Service Manuals, User Guides, Schematic Diagrams or docs for : Keithley 601 601A(Model601)

<< Back | Home

Most service manuals and schematics are PDF files, so You will need Adobre Acrobat Reader to view : Acrobat Download Some of the files are DjVu format. Readers and resources available here : DjVu Resources
For the compressed files, most common are zip and rar. Please, extract files with Your favorite compression software ( WinZip, WinRAR ... ) before viewing. If a document has multiple parts, You should download all, before extracting.
Good luck. Repair on Your own risk. Make sure You know what You are doing.




Image preview - the first page of the document
601A(Model601)


>> Download 601A(Model601) documenatation <<

Text preview - extract from the document
MODEL 601 ELECTROMETER                                                                                   CONTENTS



                                         TABLE OF CONTENTS


Section                                           Page    Section                                                Page

1.   GENERALDESCRIPTION .                            1          5-4.   Troubleshooting   . .   . .                 31
     l-l. General        . . .                       1          5-5.   Procedures to Guide
     1-2. Features         .                         1                   Troubleshooting   . . .   .               31
     1-3. Specifications     ,                       2
     l-4. Applications     . .                            6.    CALIBRATION . . .            . .   .   .           35
     1-5. Accessories    . .                        2           6-l.  General       . . , . . .    .   . . .       35
     1-6. Equipment Shipped                         4           6-2.  Calibration      Schedule    .   . .         35
                                                                6-3.  Grid Current Check .         .   . . .       35
2.   OPERATION .      . .    . .        . . .    5              6-4.  DC Amplifier      Balance
     2-l.  Front Panel Controls                                         Adjustment      . . . .    . . . .         36
              and Terminals . . . .       . .    5              6-5.  High-Megohm Resistor
     2-2.  Rear Panel Controls                                           Verification      . . .   . .    . .      36
              and Terminals .       . .   . .    6              6-6.  Meter Zero Calibration         .    .        36
     2-3.  Input Connections     . . .    . .    7              6-J.  Accuracy Check . . .         , ,    . .      37
     2-4.  Preliminary    Procedures      . .    8              6-8.  Drift Check . . . .          . .    . .      38
     2-5.  Voltage Measurements       .   . .    9
     2-6.  Off Ground Voltage                             7.    ACCESSORIES . .       . . . .    . .               43
             Measurements      . . . .    . . 11                J-l.  Model 6011 Input Cable . .                   43
     2-7.  Current Measurements       .   . . 12                J-2.  Model 1531 Gripping Probe             .      43
     2-8.  Resistance Measurements        . * 14                J-3.  Model 6301 Guarded Probe .            .      43
     2-9.  Charge Measurements . .        . . 17                7-4. Model 6012 Triaxial-to-
     2-10. Recorder Outputs      . . .    . . . 17                      Coaxial Adapter        . . .               43
     2-11. Unity Gain Output . . .        . . 18                7-5. Model 6102A Voltage Divider
                                                                        Probe , , . . , . , . .             .      44
3.   APPLICATIONS . . . . . . . . .         . *    21           7-6. Model 6103A Voltage Divider
     3-l.  General . . . : . . . . ,        . .    21                   Probe , .       . . . . , .         .      44
     3-2.  current source     , . . . .     . .    21           J-7.  Model 6104 Test Shield . .                   44
     3-3.  Static Charge Measurements         .    21           7-8. Model 6105 Resistivity
     3-4.  Capacitance Measurements         . .    21                   Adapter   . . . . . . . .           .~     45
     3-5.  current Integrator       . . .   . .    22           7-9. Models 2501 and 2503 Static
     3-6.  Potentiometric    Voltage                                    Detector Probes      , . . .        .     46
             Measurements     . . . . ,     . .    22           J-10. Rack Mounting     . . . , . .         .     46
     3-7. Measuring Diode                                       J-11. Model 6013 pH Electrode
             Characteristics     . . . .    , .    22                   Adapter       . . . . . . .         .     47
     3-8.  Peak-Reading Voltmeter       .   . .    23           J-12. MQdel 370 Recorder . . . .            .     4,8
                                                                                    \
4.   CIRCUIT DESCRIPTION . . . ,     .             25     8.    REPLACEABLEPARTS .        . . . . .         .      51
     4-l.  General . .     . . . .   .             25           8-1.  Replaceable Parts List . .            .      51
     4-2.  Voltmeter Opiration   ,   .             25           8-2.  How to Order Parts . . . .                   51
     4-3.  Voltmeter Circuit   . .   .             26                 Model 601 Replaceable
     4-4.  Ammeter Operation . .     .             27                   parts List . . . . . . .            ,      52
     4-5.  Ohmmeter Operation    .   .             28                 Schematic Diagram 194283              .      58
     4-6.  Coulombmeter Operation                  29
                                                         ,*    Change Notice    . . . . . . . . .last            page
5.   SERVICING . . . . . . .   .     .             31
     5-l.  General . . . . . . .     .             31     *    Yellow Change Notice sheet is included
     5-2.  Servicing  Schedule       .             31          only for instrument     modifications
     5-3.  Parts Replacement , .     ,             31          affecting  the instruction      manual.

0667R                                                                                                               i
GENERALDESCRIPTION                                                          MODEL 601 ELECTROMETER




                                                                        -




        XGURE 1.     Keithley   Instruments   Model 601 Electrometer.




ii                                                                                          9666
MODEL 601 ELECTROMETER                                                                 GENERALDESCRIPTION



                                SBCTION     1.   GENERAL        DESCRIPTION


l-l.    GENEXAL.

   a. The Keithley Model 601 Electrometer      is a versatile   battery  operated instrument which
measures wide ranges of dc voltages,    currents,   resistances   and charges.    It is a highly
refined dc vacuum tube voltmeter   that uses an electrometer      tube input to provide greater
than 1014 ohm input resistance.    The Model 601 has all the capabilities        of conventional
VTVMs, but Lt can also make many more measurements without loading circuits.

  b. The Electrometer      has nine volta e ranges from 0.001 volt full scale to 10 volts,
twenty-eight    current ranges from 10' l$ ampere full scale to 0.3 ampere, twenty-three   linear
resistance   ranges from 100 ohms full scale to 1013 ohms, and thirteen   charge ranges from
10-12 coulomb full scare to 10-6 coulomb.

