Basic Test Procedures

        This section gives examples of how to set up and use the
        CTS850 SDH/PDH Test Set for several common telecommunication
        network applications:
        H Network continuity checking
        H Transmission signal quality testing
        H Fault tolerance testing
        H Performance monitoring
        H Jitter and wander testing
        H Video timing quality testing


Setting Up the CTS850
        This section presents setup instructions for the CTS850 in tables.
        Perform the steps reading from left to right in the table (see example
        below). The word none in a cell indicates that no action is required.


         Press Menu        Select Menu       Highlight
         Button            Page              Parameter         Select Choice
         Begin here with   Step 2            Step 3            Step 4
         Step 1
                                             Step 5            Step 6
                                             Step 7            Step 8, CTS850
                                                               Setup is complete


        Menu buttons are located on the instrument front panel. Select menu
        pages with the buttons below the display. Use the knob to highlight a
        parameter; then use the buttons at the right to select a choice. Many
        setups require several iterations of highlighting parameters and
        selecting choices. Some setups may require more than one menu
        button or menu page selection as well.


CTS850 SDH/PDH Test Set User Manual                                          3 1
Basic Test Procedures




         The first step in each application setup initializes the CTS850 to a
         known state.


Network Continuity Checking
         With the AUTOSCAN feature of the CTS850, it is easy to make a
         quick continuity check. The continuity check verifies that a SDH
         network element (NE), or portion of the network, is correctly set up
         and provisioned and that the signal passes through intact. In this
         example, the CTS850 transmits an STM-1E signal into an NE. Then
         the CTS850 receives the output from the NE and verifies that the
         AU-4, including its payload patterns, is intact.
         1. Connect the CTS850 to the NE as shown in Figure 3 1.


                        CTS 850 SDH/PDH Test Set




                              All Signals are
                                     STM-1E

                                                    One or More SDH NEs




         Figure 3 1:Setup to Check Network Continuity




3 2                                     CTS850 SDH/PDH Test Set User Manual
                                                               Basic Test Procedures




        2. Set up the CTS850 as follows:


         Press Menu         Select Menu        Highlight
         Button             Page               Parameter            Select Choice
         TEST SETUPS        RECALL             DEFAULT              Recall Setup
                            INSTRUMENT         FACTORY
                            SETUPS             SETTINGS
         TRANSMIT           TRANSMIT           Transmit Rate        STM-1E
                            SETTINGS


        3. Press the AUTOSCAN button. The receiver locks onto the output
           from the NE, displays the signal structure, and shows information
           about the content of AU #1.
        4. To analyze the payload pattern, select Show Payload. The
           CTS850 detects a PRBS of length 223 1.
        5. Check the J1 Path Trace for any message.
        6. Check the front-panel status lights to verify that there are no
           errors, alarms, or failures.
        7. Select EXIT to leave the AUTOSCAN dialog box.


Transmission Signal Quality Testing

        Measuring Bit Error Rate
        The bit error rate (BER) test is one of the best ways to measure the
        quality of an SDH transmission link. You can use the CTS850 to
        make a BER test on a two-way link that is connected in a loop-back
        configuration. If this is a test you perform often, save it as a pass/fail
        test for easy recall.
        1. Connect the CTS850 to a network as shown in Figure 3 2.




CTS850 SDH/PDH Test Set User Manual                                                3 3
Basic Test Procedures




             CTS 850 SDH/PDH Test Set




                  All Signals are
                         STM-1E
                                           One or More SDH NEs    NE




Figure 3 2:Setup to Check a Transmission Link




3 4                                     CTS850 SDH/PDH Test Set User Manual
                                                             Basic Test Procedures




        2. Set up the CTS850 as follows:


         Press Menu        Select Menu       Highlight
         Button            Page              Parameter            Select Choice
         TEST SETUPS       RECALL            DEFAULT              Recall Setup
                           INSTRUMENT        FACTORY
                           SETUPS            SETTINGS
         TRANSMIT          TRANSMIT          Transmit Rate        STM-1E
                           SETTINGS
                                             Payload              Bulk Fill
                                                                  (Equipped)
                                             Tx/Rx Setup          Coupled
         TEST SETUPS       TEST CONTROL      Test Duration        15 Minutes
         RESULTS           SDH RESULTS       none                 SDh Anomalies


