LECROY WS Series Remote Control


>> Download LECROY WS Series Remote Control documenatation <<

Text preview - extract from the document
                   WAVESURFER OSCILLOSCOPE
                   Errata Sheet for Remote Control Manual
Most of the commands contained in the Remote Control Manual for the WaveMasterTM DSO and other
X-StreamTM instruments apply also to the WaveSurfer oscilloscope. There are also two new WaveSurfer
scope commands that are not in the manual. The Remote Control Manual can be found in the Technical
Library at LeCroy's Web site http://www.lecroy.com.
Please refer to the following table for additions, exclusions, and modifications.

        Command                   Status                                       Notes
   COMBINE_CHANNELS            Not Applicable
                                                Allowed modes: CUST [,STAT], FAIL, HREL [,ABS, DELTA,
    CURSOR_MEASURE               Modified
                                                SLOPE], OFF, PASS, VREL [,ABS, DELTA]
                                                Allowed traces: F1, F5, F6, F7, F8, TA, C1, C2, C3, C4, M1, M2,
       CURSOR_SET                Modified       M3, M4
                                                F5 to F8 correspond to Z1 to Z4, respectively.
     CURSOR_VALUE?               Modified       Allowed traces: F1, F5, F6, F7, F8, TA, C1, C2, C3, C4
         CURSORS               Not Applicable
                                                Allowed function is F1 only.
          DEFINE                 Modified
                                                Not all function parameters are allowed.
    FUNCTION_RESET               Modified       Allowed functions: F1 or TA.
           GRID                  Modified       Allowed arguments: SINGLE, XYONLY, XYSINGLE
                                                Allowed prefixes:

      HOR_MAGNIFY                Modified       TA or F1 for the Math trace
                                                Z1 to Z4 for F5 to F8
                                                F2 to F4, TB to TD are not valid
                                                Allowed prefixes:

      HOR_POSITION               Modified       TA or F1 for the Math trace
                                                Z1 to Z4 for F5 to F8
                                                F2 to F4, TB to TD are not valid
                                                This command sets the input impedance.
                                                Syntax:
        IMPEDANCE                  New
                                                :IMPEDANCDE 
                                                := A1M, D1M, D50, GND, 0VL
                                                   For external trigger: D50, D1M, GND
                                                Allowed prefixes:
                                                C1 to C4 for channel traces
           INSP?                 Modified       M1 to M4 for memory traces
                                                TA or F1 for the Math trace
                                                Z1 to Z4 for F5 to F8


                                                                                                                  1
WAVESURFER OSCILLOSCOPE
                                            F2 to F4, TB to TD are not valid
        MEMORY_SIZE        Not Applicable
                                            Allowed parameter types: CUST, OFF
        PARAMETER            Modified
                                            Allowed parameter readout: STAT, OFF
      PARAMETER_CLR          Modified       Pass/Fail is not applicable.
     PARAMETER_CUSTOM        Modified       Only parameter columns 1 to 6 allowed.
                                            Pass/Fail is not applicable.
     PARAMETER_DELETE        Modified
                                            Only parameter columns 1 to 6 allowed.
                                            Allowed parameter type: CUST
    PARAMETER_STATISTICS     Modified
                                            Allowed parameters: P1 to P6
                                            Allowed traces: F1, F5, F6, F7, F8, TA, C1, C2, C3, C4
     PARAMETER_VALUE?        Modified
                                            Not allowed: CUST7, CUST8
                                             argument is not valid.
         PASS_FAIL           Modified
                                             settings allowed: AllTrue, AllFalse
                                            Allowed traces: F1, F5, F6, F7, F8, TA, C1, C2, C3, C4, M1, M2,
      PER_CURSOR_SET         Modified
                                            M3, M4
        PERSIST_SAT          Modified       Trace selection is not allowed.
       PERSIST_SETUP         Modified       Allowed mode: ALL
       SAMPLE_CLOCK          Modified       Allowed states: INTERNAL, ECL, TTL, 0V
         SEQUENCE          Not Applicable
        STORE_SETUP          Modified       F2 to F4, TB to TD are not valid
           TRACE             Modified       F2 to F4, TB to TD are not valid
                                            ETM10 is not valid
       TRIG_COUPLING         Modified
                                            New values: AC, DC, HF, HFREJ, LFREJ
         TRIG_LEVEL          Modified       ETM10 is not valid
        TRIG_SELECT          Modified       New types: TV, RUNT, SLEW
        TRIG_SLOPE           Modified       ETM10 is not valid
                                            The TRIG_WINDOW command sets the window amplitude in
                                            volts on the current Edge trigger source. The window is
                                            centered around the Edge trigger level.
        TRIG_WINDOW            New
                                            Syntax:
                                            TRIG_WINDOW 
                                            := 0 to 25 V
       VERT_MAGNIFY          Modified       F2 to F4, TB to TD are not valid
       VERT_POSITION         Modified       F2 to F4, TB to TD are not valid
         WAVEFORM            Modified       F2 to F4, TB to TD are not valid