        The Model     601 offers complete line isolation and excellent    off-ground    measuring capa-
bi;ity.    Up to    1500 volts may be applied between the input low terminal and the case, and
safe operation      is assured with the case grounded.   A three-terminal      input allows complete
guarding of the       high terminal.

   d. The Model 601 employs matched electrometer   input tubes followed by three differen-
tial transistor amplifier  stages and a transistor   output stage.   A large amount of nega-
tive feedback is used for stability  and accuracy.

l-2.    FEATURES.

   a.   Voltmeter    accuracy   is *l% of full   scale,   exclusive   of noise   and drift.

  b. Zero drift   of the Model 601 is 200 microvolts   per hour maximum averaged over any 24-
hour period after warm-up.    During the 2-hour warm-up, zero drift  is no more than 2 milli-
volts after the first   hour.

   c. Two amplifier   outputs are available.        A switch on the rear panel allows either
+l volt or fl milliampere     for full-scale    meter deflection.     The-current output is variable
25% with 1.4-kilohm   recorders.     The unity-gain     amplifier output is equal to the input
voltage within 50 ppm or 100 inicrovolts,       exclusive of zero drift.

   d. Current measurements can be made by one of two methods: the normal method in                which
&he current is determined by measuring the voltage drop across a resistor       shunting           the
oxput, or the fast method in which negative feedback is applied through the shunt                resis-
tor.    The former method minimizes noise.  The latter    method reduces the input drop           and
greatly   increases the response speed on the low-current    ranges.
   e. The lOOO-hour life of the batteries         enables "sage in Long-term experiments without
interruptions  for-recharging.      Battery life is maintained even when the l-milliampere      re-
corder output is used.      For further    convenience, battery   condition is readily  checked on
the panel meter,




056JR                                                                                                     1
GENERALDESCRIPTION                                                                                MODEL 601 ELECTROMETER



1-3.   SPECIFICATIONS.

AS A VOLTMETER:
RANGE: .OOl volt      full     scale     to 10 volts     in nine         lx and 3x ranges.

ACCURACY: ~1% of full          scale on all       ranges exclusive           of noise      and drift.

ZERO DRIFT: After l-hour warm-up no more than 2 millivolts  in the second hour,                                        and in
any subsequent 24-hour period, the average drift will not exceed 200 microvolts                                        per hour.

METER NOISE:   Cl0 microvolts            maximum with       input    shorted     on most sensitive           range.

INPUT IMPEDANCE: Greater than 1014 ohms shunted by 20 picofarads.                                  Input     resistance       may
also be selected in decade steps from 10 to 1011 ohms.

AS AN AMMETER:

RANGE: LO-14 ampere full              scale   to 0.3 ampere in twenty-eight                lx and 3x ranges.

ACCURACY: *z% of full scale on 0.3 to lo-11 ampere ranges using the smallest available
Multiplier Switch setting; 24% of full scale on 3 x lo-12 to lo-14 ampere ranges.

METER NOISE:   Less than i3 x 10'15 ampere.

GRID CURRENT: Less than 2 x LO-l4 ampere.
AS AN OHMMETER:

RANGE: 100 ohms full          scale     to 1013 ohms in twenty-three              linear     lx and 3x ranges.

ACCUR4CY: 23% of full          scale on 100 to 108 ohm ran es using the largest                            available      Multi-
plier Switch setting:          ;t5% of full scale on 3 x 108 to 1013 ohm ranges.

AS A COULOMBMETER:

RANGE: LO-12 coulomb full              scale to lo-6     coulomb in thirteen             lx and 3x ranges.

ACCURACY: i5% of full scale on all                ranges.        Drift     due to grid      current     does not exceed
2 x lo-14 coulomb per second.

AS AN AMPLIFIER:
INPUT IMPEDANCE: Greater than 1014 ohms shunted by 20 picofarads.                                  Input     resistance       may
also be selected in decade steps from 10 to 1011 ohms.

OUTPUTS: Unity-gain          output     and either     voltage      or current     recorder       output.

UNITY-GAIN OUTPUT: At dc, output is equal to input within 50 ppm or 100 microvolts,  ex-
clusive of zero drift. Up to 0.1 milliampere may be drawn. Output polarity   is same as
input polarity




                                                                                                                             0366R
MODEL 601 ELECTROMETER                                                                                            GENERALDESCRIPTION



VOLTAGE RECORDER. OUTPUT: +l volt                  for full-scale           input.        Internal       resistance      is 910 ohms.
Output polarity is opposite input                  polarity.

   Gain:    0.1,   0.33,     etc.     to 1000.

   Frequency Response (Within 3 db):  dc to 100 cps at a gain of 1000, rising                                            to 50 kc st
   a gain of 3.3, decreasing to 1 kc at a gain of 0.1.

   Noise:   Less than 2% rms of full                 scale     at a gain of 1000, decreasing                      to less than 0.5%
   at gains below 10.

CURRENTRECORDEROUTPUT: il                 milliampere          for     full-scale       input,       variable      +5% with    1400-ohm
recorders.

GENERAL:

ISOLATION: Input low to case:   greater than 1010 ohms shunted by 0.0015 microfarad.                                                    I*-
put low may be floated up to k1500 volts with respect to case.

POLARITY: Meter switch selects left-zero    (positive                               or negative)       or center-zero         scales.
Meter switch does not reverse polarity   of outputs.

CONNECTORS: Input:            Teflon-insulated triaxial  Dage 33050-l. Lo: binding post.   Voltage
or current output:           Amphenol 80-PC2F. Unity gain output, case ground:  binding posts.

BATTERY CHECK: Condition               of all     batteries      may be checked with                 front      panel controls.

BATTERIES: Four 2N6 (or 246, VS305 or NEDA 1602);                              three     TR286 (or E286);             two RM42R (or
E42N or HG42R). 1000 hours battery life.

DIMENSIONS, WEIGHT:           10-l/2     inches     high x 6-5/b           inches      wide x 10 inches            deep; net weight,
14 pounds.

ACCESSORIESSUPPLIED:                Model 6011 Input          Cable;     mating       output     connector.


l-4.    APPLICATIONS.

   a. Voltmeter applications     include directly    measuring potentials    across pH electrodes,
piezoelectric    crystals, capacitors,   electro-chemical    cells and biological    membranes.
The Model 601 is also useful as a null detector with potentiometers           or bridges in high im-
pedance applications.

   b. As a picoammeter the Model 601 can br used with photomultiplier   tubes, flame and
beta ray ionization  detectors, lithium ion drift detectors and gas chromatographs.   It is
also useful in nuclear studies , plasma physics and vacuum research.

   c. As an ohmmeter the Electrometer       is ideal for measuring insulation    resistance    and
resistor    voltage coefficients.    It is useful for measuring "in circuit"     resistances,
since the Model 601 low terminal       can be isolated  from case ground.    Volume and surface re-
sistivities     can be measured with the Model 601 and the Model 6105 Resistivity        Adapter.

   d.   In addition        to measuring         charge directly            other      coulombmeter           uses are measuring



