Service Manuals, User Guides, Schematic Diagrams or docs for : Keithley 6522 6522RevA_DocSpec

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
6522RevA_DocSpec


>> Download 6522RevA_DocSpec documenatation <<

Text preview - extract from the document
Model 6522 Low Current/Voltage Scanner
CHANNELS PER CARD: 10.
FUNCTIONS: Volts, Amps.                                                             H              H
CONTACT CONFIGURATION:                                                                                 IN 1
                                                                                                   L




                                                                                        IN 2...9
  Voltage Mode: Single pole break-before-make for signal HI input. Guard        OUT L
                                                                                                   G
     is common for all 10 channels and output.
  Current Mode: Single pole break-before-make for signal HI input. Signal           G
     LO is common for all 10 channels and output. When a channel is off,                           H
     signal HI input is connected to signal LO via a relay. All channels con-                          IN 10
                                                                                                   L
     nect to signal LO in a make-before-break configuration between
     channel closures so that the current path is never interrupted.                               G
  All relays open when power is off.
SCAN SPEED:
  1.5 channels/second with default settings.1, 2
  43 channels/second maximum speed.1, 2, 3
  1 When used with 6517 Electrometer.

  2 Scan speeds are for fixed range readings for 200V and 20mA 6517

    ranges, into data buffer.
  3 6517 Filter off, speed fast, settling time 0 seconds.

CONNECTOR TYPE: Inputs: Triaxial. Outputs: Triaxial.
SIGNAL LEVEL: 200V, 500mA, 10VA (resistive load).
CONTACT LIFE: >106 closures at maximum signal level; >107 closures at
  low signal levels.
CONTACT RESISTANCE: <1.
CONTACT POTENTIAL: <200