        3. Press the START/STOP button.
        4. While the test is running, the CTS850 counts errors and errored
           seconds and then calculates the error ratio for each error type or
           errored seconds.
        5. After 15 minutes have elapsed, the test stops automatically and
           the measured results are held on the SDH RESULTS page until
           another test is started.
        6. When the test is complete, you can store the test results to disk or
           print out a hardcopy. To save the test results to disk, you must
           enter a name for the disk file as follows:


         Press Menu        Select Menu       Highlight
         Button            Page              Parameter            Select Choice
         RESULTS           SAVE RESULTS      Name                 EDIT NAME


        7. Enter an eight-character name for the file.




CTS850 SDH/PDH Test Set User Manual                                              3 5
Basic Test Procedures




         8. If you wish, you can also enter a description of the test results by
            highlighting Description and selecting EDIT TEXT.
         9. Save the test results to disk as follows:


          Press Menu        Select Menu       Highlight
          Button            Page              Parameter         Select Choice
                                              Save to Disk      Save Current


         Testing Mapping and Demapping (All PDH rates)
         The following map/demap procedure applies to all PDH rates in the
         CTS 850 test set: 2 Mb/s, 8 Mb/s, 34 Mb/s 45 Mb/s and 140 Mb/s.
         The example is written in terms of a 2 Mb/s signal, but the basic test
         procedure would be the same in all other PDH rates.
         To test that a 2 Mb/s signal is correctly mapped into an SDH signal,
         you would use a Bit Error Rate test. However, testing mapping/
         demapping requires a different test setup than that described on
         page 3 3.
         In this example, the Transmitter is generating a 2 Mb/s test signal
         which is mapped into a STM 1 signal by the Add/Drop Mux. The
         Receiver monitors the STM 1 signal received from the Add/Drop
         Mux, demaps the 2 Mb/s test signal and verifies the test pattern
         received.
         1. Connect the CTS850 to an Add/Drop Multiplexer (ADM) as
            shown in Figure 3 3.




3 6                                    CTS850 SDH/PDH Test Set User Manual
                                                     Basic Test Procedures




             CTS 850 SDH/PDH Test Set




                                                         2 Mb/s or any
                                                         PDH rate



                                  STM 1
                                                    2 Mb/s or
                                                  any PDH rate
                                                  Add/Drop Mux



        Figure 3 3:Setup to Test Mapping and Demapping




CTS850 SDH/PDH Test Set User Manual                                      3 7
Basic Test Procedures




         2. Set up the CTS850 as follows:


          Press Menu       Select Menu       Highlight
          Button           Page              Parameter         Select Choice
          TEST SETUPS      RECALL            DEFAULT           Recall Setup
                           INSTRUMENT        FACTORY
                           SETUPS            SETTINGS
          TRANSMIT         TRANSMIT          Transmit Rate     2 Mb/s or any
                           SETTINGS                            PDH rate
          RECEIVE          RECEIVE           Receive Rate      STM-1
                           SETTINGS
                                             Structure         TU 12Async or
                                                               applicable TU x
                                             TU Under Test     Set to match how
                                                               the NE is
                                                               provisioned
                                             Payload           2 Mb/s Unframed
                                                               or any PDH rate
          TEST SETUPS      TEST CONTROL      Test Duration     15 Minutes
          RESULTS          SDH RESULTS       none              Defects


         3. Press the START/STOP button.
         While the test is running, the CTS850 counts errors and errored
         seconds and then calculates the error ratio for each error type.
         After 15 minutes have elapsed, the test stops automatically and the
         measured results are held on the SDH RESULTS page until another
         test is started.




3 8                                   CTS850 SDH/PDH Test Set User Manual
                                                            Basic Test Procedures




        4. When the test is complete, you can store the test results to disk or
           print out a hardcopy. To save the test results to disk, you must
           enter a name for the disk file as follows:


         Press Menu        Select Menu       Highlight
         Button            Page              Parameter           Select Choice
         RESULTS           SAVE RESULTS      Name                EDIT NAME


        5. Enter an eight-character name for the file.
        6. If you wish, you can also enter a description of the test results by
           highlighting Description and selecting EDIT TEXT.
        7. Save the test results to disk as follows:


         Press Menu        Select Menu       Highlight
         Button            Page              Parameter           Select Choice
                                             Save to Disk        Save Current