0366R                                                                                                                                     3
GENERALDESCRIPTION                                                                  MODEL 601 ELECTROMETER



charge current over a period and obtaining  integral             curves of time varying         currents.          The
Electrometer  can also be used as a charge amplifier             to measure piezoelectric         crystal         out-
puts.

l-5.        ACCESSORIES. (Also   see Section   7).

      a.     Model 6011 Input Cable has 30 inches of low noise       triaxFa1     cable with      mating     triax-
ial        input connector and three alligator  clips.
       Model 6013 pH Electrode Adapter has a Z-foot cable and triaxial
      b.                                                                 connector and accepts
Leed & Northrop   and Beckman pH electrode  connectors. The Adapter allows accurate and con-
venient pH potential   measurements with the Model 601.

   c. Model 6301 is a high impedance guarded probe with a 3-foot connecting triaxial
cable that allows measurements to be made more conveniently.   The Model 6301 has an insu-
lation resistance  of over 1014 ohms. Using the Probe does not effect any Model 601
specifications.

  d. Model 1531 is a gripping     probe with a 3-foot connecting            triaxial     cable.         The Model
1531 has an insulation resistance     of over lOlo ohms.

   e. Model 4005 Rack Mounting Kit adapts the Model 601 to rack mounting 10-l/2 inches
high x 19 inches wide.  The Kit can accommodate two Model 601's side-by-side   by using an
additional  cover.

   f. Model 6012 Triaxial-to-Coaxial      Adapter permits using the Model 601 with all Keithley
electrometer accessories     having uhf type coaxial connectors.    These include the Models
6102A, 6103A, 6104, 6105, 2501 and 2503: Since circuit        low and case ground are connected
using the Adapter, the Model 601 should not be floated.

  g.  Models 6102A and 6103A voltage divider probes, described                  in Section     7, facilitate
measurements and extend the voltage range to 10 kilovolts.

   h. Model 6104 Test Shield is suitable              for resistance measurements with         either      2 or
3-terminal guarded connections, as well              as voltage and current tests.

  i. Model 6105 Resistivity   Adapter is a guarded test fixture  for measuring volume and
surface resistivities  of materials  when used with the Model 601 and the Keithley Model
240A High Voltage Supply.

  j.  Models 2501 and 2503 Static Detector Probes are capacitive  voltage dividers                          with      a
10,OOO:l ratio, when used with the Probe 3/8 inch from the charged surface.
  k. Model 370 Recorder is uniquely compatible with the Model 601 as well as other
Keithley microvoltmeters,      electrometers and picoammeters.      The recorder is a high quality
economical instrument     that maximizes the performance of the Model 601, and many other
Keithley  instruments,    even in the most critical   applications.

 l-6.  EQUIPMENTSHIPPED. The Model 601 Electrometer     is factory calibrated   and shipped
 with all components in place.   The shipping carton also contains the Instruction    Manual,
'Model 6011 Input Cable and mating output plug.




4                                                                                                              0667R
MODEL 601 ELECTROMETER                                                                     OPERATION



                                       SECTION 2.      OPERATION


2-l.    FRONT PANEL CONTROLSAND TERMINALS (See Figure        2)

   a. Range Switch.   The Range Switch selects the measuring mode and the range.            It is
divided into a VOLTS position,    11 AMPERESranges, eight OHMSranges and four COULOMBS
ranges.  A line above the dial skirt indicates     the range used. Full-scale       sensitivity
for all ranges is the Range Switch setting    times the Multiplier     Switch setting.      The 10 or
3 of the top meter scale corresponds to the full-scale      deflection   for the range selected;
for example, on the l-volt   range, the needle is at 10 for a l-volt       input.

  b. Multiplier     Switch.  The Multiplier       Switch determines the voltage sensitivity     of the
dc amplifier    and sets the full-scale     voltage range when the Range Switch is set to       VOLTS.
The Multiplier    Switch may also be used to multiply        the AMPERES(3x maximum setting     above
10a3), OHMSand COULOMBS     ranges on the Range Switch.         A line above the dial skirt   indicates
the range used.

   c. METER Switch.     The Switch has six positions:       the spring-returned   BATTERY CHECK
position  allows checking of the battery condition       with the Multiplier    Switch. POWER  OFF
shuts off the instrument.      OFF disconnects   only the meter during recorder operation.      The
+ and - positions   determine the polarity     of the meter.    CENTER ZERO sets the instrument
for center zero operation    (lower meter scale).

   d. ZERO Controls.    Two ZERO Controls are
on the front panel:    a MEDIUM Switch (outer
knob) and a lo-turn   FINE potentiometer
(center knob).    These allow precise meter
zeroing.

   e. ZERO CHECK Button.      Depressing the
Button effectively     removes all input signal
from the instrument by shunting the input
and amplifier    through 10 megohms. This
allows meter zeroing on any range.       The
Button is locked in the zero check position
when the line is horizontal.

   f.   FEEDBACKSwitch,      The FAST and NORMAL
positions   of the Switch determine the feed-
back connections within the instrument.
With the Switch at FAST, current measure-
ments are made with the range resistors       in
the feedback network; this results      in lower
input voltage drops and faster response
speeds.    The FAST position    is also used for
coulomb measurements, and~to increase re-
sponse speed. When the Switch is in NORMAL,
the range resistors    shunt the input.
                                                       FIGURE 2. Model 601 Front Panel Controls
   g. INPUT Receptacle.        The INPUT Recepta-      and Terminals.  Circuit  designations refer
cle is a Teflon-insulated       triaxial    type       to Replaceable Parts List and the Schematic
connector.   Its center     terminal     is the cir-   Diagram.


0666R                                                                                                5
OPERATION                                                                         MODEL 601 ELECTROMETER



cuit high; the inner shield is circuit        low
(circuit    ground); the outer shield is case
ground.     (See Figure 4.)    A shield cap is
provided.      The LO Terminal, below the
Receptacle,     is connected to circuit  low;
it is connected to case ground only if the
shorting    link on the rear panel is connect-
ed.

2-2.   REAR PANEL CONTROLSAND TERMINALS
       (See Figure 3).

   a. COARSEZERO SWITCH. The COARSEZERO
Switch has 11 positions  to extend the zero-
ing capability of the front panel ZERO Con-
trols.

   d. OUTPUT Switch.        The Switch is a two-
position  slide switch for the output.         In
the 1 MA position,     the instrument will
drive l-milliampere      recorders.   In the 1 V
position,    the output is 1 volt for full-
scale meter deflection.        Source resistance
is approximately     1090 ohms.

