Text preview - extract from the document
www.thinkSRS.com                                                                                                 1



Gas Correction Factors for
Bayard-Alpert Ionization Gauges

The sensitivity factor, Sg, supplied by gauge manufacturers (usually in Torr-1), is valid only for the gas for
which it is specified and the readout of the controller provides a direct pressure reading only for that
specific gas. The standard gas, used by the entire industry for gauge specification, is nitrogen and, unless
gas correction factors are applied, all readings are considered to be 'nitrogen-equivalent pressures'.


In This Application Note
Gas Correction Factors                           3
   Nominal Gas Correction Factors for Common
   Gases                                         4

References                                       5




(408)744-9040                                                                 Stanford Research Systems
www.thinkSRS.com
2   Gas Correction Factors




    Stanford Research Systems        (408)744-9040
                                www.thinkSRS.com
                                                                Gas Correction Factors                3



Gas Correction Factors
          The sensitivity factor, Sg, supplied by gauge manufacturers (usually in Torr-1), is only
          valid for the gas for which it is specified and the pressure readout of a BAG controller
          provides a direct reading only for that specific gas. The standard gas, used by the entire
          industry for gauge specification, is nitrogen and, unless gas correction factors are applied,
          all readings are considered to be 'nitrogen-equivalent pressures'.

          Nominal relative sensitivity factors, Rg, to convert nitrogen-equivalent readings into
          direct pressure readouts for gases other than nitrogen, are available from most gauge
          manufacturers and from the public vacuum literature. A summary table and reference list,
          is included in this application note. For gases where little or no data are available, it has
          been shown that a reasonable approximation to the relative sensitivity factor, Rg, can be
          obtained from the ratio of ionization cross sections at 150 eV of electron energy. Several
          ionization cross section tables are also available in the scientific literature.

          Once the relative sensitivity factor is known, direct pressure readings are calculated from
          the straightforward mathematical equation:

                   P = [ Ic / (Sg 