Fault Tolerance Checking
        Fault tolerance testing (sometimes called stimulus/response testing or
        stress testing) is used to ensure that a network responds correctly to
        various fault conditions. Though there are many additional fault
        conditions you can test with the CTS850, this section covers three
        common examples:
        H Response to errors and alarms
        H Response to pointer movements
        H Response to line frequency offset

        Response to Errors and Alarms
        This example uses two CTS850 testers to simultaneously check the
        upstream and downstream responses to an error or alarm condition.
        You can use a single CTS850 to do the same thing by alternately



CTS850 SDH/PDH Test Set User Manual                                             3 9
Basic Test Procedures




         connecting the upstream and downstream signals to the RECEIVE
         input.
         Table 3 1 shows the responses expected from Line Terminating
         Equipment (LTE), such as an STM-4/STM-1 multiplex section
         terminating element, when presented with three possible error and
         alarm conditions.

         Table 3 1:LTE Responses to Errors and Alarms

          Transmitted Error or    Expected Upstream          Expected Down-
          Alarm                   Response                   stream Response
          RS B1 Error             MS REI                     none
          MS AIS                  MS RDI                     AU AIS
          HP RDI                  none                       HP RDI
          LOS                     MS RDI                     AU AIS



         1. Connect two CTS850s to an LTE as shown in Figure 3 4.


       CTS 850 SDH/PDH Test Set                         CTS 850 SDH/PDH Test Set




           All Signals are
                  STM-1E
                                                                Downstream

                                                  LTE
                                  Upstream



Figure 3 4:Setup to Check Error and Alarm Response



3 10                                     CTS850 SDH/PDH Test Set User Manual
                                                           Basic Test Procedures




        2. To check the LTE response to an RS B1 error, set up the upstream
           CTS850 as follows:


         Press Menu      Select Menu       Highlight
         Button          Page              Parameter            Select Choice
         TEST SETUPS     RECALL            DEFAULT              Recall Setup
                         INSTRUMENT        FACTORY
                         SETUPS            SETTINGS
         TRANSMIT        TRANSMIT          Transmit Rate        STM-1E
                         SETTINGS
                                           Tx/Rx Setup          Coupled
         TRANSMIT        Defects &         Error Type           RS B1
                         Anomalies
                                           Error Rate set to    1.0E 5
         RESULTS         SDH RESULTS       none                 SDH Anomalies




CTS850 SDH/PDH Test Set User Manual                                            3 11
Basic Test Procedures




         3. Set up the downstream CTS850 as follows:


          Press Menu      Select Menu       Highlight
          Button          Page              Parameter           Select Choice
          TEST SETUPS     RECALL            DEFAULT             Recall Setup
                          INSTRUMENT        FACTORY
                          SETUPS            SETTINGS
          RECEIVE         RECEIVE           Receive Rate        STM-1E
                          SETTINGS
          RESULTS         SDH RESULTS       none                Defects


         4. Press the START/STOP buttons on both the upstream and
            downstream CTS 850 units so that the START/STOP lights are
            on.
         5. Verify that RS B1 BIP errors are counted in the SDH RESULTS
            display of the upstream CTS850.
         6. Verify there are no errors counted in the SDH RESULTS display
            of the downstream CTS850.
         7. Press the START/STOP buttons on both the upstream and
            downstream CTS850s so that the START/STOP lights are off.
         8. To check the LTE response to an MS AIS alarm, change the
            upstream CTS850 as follows:


          Press Menu      Select Menu       Highlight
          Button          Page              Parameter           Select Choice
          TRANSMIT        Defects &         Error Rate set to   None
                          Anomalies
                                            Transmit Alarm      MS AIS
                                            set to
          RESULTS         SDH RESULTS       none                SDH Defects




3 12                                  CTS850 SDH/PDH Test Set User Manual
                                                        Basic Test Procedures




        9. Change the downstream CTS850 as follows:


         Press Menu      Select Menu      Highlight
         Button          Page             Parameter           Select Choice
         Results         Main Results     SDH Defects         none


        10. Press the START/STOP buttons on both the upstream and
            downstream CTS850 units so that the START/STOP lights are on.
        11. Verify that MS AIS alarm seconds are counted in the SDH
            RESULTS display of the upstream CTS850.
        12. With the downstream CTS850, verify that the displayed K1 or K2
            bytes are not all zeros.
        13. Press the START/STOP buttons on both the upstream and
            downstream CTS850s so that the START/STOP lights are off.
        14. To check the LTE response to a HP RDI alarm, change the
            upstream CTS850 as follows:


         Press Menu      Select Menu      Highlight
         Button          Page             Parameter           Select Choice
         TRANSMIT        Defects &        Error Rate set to   None
                         Anomalies
                                          Transmit Alarm      HP RDI
                                          set to
         RESULTS         SDH RESULTS      none                Defects




CTS850 SDH/PDH Test Set User Manual                                      3 13
Basic Test Procedures




         15. Change the downstream CTS850 as follows:


          Press Menu        Select Menu       Highlight
          Button            Page              Parameter         Select Choice
          RESULTS           SDH RESULTS       none              Anomalies


         16. Press the START/STOP buttons on both the upstream and
             downstream CTS850 units so that the START/STOP lights are on.
         17. Verify that the HP RDI light is on, on the downstream CTS850.

         Response to Pointer Movements
         Pointers allow the SDH network to multiplex signals from indepen-
         dent sources, even if the signal clock rates differ slightly from each
         other. Within the allowed range, pointer movements should have no
         adverse affect on the performance of the network.
         This example transmits a signal with frequent pointer movements to
         an NE and checks that the downstream signal is error free.
         1. Connect the CTS850 to an NE as shown in Figure 3 5.




3 14                                  CTS850 SDH/PDH Test Set User Manual
                                                      Basic Test Procedures




                     CTS 850 SDH/PDH Test Set




                                                          NE
                               All Signals are
                                      STM-1E




        Figure 3 5:Setup to Check Pointer Movements




CTS850 SDH/PDH Test Set User Manual                                   3 15
Basic Test Procedures




         2. Set up the CTS850 as follows:


          Press Menu       Select Menu       Highlight
          Button           Page              Parameter         Select Choice
          TEST SETUPS      RECALL            DEFAULT           Recall Setup
                           INSTRUMENT        FACTORY
                           SETUPS            SETTINGS
          TRANSMIT         TRANSMIT          Transmit Rate     STM-1E
                           SETTINGS
                                             Tx/Rx Setup       Coupled
          TRANSMIT         POINTERS &        Pointer Control   Continuous
                           TIMING
          RESULTS          SDH RESULTS       none              Defects


         3. Press the START/STOP button on the CTS850 so that the
            START/STOP light is on.
         4. Verify that there are no errors in the SDH RESULTS display.
            When the front-panel ERROR light is not on, it also indicates
            there are no errors in the received signal.
         5. To check for received pointer movements, change the CTS850
            setup as follows:


          Press Menu       Select Menu       Highlight
          Button           Page              Parameter         Select Choice
          RESULTS          SDH RESULTS       none              Pointers


         6. Verify that the CTS850 is counting positive pointer justifications.
            Depending on the clock rate of the NE, some negative pointer
            justifications are also possible.




3 16                                  CTS850 SDH/PDH Test Set User Manual
                                                        Basic Test Procedures




        Response to Line Frequency Offset
        Pointers also accommodate differences in line frequency from one
        NE to another. In this example, the CTS850 transmits a signal at an
        offset line frequency to an NE and measures the response.
        1. Connect the CTS850 to an NE as shown in Figure 3 6.


                      CTS 850 SDH/PDH Test Set




                                                            NE
                                All Signals are
                                       STM-1E




        Figure 3 6:Setup to Check Line Frequency Offset Response




CTS850 SDH/PDH Test Set User Manual                                      3 17
Basic Test Procedures




         2. Set up the CTS850 as follows:


          Press Menu        Select Menu       Highlight
          Button            Page              Parameter          Select Choice
          TEST SETUPS       RECALL            DEFAULT            Recall Setup
                            INSTRUMENT        FACTORY
                            SETUPS            SETTINGS
          TRANSMIT          TRANSMIT          Transmit Rate      STM-1E
                            SETTINGS
                                              Tx/Rx Setup        Coupled
          TRANSMIT          POINTERS &        Pointer/Timing     Frequency Offset
                            TIMING            Mode
                                              Offset Mode        Line
                                              Frequency Offset   Stress +4.6 ppm
          RESULTS           SDH RESULTS       none               Network Defects
                                                                 Section Defects
                                                                 Path Defects