   C. 1 MA CAL Control.   The Control varies            `LtiUKli 3. Model   bUL    Kear   hnel   COntrols
the output from 0.95 to 1.05 milliampere               and Terminals.
for 1400-ohm recorders,  so the recorder
scale will correspond with the Electrometer
meter.

   d. OUTPUTReceptacle.        A 2-terminal    microphone-type  receptacle  provides 1 volt or 1 mil-
liampere for full-scale      meter deflection.      Pin No. 2 is at circuit  low when the FEEDBACK
Switch is at NORMAL. Neither terminal is at ground when the FEEDBACKSwitch is at FAST.
Both terminals.are    isolated    from case ground.

   e. Xl OUTPUTand OHMSGUARDTerminals.       The potential   between the Xl OUTPUTTerminal
and the OHMSGUARDTerminal (circuit     low when the FEEDBACKSwitch is in NORMAL) is equal
to the input voltage with 0.005% linearity   or 100 microvolts.     When the FEEDBACKSwitch
is at FAST, the Xl OUTPUTTerminal is at circuit    low and the OHMSGUARDTerminal is float-
ing.
  f.   LO Terminal. A black terminal allows connection to the input                low connection.          Con-
necting the LO and CASE GROUND  Terminals puts both at case ground.

   g. CASE GROUND  Terminal.  A blue termitial is connected directly              to the outside cabinet
of the Model 601 and the outside shell of the input connector.       It           is connected to nothing
else within the instrument.

                                                    NOTE
       If the Model 601 will be stored for a long period, remove batteries.   Also, make
       sure the METER Switch is at POWER  OFF when the instrument is not being used.


6                                                                                                           0666R
MODEL 601 ELECTROMETER                                                                            OPERATION


7
                           Lead                               Circuit                JlOl   Terminal

            heavy wire with red clip cover             Input (circuit)   high         center
            thin wire with black clip cover            Input (circuit)   low          Inner shield
            thin wire with blue clip cover             Case Ground                    Outer shield

TABLE 1.        Color Coding of Alligator    Clips   for Model 6011 Input   Cable.


    2-3.    INPUT CONNECTIONS.

  a. The accessories          described in Section 7 are designed to increase the accuracy and
convenience of input         connections.  Use them to gain the maximum capability of the Model
601.

   b. The Mo4el 6011 Input Cable, supplied with the instrument,        facilitates         input connec-
tions.    Table 1 contains the color coding of the alligator    clips.      The high terminal is
shielded by the inner braid of the triaxial    cable up to the miniature         alligator      clip. If
the unshielded   clip causes pick up from near-by electric   fields,     remove it and connect
the shielded lead directly    to the source.
                                                     NOTE
           Techniques and applications   are thoroughly     discussed in the brochure,   Electro-
           meter Measurements, by Joseph F. Keithley.        It is available from Keithley Instru-
           ments, Inc., or its representatives.

   C. Carefully  shield the input connection and the source being measured, since power
line frequencies   are well within the pass band of the Electrometer.    Unless the shiefding
is thorough, any alteration    in the electrostatic field near the input circuitry  will cause
definite  meter disturbances.

    d. Use high resistance,       low-loss materials        - such as Teflon (recommended), polyethy-
lene or polystyrene      - for insulation.          The insulation   leakage resistance     of test fixtures
and leads should be several orders magnitude higher than the internal                 resistance    of the
SOlL?ZCe.    Excessive leakage reduces the accuracy of readings from high impedance sources.
Triaxial     or coaxial cables used should be a low-noise type which employ a graphite                or
other conductive coating between the dielectric               and the surrounding   shield braid.     Amphen-
ol-Borg Electronics      Corporation,     Microdot,    Inc., and Simplex Wire and Cable Company make
satisfactory     types.   Using the supplied Model 6011 Input Cable is j simple way to insure
good input connections,

                                                     NOTE

           Clean, dry connections     and cables are very important     to maintain the value of all
           insulation   materials.    Even the best insulation    can be compromised by dust, dirt,
           solder flux, films of oil or water vapor.         A good cleaning agent is methyl alcohol,
           which dissolves most common dirt without chemically         attacking   the insulation.  Wash
           salt solutions   with distilled    water before using the cleaning agent.

   e. Any change in the capacitance       of the measuring circuit    to low will cause extraneous
disturbances.      Make the measuring setup as rigid as possible,      and tie down connecting
cables to prevent their movement. A continuous vibration           may appear at the output as a
sinusoidal    signal,  and other precautions   may be necessary to isolate     the instrument  and

0666R                                                                                                      7
OPERATION                                                                            MODEL 601 ELECTROMETER



the connecting    cable    from the vibration.

   f.   For low impedance measurements -
below 108 ohms or above 10m8 ampere - un-
shielded leads may be used. However, keep
the leads short.

   g. When measuring currents    LO-l4 ampere
or less with the FEEDBACKSwitch at FAST,
sOme insulators  - such as Teflon - may
produce random signals which show up as
erratic  meter deflections.   Insulation
used in the Model 601 is carefully      selected
to minimize these signals.                               L
                                                        FIGURE 4. Model 601 Triaxial    Incut Receo
   h. It is usual,ly better to connect the              tac1e.   The center terminal is circuit  0;
Model 601 to the circuit      only when a reading       input high; the inner shield is circuit
is being made. In some cases, the grid cur-             or input low; the outer shield is case
rent can charge the external test circuitry.            ground.
One example of this occurs when measuring a
capacitor's  leakage   resistance     by observing
the decay of the terminal voltage.          If the leakage current is less than the grid current,
there may be no decay.of the terminal voltage when the Electrometer            is left connected
across the capacitor's     terminals.

                                                     NOTE

       Keep the shielded     cap on the INPUT Receptacle       when the Electrometer        is not in
       a circuit.

  i.   The Model 6012 Triaxial to Coaxial Adapter            enables using coaxial cables and acces-
sories with the Model 601 by adapting the triaxial             INPUT connector to the uhf coaxial type.

                                                    NOTE

       The Adapter connects circuit   low to case ground.   Do not float the Model 601
       when using the Adapter.    The instrument chassis will be at the same potential
       as the input low.
2-4.    PRELIMINARY PROCEDURES.

    a.   Check battery    condition    by setting
the METER Switch to the BATTERY CHECK po-
                                                         I Multiplier    position           Battery    Checked I
sition.     Rotate the Multiplier       Switch                    .OOl                        B201    & B202
through the .OOl to 0.1 positions,           and ob-              .003                        B203
serve the meter readings.           The meter should              .Ol                         B204    & B205
read one-half of full scale or mwre for each                      .03                         B206    & B207
Multiplier    Switch position.        Table 2 shows             0.1                           R208    h 732n9
the batteries     checked by position.        If the
reading for any battery        is below half             TABLE 2. Multiplier          Switch Positions     for
scale, replace all batt'eries.          Note new bat-    Checking Condition         of Batteries
teries may cause the Model 601 to drift           more
than normal for at least 72 hours due to
change in battery      terminal voltage.


