                                                                                                                 *TB 9-6625-2005-35
                                                                                               SUPERSEDED COPY DATED 24 APRIL 1981

                                    DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

                                                   CALIBRATION PROCEDURE FOR
                                                       SPECTRUM ANALYZER
                                                  TEKTRONIX, TYPE 7L5 (PL-1391/U)
                                                       W/PLUG-IN L1 AND L3

                                     Headquarters, Department of the Army, Washington, DC
                                                         16 June 1988

                                        Approved for public release; distribution is unlimited.

                                              REPORTING OF ERRORS
        You can help improve this publication by calling attention to errors and by recommending improvements
        and stating your reasons for the recommendations. Your letter or DA Form 2028, Recommended
        Changes to Publications, should be mailed directly to Commander, U.S. Army TMDE Support Group,
        ATTN: AMXTM-LPP, Redstone Arsenal, AL 35898-5400. A reply will be furnished directly to you.

                                                                                                                      Paragraph   Page
SECTION        I.   IDENTIFICATION AND DESCRIPTION
                    Test instrument identification ...........................................................           1         2
                    Forms, records, and reports.............................................................             2         2
                    Calibration description.....................................................................         3         2

              II.   EQUIPMENT REQUIREMENTS
                    Equipment required.........................................................................          4         3
                    Accessories required.......................................................................          5         3

             III.   CALIBRATION PROCESS
                    Preliminary instructions ...................................................................         6         5
                    Equipment setup.............................................................................         7         6
                    Readout verification (L3 plug-in only)................................................               8         7
                    Calibrator frequency and amplitude..................................................                 9         7
                    Sweep triggering.............................................................................       10        10
                    Dot frequency range and accuracy...................................................                 11        11
                    Display dynamic range ....................................................................          12        13
                    Display flatness ..............................................................................     13        16
                    Frequency span accuracy and linearity.............................................                  14        18
                    Reference level accuracy (for L3 plug-in only)...................................                   15        19
                    Sweep rate accuracy.......................................................................          16        21
                    Displayed frequency stability ...........................................................           17        22
                    Final procedure...............................................................................      18        23

*This bulletin supersedes TB 9-6625-2005-35, 24 April 1981, including all changes.




                                                                                                                                         1
                                                                                                     TB 9-6625-2005-35

                                                        SECTION I

                                         IDENTIFICATION AND DESCRIPTION

1. Test Instrument Identification.          This bulletin        2.   Forms, Records, and Reports
provides instructions for the calibration of Spectrum
Analyzer, Tektronix, Type 7L5 (PL-1391/U) w/Plug-in L1                      a. Forms, records, and reports required for
and L3. The manufacturer's manual was used as the                calibration personnel at all levels are prescribed by TB
prime data source in compiling these instructions. The           750-25.
equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.                               b. Adjustments       to   be   reported     are
                                                                 designated (R) at the end of the sentence in which they
           a. Model Variations. None.                            appear. When adjustments are in tables, the (R) follows
                                                                 the designated adjustment. Report only those
          b. Time and Technique. The time required               adjustments made and designated with (R).
for this calibration is approximately 6 hours, using the
dc and low frequency technique.                                  3. Calibration Description.        TI parameters    and
                                                                 performance specifications which pertain to         this
                                                                 calibration are listed in table 1.


                                            Table 1. Calibration Description

      Test instrument                                                 Performance
        parameters                                                    specifications

 Calibrator                    Frequency range: 500 kHz 