         3. Press the START/STOP button on the CTS850 so that the
            START/STOP light is on.
         4. Verify that there are no errors in the SDH RESULTS displays.
            When the front-panel ERROR light is not on, it also indicates
            there are no errors in the received signal.
         5. To check that the NE is generating pointer movements, change
            the CTS850 setup as follows:


          Press Menu        Select Menu       Highlight
          Button            Page              Parameter          Select Choice
          RESULTS           SDH RESULTS       none               Pointers


         6. Verify that negative pointer justifications are being counted.




3 18                                   CTS850 SDH/PDH Test Set User Manual
                                                           Basic Test Procedures




Performance Monitoring
        When a new line is installed you may want to monitor performance
        for a day or so to be sure it is operating correctly. Or, if problems are
        suspected on a line, you may want to run a long test to determine the
        cause of the problem. In either case, the CTS850 makes it easy to
        monitor an electrical or optical line without the need for splitters.
        During a test, the CTS850 simultaneously takes all performance
        measurements, analyzes the results according to the G.821 standard,
        and displays the measurements and analysis in your choice of three
        formats:
        H In a brief summary
        H In detail by type (defect, anomaly, alarm, failure)
        H In detail by layer (regenerator section, multiplexer section, path)
        In this example, the CTS850 is placed directly in line with a live
        optical signal to monitor performance in an unintrusive mode
        (Through Mode).
        1. Connect the CTS850 in line with an optical signal as shown in
           Figure 3 7.




CTS850 SDH/PDH Test Set User Manual                                           3 19
Basic Test Procedures




                          CTS 850 SDH/PDH Test Set




                     NE                                         NE
                                    All Signals are STM-1



         Figure 3 7:Setup to Monitor Performance


         2. Set up the CTS850 as follows:


          Press Menu        Select Menu         Highlight
          Button            Page                Parameter      Select Choice
          TEST SETUPS       RECALL              DEFAULT        Recall Setup
                            INSTRUMENT          FACTORY
                            SETUPS              SETTINGS
          RECEIVE           RECEIVE             Receive Rate   STM-1
                            SETTINGS
          TRANSMIT          TRANSMIT            Tx/Rx Setup    Through Mode
                            SETTINGS
          RESULTS           ERROR               none           SDH G.821
                            ANALYSIS                           Analysis


         3. Press the START/STOP button on the CTS850 so that the
            START/STOP light is on.
         4. As the test is running, the analysis of Regenerator Section errors
            continually updates. You can also look at the analysis of


3 20                                   CTS850 SDH/PDH Test Set User Manual
                                                        Basic Test Procedures




           Multiplexer Section, Path, or Payload errors by pressing the
           buttons corresponding to those choices.
        5. To see network performance data since the start of the test
           presented in graphical form, change the CTS850 setup as follows:


         Press Menu         Select Menu    Highlight
         Button             Page           Parameter         Select Choice
         RESULTS            HISTORY        none              none
                            GRAPHS


        6. The CTS850 history coverage is based on storage interval which
           is determined by TEST SETUP. You can choose history graphs of
           the measured performance indicators in the following categories:
           H Error count (any error you choose)
           H Errored seconds (any error you choose)
           H Pointer justifications
           H Pointer value
           H All failures
           H All alarms




CTS850 SDH/PDH Test Set User Manual                                       3 21
Basic Test Procedures




         7. For example, to change the bottom graph to display the history of
            Path B3 Errored Seconds, change the CTS850 setup as follows:


          Press Menu       Select Menu       Highlight
          Button           Page              Parameter            Select Choice
          RESULTS          HISTORY           none                 Change Bottom
                           GRAPHS
                                             turn the knob to     Path B3 Errored
                                             select the desired   Secs
                                             graph
                                                                  Select Graph


         8. Turning the knob moves the cursor from one bar to the next. The
            specific measurement values at the cursor location are shown in
            the display.
         9. Once the graph fills the width of the display, you can pan through
            the entire test using the knob.
         10. You can compress or expand the time scale with the Zoom Out or
             Zoom In action buttons. Press the button next to Zoom Out to
             change the time scale to 5 minutes per bar.




3 22                                 CTS850 SDH/PDH Test Set User Manual
                                                           Basic Test Procedures




Jitter/Wander Testing
        With the CTS850 jitter generation and analysis capability, it is easy
        to run the following automated ITU jitter conformance tests:
        H Jitter Tolerance
        H Jitter Transfer
        H Output Jitter (Jitter Generation)
        H Pointer Jitter
        This section shows examples of the jitter tolerance, jitter transfer and
        video timing quality tests.

