LECROY AP034-OM-E-03 OPERATION


>> Download LECROY AP034-OM-E-03 OPERATION documenatation <<

Text preview - extract from the document
                                                                        2SHUDWLRQ



2SHUDWLRQ
                                                        Caution
                                The input circuits in the AP034 Active Differential
                                Probe incorporate components that protect the
                                probe from damage resulting from electrostatic
                                discharge (ESD). Keep in mind that this is an
                                active probe, and it should be handled carefully
                                to avoid damage. When using the AP034 Active
                                Differential Probe, you are advised to take
                                precautions against potential instrument damage
                                due to ESD.
CONNECTING THE PROBE TO THE TEST INSTRUMENT
                        When you are using the AP034 Active Differential Probe with a
                        LeCroy Oscilloscope equipped with ProBus, attach the probe
                        output connector to the oscilloscope input connector. The
                        oscilloscope will recognize the probe, set the oscilloscope input
                        termination to 50  DQG DFWLYDWH WKH SUREH FRQWURO IXQFWLRQV LQ
                        the user interface.
                        To use the AP034 Active Differential Probe with instrumentation
                        not equipped with a ProBus interface, it is necessary to use the
                        ADPPS Power Supply. Attach the ADPPS connector to the probe
                        output connector. The output connector of the ADPPS is a
                        standard male BNC that can be directly connected to another
                        instrument. If necessary, the output of the ADPPS can be
                        interconnected with a 50  FRD[LDO FDEOH 7R PLQLPL]H WKH HIIHFWV
                        of skin loss, this cable should be 1 m or less in length. The
                        AP034 Active Differential Probe is designed to drive a 50  ORDG
                        The gain will be uncalibrated if the output is not correctly
                        terminated. If you are using the probe with an instrument with a
                        high input impedance, place a 50  %1& LQOLQH WHUPLQDWRU RQ WKH
                        instrument input before attaching the ADPPS.
CONNECTING THE PROBE TO THE TEST CIRCUIT
                        At the probe tip, two inputs and a ground connection are
                        available for connecting the probe to a circuit under test. For
                        accurate measurements, both the + and 