 8                                                                                                        0766R
MODEL 601 ELECTROMETER                                                                                  OPERATION



  b.    Set the controls    as follows:

                      METER Switch                           POWER OFF
                      Range Switch                           VOLTS
                      Multiplier   Switch                    1
                      FEEDBACK Switch                        NORMAL
                      ZERO CHECK Button                      LOCK

  c. Turn the METER Switch to CENTER ZERO. Within ten seconds, the meter needle should
come to the center zero position.   If not, adjust to meter zero with the MEDIUM and FINE
ZERO Controls.   Normally, there is no need to use the COARSEZERO Switch.

  d. After a few moments increase the voltage sensitivity by advancing                        the Multiplier
Switch to .3, .l, etc. Continue zeroing with the FINE ZERO Control.

   e. After long periods of storage or after a" overload,   the Model 601 may drift    exces-
sively.   The electrometer tubes are shock mounted; however, a severe jolt to the Electro-
meter may cause a zero offset.   This is corrected with the Zero Controls.   Drifting,
though, can occur for several hours.
                                                      NUTE

       If the Model 601 has been stored for some time, the grid current will exceed the
       specification   when first
                       ---I         used then decrease to below the specified  amount after
       one or two hours of use. This is a" inherent      characteristic of the electrometer
       tube; the instrument     is not faulty.

   f. Although the grid current of the Electrometer   is much below that found in conven-
tional voltmeters,   it can be observed on the meter.   A small voltage results   from the
grid current charging the input capacitance,   and the Electrometer    appears to drift when
the input is open. Use the ZERO CHECK Button to discharge the build-up.

    g. Follow the particular    procedures in paragraph 2-5 to 2-9 for measuring voltage,    cur-
rent, resistance    and charge.   When using Multiplier Switch settings   --     3 and 1 in the
                                                                          of 10, -----
voltage,   current,  resistance  and charge measuring modes, make sure the Output Switch is
                                                              ---
----- to IV if the output --- notconnected
set                          is              to a load.  Otherwise,   the meter shows a loading
effect.    When the output is connected to a load, this effect is not present.

                                                      NOTE

       Using the center    zero scales    decreases     accuracy    0.5% because     Lhe scale     spa" is
       shorter.

2-5.    VOLTAGEMEASUREMENTS.

  a.    The Model 601 can be used to measure voltages              several   ways.
     1. In the normal method - FEEDBACK Switch at NORMAL - the unknown voltage                            is co"-
  netted to the INPUT Receptacle. Input impedance is iO-l4 ohms, 20 pf.

     2. To reduce the slowing effects of input capacity,                 use the fast       method to measure
  the voltage.   A guarded circuit  is possible this way.

     3. To measure Low impedance sources,             the Model 601 input      resistance      can be decreased
  if desired.

0667R                                                                                                               9
OPERATION                                                                                      MODEL 601 ELECTROMETER



       4,.     Accessory     probes extend the Model 601's         range to 10 kilovolts.
                                                           NOTE

        Locking the ZERO CHECK Switch places 10 megohms acro.ss input high and low, which
        may temporarily cause instability  in some types of high impedance sources.

  b.         Normal Method Voltage         Measurements.

       1.      Follow   the instructions       of paragraph   Z-4..     Set the controls        as follows:

                                  METER Switch                     CENTER ZERO
                                  Range Switch                     VOLTS
                                  Multiplier   Switch              10
                                  FEEDBACK Switch                  NORMAL
                                  ZERO CHECK Button                LOCK

       2. Connect the unknown voltage to the INPUT Receptacle;    unlock                        the ZERO CHECK Button.
     Set the METER Switch to + or -, as necessary.   Increase sensitivity                         with the Multiplier
     Switch.  Recheck zero setting  after increasing sensitivity.

       3.      For off-ground      measurements,     see paragraph      2-6.

  C. Fast Method Voltage Measurements.   This method reduces                          the slowing     effects    of input
capacity and allows guarded voltage measurements.

       1.      Follow   the instructions       of paragraph   Z-4,.     Set the controls        as follows:

                                  METER Switch                     CENTER ZERO
                                  Range Switch                     VOLTS
                                  Multiplier   Switch              10
                                  FEEDBACK Switch                  FAST
                                  ZERO CHECK                       LOCK

       2.      Connect CASE GROUNDTerminal           to OHMSGUARDTerminal,            using    the shorting     link.

     3. Connect the unknown voltage t.o the high (center)     terminal of the INPUT Receptacle
  and to the CASE GROUND   Terminal on the rear panel.    Use the LO Terminal as a guard be-
  tween circuit  high and low. Unlock the ZERO CHECK Button.        Set the METER Switch to +
  or -, as necessary.   Increase sensitivity   with the Multiplier    Switch.  Recheck zero set-
  ting after increasing   sensitivity.

       4..     To make off      ground voltage     measurements,      see paragraph     2-6.

   d. Low Impedance Measurements.     To decrease input resistance,      set the Range Switch to
one of the AMPERESranges.     The input resistance     is now the reciprocal    of the current
range.   For instance, to obtain a" input resistance      of 107 ohms, set the Range Switch to
the 10s7 AMPERESrange.    Set the full-scale    voltage range with the Multiplier      Switch.
Operating procedures are the same as subparagraph b. At lower input resistances,            the
Model 601 will not run off scale as easily in the presence of excessive ac fields.             This
occurs only when the input is left open.




10                                                                                                                      0567R
MODEL 601 ELECTROMETER                                                                       OPERATION


   e. To measure sources more than 10 volts,        use one of two divider   probes.   The Model
6102A 1O:l Divider Probe extends the Model 601's range to 100 volts;          overall accuracy is
?3% and input resistance       is lOLo ohms. The Model 6103A 1OOO:l Divider Probe extends the
Model 601's range to 10 kilovolts;        overall accuracy is f5% and input resistance    is 1012
ohms. Follow the same operating        procedures with the dividers   as in subparagraph b. The
Model 6012 Triaxial-to-Coaxial       Adapter must be used with the Models 6102A and 6103A
Divider Probes.     Note, however, using the Adapter connects circuit      low to case ground; do
not float the Electrometer.        The full-scale  voltage range is the divider ratio times the
Multiplier  Switch setting.



        Operating the Model 601 nwre than 1500 volts off ground may permanently damage
        the instrument.    Isolation      between circuit    low and ground may break down some-
        where Fn the circuit,      putting    the case at an off-ground    potential. Since these
        breakdowns are very difficult         to locate,  it might not be possible to float the
        instrument  safely again.