        Testing Jitter Tolerance
        The CTS850 generates a signal that contains jitter at various
        frequencies and feeds it to the network element. Simultaneously, the
        CTS850 receives the output from the network element and analyzes
        its performance.
        You can specify several parameters for the test:
        H Mask type for jitter analysis
        H Start frequency
        H End frequency
        H Number of frequency samples
        H Tolerance criteria
        The results are displayed in numerical and graphical form.
        The CTS850 provides two methods to measure the tolerance of a
        network element to jitter applied at its input:
        H Onset-of-errors method
        H 1 dB power penalty (BER) method
        These methods are fully specified and described in the ITU-T
        Recommendation O.171 test set standard.




CTS850 SDH/PDH Test Set User Manual                                         3 23
Basic Test Procedures




         Onset-of-Errors Method. The onset-of-errors method measures the
         error output of a network element when jitter is applied to its input.
         If there are two or more Errored Seconds in a 30-second time
         interval, then the CTS850 plots the jitter frequency and amplitude on
         a graph and compares it to a mask you can choose.
         1. Connect the CTS850 to the network element as shown in
            Figure 3 8.


                        CTS 850 SDH/PDH Test Set




                                   These two                  NE
                                   signals are
                                     STM 1E



         Figure 3 8:Setup to Check Onset-of-Errors Jitter Tolerance


         2. Set the transmit and receive signals to STM-1E.




3 24                                   CTS850 SDH/PDH Test Set User Manual
                                                              Basic Test Procedures




        3. Set up the CTS850 as follows:


         Press Menu        Select Menu        Highlight
         Button            Page               Parameter            Select Choice
         TRANSMIT          JITTER &           Output Jitter        Line
                           WANDER
         TEST SETUPS       JITTER TESTS       Test Type            Jitter Tolerance
                                              Test Control         Edit Setup
                                              Mask Type            as appropriate
                                              Start Frequency      as appropriate
                                              End Frequency        as appropriate
                                              Number Freq          as appropriate
                                              Samples
                                              Tolerance Criteria   Onset of Errors


        4. Press the START/STOP button on the CTS850 so that the
           START/STOP light is on.
        As the test is running, the jitter tolerance graph is displayed on the
        screen. The results of each frequency point will also be displayed.
        The test stops automatically after it has completed. The CTS 850
        will also display whether the test passed or failed.




CTS850 SDH/PDH Test Set User Manual                                              3 25
Basic Test Procedures




         1 dB Power Penalty Method. The 1 dB power penalty method uses the
         BER to measure the amount of jitter that is equivalent to a 1 dB loss
         of signal power at the optical input of a network element.
         1. Connect the CTS850 to the network element as shown in
            Figure 3 9.


                    CTS 850 SDH/PDH Test Set




                                                          All Signals are STM-1



                               Variable
                                optical                       NE
                              attenuator



         Figure 3 9:Setup to Check 1 dB Power Penalty Jitter Tolerance




3 26                                   CTS850 SDH/PDH Test Set User Manual
                                                           Basic Test Procedures




        2. Set up the CTS850 as follows:


         Press Menu       Select Menu      Highlight
         Button           Page             Parameter            Select Choice
         TRANSMIT         JITTER &         Jitter Output        Line
                          WANDER
         TEST SETUPS      JITTER TESTS     Test Type            Jitter Tolerance
                                           Test Control         Edit Setup
                                           Mask Type            as appropriate
                                           Start Frequency      as appropriate
                                           End Frequency        as appropriate
                                           Number Freq          as appropriate
                                           Samples
                                           Tolerance Criteria   BER
                                           Record Error         Measure BER
                                           Threshold


        3. Observe the Current BER and Recommended BER values in the
           display. Increase the optical attenuation until the Current BER
           reaches or exceeds the Recommended BER.
        4. When the Current BER exceeds the Recommended BER:


         Press Menu       Select Menu      Highlight
         Button           Page             Parameter            Select Choice
         TEST SETUPS      JITTER TESTS     Record Error         Record BER
                                           Threshold


        5. Reduce the optical attenuation by 1 dB.
        6. Press the START/STOP button on the CTS850 so that the
           START/STOP light is on.




