or humidity when retrieved over GPIB or RS232.

          Resolution:
          The issue has been repaired.

PR37871 Models affected:
          Model 6517B

          Symptom:
          Clock time does not change between AM and PM until 1:00.

          Resolution:
          The issue has been repaired.

PR37889 Models affected:
          Model 6517B

          Symptom:
          External feedback function does not work properly.

          Resolution:
          The issue has been repaired.

PR37907 Models affected:
          Model 6517B

          Symptom:
          Math percent function does not allow values over 5E9.


6517B-801-FRP Rev. K / May13, 2010                                                                    7
A03 Release - Critical fixes                                                         6517B Firmware Release Notes


           Resolution:
           The issue has been repaired.

PR38200 Models affected:
           Model 6517B

           Symptom:
           :sour:volt:range MIN|MAX|DEF arguments do not work properly.

           Resolution:
           The issue has been repaired.

PR38394 Models affected:
           Model 6517B

           Symptom:
           Self test query (*TST?) reports failure (1).

           Resolution:
           The issue has been repaired.

PR38397 Models affected:
           Model 6517B

           Symptom:
           Resistance measurements show different current values in resistance function when using the "next"
           display to show measured current, and in current function when measuring current.

           Resolution:
           The issue has been repaired.

PR38398 Models affected:
           Model 6517B

           Symptom:
           Pressing Local causes voltage source to go to zero volts, but output is still on. Sending a device clear
           command causes similar behavior.

           Resolution:
           The issue has been repaired.

PR38441 Models affected:
           Model 6517B

           Symptom:
           Can't change temperature units from degrees C.

           Resolution:

8                                                                            6517B-801-FRP Rev. K / May 13, 2010
6517B Firmware Release Notes


          The issue has been repaired.


Enhancements
No enhancements have been made in A03.

Non-critical fixes
PR38202 Models affected:
          Model 6517B

          Symptom:
          During calibration, the current offset is excessive if the current offset calibration steps fail.

          Resolution:
          The issue has been repaired.


Known issues
There are no known issues in A03.




6517B-801-FRP Rev. K / May13, 2010                                                                            9