2-6.     OFF GROUNDVOLTAGEMEASUREMENTS.

   a. The Model 601 can measure an unknown voltage whose low is up to             1500 volts off
ground while its own case is at ground.    This allows safe operation  of          the Electrometer.
Its operation  is the same as given in paragraph 2-5, except for input            connections  and
some added cautions.   These differ,  depending upon the FEEDBACKSwitch             setting.

     1. FEEDBACKSwitch set to NORMAL. Disconnect the shorting       link between the LO and
  CASE GROUND  Terminals on the rear panel.     Make sure the Model 601 case is securely con-
  nected to an earth ground, and that the low of the unknown voltage is less than 1500
  volts dff ground.     Connect the unknown voltage directly  to the INPUT Receptacle.   Opel-~
  ate the Model 601 as described for normal method voltage measurements.
     2. FEEDBACKSwitch set to FAST. Do not use the shorting       link. Make sure the Model
  601 case is securely connected to an earth ground, and that the low of the unknown vol-
  tage is less than 1500 volts off ground.     Ground the outer shield of the INPUT Receptacle.
  Connect the high of the unknown voltage to the canter terminal of the INPUT Receptacle.
  Connect the low to the GUARDTerminal.      Use the inner shield of the INPUT Receptacle as
  a guard.   Operate the Model 601 as described for fast method voltage measurements.

                                                   NOTE
        When the Model 601 is off ground, make sure the shell of a mating plug to the
        OUTPUT Receptacle is not connected to either pin in the Receptacle.
  b. If the Model 6012 Adapter is used, do not float the Model 601. The Adapter connects
the input low to the case ground, so that the Model 601 chassis and controls are at the
same potential  as the Low of the unknown source.

                                                   NOTE
        Use only an insulated   blade screwdriver    to adjust the COARSEZERO Switch and 1 MA
        CAL Cdntrol when floating    the Model 601. An ordinary    screwdriver could short the
        circuit  low to case ground, creating     a shock hazard and damaging the external cir-
        cuitry.



06668                                                                                                  II
OPERATION                                                                      MODEL 601 ELECTROMETER




            r -------------                  1
                                                             r----------`--

                                   R
            I                                    I           I
                                                                                  Rin
            I EE                                 I           I
                                                             I                  Imeter
            I
                                                 I
                                                             I
            I




FIGURE 5. Annneter Circuit   Loading.   Current sources may be considered a voltage (E) in
sarias with a resistance   (R).   The current with no annneter attached is I=E/R.         When an
ammeter is attached to measure the current,      the effective   input resistance    of the ammeter
Rin is in series with the source resistance      (R).   The current in the complete circuit       is
 less and Imeter = E/(R + Rtn).    If the effective    ammeter input resistance     is small com-
pared to R, Imeter+ - I and the error introduced      by circuit   loading is negligible.



2-7.    CURRENTMEASUREMENTS.
   a.   The Model 601 can measure currants           three       ways.

      1. In the normal method - used on any range - the currant is determined by measur-
   ing the voltage drop across a resistor   shunting the amplifier  input. This method is
   useful when lower noise is more important   than faster response speeds or if some damp-
   ing is needed.

      2.  In the fast method - for use only             below the 10V5 ampere range - the shunt resis-
   tor is between the amplifier   output and          input in the feedback loop.    This circuit
   largely neutralizes  the effect of input           capacity and greatly  increases the response
   speed. Also, the input voltage drop is             reduced to a maximum of one millivolt     on any
   range.

      3.   For galvanometric    current measurements, the Model 601 acts as a null indicator
   between a very accurate current source and the unknown current source.         Its off ground
   operating   capability    makes it ideal for this application since the reference source and
   unknown may both have a common grounded terminal.

   b. Rise tima varies primarily    with the current range, the input capacity and the method
used. On most ranges, the rise time in the fast mode is less than one second with 50
picofarads   across the input.   Even with much larger shunt capacities,     the negative feed-
back maintains a short rise time.      Given a choice, it is better to placg the Electrometer
nearer to the current source than to the data reading instrument.        Transmitting    the input
signal through long cables greatly     decreases the response speed and increases noise due
to the cable capacitance.
   C. To measure from a source with both terminals  off ground in either method, remove
the link between the LO and CASE GROUND  Terminals on the rear panel.   Connect the unknown
current to the INPUT .Receptacle.  The source must be less than *1500 volts off ground
(see paragraph 2-6).

12                                                                                               0766R
MODEL 601 ELECTROMETER                                                                                 OPERATION



  d.        Normal Method (0.3     to lo-l4     ampere ranges).

       1.     Follow   the instructions       of paragraph    2-4.    Set the controls   as follows:

                             METER Switch                            CENTER ZERO
                             Range Switch                            10-l AMPERES
                             Multiplier   Switch                     1
                             FEEDBACK Switch                         NORMAL
                             ZERO CHECK Button                       LOCK
  Connect the unknown current to the INPUT Receptacle and unlock the ZERO CHECK Button.
  Set the METER Switch to + or -, as necessary.       Increase the sensitivity  with the Range
  Switch and the Multiplier     Switch.  Do not set the Multiplier  Switch higher than 3 for
  Range Switch settings     10m3 and above. Check zero with the ZERO CHECK Button.

     2. Full-scale   current range is the settings    of the Range Switch times the Multiplier
  Switch.    Use the smallest Multiplier  Switch setting    possible to minimize input voltage
  drop and obtain the best accuracy.     The input Fesistor    varies with the Range Switch set-
  ting, from 10 ohms at 10-l AMPERES to loll ohms for lo-11 AMPERES. Input voltage drop
  is the percentage of full scale that the meter reads times the Multiplier        Switch setting.

                                                         NOTE

        On the low current ranges, balance cut the grid current with the Zero Controls
        or subtract the value from the reading.  To find the aawunt of grid current,
        cap the INPUT Receptacle and read the meter.

  e.        Fast Method (ranges     below lOa      ampere).

       1.     Follow   the instructions       of paragraph    2-4.    Set the controls   as follows:

                             METER Switch                            CENTER ZERO
                             Range Switch                            10-6 AMPERES
                             Multiplier   Switch                     1
                             FEEDBACK Switch                         FAST
                             ZERO CHECK Button                       LOCK

  Connect the unknown source to the INPUT Receptacle and unlock the ZERO CHECK Button.
  Set the METER Switch to + or -, as necessary.   Increase the sensitivity with the
  Range Switch and the Multiplier  Switch.  Do not set the Range Switch to 10e5 AMPERESor
  higher.  Check zero with the ZERO CHECK Button.

                                                         NOTE

        Use only the ZERO CHECK Button to check zero for the fast method.                    Do not short
        the input, because this will remove the feedback from the circuit.