CTS850 SDH/PDH Test Set User Manual                                           3 27
Basic Test Procedures




         The CTS850 applies progressively increasing jitter to the input until
         the Current BER reaches the threshold you set. At that time, the test
         stops automatically and the results are displayed.

         Testing Clock-to-Line Jitter Transfer
         This test measures the transfer function for jitter in the output of a
         network element relative to jitter in its clock. The CTS850 generates
         a 2 MHz G.703 clock signal containing jitter that is connected to the
         network element. The CTS850 also generates a jitter-free test pattern
         connected to the input of the network element. Then, the CTS850
         receives the output from the network element and analyzes its
         performance.
         You can specify several parameters for the test:
         H Mask type for jitter analysis
         H Start frequency
         H End frequency
         H Number of frequency samples
         The results are displayed in numerical and graphical form.
         1. Connect the CTS850 to the network element as shown in
            Figure 3 10.




3 28                                   CTS850 SDH/PDH Test Set User Manual
                                                             Basic Test Procedures




                                                  2 MHz G.703 clock
                                                   from J/CLK OUT

                      CTS 850 SDH/PDH Test Set




                                                                       Clock input


                                                                  NE
                                All Signals are
                                       STM-1E




        Figure 3 10:Setup to Check Clock-to-Line Jitter Transfer




CTS850 SDH/PDH Test Set User Manual                                                  3 29
Basic Test Procedures




         2. Set up the CTS850 as follows:


          Press Menu        Select Menu        Highlight
          Button            Page               Parameter          Select Choice
          TRANSMIT          JITTER &           Jitter Output      MHz G.703
                            WANDER
          RECEIVE           JITTER &           Jitter Input       Line
                            WANDER             Source
          TEST SETUPS       JITTER TESTS       Test Type          Jitter Transfer
                                               Test Control       Edit Setup
                                               Mask Type          as appropriate
                                               Start Frequency    as appropriate
                                               End Frequency      as appropriate
                                               Number Freq        as appropriate
                                               Samples


         3. Press the START/STOP button on the CTS850 so that the
            START/STOP light is on.
         As the test is running, the jitter transfer graph is displayed on the
         screen. The test stops automatically after it has completed.




3 30                                   CTS850 SDH/PDH Test Set User Manual
                                                           Basic Test Procedures




Video Timing Quality Testing
        Compressed digital video signals transmitted over a network can be
        impaired by low-frequency jitter or frequency drift that may be
        acceptable in other applications. One source of this jitter is pointer
        movement used to synchronize SDH payloads across a network. You
        can use the CTS850 to analyze jitter in a PDH signal due to pointer
        movement in the SDH signal. Two new measurements, Frequency
        Drift Rate and Pointer Hit Seconds, help show this effect.
        The CTS850 generates a SDH signal containing a sequence of
        pointer movements. The CTS850 analyzes the timing of the dropped
        34 Mb/s signal.
        1. Connect the CTS850 to the network element as shown in
           Figure 3 11.


            CTS 850 SDH/PDH Test Set




                                                       34 Mb/s




                                 STM-1
                                                      34 Mb/s
                                                    Add/Drop Mux



        Figure 3 11: Setup to Check Video Timing Quality




CTS850 SDH/PDH Test Set User Manual                                        3 31
Basic Test Procedures




         2. Set up the CTS850 as follows:


          Press Menu        Select Menu       Highlight
          Button            Page              Parameter          Select Choice
          TRANSMIT          TRANSMIT          Transmit Rate      STM-1E
                            SETTINGS
                                              Structure          TU3 Async
          TRANSMIT          JITTER &          Jitter Output      Line
                            WANDER
          TRANSMIT          POINTERS &        Pointer/Timing     Pointer
                            TIMING            Mode               Sequences
                                              Pointer Type       AU Pointer
                                              Sequence Type      Periodic 87 3
          RECEIVE           RECEIVE           Receive Rate       34 Mb/s
                            SETTINGS
          RESULTS           SDH RESULTS       none               Jitter


         3. Press the POINTER ACTION button to start the pointer
            sequence.
         4. Press the START/STOP button on the CTS850 so that the
            START/STOP light is on.
         As the test is running, results are displayed on the screen. The test
         stops automatically after it has completed.




3 32                                   CTS850 SDH/PDH Test Set User Manual