     2. The full-scale     current range is the Range Switch setting        times the Multiplier
  Switch setting.      When selecting  the Multiplier    Switch setting,    remember small settings
  permit lower current source resistance,        and larger settings     improve instrument    zero
  stability.    Check the caution in subparagraph 3a below.




066711                                                                                                      13
OPERATION                                                                    MODEL 601 ELECTROMETER



     3. With the fast method,      the input   drop is reduced and the response     speed is in-
  creased at least 100 times.       However,   follow these precautions:

          4 The internal    impedance of the unknown current scnxce should not be less than
       0.1 of the value of the feedback resistor    being used. otherwise,     adequate feedback
       voltage cannot be developed at the input, and zero instability      results.    The feedback
       resistor  value is the reciprocal   of the AMPERESrange of the Range Switch.       Also, the
       voltage drop across the current source should be at least 100 times the voltage drop
       acrOsS the Model 601.
         b) The low side (Pin No. 2) of the OUTPUT Receptacle is no          longer connected to the
       low side of the INPUT Receptacle.   Therefore,   do not allow the       low side of a recorder
       to be electrically  connected, such as through the ground lead        of a power cord, to the
       low side of the current source.   Another alternative   is "sing      the unity-gain  output.
       (See paragraph Z-11.)
        c) Do not use the fast method to measure capacitance      unless "sing a very stable vol-
    tage supply.     Connecting a capacitor  to the input changes the circuit    to a differentia-
.   to;,   resulting  in extreme sensitivity   to very small voltage transients.
                                                   I
  f. Galvanometric     Method.

    1. Operate the Model 601 as a pico-
  ammeter in the fast method of operation.
  Use an accurate reference  current source
  to buck out the unknown current source.
  Connect as shown in Figure 6.

     2. Set the METER Switch to CENTER
  ZERO and "se the higher current ranges.             FIGURE 6. Measuring Current by the Galvan-
  Adjust the buckout current to indicate              ometric Method.       Use an accurate reference
  null on the Model 601. Increase the                 current source to buck Out the unknown cur-
  Electrometer's    sensitivity  as needed.           rent source, I,.       The Model 601, on its
  When the Model 601 is as close to null              current ranges, serves as a null detector.
  as possible,   the known reference    cur-          Use a uhf-tee fitting      and Model 6012 Adapter
  rent source equals the unknown source               at the Model 601 input.       Connect the Elec-
  * the Model 601 current readings.                   trometer to the two sources with coaxial
                                                      cable.    Select cable carefully     for very low
2-S.     RESISTANCE MEASUREMENTS.                     currents    (see paragraph Z-3).     For off
                                                      ground measurements, "se triaxial       cable and
  a. The Model 601 can measure resistan-              connectors,     grounding the outer shield.
ces by three methods.
      1. In the normal or two-terminal    method (ammeter-voltmeter),   the Electrometer  measures
  the voltage drop across the unknown sample as a known, constant current flows through
  it.    The voltage drop is proportional   to the resistance   of the sample.   This method is
  the simplest for the 100 to loLL ohm ranges.

    2. Above 10L1 ohms or to prevent leakage, the guarded method is better.       It results
  in faster response speeds and also nullifies   leakage errors acress the Electrometer      in-
  Put I since the potential acress the input terminal   is small.

    3.     In the preceding methods, the voltage across the sample cannot be arbitrarily
  set.     In some cases, as in measuring capacitor  leakage, these methods involve much more

                                                                        ,a
14                                                                                                 06678
MODEL 601 ELECTROMETER                                                                                             OPERATION



 time than if a larger voltage could be applied.   In the external voltage method the
 Model 601 is used as a fast picoammeter.  The unknown resistance    sample is connected to
 an external known voltage source and the current through the sample is measured.      Either
 the normal or fast method may be used. The resistance   is calculated   from the readings.

                                                              NOTE

       Discharge any capacitor  before removing it                          from the circuit.     Depressing the
       ZERO CHECK Button shorts the input through                           a lo-megohm resistor,    providing  a
       discharge path.


 b.        Normal Method (100 to 10"                ohm ranges).

      1.     Follow   the instructions         of paragraph          2-4.      Set the controls      as follows:

                               METER Switch                                   +
                               Range Switch                                   105 OHMS
                               Multiplier   Switch                            1
                               FEEDBACK Switch                                NORMAL
                               ZERO CHECK Button                              LOCK

 Connect the resistance  sample to the INPUT Receptacle.                               Unlock     the ZERO CHECK Button.
 Check zero with only the ZERO CHECK Button.

                                                              NOTE

       Do not open circuit the Electrometer  on the OHMS ranges;                           the input will develop
       up to 10 volts due to its constant current characteristic.                             Keep the input
       shorted or the ZERO CHECK Button locked.

   2. The full-scale  ohms range is the Range Switch                              setting  times the Multiplier  Switch
 setting. Use the largest Multiplier   Switch setting                             possible   to obtain the best accuracy.

    3. Before making a final reading, manipulate the Multiplier      and Range Switches, so
 the sample is tested at a number of test potentials.   The applied test voltage is the
  percentage of full scale that the meter reads times the Multiplier     Switch setting.

      4.     When the test       current     is applied,      high     termipal     of the INPUT Receptacle        is negative.

                                                              NOTE

                      Shield     the input     if     the resistance          sample exceeds LO8 ohms.

 c.        Guarded Method (to        1014 ohm ranges).

      1.     Follow   the instructions         of paragraph        2-4.        Set the controls      as follows:
                               METER Switch
                               Range Switch                                   loll OHMS
                               Multiplier   Switch                            1
                               FEEDBACK Switch                                FAST
                               ZERO CHECK Button                              LOCK



0667R                                                                                                                      15
OPERATION                                                                                 MODEL 601 ELECTROMETER


     Connect the low impedance side of the resistance  sample to the Model 601 GUARDTerminal,
     and the high impedance side to the center terminal of the INPUT Receptacle.  Unlock the
     ZERO CHECK Button.

          2.     Read the resistance     as outlined   for   the normal method,       subparagraph   b.

       3. The low terminal of the INPUT Receptacle is now a driven guard.    It may be used
     to minimize the slowing effects of capacity between high and low and errors due to leak-
     age resistance between high and low.

        4. The Model 6011 Input Cable, supplied with the Model 601, provides a convenient
     means of making guarded resistance measurements.   Connect the shorting  link between the
     CASE GROUNDand GUARDTerminals on the rear panel.     This allows the CASE GROUNDor blue
     test lead terminal to be connected to the low impedance side `of the unknown resistance.
     The inner shield or the black test clip is the GUARDTerminal.

     d.        External   Voltage   Method (to   1017 ohms).

        1. Turn the ZERO CHECK Switch to LOCK. Connect the sample between the High Terminal
     of the INPUT Receptacle and the power supply.      (See Figure 7,.) Put a switch in the
     high voltage line to ground the low impedance end of the sample when it is disconneCtad
     from the potential.    Before disconnecting,   make sure to lock the ZERO CHECK Button.  If
     the Power Supply must be floating,    remove the link between the CASE GROUNDand LO Ter-
     minals and connect the CASE GROUND    Terminal to an earth ground.

        2. Set the FEEDBACKSwitch to NORMAL. Usually this method is best, since                            instabilit-
     ies can arise for resistance samples less than 0.1 the value of the feedback                          resistor.

       3. Apply a potential to the sample before releasing      the ZERO CHECK Button.  Set the
     Range Switch to .3 AMPERESand increase sensitivity    until a reading is obtained.

    4. If the potential      applied is at                     I                                                         I
 least 100 times the full-scale        input
 drop (Multiplier     Switch setting),     the



                                                             ,I r-
 resistance    is equal to the applied poten-
 tial divided by the current reading.           The
 high voltage sensitivity      of the Model 601,
 therefore,    permits external.voltages       of                                 I       -               601    1
 0.1 volt or more to be used.

    5. If the potential  applied is less
 &han 100 times the input drop, the resis-
 tance is equal to the difference   between
 the applied potential  and the input drop,
 all divided by the current reading.
                                                               FIGURE 7. Measuring Resistance by the Ex-
    6. If.the     current is read by the fast                  ternal Voltage Method.      A potential    from a
 method, the input drop is so slight     that                  known source, V, is applied to the unknown
 it need not be included in the calcula-                       resistance   sample, R,. The Model 601 mea-
 tion.     If the capacity shunted across the                  sures the current through s, from which
 sample is large, such as encountered in                       the resistance   is calculated.      Switch S
 capacitor    leakage measurements, the fast                   grounds R when no potential       is applied.
 method increases response speed and this                      Note in ai2eve figure the power supply is
 connection     is recommended.                                floating.


16
MODEL 601 ELECTROMETER                                                                         OPERATION



2-9.     CHARGEMEASUREMENTS.

  a.     Follow   the instructions    of paragraph    2-4,.   Set the controls   as follows:

                          METER Switch                         CENTER ZERO
                          Range Switch                        10-7 COULOMBS
                          Multiplier   Switch                 .Ol
                          FEEDBACK Switch                     FAST
                          ZERO CHECK Button                   LOCK

Unlock the ZERO CHECK Button and then connect the unknown source to the INPUT Receptacle.
If the Electrometer  reads off scale, increase the Multiplier        Switch setting.      If the sen-
sitivity  is not enough, decrease the Multiplier      Switch setting     until the reading is on
scale.   Changing the Multiplier  Switch setting    does not affect the transfer        of charge
from the unknown source to the instrument.       If increasing    sensitivity    with the Multiplier
Switch does not bring the reading on scale, increase sensitivity            with the Range Switch
and repeat the above steps.

  b. The full-scale  charge range is the Range Switch setting    times the Multiplier            Switch
setting. Grid current contributes   2 x 10-14. coulomb per second maximum.

                                                     NOTE

        Because of the instrument's  RC time constant, wait 20 seconds after discharginlg
        internal capacitance  on the LOS7 coulomb range before making another measurement.
        On the 10-S coulomb range, wait at least two seconds.

2-10.     RECORDEROUTPUTS.
   a. For recording      with  the Model 601, "se the Keithley   Model 370 Recorder for ease, econ-
omy, versatility      and performance.     The Model 370 is a pen recorder with 10 chart speeds and
1% linearity.       It can float up to l 500 volts off ground.    The Model 370's input cable has a
connector which mates directly         with the OUTPUT Connector on the Model 601; this avoids
interface     problems often encountered between a measuring instrument     and a recorder.    The
Model 601 OUTPUT, when set to IMA Position,         will drive the 370; no preamplifier   is needed.

  b. Other recorders,   oscilloscopes       and similar    instruments      can be used with the Model
601. The Model 601 has two variable         outputs,    fl volt and *l milliampere,      to amplify sig-
nals within l/2% for recorders,     oscilloscopes       and similar    instruments.    These can be used
on all ranges of the Model 601.

                                                     NOTE                                          .
        The Model 601 may be used with the FEEDBACK Switch in FAST position   with other
        instruments.   However, make sure there is no common ground between low terminals
        of the Electrometer  and the other instrument.

  c. l-Volt   output.  Connect oscilloscopes     and pen recorder amplifiers     to the OUTPUT
Receptacle.   Pin no. 1 is the negative terminal     and pin no. 2 is grounded when the FEED-
BACK Switch is set to NORMAL. Set the OUTPUT Switch to 1 V. The Model 601 output is now
*l volt for full-scale  meter deflection     on any range.   Internal resistance     is 910 ohms.
The frequency response (*3 db) is dc to 200 cps at a gain of 1000, rising          to 50 kc at a
gain of 3.3, and decreasing   to 1 kc at a gain of 0.1.     Noise is less than 2% rms of full
scale at a gain of 1000, decreasing     to 0.5% at gains below 10. The METER Switch does not
reverse the output polarity.

0667R                                                                                                  17
OPERATION                                                                         MODEL 601 ELECTROMETER



                       NOTE

      Neither terminal of the OUTPUT
      Receptacle should be at case
      ground potential  if the instru-
      ment is used off ground.     Make
                                                                            L              1 kR
      sure the shell of any mating                            M^rl.c.l
                                                               Lx"UrL

      plug is not connected to either                         601
      terminal  in the Receptacle.     The                                                             >
                                                              output                   t
      shorting  link should not be con-                        (Tlcl?I                 z
      nected between LO and CASE GROUND.                                                          Recorder
      Use a recorder with an input iso-
      lated from ground when making off
      ground measurements.
    d. l-Milliampere       Output.    Connect 1-mil-   b                                                     .
liampere instruments         to the OUTPUTRecept-        FIGURE 8. Divider Circuits         Across Model
acle.      Pin no. 1 Fs the negative terminal.           601 Output for Driving 50 and lOO-Millivolt
Set the OUTPUT Switch to 1 MA. The output                Recorders.   Use 1% resistors        in the divid-
is approximately        1 milliampere    for full-       ers for 50-millivolt      recorders,    resistor    R
scale meter deflection          on any range.      For   is 5Ofi; for 100-millivolt,       R is 1OOQ.
exact output, adjust the meter on the .003-
volt range with the FINE ZERO Control for
full-scale     deflection.       Then adjust the 1 MA CAL Control until      the recorder reads full
scale.      Check the recorder and meter zero and repeat



◦ Jabse Service Manual Search 2024 ◦ Jabse PravopisonTap.bg ◦ Other service manual resources online : FixyaeServiceinfo